VentriClear Ventricular Drainage Catheter Set with Cook Spectrum Antibiotic Impregnation, VentriClear II Ventricular Drainage Catheter
Applicant
Cook Incorporated
Product Code
NHC · Neurology
Decision Date
Apr 8, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VentriClear Ventricular Drainage Catheter Set with Cook Spectrum Antibiotic Impregnation and the VentriClear II Ventricular Drainage Catheter have been designed for obtaining access to a ventricular cavity of the brain for short-term use to externally drain fluid for the purpose of relieving elevated intracranial pressure or fluid volume.
Device Story
Ventricular drainage catheters provide access to brain ventricular cavities for short-term external fluid drainage; used to relieve elevated intracranial pressure or fluid volume. Catheters are 9 French, 33 cm long, silicone elastomer, impregnated with minocycline and rifampin to provide antimicrobial properties. Radiopaque tip aids radiographic visualization; depth markings facilitate placement. Set includes stainless steel stylet/obturator, tunneling trocar, Luer lock components, and silicone winged tie-down. Used in clinical settings by physicians. Output is external drainage of cerebrospinal fluid; clinical decision-making relies on monitoring intracranial pressure and fluid volume. Benefit is reduction of intracranial pressure while mitigating infection risk via antimicrobial impregnation.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
9 French silicone elastomer catheter; 33 cm length; radiopaque tip; 16 or 32 sideports; closed-end configuration; impregnated with minocycline and rifampin. Sterilized via Ethylene Oxide (EtO).
Indications for Use
Indicated for patients requiring short-term access to the brain's ventricular cavity for external fluid drainage to relieve elevated intracranial pressure or fluid volume. Contraindicated in patients with known allergy to tetracyclines or rifampin, and in pregnant women.
Regulatory Classification
Identification
A ventricular catheter is a device used to gain access to the cavities of the brain for injection of material into, or removal of material from, the brain.
VentriClear II Ventricular Drainage Catheter Set (K071640)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 8, 2016
Cook Incorporated Mr. Steven Lawrie, MS, MA, RAC Regulatory Affairs Specialist 750 Daniels Way Bloomington, IN 47404
Re: K160223
Trade/Device Name: VentriClear Ventricular Drainage Catheter Set with Cook Spectrum Antibiotic Impregnation and VentriClear II Ventricular Drainage Catheter Regulation Number: 21 CFR 882.4100 Regulation Name: Ventricular Catheter Regulatory Class: Class II Product Code: NHC Dated: March 9, 2016 Received: March 10, 2016
Dear Mr. Steven Lawrie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -5D/Δ
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160223
#### Device Name
VentriClear Ventricular Drainage Catheter Set with Cook Spectrum Antibiotic Impregnation VentriClear II Ventricular Drainage Catheter
#### Indications for Use (Describe)
The VentriClear Ventricular Drainage Catheter Set with Cook Spectrum Antibiotic Impregnation and the VentriClear II Ventricular Drainage Catheter have been designed for obtaining access to a ventricular cavity of the brain for short-term use to externally drain fluid for the purpose of relieving elevated intracranial pressure or fluid volume.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/14 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in large, white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font on a red background.
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA 812.339.2235 TOLL BREE 800.457.4500 WWW.COOKMEDICAL.COM
# 510(k) Summary
# VentriClear® Ventricular Drainage Catheter Set with Cook® Spectrum® Antibiotic Impregnation VentriClear® II Ventricular Drainage Catheter 21 CFR §882.4100 Date Prepared: February 23, 2016
## Submitted By:
| Applicant: | Cook Incorporated |
|-----------------------|------------------------------------------|
| Contact: | Steven Lawrie |
| Applicant Address: | 750 Daniels Way<br>Bloomington, IN 47404 |
| Contact Phone Number: | (812) 335-3575 x104518 |
| Contact Fax Number: | (812) 332-0281 |
#### Device Information
| Trade Name: | VentriClear® Ventricular Drainage Catheter Set<br>with Cook® Spectrum® Antibiotic Impregnation<br>VentriClear® II Ventricular Drainage Catheter |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Ventricular Catheter |
| Classification Name: | Catheter, Ventricular (Containing Antibiotic Or<br>Antimicrobial Agents)<br>NHC (21 CFR §882.4100) |
| Class: | Class II |
## Predicate Devices:
- K011812, Spectrum® Ventricular Catheter .
- K071640, VentriClear® II Ventricular Drainage Catheter Set .
## Device Description:
The VentriClear® Ventricular Drainage Catheter Set with Cook® Spectrum® Antibiotic Impregnation consists of a VentriClear® Ventricular Drainage catheter and other components that are used to facilitate the procedure. The VentriClear® Ventricular Drainage Catheter is a 9 French catheter designed with a radiopaque tip which aids in radiographic recognition and has a closed-end configuration with 16 standard and 16 small sideports. The VentriClear® Ventricular Drainage Catheter has markings at 1 cm
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Cook Medical logo. The word "COOK" is in white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font on a red background.
increments to aid in determining depth and ease of placement and is 33 cm in length. The VentriClear® Ventricular Drainage Catheter consists of silicone elastomer impregnated with the antimicrobial agents minocycline and rifampin.
The VentriClear® II Ventricular Drainage Catheter consists of a VentriClear® II Ventricular Drainage catheter and other components that are used to facilitate the procedure. The VentriClear® II Ventricular Drainage Catheter is a 9 French catheter designed with a radiopaque tip which aids in radiographic recognition and has a closedend configuration with 16 sideports. The VentriClear® II Ventricular Drainage Catheter has markings at 1 cm increments, starting at the 3 cm mark and extending to 33 cm, to aid in determining depth and ease of placement and is 33 cm in length. The VentriClear® II Ventricular Drainage Catheter consists of silicone elastomer impregnated with the antimicrobial agents minocycline and rifampin.
# Indications for Use:
The VentriClear® Ventricular Drainage Catheter Set with Cook® Spectrum® Antibiotic Impregnation and the Ventriclear® II Ventricular Drainage Catheter have been designed for obtaining access to a ventricular cavity of the brain for short-term use to externally drain fluid for the purpose of relieving elevated intracranial pressure or fluid volume.
The indications for use are same as the predicate devices, the Spectrum® Ventricular Catheter (K011812) and the VentriClear® II Ventricular Drainage Catheter Set (K071640).
## Comparison to Predicate:
The VentriClear® Ventricular Drainage Catheter Set with Cook® Spectrum® Antibiotic Impregnation and the VentriClear® II Ventricular Drainage Catheter are identical to the predicate devices, the Spectrum® Ventricular Catheter (K011812) and the VentriClear® II Ventricular Drainage Catheter Set (K071640), with the exception of the addition of a contraindication. The following contraindication is added to the IFU for the VentriClear Ventricular Drainage Catheter Set:
"VentriClear catheters are contraindicated for use in pregnant women."
Additionally, the following contraindication is added to the IFU for the VentriClear® II Ventricular Drainage Catheter:
"The VentriClear II is contraindicated for use in pregnant women."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Cook Medical logo. The word "COOK" is in white, bold letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, smaller, bold letters on a red background.
Table 1 below provides a detailed comparison of the subject and predicate devices.
Table 1: Comparison Table
| | PREDICATE DEVICES | PROPOSED DEVICE | | |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Spectrum®<br>Ventricular Catheter<br>(K011812) | VentriClear® II<br>Ventricular Drainage<br>Catheter Set<br>(K071640) | VentriClear®<br>Ventricular Drainage<br>Catheter Set with<br>Cook® Spectrum®<br>Antibiotic<br>Impregnation and the<br>VentriClear® II<br>Ventricular Drainage<br>Catheter | |
| Regulation Number | 882.4100 | 882.4100 | 882.4100 | |
| Product Code | NHC | NHC | NHC | |
| Classification | Catheter, Ventricular<br>(Containing Antibiotic<br>Or Antimicrobial<br>Agents) | Catheter, Ventricular<br>(Containing Antibiotic<br>Or Antimicrobial<br>Agents) | Catheter, Ventricular<br>(Containing Antibiotic<br>Or Antimicrobial<br>Agents) | |
| Indications for Use | For obtaining access to<br>a ventricular cavity of<br>the brain for short-term<br>use to externally drain<br>fluid for the purpose of<br>relieving elevated<br>intracranial pressure or<br>fluid volume. | For obtaining access to<br>a ventricular cavity of<br>the brain for short-term<br>use to externally drain<br>fluid for the purpose of<br>relieving elevated<br>intracranial pressure or<br>fluid volume. | Identical | |
| Contraindications | • This device is<br>contraindicated in<br>patients having allergy<br>or history of allergy to<br>tetracyclines or<br>rifampin.<br>NOTE: Because the<br>VentriClear Ventricular<br>Drainage Catheter is<br>impregnated with a<br>combination of the<br>antimicrobial agents<br>minocycline (a<br>derivative of<br>tetracycline) and<br>rifampin (a derivative of<br>rifamycin B), their<br>associated<br>contraindications,<br>warnings and<br>precautions (Physician's<br>Desk Reference) should<br>be considered when<br>using this device, | • This device is<br>contraindicated in<br>patients having allergy<br>or history of allergy to<br>tetracyclines or<br>rifampin.<br>NOTE: Because the<br>VentriClear Ventricular<br>Drainage Catheter is<br>impregnated with a<br>combination of the<br>antimicrobial agents<br>minocycline (a<br>derivative of<br>tetracycline) and<br>rifampin (a derivative<br>of rifamycin B), their<br>associated<br>contraindications,<br>warnings and<br>precautions (Physician's<br>Desk Reference) should<br>be considered when<br>using this device, | • This device is<br>contraindicated in<br>patients having allergy<br>or history of allergy to<br>tetracyclines or<br>rifampin.<br>NOTE: Because the<br>VentriClear Ventricular<br>Drainage Catheter is<br>impregnated with a<br>combination of the<br>antimicrobial agents<br>minocycline (a<br>derivative of<br>tetracycline) and<br>rifampin (a derivative of<br>rifamycin B), their<br>associated<br>contraindications,<br>warnings and<br>precautions (Physician's<br>Desk Reference) should<br>be considered when<br>using this device, | |
| | | PREDICATE DEVICES | | PROPOSED DEVICE |
| | | Spectrum®<br>Ventricular Catheter<br>(K011812) | VentriClear® II<br>Ventricular Drainage<br>Catheter Set<br>(K071640) | VentriClear®<br>Ventricular Drainage<br>Catheter Set with<br>Cook® Spectrum®<br>Antibiotic<br>Impregnation and the<br>VentriClear® II<br>Ventricular Drainage<br>Catheter |
| Contraindications<br>(continued) | | although there have<br>been no reports of<br>detected systemic levels<br>of minocycline or<br>rifampin in patients<br>receiving a VentriClear<br>Catheter. | although there have<br>been no reports of<br>detected systemic levels<br>of minocycline or<br>rifampin in patients<br>receiving a VentriClear<br>Catheter. | although there have<br>been no reports of<br>detected systemic levels<br>of minocycline or<br>rifampin in patients<br>receiving a VentriClear<br>Catheter.<br>• [VentriClear catheters<br>/ The VentriClear II] are<br>contraindicated for use<br>in pregnant women. |
| | Material | Silicone | Silicone | Identical |
| | Size | 9 Fr | 9 Fr | Identical |
| | Length | 33 cm | 33 cm | Identical |
| | Sideports | 32 | 16 | 32 (VentriClear), 16<br>(VentriClear II) |
| Catheter | End-hole | Closed | Closed | Identical |
| | Depth<br>Marks | Yes | Yes | Identical |
| | Tip | Radiopaque | Radiopaque | Identical |
| | Antimicro-<br>bial Agents | Minocycline/Rifampin | Minocycline/Rifampin | Identical |
| Set Components | | Pre-loaded stainless<br>steel stylet/obturator,<br>stainless steel tunneling<br>trocar with proximal<br>fittings, Luer Lock Plug,<br>Female Luer Lock<br>Adapter, Snap Fit Cap,<br>and Silicone Winged<br>Tie Down | Pre-loaded stainless<br>steel stylet/obturator,<br>stainless steel tunneling<br>trocar with proximal<br>fittings, Luer Lock<br>Plug, Female Luer Lock<br>Adapter, Snap Fit Cap,<br>and Silicone Winged<br>Tie Down | Identical |
| Packaging | | Tyvek/polyester pouch | Tyvek/polyester pouch | Identical |
| Sterilization | | EtO | EtO | Identical |
| Shelf Life | | 2 years | 2 years | Identical |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, white, sans-serif letters. Below the word "COOK" is the word "MEDICAL" in smaller, white, sans-serif letters on a red background.
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA PHONE: 812.339.2235 TOLL PREE: 800.457.4500 WWW.COOKMEDICAL.COM
# Table 1: Comparison Table (continued)
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, bold letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, smaller letters on a red background. The red background is a rectangle shape.
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA HONE: 812.339.2235 TOLL FREE 800.457.4500 WWW.COOKMEDICAL.COM
# Nonclinical Testing:
There were no device changes, so no additional testing was required to support the determination of substantial equivalence to the predicate devices.
# Conclusion:
The subject devices are substantially equivalent to the predicate devices, the Spectrum® Ventricular Catheter (K011812) and the VentriClear® II Ventricular Drainage Catheter Set (K071640).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.