← Product Code [HCD](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HCD) · K102101

# SURGIVISION MR COMPATIBLE VENTRICULAR CANNULA (K102101)

_Surgivision, Inc. · HCD · Jan 26, 2011 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HCD/K102101

## Device Facts

- **Applicant:** Surgivision, Inc.
- **Product Code:** [HCD](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HCD.md)
- **Decision Date:** Jan 26, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4060
- **Device Class:** Class 1
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The MR Compatible Ventricular Cannula is intended for injection of Cytarabine or removal of CSF from the ventricles during intracranial procedures. The device is not intended for implant. The device is intended for "single patient use only."

## Device Story

Ventricular cannula for intracranial access; used in MRI diagnostic/surgical rooms. Device inputs: manual manipulation by physician; outputs: fluid injection (Cytarabine) or CSF aspiration. Features rigid ceramic body; non-reactive silica lumen; proximal lumen extension (3ft or 9ft) for remote operation outside MRI bore. Designed for MR compatibility; allows physician to perform procedures while patient is inside scanner. Benefits: enables real-time visualization during placement and remote fluid management without removing device or requiring non-MR safe equipment near scanner. Single-patient use; sterile.

## Clinical Evidence

Bench testing only. Includes design verification, pressure withstand, injection/aspiration flow testing, lateral tip deflection, transit testing (ASTM D4169), and accelerated aging. Biocompatibility testing per ISO 10993 (cytotoxicity, pyrogen, sensitization, toxicity). Cytarabine injection and primate CSF aspiration validated in bench studies.

## Technological Characteristics

Rigid ceramic body; non-reactive silica fluid lumen; PVC proximal extension tubing. Female luer connector. MR compatible/safe. Sterilized via radiation (ISO 11137-2). Dimensions: 16ga or 14ga; 4ft or 10ft lengths. No side holes. Manual operation.

## Regulatory Identification

A ventricular cannula is a device used to puncture the ventricles of the brain for aspiration or for injection. This device is frequently referred to as a ventricular needle.

## Predicate Devices

- Adson Cannula (9Fr x 14 cm; stainless steel)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 510(k) Summary

:

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR Part 807.92 for the SurgiVision MR Compatible Ventricular Cannula.

#### 1. Company Making the Submission:

JAN 2 6 2011

6

| Name of Owner: | SurgiVision, Inc.        |
|----------------|--------------------------|
| Address:       | 5 Musick                 |
|                | Irvine, CA 92618         |
| Telephone:     | 949-900-6833             |
| Fax:           | 949-900-6834             |
| Contact:       | Edward Waddell           |
| E-mail:        | ewaddell@surgivision.com |

#### Device Name: 2.

| Common Name:       | Ventricular Cannula                  |
|--------------------|--------------------------------------|
| Proprietary Name:  | MR Compatible<br>Ventricular Cannula |
| Classification:    | Class I                              |
| Regulation Number: | 882.4060                             |
| Product Code:      | HCD                                  |

#### Predicate Device: 3.

Adson Cannula (9Fr x 14 cm; stainless steel) (class I, 510(k)-exempt)

#### Description of Device: 4.

The Cannula has a stepped distal tip with a 30 cm rigid ceramic stylet protecting the fluid lumen while providing rigidity to the distal portion of the device. Soft this in areets the lumen in the center portion and at the distal end where it terminates at a female luer fitting. The fluid containing central lumen is manufactured from non-reactive silica.

The cannula will be marketed in the following sizes:

- 16 ga Ventricular Cannula, .008" ID x 4ft
- 16 ga Ventricular Cannula. .008" ID x 10ft
- 14 ga Ventricular Cannula, .021" ID x 4ft

14 ga Ventricular Cannula, .021" ID x 10ft

Surgi Vision, Inc.

Section 5 510(k) Submission Page #

{1}------------------------------------------------

#### 5. Intended Use Statement:

The MR Compatible Ventricular Cannula is intended for injection of Cytarabine or removal of CSF from the ventricles during intracranial procedures. The device is not intended for implant. The device is intended for "single patient use only."

## 6. Summary of the Technological Characteristics of the Device Compared to the Predicate Device

|                                                        | SurgiVision<br>Ventricular Cannula<br>(VC)                                                                                       | Predicate Device:<br>Adson Ventricular<br>Cannula                                                                     | Discussion                                                                                                                                                                                                                                                                                                                      |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification                                         | 21 CFR 882.4060                                                                                                                  | 21 CFR 882.4060                                                                                                       | Equivalent                                                                                                                                                                                                                                                                                                                      |
| Product Code                                           | HCD                                                                                                                              | HCD                                                                                                                   | Equivalent                                                                                                                                                                                                                                                                                                                      |
| Intended Use                                           | Injection of Cytarabine to the<br>ventricles of the brain or aspiration<br>of CSF from the ventricles of the<br>brain            | Injection or aspiration of fluid<br>in the ventricles of the brain                                                    | Equivalent                                                                                                                                                                                                                                                                                                                      |
| Indications for<br>Use                                 | Gains access to the brain ventricles                                                                                             | Gains access to the brain<br>ventricles                                                                               | Equivalent                                                                                                                                                                                                                                                                                                                      |
|                                                        | Allows Injection of Cytarabine into<br>the brain ventricles                                                                      | Allows injection of fluids into<br>the brain ventricles                                                               | VC validated using<br>Cytarabine                                                                                                                                                                                                                                                                                                |
|                                                        | Allows aspiration of CSF from the<br>brain ventricles                                                                            | Allows aspiration of fluids from<br>the brain ventricles                                                              | VC validated using<br>primate CSF                                                                                                                                                                                                                                                                                               |
|                                                        | Not implantable                                                                                                                  | Not implantable                                                                                                       | Equivalent                                                                                                                                                                                                                                                                                                                      |
|                                                        | Single Patient Use                                                                                                               | Reusable                                                                                                              | VC is validated for<br>single patient use.<br>Component materials<br>do not lend to<br>multiple use and re-<br>sterilization                                                                                                                                                                                                    |
| Target<br>Population                                   | Pt.s needing injection of<br>Cytarabine to the brain ventricles<br>or aspiration of CSF from the brain<br>ventricles             | Pt.s needing aspiration or<br>injection of fluids from the<br>brain ventricles                                        | VC validated using<br>Cytarabine and<br>primate CSF                                                                                                                                                                                                                                                                             |
| Anatomical Sites                                       | Brain ventricle                                                                                                                  | Brain ventricle                                                                                                       | Equivalent                                                                                                                                                                                                                                                                                                                      |
|                                                        | Operating Room                                                                                                                   | Operating Room                                                                                                        | Equivalent                                                                                                                                                                                                                                                                                                                      |
| Where Used                                             | MRI Diagnostic / Surgical Room                                                                                                   | N/A                                                                                                                   | VC device is MR<br>Compatible/ MR<br>Safe, whereas the<br>predicate device is<br>not indicated for MR<br>environments                                                                                                                                                                                                           |
| Energy used                                            | N/A                                                                                                                              | N/A                                                                                                                   | Equivalent                                                                                                                                                                                                                                                                                                                      |
| Human Factors                                          | Labeling indicates size and length                                                                                               | Labeling indicates size and<br>length                                                                                 | Equivalent                                                                                                                                                                                                                                                                                                                      |
|                                                        | Labeling contains flow vs. pressure<br>tables                                                                                    | No information provided                                                                                               | VC provides<br>additional<br>information to the                                                                                                                                                                                                                                                                                 |
|                                                        | SurgiVision<br>Ventricular Cannula<br>(VC)                                                                                       | Predicate Device:<br>Adson Ventricular<br>Cannula                                                                     | Discussion                                                                                                                                                                                                                                                                                                                      |
|                                                        |                                                                                                                                  |                                                                                                                       | physician regarding<br>flow and pressures<br>for the various sizes                                                                                                                                                                                                                                                              |
|                                                        | Can be manipulated with gloved<br>hand                                                                                           | Can be manipulated with gloved<br>hand                                                                                | Equivalent                                                                                                                                                                                                                                                                                                                      |
|                                                        | Designed to be placed through a<br>prepared opening through the skull<br>and dura into the brain ventricle                       | Designed to be placed through a<br>prepared opening through the<br>skull and dura into the brain<br>ventricle         | Equivalent                                                                                                                                                                                                                                                                                                                      |
|                                                        | Rigid section to enter the brain                                                                                                 | Rigid section to enter the brain                                                                                      | Equivalent                                                                                                                                                                                                                                                                                                                      |
| Design                                                 | Straight section to enter the brain                                                                                              | Straight section to enter the<br>brain                                                                                | Equivalent                                                                                                                                                                                                                                                                                                                      |
|                                                        | Hole at distal end for fluid<br>movement                                                                                         | Hole at distal end for fluid<br>movement                                                                              | Equivalent                                                                                                                                                                                                                                                                                                                      |
|                                                        | No side holes                                                                                                                    | Three .055" (1.4 mm) side<br>holes located 11 mm to 21 mm<br>from the distal end                                      | The VC can only<br>transfer fluids from<br>the distal end                                                                                                                                                                                                                                                                       |
|                                                        | No marking on body of device                                                                                                     | CM markings on body                                                                                                   | VC Device length is<br>provided with the<br>instructions                                                                                                                                                                                                                                                                        |
|                                                        | Length of rigid section:<br>10.5" (30 cm)                                                                                        | Length of rigid section:<br>5.5" (14 cm)                                                                              | The VC has a longer<br>length to facilitate<br>handling of the<br>device when the<br>patient is within the<br>bore of an MR<br>scanner                                                                                                                                                                                          |
|                                                        | Inside diameter:<br>.008" (0.2 mm) to<br>.021" (0.53 mm)                                                                         | Inside diameter:<br>.079" (2.0 mm)                                                                                    | Both devices are<br>moving fluids and<br>their different inside /<br>outside diameters<br>result in different<br>flow rates.                                                                                                                                                                                                    |
|                                                        | Outside diameter:<br>.065" (1.6 mm)<br>and .080" (2.0 mm)                                                                        | Outside diameter:<br>.120" (3.0 mm)                                                                                   |                                                                                                                                                                                                                                                                                                                                 |
|                                                        | No Stylet                                                                                                                        | Stylet                                                                                                                | VC has stepped<br>features for<br>dimensional<br>transitions at the tip<br>and the predicate<br>device uses a stylet to<br>create the transitions                                                                                                                                                                               |
|                                                        | SurgiVision<br>Ventricular Cannula<br>(VC)                                                                                       | Predicate Device:<br>Adson Ventricular<br>Cannula                                                                     | Discussion                                                                                                                                                                                                                                                                                                                      |
|                                                        |                                                                                                                                  |                                                                                                                       | VC does not have a<br>stylet that would have<br>to be removed prior<br>to assisted injection<br>or aspiration<br>VC allows immediate<br>un-assisted aspiration<br>of CSF from ventricle<br>(with visualization by<br>physician) during<br>placement                                                                             |
|                                                        | Lumen extension (3 foot) allows<br>remote (end of scanner bore)<br>injection/aspiration                                          | No Proximal Extension present                                                                                         | VC has an<br>incorporated<br>proximal lumen<br>extension, this<br>provides a single<br>lumen extending from<br>the rigid cannula to<br>the proximal<br>connector                                                                                                                                                                |
|                                                        | Lumen extension (9 foot) allows<br>remote (outside of scanner 5 gauss<br>line) Injection/aspiration using a<br>non-MRI safe pump | No Proximal Extension Present                                                                                         | VC has an<br>incorporated<br>proximal lumen<br>extension; this<br>provides a single<br>lumen extending from<br>the rigid cannula to<br>the proximal<br>connector. This<br>extension provides a<br>continuous lumen that<br>is MR compatible and<br>can connect to a<br>remote, not MR safe<br>pump for<br>injection/aspiration. |
| Performance                                            | Sufficiently rigid to pass through<br>brain tissue without additional<br>support                                                 | Sufficiently rigid to pass<br>through brain tissue without<br>additional support                                      | Equivalent                                                                                                                                                                                                                                                                                                                      |
| Performance                                            | Contains a channel through which<br>fluids can be removed (aspiration)<br>or placed (injection) into the<br>ventricle            | Contains a channel through<br>which fluids can be removed<br>(aspiration) or placed (injection)<br>into the ventricle | Equivalent                                                                                                                                                                                                                                                                                                                      |
|                                                        | Flow rate of:<br>0.3 ml/hr (.008" I.D.) to<br>25.0 ml/hr (.021" I.D.)<br>at 0.7 PSI                                              | Flow rate of 16,000 ml/hr<br>at 0.7 PSI                                                                               | Both devices are<br>moving fluids and the<br>different flow rates<br>result only from their<br>different diameters                                                                                                                                                                                                              |
| Materials                                              | Rigid body:<br>Ceramic                                                                                                           | Rigid body:<br>Stainless Steel or Sterling Silver                                                                     | Both devices use<br>rigid material for the<br>body, VC provides a                                                                                                                                                                                                                                                               |
|                                                        | SurgiVision<br>Ventricular Cannula<br>(VC)                                                                                       | Predicate Device:<br>Adson Ventricular<br>Cannula                                                                     | Discussion                                                                                                                                                                                                                                                                                                                      |
|                                                        |                                                                                                                                  |                                                                                                                       | rigid body that is MR<br>Compatible due to<br>ceramic material                                                                                                                                                                                                                                                                  |
|                                                        | Stylet: N/A                                                                                                                      | Stylet:<br>Stainless Steel or Sterling Silver                                                                         | VC has features that<br>provide rigidity and<br>tip transition which<br>are accomplished by<br>the stylet                                                                                                                                                                                                                       |
|                                                        | Through lumen:<br>Polymer covered silica                                                                                         | Through lumen:<br>Stainless Steel or Sterling Silver                                                                  | Both devices have a<br>continuous through<br>lumen but the VC is<br>MR compatible                                                                                                                                                                                                                                               |
|                                                        | Lumen extension (outer support<br>tubing): PVC extrusion                                                                         | No lumen extension.                                                                                                   | VC is provided with<br>lumen extension                                                                                                                                                                                                                                                                                          |
|                                                        | Proximal Connector:<br>Female Luer connector                                                                                     | Proximal Connector:<br>Female Luer connector                                                                          | Equivalent                                                                                                                                                                                                                                                                                                                      |
| Biocompatibility                                       | Tissue contact tested per ISO<br>10993: Biological Evaluation of<br>Medical Devices                                              | Presumed                                                                                                              | Equivalent                                                                                                                                                                                                                                                                                                                      |
| Compatibility<br>with environment<br>and other devices | Safe in a 1.5T MRI environment                                                                                                   | Not safe in a 1.5T MRI<br>environment                                                                                 | VC device is<br>designed for MR<br>environment                                                                                                                                                                                                                                                                                  |
|                                                        | Female Luer connector on<br>proximal end fits all male luer<br>connectors (e.g. syringe tips)                                    | Female Luer connector on<br>proximal end fits all male luer<br>slip connectors (e.g. syringe<br>tips)                 | Equivalent                                                                                                                                                                                                                                                                                                                      |
| Sterility                                              | Yes per ANSI/AAMI/ISO 11137-<br>2: Sterilization of health care<br>products -- Radiation                                         | Provided non-sterile and<br>sterilized on-site                                                                        | Equivalent                                                                                                                                                                                                                                                                                                                      |
| Electrical Safety                                      | N/A                                                                                                                              | N/A                                                                                                                   | Equivalent                                                                                                                                                                                                                                                                                                                      |
| Mechanical safety                                      | N/A                                                                                                                              | N/A                                                                                                                   | Equivalent                                                                                                                                                                                                                                                                                                                      |
| Chemical Safety                                        | Silica lumen non-reactive                                                                                                        | Stainless steel                                                                                                       | Equivalent                                                                                                                                                                                                                                                                                                                      |
| Thermal Safety                                         | MRI Safe. All brain contacting<br>components tested safe in a 1.5T<br>environment                                                | Not MRI safe                                                                                                          | VC use is not<br>restricted to non-MRI<br>environment                                                                                                                                                                                                                                                                           |
| Radiation safety                                       | N/A                                                                                                                              | N/A                                                                                                                   | Equivalent                                                                                                                                                                                                                                                                                                                      |

SurgiVision, Inc.

Section 5 510(k) Submission Page #

-7.6

{2}------------------------------------------------

K102101

## 510(k) (Traditional) Submission Section 5, 510(k) Summary

SurgiVision, Inc.

.

Section 5 510(k) Submission Page #

3 of 6

{3}------------------------------------------------

K102101

# 510(k) (Traditional) Submission

SurgiVision, Inc.

.

Section 5 510(k) Submission Page #

26

{4}------------------------------------------------

## 长10210 |

## 510(k) (Traditional) Submission Section 5, 510(k) Summary

Summarizing the differences noted above:

- The SurgiVision MR Compatible Ventricular Cannula (VC) is single patient use and is . not re-sterilizable.
- The VC is provided sterilized and pyrogen free in Tyvek packaging. .
- The VC is MR safe and can be used within an MR environment. ●
- The VC Is MN sale and can oe association to the physician regarding flow rates ● as a function of pressure. The predicate provides no such information.
- The VC has no side holes in the lumen .
- The VC has no depth marking on the barrel as the physician will be adjusting location . with direct visualization under MR using a stereotactic frame
- While and the rigid section of the VC is longer than the predicate for ease of . handling within the MR bore.

SurgiVision, Inc.

Section 5 510(k) Submission Page #

{5}------------------------------------------------

K102101

- . The VC does not require a stylet for entry to the ventricles that requires removal prior to injection or aspiration.
- The VC has an integral extension for injection or aspiration external to the bore. .
- t The VC and predicate flow rates differ because of the different lumen sizes and extension.

These recognized differences between the VC and the predicate are intended to meet the needs of the MR work environment and do not represent changes that effect the safety or effectiveness of the device.

## 7. Testing:

Testing to applicable standards has been completed with acceptable outcomes. The following testing has been performed:

- Sterilization and Shelf Life, including sterilization validation using the . VDmax25 procedure of ISO 11137-2.
- Biocompatibility Testing including cytotoxicity, material mediated . pyrogen, ISO maximization study, intracutaneous toxicity and systemic toxicity with acceptable results.
- Performance Testing Bench, including design verification testing, . comparison testing with the predicate Adson Cannula, pressure withstand testing and injection/aspiration testing, lateral tip deflection testing, transit testing in conformance to D4169 and accelerated aging testing.
- Cytarabine injection testing and aspiration of primate CSF was completed . with acceptable results..

These tests demonstrated that the MR Compatible Ventricular Cannula functions as intended and is substantially equivalent to legally marketed predicate device.

#### Rx or OTC: 8.

The MR Compatible Ventricular Cannula is an Rx prescription device per 21 CFR Part 801, Subpart D.

## 9. Substantial Equivalence:

The MR Compatible Ventricular Cannula has the same indications for use as the The Mice device, Adson Cannula. While there are technological differences between predical cornect, Notentricular Cannula and the Adson Cannula, such as the device matcrial, construction and dimensions, these differences do not raise new types of material, consultion and entions when all listed warnings and cautions are followed. The MR Compatible Ventricular Cannula is substantially equivalent to the marketed Adson Cannula.

SurgiVision, Inc.

Date: 1/22/21

Edward Waddell, Director of Regulatory Affairs

Surgi Vision, Inc.

Section 5 510(k) Submission Page #

006

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The eagle is black, and the text is also black.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

SurgiVision, Inc. c/o Mr. Edward Waddell Director, Regulatory Surgi Vision, Inc. 5 Musick Irvine, CA 92618

JAN 26 2011

Re: K102101

Trade/Device Name: SurgiVision MR Compatible Ventricular Cannula Regulation Number: 21 CFR 882.4060 Regulation Name: Ventricular cannula Regulatory Class: Class I Product Code: HCD Dated: January 17, 2011 Received: January 18, 2011

Dear Mr. Waddell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

:

:

.

 

. .

and the contraction of the comments of the comments of the comments of

 : 

. . . . . . . .

. . . . .

{7}------------------------------------------------

Page 2 - Mr. Edward Waddell

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

. . . . . . . .

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Indications for Use Statement

KI02101 510(k) Number (if known):

Device Name: SurgiVision, Inc. MR Compatible Ventricular Cannula Indications for Use:

The MR Compatible Ventricular Cannula is intended for injection of Cytarabine or removal of CSF from the ventricles during intracranial procedures. The device is not intended for implant. The device is intended for "single patient use only."

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

R. Peterson

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K102101

Page __ of _

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HCD/K102101](https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HCD/K102101)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
