The DORO® Sterile Disposable Skull Pin is utilized in a skull clamp and is applied to the patient's skull to hold their head and neck securely in a particular position when rigid fixation is desired. The skull clamp is indicated for use in open and percutaneous craniotomies as well as spinal surgery when rigid fixation is necessary.
Device Story
Disposable skull pins for use with skull clamps to provide rigid head and neck fixation during neurosurgical and spinal procedures. Available in stainless steel (MR unsafe) and titanium (MR conditional) versions for adult and pediatric patients. Pins are inserted into the rocker arm and opposite side of a skull clamp. Titanium version allows for intra-operative X-ray, CT, or MR imaging. Device is supplied sterile via gamma irradiation for single-patient use. Healthcare providers (surgeons) install pins into headrest systems to stabilize the patient's head, facilitating surgical access and precision.
Clinical Evidence
Bench testing only. Testing included axial load and creep tests, shear tests, transport validation, packing integrity, system handling, biocompatibility (cytotoxicity and chemical analysis), color coding verification, sterility testing (VDmax and dye penetration), and MR compatibility for titanium models. All tests passed.
Technological Characteristics
Materials: Stainless Steel 1.4034 or Titanium Grade 5 (Ti-6Al-4V) with PA6 polymer base. Sensing/Actuation: Mechanical fixation via skull clamp. Energy: None. Form factor: Adult/Pediatric pins. Connectivity: None. Sterilization: Gamma irradiation. Software: None.
Indications for Use
Indicated for use in combination with a skull clamp in open and percutaneous craniotomies as well as spinal surgery when rigid fixation is necessary. Not intended for use on infants or children under 5 years of age.
Regulatory Classification
Identification
A neurosurgical head holder (skull clamp) is a device used to clamp the patient's skull to hold head and neck in a particular position during surgical procedures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 2, 2020
Pro Med Instruments GmbH Muhammad Zubair Regulatory Affairs Manager Boetzinger Str. 38 Freiburg, 79111 De
### Re: K193438
Trade/Device Name: DORO Sterile Disposable Skull Pins (Stainless Steel), DORO Sterile Disposable Skull Pins (Titanium) Regulation Number: 21 CFR 882.4460 Regulation Name: Neurosurgical Head Holder (Skull Clamp) Regulatory Class: Class II Product Code: HBL Dated: April 29, 2020 Received: May 4, 2020
### Dear Muhammad Zubair:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce. Ph.D. Acting Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193438
Device Name
DORO Sterile Disposable Skull Pins (Stainless steel & Titanium)
Indications for Use (Describe)
Indicated for use in combination with a skull clamp in open and percutaneous craniotomies as well as spinal surgery when rigid fixation is necessary.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### DATE:
June 2, 2020
## DOCUMENT NUMBER:
K193438
APPLICANT:
Pro Med Instruments GmbH Bötzinger Straße 38 79111 Freiburg im Breisgau Germany Tel: + 49 (0) 761 384 222 10 Fax: +49 (0) 761 384 222 81 E-Mail: pmi@pmisurgical.com
## CONTACT PERSON:
Name: Muhammad Zubair Position: Regulatory Affairs Manager Tel.: +49 761 384 222 67 E-Mail: ra@pmisurgical.com
{4}------------------------------------------------
#### 1. Device Name
| Trade Name: | DORO® Sterile Disposable Skull Pins |
|-----------------------------|-------------------------------------------|
| Common Name: | Skull Pins |
| Device Classification Name: | Holder, head, neurosurgical (skull clamp) |
#### 2. Classification / Product Code
DORO® Sterile Disposable Skull Pins can be classified according to following device name and product code:
| Device | Regulation<br>Description | Regulation<br>Medical<br>Specialty | Review<br>Panel | Product<br>Code | Regulation<br>Number | Device<br>Classification |
|-------------------------------------------------|-----------------------------------------------|------------------------------------|-----------------|-----------------|----------------------|--------------------------|
| Holder, head,<br>neurosurgical<br>(skull clamp) | Neurosurgical<br>head holder<br>(skull clamp) | Neurology | Neurology | HBL | 882.4460 | 2 |
#### Predicate Device 3.
### Subject Device 1: DORO® Sterile Disposable Skull Pins (Stainless Steel)
| Subject Device | Predicate Device | 510(k) Number | 510(k) Holder |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|---------------|--------------------------------------|
| DORO® Sterile<br>Disposable Skull Pins<br>(Stainless Steel)<br><br>Model:<br><br>1. (3006-00<br>Stainless Steel,<br>Adult, Blue)<br>2. (3006-50<br>Stainless Steel,<br>Adult, Black) | TZ Skull Pin<br>Adult | K163322 | TZ<br>Medical, Inc. |
| 3. (3006-10<br>Stainless Steel,<br>Pediatric,<br>Yellow) | Mayfield Child<br>Disposable Skull Pin | K941558 | OHIO MEDICAL INSTRUMENT<br>CO., INC. |
{5}------------------------------------------------
| Device | Predicate Device | 510(k) Number | 510(k) Holder |
|-------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|------------------------|-------------------------------------------------------------------------|
| DORO® Sterile Disposable Skull Pins (Titanium)<br>Model:<br>1. (3006-20 Titanium, Adult, Orange)<br><br>2. (3006-30 Titanium, Pediatric, Green) | Limited Artifact Skull Pin<br><br>Mayfield Child Disposable Skull Pin | K073163<br><br>K941558 | Dinkier Surgical Devices, Inc.<br><br>OHIO MEDICAL INSTRUMENT CO., INC. |
### Subject Device 2: DORO® Sterile Disposable Skull Pins (Titanium)
#### 4. Device Description
DORO® Sterile Disposable Skull Pins are a disposable device. These are delivered sterile (gamma-sterilization), ready for use, and are intended to be disposed after one use.
DORO® Sterile Disposable Skull Pins are available in 2 different types (Stainless steel & Titanium) and 2 different sizes (Adult & Pediatric).
#### DORO® Sterile Disposable Skull Pins (Stainless Steel) 4.1.
DORO® Sterile Disposable Skull Pins are used together with the DORO ® Headrest Systems, intended as a neck and head support to stabilize the patient's head during neurosurgical procedures.
Two DORO® Sterile Disposable Skull Pins are inserted into the rocker arm side of the skull clamp and one single skull pin is inserted in the opposite side.
#### DORO® Sterile Disposable Skull Pins (Titanium) 4.2.
DORO® Sterile Disposable Skull Pins are used together with the DORO ® Headrest System, intended as a neck and head support to stabilize the patient's head during neurosurgical procedures.
Two DORO® Sterile Disposable Skull Pins are inserted into the rocker arm side of the skull clamp and one skull pin is inserted in the opposite side.
DORO® Sterile Disposable Skull Pins (Titanium) are used when Intra-Operative X-Ray-, CTor MR Imaging of the patient is used.
{6}------------------------------------------------
## 5. Indications for Use
The DORO® Sterile Disposable Skull Pin is indicated for use in combination with a skull clamp in open and percutaneous craniotomies as well as spinal surgery when rigid fixation is necessary.
#### Technological Characteristics 6.
The technological characteristics of DORO® Sterile Disposable Skull Pins are the same as the technological characteristics of the predicate device.
## 6.1. DORO® Sterile Disposable Skull Pins (Stainless Steel)
| Company | pro med instruments GmbH<br>(New Device) | TZ Medical, Inc.<br>(Predicate Device) | |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | DORO® Sterile Disposable Skull Pins<br>(Stainless steel) | TZ Skull Pin Adult | |
| Regulation Number | 882.4460 | 882.4460 | |
| Class | 2 | 2 | |
| Code | HBL | HBL | |
| 510(k) number | K193438 | K163322 | |
| Intended Use | The DORO® Sterile Disposable Skull Pin is<br>utilized in a skull clamp and is applied to the<br>patient's skull to hold their head and neck<br>securely in a particular position when rigid<br>fixation is desired. The skull clamp is<br>indicated for use in open and percutaneous<br>craniotomies as well as spinal surgery when<br>rigid fixation is necessary. | The TZ Medical skull pin is placed in a<br>clamp and is applied to the patient's skull<br>to hold their head and neck securely in a<br>particular position when rigid fixation is<br>desired. The clamp is indicated for use in<br>open and percutaneous craniotomies as<br>well as spinal surgery when rigid fixation<br>is necessary. | |
| Indication for Use | Indicated for use in combination with a skull<br>clamp in open and percutaneous craniotomies<br>as well as spinal surgery when rigid fixation is<br>necessary. | Indicated for use in open and<br>percutaneous craniotomies as well as<br>spinal surgery when rigid fixation is<br>necessary. | |
| Material | Stainless Steel 1.4034 with a polymer base<br>PA6 | 316L (1.4404 ) Stainless Steel | |
| Size | Adult | Adult | |
| Dimension | | | |
| Tip to Shoulder<br>Distance | Adult: 0.5 Inches | Adult: 0.5 Inches | |
| Width of back | Adult: 0.31 inches | Adult:0.31 inches | |
| O-ring | 3006-00<br>Stainless Steel,<br>Adult, Blue<br>3006-50<br>Stainless Steel,<br>Adult, Black | No O-Ring<br>Nitrile butadiene<br>rubber (NBR) O-<br>Ring<br>Silicone O-Ring | |
| | pro med instruments GmbH<br>(New Device) | | TZ Medical, Inc.<br>(Predicate Device) |
| Company | | | |
| Manufacturing | 3006-00<br>Stainless Steel,<br>Adult, Blue<br><br>3006-50<br>Stainless Steel,<br>Adult, Black | The pin insert is<br>machined according to<br>the dimensions<br>specified, and cleaned<br>afterwards. The Pin<br>collet is injection-<br>molded around the<br>Pin insert in controlled<br>environmental<br>conditions.<br><br>The pin is machined<br>according to the<br>dimensions specified,<br>and cleaned<br>afterwards. Assembly<br>of the O-ring takes<br>place in controlled<br>environmental<br>conditions. | The pin is machined to the dimensions<br>specified. O-Ring is placed on device when<br>manufacturing and cleaning is completed. |
| Preparation for<br>surgery | None, the pins are supplied sterile in a 3-pack<br>pouch ready for use | | None, the pins are supplied sterile in a 3-<br>pack pouch ready for use |
| Method of Use | Typically , three pins are installed in<br>receptacles of the head holder | | Typically, three pins are installed in<br>receptacles of the head holder |
| Clamp<br>Compatibility | DORO® Sterile Disposable Skull Pins are<br>compatible with all the models of Doro brand<br>skull clamps | | V. Mueller Spetzler Skull Clamp-Model #<br>M-1500<br>Integra Mayfield Skull Clamp- Model # A-<br>1059<br>Integra Mayfield Skull Clamp- Model # A-<br>2000 |
| MR Compatibility | MR Unsafe | | MR Unsafe |
| Method of<br>Sterilization | Gamma Irradiation | | Ethylene Oxide |
| Use | Disposable; Single-Patient Use only | | Disposable; Single-Patient Use only |
| Packaging | Medical grade packaging (sealed<br>TYVEK/Blister) | | Medical grade packaging (sealed<br>TYVEK (flexible film pouches) |
| Company | pro med instruments GmbH<br>(New Device) | OHIO MEDICAL INSTRUMENT CO.,<br>INC.<br>(Predicate Device) | |
| Device Name | DORO® Sterile Disposable Skull Pins<br>(Stainless steel) | Mayfield Child Disposable Skull Pin | |
| Regulation Number | 882.4460 | 882.4460 | |
| Class | 2 | 2 | |
| Code | HBL | HBL | |
| 510(k) number | K193438 | K941558 | |
| Intended Use | The DORO® Sterile Disposable Skull Pin is<br>utilized in a skull clamp and is applied to the<br>patient's skull to hold their head and neck<br>securely in a particular position when rigid<br>fixation is desired. The skull clamp is<br>indicated for use in open and percutaneous<br>craniotomies as well as spinal surgery when<br>rigid fixation is necessary. | Information not publically available in 510<br>(k) summary | |
| Indication for Use | Indicated for use in combination with a skull<br>clamp in open and percutaneous craniotomies<br>as well as spinal surgery when rigid fixation is<br>necessary. | Information not publically available in 510<br>(k) summary | |
| Recommended Age<br>of use | The disposable skull pin is not intended for<br>use on infants or children under 5 years of<br>age. | The disposable skull pin is not intended for<br>use on infants or children under 5 years of<br>age. | |
| Material | Stainless Steel 1.4034 with a polymer base<br>PA6 | Information not publically available in 510<br>(k) summary | |
| Size | Pediatric | Pediatric | |
| Dimension | | | |
| Tip to Shoulder<br>Distance | Pediatric: 0.4 Inches | Information not publically available in 510<br>(k) summary | |
| Width of back | Pediatric: 0.31 inches | Information not publically available in 510<br>(k) summary | |
| O-ring | 3006-10<br>Stainless Steel,<br>Pediatric, Yellow | Information not publically available in 510<br>(k) summary | |
### Adult:
{7}------------------------------------------------
{8}------------------------------------------------
## Pediatric:
{9}------------------------------------------------
| Company | pro med instruments GmbH<br>(New Device) | OHIO MEDICAL INSTRUMENT CO.,<br>INC.<br>(Predicate Device) | |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Manufacturing | 3006-10<br>Stainless Steel,<br>Pediatric, Yellow<br><br>The pin insert is<br>machined according to<br>the dimensions<br>specified, and cleaned<br>afterwards. The Pin<br>collet is injection-<br>molded around the<br>Pin insert in controlled<br>environmental<br>conditions. | Information not publically available in 510<br>(k) summary | |
| Preparation for<br>surgery | None, the pins are supplied sterile in a 3-pack<br>pouch ready for use | Information not publically available in 510<br>(k) summary | |
| Method of Use | Typically, three pins are installed in<br>receptacles of the head holder | Information not publically available in 510<br>(k) summary | |
| Clamp<br>Compatibility | DORO® Sterile Disposable Skull Pins are<br>compatible with all the models of Doro brand<br>skull clamps | Information not publically available in 510<br>(k) summary | |
| MR Compatibility | MR Unsafe | Information not publically available in 510<br>(k) summary | |
| Method of<br>Sterilization | Gamma Irradiation | Information not publically available in 510<br>(k) summary | |
| Use | Disposable; Single-Patient Use only | Information not publically available in 510<br>(k) summary | |
| Packaging | Medical grade packaging (sealed<br>TYVEK/Blister) | Information not publically available in 510<br>(k) summary | |
| Company | pro med instruments GmbH<br>(New Device) | Dinkier Surgical Devices, Inc.<br>(Predicate Device) | |
| Device Name | DORO® Sterile Disposable Skull Pins<br>(Titanium) | Limited Artifact Skull Pin | |
| Regulation Number | 882.4460 | 882.4460 | |
| Class | 2 | 2 | |
| Code | HBL | HBL | |
| 510(k) number | K193438 | K073163 | |
| Intended Use | The DORO® Sterile Disposable Skull Pin<br>is utilized in a skull clamp and is applied<br>to the patient's skull to hold their head<br>and neck securely in a particular<br>position when rigid fixation is desired.<br>The skull clamp is indicated for use in<br>open and percutaneous craniotomies as<br>well as spinal surgery when rigid fixation<br>is necessary. | Limited Artifact Skull Pins are used with a<br>head holder device that is placed on the<br>patients'skull to hold their head and neck in a<br>particular position during surgical procedures<br>when rigid skeletal fixation is desired. | |
| Indication for Use | Indicated for use in combination with<br>skull clamp in open and percutaneous<br>craniotomies as well as spinal surgery<br>when rigid fixation is necessary. | Indicated for use in open and percutaneous<br>craniotomies and spinal surgeries when rigid<br>skeletal fixation is necessary and during intra-<br>operative use within a MR environment rated<br>up to and including a 1.5 Tesla magnetic Field | |
| Material | Titanium Grade 5 (Ti-6Al-4V) with a<br>polymer base PA6 | The pin tip is Titanium 6Al-4V and the<br>polymer base is PEEK with 10% carbon filler | |
| Dimension | | | |
| Tip to Shoulder<br>Distance | Adult: 0.5 Inches | Information not publically available in 510 (k)<br>summary | |
| Width of back | Adult: 0.31 inches | Information not publically available in 510 (k)<br>summary | |
| O-ring | 3006-20<br>Titanium, Adult,<br>Orange)<br>No O-Ring | O-Ring | |
| Manufacturing | The pin insert is machined according to<br>the dimensions specified, and cleaned<br>afterwards. The Pin collet is injection-<br>molded around the Pin insert in<br>controlled environmental conditions. | The pin tip is machined to the dimensions<br>listed on the pin tip drawing.<br>The pin tip is injection molded to the<br>dimensions listed on the pin base drawing. | |
| Preparation for<br>surgery | None, the pins are supplied sterile in a<br>3-pack pouch ready for use. | None, the pin are supplied sterile in a 3-pack<br>pouch ready for use. | |
| Method of Use | Typically , three pins are installed in | Typically , three pins are installed in | |
| Company | pro med instruments GmbH<br>(New Device) | Dinkier Surgical Devices, Inc.<br>(Predicate Device) | |
| Clamp<br>Compatibility | DORO® Sterile Disposable Skull Pins are<br>compatible with all the models of Doro<br>brand skull clamps | Information not publically available in 510 (k)<br>summary | |
| MR Compatibility | MR Conditional<br>Main magnetic field strength: 1.5 Tesla or 3 Tesla | MR Conditional | |
| Method of<br>Sterilization | Gamma Irradiation | radiation sterilized | |
| Use | Disposable; Single-Patient Use only | | |
| Packaging | Medical grade packaging (sealed<br>TYVEK/Blister) | Medical grade packaging (sealed<br>TYVEK/flexible film pouches) | |
| Company | pro med instruments GmbH<br>(New Device) | OHIO MEDICAL INSTRUMENT CO., INC.<br>(Predicate Device) | |
| Device Name | DORO® Sterile Disposable Skull Pins<br>(Titanium) | Mayfield Child Disposable Skull Pin | |
| Regulation Number | 882.4460 | | |
| Class | 2 | 2 | |
| Code | HBL | HBL | |
| 510(k) number | K193438 | K941558 | |
| Intended Use | The DORO® Sterile Disposable Skull Pin<br>is utilized in a skull clamp and is applied<br>to the patient's skull to hold their head<br>and neck securely in a particular<br>position when rigid fixation is desired.<br>The skull clamp is indicated for use in<br>open and percutaneous craniotomies as<br>well as spinal surgery when rigid fixation<br>is necessary. | Information not publically available in 510 (k)<br>summary | |
| Indication for Use | Indicated for use in combination with<br>skull clamp in open and percutaneous<br>craniotomies as well as spinal surgery<br>when rigid fixation is necessary. | Information not publically available in 510 (k)<br>summary | |
| Recommended Age<br>of Use | The disposable skull pin is not intended<br>for use on infants or children under 5<br>years of age. | The disposable skull pin is not intended for<br>use on infants or children under 5 years of<br>age. | |
| Material | Titanium Grade 5 (Ti-6AI-4V) with a<br>polymer base PA6 | Information not publically available in 510 (k)<br>summary | |
| Size | Pediatric…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.