K231403 · Evonos GmbH & Co. KG · HBF · Mar 6, 2024 · Neurology
Device Facts
Record ID
K231403
Device Name
evoDrill Cranial Perforator
Applicant
Evonos GmbH & Co. KG
Product Code
HBF · Neurology
Decision Date
Mar 6, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4305
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The evoDrill Cranial Perforators are single-use surgical devices for cranium perforation. The perforators with the 3mm step automatically release and stop perforating at cranial bone thickness of at least 3mm. The perforators with the 1mm step automatically release and stop perforating at cranial bone thickness of at least 1mm.
Device Story
Single-use surgical cranial perforator; used for trephining cranial bone. Driven by pneumatic or electric surgical motor (800-1500 RPM). Features automatic spring-loaded release mechanism; inner drill coupled to housing via axial force; resistance from bone maintains coupling; upon piercing, resistance drops below spring force, causing automatic decoupling and stopping of perforation. Used in clinical/surgical settings by surgeons. Benefits patient by preventing over-penetration of skull bone.
Clinical Evidence
Bench testing only. Evaluated automatic stop mechanism, drilling time, and device geometry. Biocompatibility testing performed per ISO 10993. Sterilization validation performed per ISO 11137 and ISO 11607. All testing met acceptance criteria.
Technological Characteristics
Materials: Plastic, stainless steel, aluminum. Principle: Mechanical spring-loaded automatic decoupling. Energy: Pneumatic or electric motor (800-1500 RPM). Form factor: Single-use drill with Hudson fitting. Sterilization: Radiation (SAL 10^-6).
Indications for Use
Indicated for cranium perforation in patients requiring burr holes. Available in variants for cranial bone thickness of at least 1mm or at least 3mm.
Regulatory Classification
Identification
Powered compound cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments employ a clutch mechanism to disengage the tip of the instrument after penetrating the skull to prevent plunging of the tip into the brain.
{0}------------------------------------------------
March 6, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Evonos GmbH & Co. KG Jörg Mans CEO Stockacher Strasse 134 Tuttingen, Baden-Württemberg 78532 Germany
# Re: K231403
Trade/Device Name: evoDrill Cranial Perforator Regulation Number: 21 CFR 882.4305 Regulation Name: Powered Compound Cranial Drills, Burrs, Trephines, And Their Accessories Regulatory Class: Class II Product Code: HBF Dated: December 8, 2023 Received: December 8, 2023
# Dear Jörg Mans:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Pierce -S
Sincerely, Digitally signed by Adam D.
Adam D. Pierce -S Date: 2024.03.06 17:22:17 -05'00'
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231403
Device Name evoDrill Cranial Perforator
#### Indications for Use (Describe)
The evoDrill Cranial Perforators are single-use surgical devices for cranium perforation. The perforators with the 3mm step automatically release and stop perforating at cranial bone thickness of at least 3mm. The perforators with the 1mm step automatically release and stop perforating at cranial bone thickness of at least 1mm.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
Date Prepared: March 6, 2024
## Owner's Name (21 CFR 807.92(a)(1):
| Company Name: | evonos GmbH & CO. KG |
|-----------------------------|------------------------|
| Company Address: | Stockacher Strasse 134 |
| D-78532 Tuttlingen, Germany | |
| Contact Person: | Jörg Mans |
|--------------------|---------------------|
| Company Telephone: | + 49 7461-965774-50 |
| Company Fax: | + 49 7461-965774-89 |
| Email-Address: | j.mans@evonos.de |
#### Trade Name, Common Name, Classification (21 CFR 807.92(a)(2))
| Device Trade Name: | evonos evoDrill Cranial Perforator |
|-----------------------------|------------------------------------------------------------|
| Device Common Name: | Disposable Cranial Perforator |
| Device Classification Name: | Drills, Burrs, Trephines & Accessories (Compound, Powered) |
| Device Classification: | Class II |
| Classification Code: | HBF |
| Regulation Number: | 21 CFR 882.4305 |
## Predicate Device Names (21 CFR 807.92(a)(3))
- Acra Cut Automatic Cranial Drill (DGR-1 #200-241; DGR-II #200-251) (K892866) (Primary) -
- -Codman Disposable Perforators 26-1221 (K933894)
{5}------------------------------------------------
#### DEVICE DESCRIPTION
The evonos evoDrill cranial perforator are single-use for trephining the cranial bone. The perforators have an automatic stop mechanism that stops the trephination process as soon as the skull bone has been pierced. The two variants evoDrill cranial perforator 3-flutes and evoDrill cranial perforator 4-flutes are available in two versions that are suitable for different bone thicknesses.
The first version is used from bone thicknesses of 3mm, the second version is used for trepanation with bone thicknesses between 1mm and 3mm. Both variants are available in four different diameters.
| Item Number | Item Description | Inner<br>Diameter | Outer<br>Diameter | Bone<br>Thickness |
|-------------|-------------------------------------------------|-------------------|-------------------|-------------------|
| 26-000002 | evoDrill Cranial Perforator 9/6/1mm, 3-flutes | 6mm | 9mm | 1-3mm |
| 26-000003 | evoDrill Cranial Perforator 9/6/3mm, 3-flutes | 6mm | 9mm | > 3mm |
| 26-000004 | evoDrill Cranial Perforator 11/7/1mm, 3-flutes | 7mm | 11mm | 1-3mm |
| 26-000005 | evoDrill Cranial Perforator 11/7/3mm, 3-flutes | 7mm | 11mm | > 3mm |
| 26-000006 | evoDrill Cranial Perforator 13/9/1mm, 3-flutes | 9mm | 13mm | 1-3mm |
| 26-000007 | evoDrill Cranial Perforator 13/9/3mm, 3-flutes | 9mm | 13mm | > 3mm |
| 26-000008 | evoDrill Cranial Perforator 14/11/1mm, 3-flutes | 11mm | 14mm | 1-3mm |
| 26-000009 | evoDrill single use, Ø14/11mm, 3mm, 3-flutes | 11mm | 14mm | > 3mm |
| Item Number | Item Description | Inner<br>Diameter | Outer<br>Diameter | Bone<br>Thickness |
|-------------|-------------------------------------------------|-------------------|-------------------|-------------------|
| 26-409061 | evoDrill Cranial Perforator 9/6/1mm, 4-flutes | 6mm | 9mm | 1-3mm |
| 26-409063 | evoDrill Cranial Perforator 9/6/3mm, 4-flutes | 6mm | 9mm | > 3mm |
| 26-411071 | evoDrill Cranial Perforator 11/7/1mm, 4-flutes | 7mm | 11mm | 1-3mm |
| 26-411073 | evoDrill Cranial Perforator 11/7/3mm, 4-flutes | 7mm | 11mm | > 3mm |
| 26-413091 | evoDrill Cranial Perforator 13/9/1mm, 4-flutes | 9mm | 13mm | 1-3mm |
| 26-413093 | evoDrill Cranial Perforator 13/9/3mm, 4-flutes | 9mm | 13mm | > 3mm |
| 26-414111 | evoDrill Cranial Perforator 14/11/1mm, 4-flutes | 11mm | 14mm | 1-3mm |
| 26-414113 | evoDrill Cranial Perforator 14/11/3mm, 4-flutes | 11mm | 14mm | > 3mm |
Table 1: Item Description
{6}------------------------------------------------
The evonos evoDrill cranial perforator per CFR, Part 882.4305, is a bone cutting and drilling instrument driven by a pneumatic or electric surgical motor in order to drill burr holes through the skull of a patient. The evonos evoDrill cranial perforator are devices similar in design and construction to other cranial perforators currently on the market; (e.g.: Acra Cut Automatic Cranial Drill DGR-1 #200-241; DGR-II #200-251 and Codman Disposable Perforators 26-1222, 26-1223). The evonos evoDrill cranial perforator requires a motor and an attached or integrated speed reducer in order to run the perforators at a speed range of 800 RPM to 1500 RPM.
Image /page/6/Picture/2 description: The image shows a medical drill being used on a bone. The drill is labeled "evoDrill Made in Germany 14-11-1". The drill is silver with a blue handle. The bone is tan and brown.
Picture 1: Schematic illustration of the bone perforation
### INDICATIONS FOR USE
The evoDrill Cranial Perforators are single-use surgical devices for cranium perforation. The perforators with the 3mm step automatically release and stop perforating at cranial bone thickness of at least 3mm. The perforators with the 1mm step automatically release and stop perforating at cranial bone thickness of at least 1mm.
| Device | evoDrill | Codman Cranial Perforator | AcraCut Cranial Perforator |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | evonos GmbH & Co. KG | Codman | AcraCut |
| 510(k) Number | K231403 | K933894 | K892866 |
| | (subject device) | (Predicate Device) | (Predicate Device) |
| Indications For<br>Use | The evoDrill Cranial Perforators<br>are single-use surgical devices<br>for cranium perforation. The<br>perforators with the 3mm step<br>automatically release and stop<br>perforating at cranial bone<br>thickness of at least 3mm. The<br>perforators with the 1mm step<br>automatically release and stop<br>perforating at cranial bone<br>thickness of at least 1mm. | The Codman disposable<br>Perforator is for use in<br>perforating the cranium. When<br>properly used, it is designed to<br>automatically disengage once<br>perforation is accomplished and<br>when pressure is removed from<br>the drill point. | The Acra Cut DGR-I, DGR-II and<br>DGR-0 are designed to<br>automatically release and stop<br>upon penetration of bone that is<br>at least 3mm thick. The DGR-II<br>perforator is for use at thin skull<br>areas such as pediatric,<br>temporal and suboccipital areas.<br>Thea are designed to<br>automatically release and stop<br>upon penetration of bone as<br>thin as 1mm. |
| Material | Plastic/ stainless steel /<br>Aluminium | Plastic/ stainless steel | Plastic/ stainless steel |
| Fittings | Hudson | Hudson | Hudson |
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
{7}------------------------------------------------
| Device | evoDrill | Codman Cranial Perforator | AcraCut Cranial Perforator |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | evonos GmbH & Co. KG | Codman | AcraCut |
| 510(k) Number | K231403 | K933894 | K892866 |
| | (subject device) | (Predicate Device) | (Predicate Device) |
| Prinicpal of<br>operation | similar | similar | similar |
| Cutting<br>Performance | similar | similar | similar |
| Burr hole<br>diameter | 14/11mm; 13/9mm; 11/7mm;<br>9/6mm | 14/11 mm; 11/8mm; 9/6mm | 14/11mm; 13/9mm; 11/7mm |
| Bone Thicknesses | 1mm and thicker | 3mm and thicker | 1mm and thicker |
| Release<br>mechanism | Automatic spring-loaded release<br>mechanism; The inner drill is<br>coupled with the housing once<br>an axial force is applied on the<br>fitting and the tip of the drill has<br>to overcome resistance (bone),<br>so that the resistance is higher<br>than the spring force of the<br>couple mechanism. Once the<br>inner drill has pierced the skull<br>bone, the resistance drops so<br>that the spring pressure is<br>higher than the resistance and<br>the inner drill is decoupled | Similar<br>Automatic spring-loaded release<br>mechanism; The inner drill is<br>coupled with the housing once<br>an axial force is applied on the<br>fitting and the tip of the drill has<br>to overcome resistance (bone),<br>so that the resistance is higher<br>than the spring force of the<br>couple mechanism. Once the<br>inner drill has pierced the skull<br>bone, the resistance drops so<br>that the spring pressure is<br>higher than the resistance and<br>the inner drill is decoupled | Similar<br>Automatic spring-loaded release<br>mechanism; The inner drill is<br>coupled with the housing once<br>an axial force is applied on the<br>fitting and the tip of the drill has<br>to overcome resistance (bone),<br>so that the resistance is higher<br>than the spring force of the<br>couple mechanism. Once the<br>inner drill has pierced the skull<br>bone, the resistance drops so<br>that the spring pressure is<br>higher than the resistance and<br>the inner drill is decoupled |
| Rotation speed | 800 – 1500 rpm | 800 – 1200 rpm | 800 – 1200 rpm |
| How supplied | Sterile<br>Sterilized by Radiation | Sterile<br>Sterilized by Radiation | Sterile<br>Sterilized by Radiation |
| Number of<br>Cutting Flutes | 3 / 4 | 4 | 4 |
Table 2: Device Comparison
Based on the information summarized above, the proposed evoDrill Cranial Perforators are similar to the currently marketed predicate and reference devices in indications for use, design, principle of operation and device material. All perforators are supplied in sterilized using radiation and for single use only.
### PERFORMANCE TESTING
The following performance data were provided in support of the substantial equivalence determination:
| Test | Method | Results |
|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | The device is classified as an externally<br>communicating devices in contact with tissue/bone<br>for a limited duration (≤24hr). Biocompatibility<br>testing was addressed in accordance with ISO 10993<br>as described in FDA's Biocompatibility Guidance. | All biocompatibility testing confirmed that<br>the device fulfills the biocompatibility<br>requirements considering contact area and<br>duration. |
| Sterilization and<br>packaging validation<br>testing | Sterilization validation testing in accordance with ISO<br>11137-1, ISO 11137-2, ISO 11607-1 and ISO 11607-2. | All tested specimen confirm that devices<br>are sterile and the SAL of 10-6 is met.<br>Shelf-Life validation confirmed the integrity<br>of the product and packaging of the defined<br>shelf-life, labeled on the device. |
{8}------------------------------------------------
| Test | Method | Results |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cutting Performance | The performance of the subject device was evaluated for automatic stop function mechanism, maximum number of uses, drilling time and device geometry for establishing substantial equivalency to the predicate. | All tested specimen confirmed that the overall cutting performance related to the autostop function mechanism, drilling time and surface structure fulfilled the acceptance criteria |
Table 3: Performance Testing
## CONCLUSION
Based on the similarities of the intended use/indications for use, technological and functional characteristics, and the results of the non-clinical performance testing, the evonos evoDrill Cranial Perforators, are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.