NeuroLine Disposable Cranial Perforator with Hudson end
K213228 · Emd Endoszkop Muszer Gyarto ES Kereskedelmi Kft. · HBF · Jun 20, 2022 · Neurology
Device Facts
Record ID
K213228
Device Name
NeuroLine Disposable Cranial Perforator with Hudson end
Applicant
Emd Endoszkop Muszer Gyarto ES Kereskedelmi Kft.
Product Code
HBF · Neurology
Decision Date
Jun 20, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4305
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
NeuroLine perforators are designed to make holes in the skull bone. The perforators have an integrated safety mechanism that releases automatically after passing through the bone and stops drilling.
Device Story
NeuroLine Disposable Cranial Perforator is a single-use, sterile cutting tool designed for cranial bone trephination. It features a Hudson-type connection for compatibility with standard surgical driver units. The device utilizes a mechanical clutch mechanism that automatically disengages when resistance to bone ceases, preventing over-penetration after passing through the skull. It is intended for use by surgeons in clinical settings. The device is available in various diameters (6/9 mm, 7/11 mm, 9/13 mm, 11/14 mm) and is selected based on skull thickness (minimum 1.5 mm or 3.0 mm). By automating the stop function, the device reduces the risk of accidental dural injury, providing a safer surgical outcome for the patient.
Clinical Evidence
Bench testing only. Performance evaluated for automatic stop function, drilling time, and device geometry. Biocompatibility testing conducted per ISO 10993. Sterilization validation performed per ISO 11137 and ISO 11607. All tests passed.
Technological Characteristics
Material: Stainless steel. Principle: Mechanical clutch-based automatic stop. Form factor: Disposable cranial drill with Hudson connection. Sizes: 6/9 mm, 7/11 mm, 9/13 mm, 11/14 mm. Sterilization: Gamma radiation.
Indications for Use
Indicated for creating holes in the skull bone for patients requiring cranial trephination.
Regulatory Classification
Identification
Powered compound cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments employ a clutch mechanism to disengage the tip of the instrument after penetrating the skull to prevent plunging of the tip into the brain.
Predicate Devices
ACRA-CUT Standard and Disposable Cranial perforator (K833266)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 20, 2022
EMD Endoszkop Muszer Gyarto es Kereskedelmi Kft. Gergo Ujvari Export Manager Bartok Bela u. 113/B Debrecen, Hajdu-Bihar 4031, Hungary
### Re: K213228
Trade/Device Name: NeuroLine Disposable Cranial Perforator with Hudson end Regulation Number: 21 CFR 882.4305 Regulation Name: Powered Compound Cranial Drills, Burrs, Trephines, And Their Accessories Regulatory Class: Class II Product Code: HBF Dated: May 13, 2022 Received: May 16, 2022
### Dear Gergo Ujvari:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213228
Device Name
NeuroLine Disposable Cranial Perforator with Hudson End
Indications for Use (Describe)
NeuroLine perforators are designed to make holes in the skull bone. The perforators have an integrated safety mechanism that releases automatically after passing through the bone and stops drilling.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## Date Prepared: June 14, 2022
### Applicant:
EMD Kft.
Bartok Bela u. 113/B Debrecen, Hungary 4031
Tel .: +36-52-486-034
Fax.: +36-52-486-035
### Contact Person:
Gergo Ujvari – International Sales Manager
E-mail: gergo.ujvari@emd.hu
Trade Name: NeuroLine Disposable Cranial Perforator with Hudson End
Common Name: Disposable Cranial Perforator
Classification name: 21CFR section 882.4305
Product Code: HBF
Device Classification: Class II
Predicate Device:
ACRA-CUT Standard and Disposable Cranial perforator
#### Reference Device:
EasyDrill Autostop Cranial Perforator
# Device Description:
The NeuroLine Disposable Cranial perforators (NLO-x/y-z) are single-use cutting devices that are packaged sterile and are used to perforate the skull bone. The perforator functions automatically, which means that it employs a clutch mechanism to automatically disenqage once perforation is accomplished and as the drill ceases to find resistance to bone.
The perforator has Hudson type connection and can connect to all the driver units with Hudson connection.
Perforators are offered in three different sizes: 6/9 mm, 7/13 mm and 11/14 mm. Sizes indicate hole diameters to be made by perforators in millimeters.
Perforators are also differentiated based on the thickness of skull bone. The NLO-x/y-3.0 perforators are used on skulls at least 3 mm thick, whereas NLO-x/y-1.5 ones are used on skulls at least 1.5 mm thick.
K833266
K141455
{4}------------------------------------------------
Image /page/4/Figure/1 description: This image shows a table with specifications for different models of tools. The table lists the models NLO-6/9-1.5, NLO-7/11-1.5, NLO-9/13-1.5, NLO-11/14-1.5, NLO-6/9-3.0, NLO-7/11-3.0, NLO-9/13-3.0, and NLO-11/14-3.0. For each model, the table provides the dimensions A, B, and C in millimeters, as well as the application for which the tool is intended. The tools with 1.5 in the model name are for use on skulls at least 1.5mm thick, while the tools with 3.0 in the model name are for use on skulls at least 3mm thick.
# Indications for Use:
NeuroLine perforators are designed to make holes in the skull bone. The perforators have an integrated safety mechanism that releases automatically after passing through the bone and stop drilling.
### Comparison of technology characteristics
| Device | NeuroLine Disposable<br>Cranial Perforator with<br>Hudson End | ACRA-CUT Standard and<br>Disposable Cranial<br>Perforator | EasyDrill Autostop Cranial<br>Perforator |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | EMD Endoszkop Muszer Gyarto<br>es Kereskedelmi Kft. | Acra-Cut Inc. | Micromar Industria e Comercio<br>Ltda. |
| 510(k) number | K213228<br>(Subject Device) | K833266<br>(Predicate Device) | K141455<br>(Reference Device) |
| Specified hole<br>sizes | 6/9 mm, 7/11 mm, 9/13mm,<br>11/14 mm | Similar | 6/9 mm, 7/11 mm, 11/14 mm |
| Indication for<br>use | NeuroLine perforators are<br>designed to make holes in<br>the skull bone. The<br>perforators have an<br>integrated safety mechanism<br>that releases automatically<br>after passing through the<br>bone and stops drilling. | Similar | The EasyDrill Cranial Perforator is a<br>sterile, single use cutting device<br>intended for performing cranial<br>bone trephination. |
| Operating<br>principles | Same | Same | Same |
| Device Material | Stainless steel | Similar | Stainless steel |
{5}------------------------------------------------
| Sterilization | Sterile<br>(Gamma Radiation) | Sterile | Sterile<br>(Gamma Radiation) |
|-------------------------|------------------------------|---------|------------------------------|
| Reusable<br>/Single Use | Single Use | Similar | Single Use |
Based on the information summarized above, the proposed NeuroLine Disposable Cranial Perforator and with Hudson end are similar to the currently marketed predicate and reference devices in indications for use, design, principle of operation and device material. All perforators are offered sterile.
# Performance Testing
The following performance data were provided in support of the substantial equivalence determination:
| Test | Method | Results |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Biocompatibility | The device is classified as an externally communicating<br>devices in contact with tissue/bone for a limited duration<br>(≤24hr). Biocompatibility testing was addressed in<br>accordance with ISO 10993 as described in FDA's<br>Biocompatibility Guidance. | Pass |
| Sterilization and packaging<br>validation testing | Sterilization validation testing in accordance with ISO<br>11137-1, ISO 11137-2, ISO 11607-1 and ISO 11607-2. | Pass |
| Bench Testing | The performance of the subject device was evaluated for<br>automatic stop function mechanism, maximum number of<br>uses, drilling time and device geometry for establishing<br>substantial equivalency to the predicate. | Pass |
#### Conclusions
Based on the similarities of the intended use/indications for use, technological and functional characteristics, and the results of the non-clinical performance testing, the NeuroLine Disposable Cranial Perforator with Hudson end, are substantially equivalent to the predicate device, ACRA-CUT Standard and Disposable Cranial perforator.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.