The Codman Disposable Perforator is for use in perforating the cranium. When properly used, it is designed to automatically disengage once perforation is accomplished and when pressure is removed from the drill point.
Device Story
Single-use, sterile, disposable cranial perforator; available in 9mm, 11mm, and 14mm sizes. Device features Hudson end for attachment to surgical drill driver. Principle of operation: mechanical engagement via pressure applied to drill point; automatic disengagement upon completion of perforation or removal of pressure. Used in neurosurgical procedures; operated by surgeons. Benefits include controlled cranial perforation and automatic disengagement to prevent unintended penetration. Device is identical to predicates except for alternate gamma irradiation sterilization cycle.
Clinical Evidence
No clinical data. Bench testing only. Validation of alternate gamma sterilization cycle performed via dose mapping study (Dose Uniformity Ratio < 2) in compliance with ISO 11137-1.
Technological Characteristics
Materials: ABS plastic (sleeve), metal components (drill body, spring, drive pin, bearing, nyliner). Energy source: mechanical (powered by external surgical drill driver). Form factor: color-coded disposable drill bits (9mm, 11mm, 14mm). Sterilization: Gamma irradiation (25kGy-50kGy, SAL 10^-6).
Indications for Use
Indicated for use in perforating the cranium.
Regulatory Classification
Identification
Powered compound cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments employ a clutch mechanism to disengage the tip of the instrument after penetrating the skull to prevent plunging of the tip into the brain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
January 20, 2019
Integra LifeSciences Corp. Jocelyn Raposo Associate Director, Regulatory Affairs 11 Cabot Blvd Mansfield, Massachusetts 02048
Re: K183581
Trade/Device Name: Codman Disposable Perforators Regulation Number: 21 CFR 882.4305 Regulation Name: Powered Compound Cranial Drills, Burrs, Trephines, And Their Accessories Regulatory Class: Class II Product Code: HBF Dated: December 20, 2018 Received: December 21, 2018
Dear Jocelyn Raposo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Matthew C. Krueger -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Indications for Use | |
| 510(k) Number (if known) | |
| K183581 | |
| Device Name | |
Codman Disposable Perforator
Indications for Use (Describe)
The Codman Disposable Perforator is for use in perforating the cranium. When properly used, it is designed to automatically disengage once perforation is accomplished and when pressure is removed from the drill point.
Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Sevices (301) 443-6740 EF
{3}------------------------------------------------
## Section 5 - 510(k) Summary
| I. Submitter | Integra LifeSciences Corp.<br>11 Cabot Blvd.<br>Mansfield, MA 02048 | |
|--------------|---------------------------------------------------------------------|----------------------------------|
| | Establishment Registration Number: 1226348 | |
| | Primary Contact:<br>Phone: | Jocelyn Raposo<br>(781) 971-5688 |
| | Secondary Contact:<br>Phone: | Jennifer Siu<br>(781) 971-5692 |
Date of Preparation: December 19, 2018
#### II. Device(s)
| Device Proprietary Name | Codman Disposable Perforators |
|---------------------------|---------------------------------------------------------------------------------|
| Common Name | Perforators |
| Classification Name | Drills, Burrs, Trephines & Accessories<br>(Compound, Powered) (21 CFR 882.4305) |
| Regulatory Classification | II |
| Product Code | HBF |
III. Predicate
The predicate devices for this submission are:
Device(s)
The predicate devices for this submission are:
| 510(k) Number | Date Cleared | Title |
|---------------|---------------|-----------------------------------------------|
| K791101 | Aug. 28, 1979 | Codman Disposable Perforator<br>(14mm) |
| K071931 | Oct. 22, 2007 | Codman Disposable Perforators<br>(11mm & 9mm) |
IV. Device Description(s) The Codman Disposable Perforators are pre-assembled, single-use, sterile devices that are designed to perforate the cranium. When properly used, the Disposable Perforators are designed to automatically disengage once perforation is accomplished and when pressure is removed from the drill point. They are designed with a Hudson end and are available in three color-coded sizes:
- 14mm (blue ABS sleeve) – product number: 26-1221
- 11mm (green ABS sleeve) product number: 26-1222 ●
- 9mm (yellow ABS sleeve) product number: 26-1223 ●
{4}------------------------------------------------
### V. Indications for Use
The Indications for Use statement of the proposed device remains identical to the predicate devices'.
| Equivalence<br>Comparison | Codman Disposable<br>Perforators (Predicates) | Codman Disposable<br>Perforators (Subject of This<br>Submission) |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Codman Disposable<br>Perforator is for use in<br>perforating the cranium.<br>When properly used, it is<br>designed to automatically<br>disengage once perforation is<br>accomplished and when<br>pressure is removed from the<br>drill point. | The Codman Disposable<br>Perforator is for use in<br>perforating the cranium.<br>When properly used, it is<br>designed to automatically<br>disengage once perforation is<br>accomplished and when<br>pressure is removed from the<br>drill point. |
#### VI. Comparison to Predicate Device
The proposed Codman Disposable Perforators are identical to the currently marketed Codman Disposable Perforators (K791101 & K071931) except for the alternate gamma irradiation sterilization cycle proposed in this 510(k). The indications for use, design, principle of operation, performance, packaging, and shelf life remain identical to the predicate devices.
| Substantial Equivalence Comparison | | |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| Characteristic | Codman Disposable<br>Perforators (Predicates) | Codman Disposable<br>Perforators (Subject of<br>This Submission) |
| Manufacturer | Codman & Shurtleff, Inc. | Identical |
| Classification Panel | Neurology | Identical |
| Classification Name | Drills, Burrs, Trephines &<br>Accessories (Compound,<br>Powered) (21 CFR<br>882.4305) | Identical |
| Indications for Use | The Codman Disposable<br>Perforator is for use in<br>perforating the cranium.<br>When properly used, it is<br>designed to automatically<br>disengage once<br>perforation is<br>accomplished and when<br>pressure is removed from<br>the drill point. | Identical |
| Operating Principle | With adequate pressure<br>applied on the perforator<br>point, engagement should<br>occur as the Hudson End<br>is rotated in a driver. Once<br>pressure is removed from<br>the drill point the device is<br>designed to disengage. | Identical |
| Implant | No | Identical |
| Single Use Only | Yes | Identical |
{5}------------------------------------------------
VII.
Data
Performance
| Special 510(k) Premarket Notification |
|---------------------------------------|
| Codman Disposable Perforators |
| Substantial Equivalence Comparison | | |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| Characteristic | Codman Disposable<br>Perforators (Predicates) | Codman Disposable<br>Perforators (Subject of<br>This Submission) |
| Design | 9 mm Perforator<br>consisting of outer and<br>inner body drill, drill<br>driver, compression<br>spring, drive pin, molded<br>sleeve, snap bearing,<br>flanged nyliner<br>11mm Perforator<br>consisting of outer and<br>inner body drill, drill<br>driver, compression<br>spring, drive pin, molded<br>sleeve, snap bearing,<br>flanged nyliner<br>14mm Perforator<br>consisting of outer and<br>inner body drill, drill<br>driver, compression<br>spring, drive pin, molded<br>sleeve, snap bearing,<br>flanged nyliner | Identical |
| Shelf Life | 5 years | Identical |
| Sterilization Method | Gamma | Identical |
| Sterility Assurance Level<br>(SAL) | 10-6 | Identical |
| Packaging | PETG Blister Tray with<br>heat-sealed lid placed into<br>a unit box | Identical |
There were no changes made that affect the Codman Disposable Perforators' indications for use, principle of operation, performance, packaging, or shelf life. The only difference between the predicate and proposed devices is the use of an alternate gamma irradiation sterilization cycle to sterilize the products. There is no change to the sterilization dose parameters (25kGy-50kGy). The products continue to meet a SAL of 10-6 with the new cycle. A full dose mapping study has been performed to ensure there are no new issues relating to the safety or effectiveness of the products.
Please see the Summary of Testing table below for acceptance criteria and testing results.
| Table Summary of Testing | | | |
|---------------------------------|------------------------------------------------------------------------------------------------|---------------------------|--------|
| Test | Test Method /<br>Purpose | Acceptance Criteria | Result |
| Full Dose Mapping<br>Validation | Establish a new<br>gamma radiation<br>sterilization cycle in<br>compliance with<br>ISO 11137-1 | Dose Uniformity Ratio: <2 | Pass |
{6}------------------------------------------------
Integra LifeSciences Codman Specialty Surgical Special 510(k) Premarket Notification Codman Disposable Perforators
#### Animal Testing:
No animal studies were required based on similarities of the proposed device to the predicate device and from the results of the sterilization validation.
### Clinical Testing:
No clinical studies were required based on similarities of the proposed device to the predicate device and from the results of the sterilization validation.
#### Conclusion:
Results from the dose mapping study have demonstrated that the alternate sterilization cycle is suitable and does not raise new issues of safety and effectiveness for the finished devices. Based on the indications for use, fundamental scientific technology and a comparison to the predicate devices, the subject Codman Disposable Perforators are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.