Surgify Halo is a sterile surgical burr indicated for shaping and removal of hard tissue and bone in Neurosurgical, Spinal, ENT, and general surgical procedures
Device Story
Surgify Halo is a sterile, single-use rotary surgical burr designed for use with high-speed surgical drill systems in hospital settings. The device features a shank for rotary motion transfer, locking features for drill attachment, and a burr head with two cutting edges. A mechanical safety system, consisting of a self-centering ring and spring, operates in two modes: a hard tissue mode where the ring allows cutting edge engagement, and a soft tissue mode where the ring shields the cutting edges to minimize accidental soft tissue contact. Operated by surgeons, the device provides precise bone and hard tissue removal while reducing chattering. It benefits patients by enhancing surgical precision and safety during bone shaping procedures.
Clinical Evidence
Bench testing only. Performance testing included functional verification of durability, chattering, and compatibility with high-speed surgical drill systems. Sterilization validation was performed per ISO 11137. In-process cleaning and passivation validations were also conducted. All tests met design input requirements.
Technological Characteristics
Single-use rotary cutting burr; metallic construction; self-centering ring and spring safety mechanism; compatible with standard high-speed surgical drill handpieces; gamma sterilization. No software or electronic components.
Indications for Use
Indicated for shaping and removal of hard tissue and bone in neurosurgical, spinal, ENT, and general surgical procedures.
Regulatory Classification
Identification
Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.
{0}------------------------------------------------
March 13, 2025
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Surgify Medical Oy % Richard Lilly Consultant to Surgify Medical Oy Avania, LLC 3031 Tisch Way, Suite 1010 San Jose, California 95128
Re: K250380
Trade/Device Name: Surgify Halo (54.085.SHD.H1); Surgify Halo (54.140.SHD.H1); Surgify Halo (54.070.NVG.H1); Surgify Halo (54.125.NVG.H1); Surgify Halo (54.000.SEE.H1) Regulation Number: 21 CFR 882.4310 Regulation Name: Powered simple cranial drills, burrs, trephines, and their accessories Regulatory Class: Class II Product Code: HBE Dated: February 7, 2025 Received: February 11, 2025
Dear Richard Lilly:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Adam D. Adam D. Pierce -S Date: 2025.03.13 Pierce -S 13:36:23 -04'00'
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K250380
Device Name
Surgify Halo (54.085.SHD.H1); Surgify Halo (54.140.SHD.H1); Surgify Halo (54.070.NVG.H1); Surgify Halo (54.125.NVG.H1); Surgify Halo (54.000.SEE.H1)
Indications for Use (Describe)
Surgify Halo is a sterile surgical burr indciated for shaping and removal of hard tissue and bone in Neurosurgical, Spinal, ENT, and general surgical procedures
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR 807.87 and 807.92.
#### I. Submitter
| Name: | Surgify Medical Oy |
|---------------------------------|-------------------------------------------------------------------------|
| Address: | Otakaari 5 |
| | 02150 Espoo, Finland |
| Phone: | +358 40 707 3200 |
| Email: | jukka.kreander@surgifymedical.com |
| Contact Person: | Jukka Kreander, CQRO |
| Date Prepared: | March 13, 2025 |
| Device Trade<br>Name: | Surgify Halo |
| Device Common<br>or Usual Name: | Burr |
| Regulation Name: | Powered simple cranial drills, burrs, trephines, and their accessories. |
| Regulation<br>Number: | 21CFR 882.4310 |
| Product Code: | HBE |
| Device Class: | Class II |
| Classification<br>Panel: | Neurology |
#### II. Predicate Device
Predicate Name and 510(k) Number: Surgify Halo, K232684.
#### III. Device Description
The Surgify Halo is a sterile-packaged, single-use rotary cutting device (a burr) made from high-grade metallic materials, designed to cut bone and other hard tissues selectively while minimizing chattering. The device is similar to conventional surgical cutting burrs but includes a ring mechanism in the burr head to reduce chattering. Like traditional burrs, the device is designed to be compatible with commonly marketed, high-speed, surgical drill systems.
{5}------------------------------------------------
The Surgify Halo comprises:
- Shank: Transfers rotary motion from the drill motor to the burr head.
- 트 Locking Features: Located at the distal end of the shank, these features securely attach the burr to the high-speed surgical drill.
- 트 Burr Head: Shaped with two cutting edges for cutting of bone and hard tissue.
- 트 Safety Mechanism: A self-centering ring and spring mechanism installed in the burr head. The moving ring, larger than the burr head, can adopt two positions:
- o First Position: (Hard Tissue Mode) The ring protrudes, preventing the cutting edges from engaging with the work material.
- Second Position: (Soft Tissue Mode) The ring extends beyond the cutting edge, O shielding it to minimize accidental soft tissues and an exposed position for cutting hard surfaces like bone.
### IV. Indications for Use
Surgify Halo is a sterile surgical burr indicated for shaping and removal of hard tissue and bone in Neurosurgical, Spinal, ENT, and general surgical procedures.
### V. Technological Characteristics
The technological characteristics of the subject Surgify Halo remain the same as the previously cleared device under K232684.
The similarities and differences do not alter the intended use of the device, nor do they affect the safety and effectiveness of the subject device to the predicate. Both the subject and predicate devices have the same intended use for shaping and removal of hard tissue and bone in Neurological, Spinal, ENT and general surgical procedures. The proposed modifications of adding new models with compatible attachments and supplier changes are verified to ensure overall performance remains the same as the predicate with acceptable results.
In summary, the subject device has the following technological characteristics which are same as the predicate device:
- 1. Intended use/Indications for Use: The intended use/indications for use of the subject device are identical to those of the predicate device i.e. for shaping and removal of hard tissue and bone in Neurological, Spinal, ENT and general surgical procedures.
- Environment of Use: The environment of use remains the same as the predicate, i.e., hospital use. 2.
- 3. Technology (Operating Principle): The Surgify Halo Burr operates by transferring rotary motion from a high-speed surgical drill motor through the shank to the burr head. equipped with dual cutting edges, is designed for precise cutting and removal of bone and hard tissue. A selfcentering mechanism comprising a ring and spring ensures that the cutting edges only engage with the target tissue when appropriate, enhancing surgical precision and safety. The burr's position can be adjusted through the ring mechanism, which allows the burr to either prevent contact with the tissue or engage it, depending on the surgical requirements.
- Design Configuration: The Surgify Halo Burr consists of a cylindrical shank with locking features 4. at the distal end that securely attach the burr to the drill handpiece for stability. The burr head
{6}------------------------------------------------
contains two precision cutting edges for efficient shaping and removal of bone and hard tissue. An end-attachment feature allows the Surgify Halo to be locked to commonly marketed high-speed surgical drill systems.
- Materials Compatibility: Indirect tissue and bone contacting materials meet biocompatibility న. requirements.
- 6. Compatible Interface: The device is compatible with commonly available surgical drill handpieces and attachments, allowing for flexible integration into existing surgical setups.
- 7. Sterility: Single-use, Gamma Sterilization
{7}------------------------------------------------
#### VI. Performance Data
Performance testing requirements were determined through the application of a risk management process, applicable FDA guidance documents and performance standards (21 CFR §882.4310). Performance testing in support of substantial equivalence determination included:
- Functional: Testing (see below) to verify that the device functions as intended, and all ● design and functional specifications are met for all models/system configurations. The relevant performance tests for the modified device included durability, chattering, and compatibility with drill systems.
| Test | Test Method Summary | Results |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| Design Verification Tests | Device performance<br>parameters.<br><br>Durability Chatter • Compatibility with the Drill Systems<br>The durability of the device if<br>the device is used at higher<br>speed (rpm) than<br>recommended.<br><br>Geometrical feature of the<br>device that influences<br>compatibility with the high-<br>speed surgical drills.<br><br>Geometrical feature of the<br>device that influence blister<br>design e.g., overall length and<br>shaft diameter | All tests fulfilled Design<br>input requirements |
- Sterilization Validation: Sterilization assessment (i.e., bioburden re-validation) per ISO ● 11137 for sterilization validation was conducted with acceptable results.
- In-process validation: Both in-process cleaning and in-process passivation validation were ● performed with acceptable results.
#### VII. Conclusions
The subject Surgify Halo device is as safe and as effective as previously cleared predicate devices (K232684), as demonstrated by performance data and risk assessment. The intended use, technological characteristics and principle of operation are substantially equivalent to the predicate device. Thus, the subject Surgify Halo device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.