K230619 · Hubly, Inc. · HBE · May 4, 2023 · Neurology
Device Facts
Record ID
K230619
Device Name
Hubly Electric Drill (H100)
Applicant
Hubly, Inc.
Product Code
HBE · Neurology
Decision Date
May 4, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Hubly Electric Drill is a single-use, sterile, disposable device intended for use on adult patients during neurosurgical procedures for drilling of cranial bone.
Device Story
Battery-powered, single-use, sterile, disposable hand drill for cranial burr holes. Used in ER, bedside, or OR by physicians. Features trigger-activated, single-speed operation; visual depth indicators (5mm, 10mm); tapered stainless steel bit for mechanical plunge prevention. Includes firmware-based auto-stop feature detecting inner skull table breakthrough; LED indicators (green for force, red for stop). Provides intracranial access for emergent conditions. Benefits include streamlined bedside access and enhanced safety via auto-stop and mechanical plunge prevention. Physician monitors depth visually and receives feedback via LEDs; auto-stop prevents dural/brain tissue damage.
Clinical Evidence
No clinical studies on human subjects. Evidence includes bench testing (battery performance, depth accuracy, bit securement, trigger function, EMC/electrical safety per IEC 60601-1/4-2) and a GLP animal study (sheep model) for design validation and summative usability, confirming safe intracranial access without dural or brain tissue damage.
Technological Characteristics
Battery-powered, single-speed, trigger-activated drill. Materials: 316 Stainless Steel (ASTM F899) bit, plastic housing. Features: Tapered bit for mechanical plunge prevention, firmware-based auto-stop. Sterilization: Ethylene Oxide (EO). Connectivity: None. Software: Moderate level of concern, Class B (IEC 62304).
Indications for Use
Indicated for adult patients undergoing neurosurgical procedures requiring cranial bone drilling.
Regulatory Classification
Identification
Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency's role in regulating and overseeing food and drug products in the United States.
May 4, 2023
Hubly Inc. Julie Byars Quality Lead 750 Warrenville Rd. Suite 303 Lisle, Illinois 60532
Re: K230619
Trade/Device Name: Hubly Electric Drill (H100) Regulation Number: 21 CFR 882.4310 Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, And Their Accessories Regulatory Class: Class II Product Code: HBE Dated: March 1, 2023 Received: March 6, 2023
Dear Julie Byars:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Digitally signed by Adam D. Adam D. Pierce -S Date: 2023.05.04 Pierce -S 08:55:18 -04'00'
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230619
Device Name Hubly Electric Drill
Indications for Use (Describe)
The Hubly Electric Drill is a single-use, sterile, disposable device intended for use on adult patients during neurosurgical procedures for drilling of cranial bone.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### A. Device Information:
| Category | Comments |
|------------------------------------|--------------------------------------------------------------------------------------------------|
| Sponsor: | Hubly, Inc.<br>Casey Grage<br>750 Warrenville Rd. Suite 303<br>Lisle, IL 60532<br>(844) 482-5942 |
| Correspondent Contact Information: | Julie Byars<br>750 Warrenville Rd. Suite 303<br>Lisle, IL 60532<br>901-848-4010 |
| Device Common Name: | Powered Cranial Drill |
| Device Regulation & Name: | 21 CFR 882.4310 - Powered simple cranial drills,<br>burrs, trephines and their accessories |
| Classification & Product Code: | Class II, HBE |
| 510(k) Number: | K222391 |
| Device Proprietary Name: | Hubly Electric Drill |
#### Predicate Device Information:
| Predicate Device: | Phasor Drill |
|----------------------------------------------------|--------------------------------------------------------------------------------------------|
| Predicate Device Manufacturer: | Biotex Inc. |
| Predicate Device Common Name: | Powered Cranial Drill |
| Predicate Device Premarket Notification<br>#: | K161704 |
| Predicate Device Classification & Name: | 21 CFR 882.4310 - Powered simple cranial drills,<br>burrs, trephines and their accessories |
| Predicate Device Classification &<br>Product Code: | Class II, HBE |
# B. Date Summary Prepared
May 4, 2023
# C. Description of Device
The Hubly Electric Drill is a battery powered cranial drill. It is single-use disposable and provided ethylene oxide sterilized. The drill is designed to create optimal burr holes, in the emergency room, at the bedside, or in the operating room. The system is designed to streamline bedside intracranial access and facilitate the treatment of emergent conditions in adult patients.
{5}------------------------------------------------
# Hubly, Inc. Hubly Electric Drill
The device is a battery-powered hand drill which can be used one-handed using either hand. The drill is trigger-activated after removal of the battery pull tab. The drill has a single speed and turns off when the trigger is released. The drill bit has depth indicators at 5 and 10mm depth, which the physician may use to visually gauge depth of penetration while drilling.
The drill also has an auto-stop feature which detects when the bit breaks through the inner table of the skull and immediately stops the drill. The device has an LED indicator which indicates to the user (green) when they are applying enough force and (red) when the drill stops.
The drill features mechanical plunge prevention with a tapered stainless steel drill may be reactivated any number of times using the trigger if the physician desires.
#### D. Indications for Use
The Hubly Electric Drill is a single-use, sterile, disposable device intended for use on adult patients during neurosurgical procedures for drilling of cranial bone.
# E. Comparison of the Technological Characteristics
| | Application Device<br>Hubly Electric Drill - K222391 | Predicate Device<br>Phasor Drill - K161704 | Impact on Substantial<br>Equivalence |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Company | Hubly, Inc. | Biotex, Inc. | N/A |
| Regulation<br>Number | 21 CFR 882.4310 | 21 CFR 882.4310 | Identical |
| Product Code | HBE - Class II | HBE - Class II | Identical |
| Intended Use &<br>Indications for<br>use | The Hubly Electric Drill is a<br>single-use, sterile, disposable<br>device intended for use on<br>adult patients during<br>neurosurgical procedures for<br>drilling of cranial bone. | The Phasor Drill is a sterile,<br>single-use, disposable<br>device intended for use on<br>adult patients during<br>neurosurgical procedures<br>for drilling of cranial bone. | Identical |
| Technology | | | |
| Power Source | Battery Power (removable) | Battery Power (removable) | Identical |
| Automatic stop | Auto-stop | None | No impact - Additional<br>safety feature |
| Mechanical<br>plunge<br>prevention | Tapered bit | Thumbscrew depth stop | No impact |
| Bit Material | Stainless Steel (316) | Stainless Steel (440A) | No impact |
| Drill activation | Trigger - On/Off | Button - On/Off | Identical |
| Drill rotation | Forward | Forward & Reverse | No impact |
| Flute<br>characteristics | 2 Straight flutes | 2-3 Twist flutes | No impact – Straight<br>flutes remove more<br>easily without reverse<br>rotation. |
{6}------------------------------------------------
# F. Summary of Supporting Data
The Hubly Electric Drill has the same intended use, patient population and classification as the predicate. At a high level, the subject and predicate devices are based on the following same technological elements:
- . Sterile packaged for single-use, and disposable
- Operates on battery power (batteries removable for disposal)
- . Two state operation, single speed
- Employ mechanical plunge prevention mechanisms for safety
- . Plastic drill housing with stainless steel drill bit(s)
The following technical characteristics of the subject device differ from the predicate device:
- Use of tapered drill bit for mechanical plunge prevention (vs thumbscrew depth stop)
- Drill bit diameter & length, flute type
- . Addition of firmware based automatic stopping feature
- . Forward-only drill functionality
- EO sterilization (vs gamma)
The Hubly drill bit is larger in diameter and tapers to the tip for mechanical plunge prevention, whereas the predicate has a thumbscrew depth stop for plunge prevention. The Hubly drill bit has straight flutes and tapers and thus is easily removed from the burr hole without reverse functionality. The applicant device has an additional firmware-based automatic stopping safety feature, and for this reason is EO sterilized.
The proposed device does not differ from the predicate in any way which would negatively impact the safety or effectiveness of the device.
# G. Discussion of Performance Testing
The following performance data were provided in support of the substantial equivalence determination.
# BIOCOMPATIBILITY TESTING
The drill bit is the only patient-contacting material used in the Hubly device. 316 Stainless Steel certified to ASTM F899 was selected for its known biocompatibility with human tissues and fluids. A biological risk assessment was conducted for the device. It considers raw material, manufacturing processing, and full endpoint biocompatibility testing performed, and determines full compliance of the device with ISO 10993-1.
{7}------------------------------------------------
# Hubly, Inc. Hubly Electric Drill
# ELECTRICAL SAFETY AND ELECTROMAGNETIC COMPATIBILTY (EMC)
Electrical safety and EMC testing were conducted on the Hubly device. The system complies with the IEC 60601-1 standard for safety and the IEC 60601-4-2 and 60601-4-2 standards for EMC.
# SOFTWARE VERIFICATION TESTING
Software verification was conducted on the firmware responsible for the Hubly Electric Drill's automatic stopping feature. The device has an equivalent expected safety profile compared to the standard of care without utilizing software mitigations. The software implements additional safety features compared to the current standard of care.
The Hubly Electric Drill software is initially classified as a Class B software system according to IEC 62304 Section 4.3a based on the possible hazards to which the software component can contribute. For this reason, the resulting possible harm due to failure of software is considered non-serious. Based on the level of concern decision process (as described in Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices), the level of concern for the Hubly Electric Drill has been determined to be "Moderate".
# BENCH TESTING
The following benchtop performance tests were conducted. Results demonstrated that the drill meets all design specifications and requirements.
- Performance - battery/continuous use
- Depth of Penetration, bone thickness variation
- . Bit securement
- . Trigger performance
- Basic Safety and Essential Performance
- . EMC Testing (Basic Safety, Essential Performance, Electromagnetic disturbances)
# ANIMAL STUDY
In the GLP animal study conducted, design validation and summative usability evaluation were conducted. A sheep model was selected for its close approximation of the characteristics of the human skull. This study demonstrated that Users (Test Device Evaluators) could safely use the Hubly Electric Drill for intracranial access without damaging the dura or brain tissue. Additionally, all in vivo requirements were met to support Usability Evaluation for the Hubly electric Drill.
# CLINICAL STUDIES
Clinical testing was not performed on human subjects.
# H. Conclusion
The predicate device was cleared based on bench testing alone. The performance and design validation testing conducted on the Hubly Electric Drill on the bench and in animals demonstrates that it performs equivalent to the predicate device that is currently marketed for the same intended use. The proposed device adds additional control and safety features, and hardware and
{8}------------------------------------------------
software verification and validation demonstrate that the Hubly drill will perform as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.