← Product Code [HBE](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HBE) · K213454

# Midas Rex MR8 Depth Stop Attachment, Midas Rex MR8 Depth Stop Tool 2.4mm, Midas Rex MR8 Depth Stop Tool 3.2mm, Midas Rex MR8 Depth Stop Tool 3.2mm Long, Midas Rex MR8 Depth Stop Tool 7.5mm (K213454)

_Medtronic Powered Surgical Solutions · HBE · Nov 23, 2021 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HBE/K213454

## Device Facts

- **Applicant:** Medtronic Powered Surgical Solutions
- **Product Code:** [HBE](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HBE.md)
- **Decision Date:** Nov 23, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.4310
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Midas Rex MR8 Depth Stop attachment and tools are indicated for the incision, cutting, removing, and drilling of soft and hard tissue during cranial surgical procedures with the intent to create a hole through the cranium to allow surgeons access to desired surgical locations and/or to facilitate insertion, placement of other surgical devices during such procedures.

## Device Story

System comprises depth stop attachments and surgical dissecting tools used with MR8 high-speed drill systems (pneumatic/electric) and Stealth Autoguide system; creates cranial access holes of controlled diameter and depth. Used in cranial surgery by surgeons. Device provides mechanical depth control during drilling; output is physical hole in cranium. Benefits include precise access to surgical sites and facilitation of device placement. Modifications from predicate include compatibility with MR8 drill systems, extended depth setting to 20mm, minor tool head geometry changes, and updated user workflow.

## Clinical Evidence

Bench testing only. No clinical data presented. Testing included product life verification, summative validation, and biocompatibility testing to demonstrate safety and effectiveness of modifications.

## Technological Characteristics

Materials: Stainless steel, aluminum, ceramic, phenolic, epoxy, chrome-coated brass, polymeric. Sensing/Actuation: Mechanical depth stop mechanism. Energy: Pneumatic or electric drill systems. Connectivity: Compatible with Stealth Autoguide system. Sterilization: Steam (reusable attachments), Gamma (single-use tools).

## Regulatory Identification

Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.

## Predicate Devices

- Midas Rex Legend Depth Stop Attachment and Tools ([K191597](/device/K191597.md))

## Reference Devices

- MR8 Drill Systems ([K181535](/device/K181535.md))
- Legend Drill Systems ([K163182](/device/K163182.md))
- Legend Drill Systems ([K170312](/device/K170312.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 23, 2021

Medtronic Powered Surgical Solutions Carrie Eddings Principal Regulatory Affairs Specialist 4620 North Beach Street Fort Worth, Texas 76137

Re: K213454

Trade/Device Name: Midas Rex MR8 Depth Stop Attachment, Midas Rex MR8 Depth Stop Tool 2.4mm, Midas Rex MR8 Depth Stop Tool 3.2mm, Midas Rex MR8 Depth Stop Tool 3.2mm Long, Midas Rex MR8 Depth Stop Tool 7.5mm Regulation Number: 21 CFR 882.4310 Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, And Their Accessories Regulatory Class: Class II Product Code: HBE, HBB, HBC Dated: October 22, 2021 Received: October 26, 2021

Dear Carrie Eddings:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K213454

#### Device Name

Midas Rex MR8 Depth Stop Attachment (MR8-ASDS01 ); Midas Rex MR8 Depth Stop Tool 2. 4mm (MR8-DS1TD24); Midas Rex MR8 Depth Stop Tool 3.2mm (MR8-DS1TD32);Midas Rex MR8 Depth Stop Tool 3.2mm Long (MR8-DS1TD32L); Midas Rex MR8 Depth Stop Tool 7.5mm (MR8-DS1TD75)

#### Indications for Use (Describe)

The Midas Rex MR8 Depth Stop attachment and tools are indicated for the incision, cutting, removing, and drilling of soft and hard tissue during cranial surgical procedures with the intent to create a hole through the cranium to allow surgeons access to desired surgical locations and/or to facilitate insertion, placement of other surgical devices during such procedures.

Type of Use (Select one or both, as applicable)

| <span style="font-family: DejaVu Sans, sans-serif">✘</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary

# November 23, 2021

- Company: Medtronic Powered Surgical Solutions 4620 North Beach Street Fort Worth, Texas 76137 Telephone Number: (817) 788-6400
- Carrie Eddings (Primary) Contact: Principal Regulatory Affairs Specialist Telephone Number: (214) 384-0606 Email: carrie eddings(@medtronic.com

Kyle Hoefling (Alternate) Senior Regulatory Affairs Manager Telephone: (760) 207-2432 Email: kyle.d.hoefling(@medtronic.com

Proprietary Trade Name: Midas Rex MR8 Depth Stop Attachment and Tools Common Name: MR8 Depth Stop System Classification Name: Powered cranial drills, burrs, trephines, and their accessories (21 CFR 882.4310) Classification: Class II Product Code: HBE Primary Predicate: Midas Rex Legend Depth Stop Attachment and Tools, K191597

### Product Description:

The Midas Rex MR8 Depth Stop attachment and tools are intended to create a cranial access hole of a known diameter and depth.

#### Indications for Use:

The Midas Rex MR8 Depth Stop attachment and tools are indicated for the incision, cutting. removing, and drilling of soft and hard tissue during cranial surgical procedures with the intent to create a hole through the cranium to allow surgeons access to desired surgical locations and/or to facilitate insertion, placement of other surgical devices during such procedures.

#### Comparison of the Technological Characteristics:

There have been no significant changes to the Midas Rex™ Legend Depth Stop Attachment and Tools since prior clearance in K191597.

The subject device, Midas Rex™ MR8™ Depth Stop Attachment and Tools, are to be used with the MR8TM High-Speed Drill Systems and the Stealth Autoguide™ System to create

{4}------------------------------------------------

cranial access holes of specific diameter and depth as previously cleared through K191597 for the Midas Rex™ Legend Depth Stop Attachment and Tools.

Changes from the predicate include:

- Compatibility with the predicate MR8 Drill Systems included within K181535 .
- . Extended depth setting to 20mm
- Minor modifications to tool head geometry ●
- Adding a new task in the user workflow .

{5}------------------------------------------------

| Feature                                          | Subject Device (Midas Rex™<br>MR8™ Depth Stop System)                                                                                                                                                                                                                                                                                                                                                       | Predicate Device (Midas Rex™ Legend™<br>Depth Stop System, K191597)                                                                                                                                                                                                                                                                                                                                | Equivalence Discussion                                                                                                                                                     |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor                                          | Medtronic Powered Surgical<br>Solutions                                                                                                                                                                                                                                                                                                                                                                     | Medtronic Powered Surgical Solutions                                                                                                                                                                                                                                                                                                                                                               | Identical                                                                                                                                                                  |
| Product Code                                     | HBE                                                                                                                                                                                                                                                                                                                                                                                                         | HBE, HBC, HBB                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                  |
| Drill System                                     | • Pneumatic<br>• Electric                                                                                                                                                                                                                                                                                                                                                                                   | • Pneumatic<br>• Electric                                                                                                                                                                                                                                                                                                                                                                          | Identical                                                                                                                                                                  |
| Operating Principle                              |                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                                                                                                                  |
| Intended Use                                     | The Midas Rex MR8 Depth Stop<br>system is intended to create a<br>cranial access hole of a known<br>diameter and depth.                                                                                                                                                                                                                                                                                     | The Midas Rex Legend Depth Stop system is<br>intended to create a cranial access hole of a<br>known diameter and depth.                                                                                                                                                                                                                                                                            | Identical                                                                                                                                                                  |
| Indications for Use                              | The Medtronic Midas Rex MR8<br>Drill System is indicated for the<br>incision, cutting, removing, and<br>drilling of soft and hard tissue<br>during cranial surgical procedures<br>with the intent to create a hole<br>through the cranium to allow<br>surgeons access to desired<br>surgical locations and/or to<br>facilitate insertion, placement of<br>other surgical devices during such<br>procedures. | The Medtronic Midas Rex Legend Drill System<br>is indicated for the incision, cutting, removing,<br>and drilling of soft and hard tissue during cranial<br>surgical procedures with the intent to create a<br>hole through the cranium to allow surgeons<br>access to desired surgical locations and/or to<br>facilitate insertion, placement of other surgical<br>devices during such procedures. | Identical                                                                                                                                                                  |
| System<br>components                             | • Electric Handpieces<br>• Pneumatic Handpieces<br>• Attachments<br>• Surgical Dissecting Tools                                                                                                                                                                                                                                                                                                             | • Electric Handpieces<br>• Pneumatic Handpieces<br>• Attachments<br>• Surgical Dissecting Tools                                                                                                                                                                                                                                                                                                    | Identical                                                                                                                                                                  |
| Patient Contacting<br>Components                 | • Attachments<br>• Surgical Dissecting Tools                                                                                                                                                                                                                                                                                                                                                                | • Attachments<br>• Surgical Dissecting Tools                                                                                                                                                                                                                                                                                                                                                       | Identical                                                                                                                                                                  |
| Attachment<br>Materials                          | Stainless Steel, Aluminum,<br>Ceramic, Phenolic, Epoxy,<br>Chrome Coated Brass, Polymeric                                                                                                                                                                                                                                                                                                                   | Stainless Steel, Aluminum, Ceramic, Phenolic,<br>Epoxy, Chrome Coated Brass, Polymeric                                                                                                                                                                                                                                                                                                             | Identical                                                                                                                                                                  |
| Attachment<br>Configuration                      | Straight, Variable                                                                                                                                                                                                                                                                                                                                                                                          | Straight, Variable                                                                                                                                                                                                                                                                                                                                                                                 | Identical                                                                                                                                                                  |
| Feature                                          | Subject Device (Midas Rex™<br>MR8™ Depth Stop System)                                                                                                                                                                                                                                                                                                                                                       | Predicate Device (Midas Rex™ Legend™<br>Depth Stop System, K191597)                                                                                                                                                                                                                                                                                                                                | Equivalence Discussion                                                                                                                                                     |
| Attachment Tube<br>Length                        | 12.8cm (tube retracted)<br>14.5cm (tube exposed)                                                                                                                                                                                                                                                                                                                                                            | 13.7cm (tube retracted)<br>14.8cm (tube exposed)                                                                                                                                                                                                                                                                                                                                                   | Equivalent<br>The attachment tube length is<br>inherent due to the overall design of<br>MR8 attachments and does not<br>impact the performance or safety of<br>the device. |
| Sterilization<br>Method – Reusable<br>Attachment | Steam                                                                                                                                                                                                                                                                                                                                                                                                       | Steam                                                                                                                                                                                                                                                                                                                                                                                              | Identical                                                                                                                                                                  |
| Surgical Dissecting<br>Tool Materials            | Tool Steel                                                                                                                                                                                                                                                                                                                                                                                                  | Tool Steel                                                                                                                                                                                                                                                                                                                                                                                         | Identical                                                                                                                                                                  |
| Dissecting Tool<br>Tip Style                     | Twist Drill                                                                                                                                                                                                                                                                                                                                                                                                 | Twist Drill                                                                                                                                                                                                                                                                                                                                                                                        | Equivalent<br>Minor tool tip modifications which<br>do not impact safety or performance.                                                                                   |
| Dissecting Tool<br>Overall Length                | 17-22cm                                                                                                                                                                                                                                                                                                                                                                                                     | 17-22cm                                                                                                                                                                                                                                                                                                                                                                                            | Identical                                                                                                                                                                  |
| Dissecting Tool<br>Head Diameter                 | 2.1 – 7.5 mm                                                                                                                                                                                                                                                                                                                                                                                                | 2.1 – 7.5 mm                                                                                                                                                                                                                                                                                                                                                                                       | Identical                                                                                                                                                                  |
| Packaging – Sterile<br>Dissecting Tools          | Individually packaged in a clear<br>plastic capped tube placed within<br>a polypoly pouch                                                                                                                                                                                                                                                                                                                   | Individually packaged in a clear plastic capped<br>tube placed within a polypoly pouch                                                                                                                                                                                                                                                                                                             | Equivalent<br>The packaging is identical except for<br>the tube cap. On the predicate device<br>it is yellow, and on the subject<br>device, the tube cap is blue.          |
| Sterilization<br>Method – Single<br>Use Tools    | Gamma with minimum radiation<br>dose of 25 kGy                                                                                                                                                                                                                                                                                                                                                              | Gamma with minimum radiation dose of 25 kGy                                                                                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                  |
| Shelf Life –<br>Dissecting Tools                 | 5 Years                                                                                                                                                                                                                                                                                                                                                                                                     | 5 Years                                                                                                                                                                                                                                                                                                                                                                                            | Identical                                                                                                                                                                  |

{6}------------------------------------------------

{7}------------------------------------------------

#### Discussion of the Performance Testing:

The intended use of the subject devices, Midas Rex™ MR8™ Depth Stop Attachments and Tools and predicate are identical, and their technological characteristics are similar. The subject devices are a line extension to the MR8 Drill System (K183515) in the same way that the predicate device was a line extension to the Legend Drill Systems (K163182 and K170312). The changes from the predicate, do not raise any new issues of safety and effectiveness.

Risk management was conducted in accordance with ISO 14971:2019 and did not identify any new risks when compared to the predicate. The following testing was conducted to demonstrate that the modifications to the subject devices are as safe and effective as the predicate.

- . Product Life Verification
- Summative Validation .
- . Biocompatibility Testing

Results of these tests demonstrate that the functionality, integrity, and safety and effectiveness of the subject devices are sufficient for their intended use, indications for use and support a determination of substantial equivalence.

#### Conclusions

The Midas Rex™ MR8™ Depth Stop Attachment and Tools have shown through comparison to be substantially equivalent to the identified predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HBE/K213454](https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HBE/K213454)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
