Stryker Elite Attachments, Stryker Heavy Duty (HD) Attachments
Applicant
Stryker
Product Code
HBE · Neurology
Decision Date
Jun 14, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4310
Device Class
Class 2
Indications for Use
The Elite and Heavy Duty Attachments are intended to be used with the Stryker Consolidated Operating Room Equipment (CORE) console and electric and pneumatic motors. When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are intended to cut bone cement in the following manner: drilling, reaming, decorticating, shaping, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose and Throat (ENT)/Otology/Neurotology/Otothinolaryngology; Craniofacial; Dental; and Endoscopic applications. The specific applications include Craniotomy, Laminotomy, Laminectomy, Minimally Invasive Surgery (MIS) Spine, Expanded Endonasal Approach (EEA)/Anterior Skull Base/Endoscopic/Transnasal/Transphenoidal, and Orthopedic Spine. When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are intended to cut teeth in the following manner: sectioning, splitting, fragmenting, extracting, removing, drilling, and reaming. When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are also usable in the preparation for the placement of screws, metal, wires, pins, and other fixation devices, or the cutting of screws, metal, wires, pins, and other fixation devices in the following manner: sectioning, smoothing or shaping of metal, and removing/rounding sharp edges.
Device Story
Stryker Elite and Heavy Duty (HD) Attachments serve as mechanical interfaces between powered surgical motors (electric/pneumatic) and cutting accessories (burs). Attachments transmit torque from motor to accessory to facilitate cutting, drilling, reaming, or shaping of bone, bone cement, teeth, and fixation hardware (screws/wires). Used in OR settings by surgeons; attachments provide a grip location for the user. Device features straight/angled configurations and color-coded bands for accessory compatibility. Modification involves updated color band material and colorant. No software or electronic processing involved.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench performance testing (colorfastness, durability) and biocompatibility evaluation per ISO 10993-1:2018.
Technological Characteristics
Materials: Stainless steel (SST) with modified color band material/colorant. Principle: Mechanical torque transmission. Configuration: Straight and angled, lengths 7-20cm. Energy: Powered by external electric/pneumatic motors. Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for patients requiring bone, bone cement, or tooth cutting, shaping, or removal during neurosurgical, spinal, ENT, craniofacial, dental, or orthopedic procedures, including craniotomy, laminectomy, and endoscopic skull base approaches.
Regulatory Classification
Identification
Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.
Predicate Devices
Stryker Elite and Heavy Duty (HD) Attachments (K143320)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 14, 2019
Stryker Paminder Khurmi, RAC Senior Regulatory Affairs Specialist 4100 E. Milham Ave Kalamazoo, Michigan 49001
Re: K190172
Trade/Device Name: Stryker Elite Attachments, Stryker Heavy Duty (HD) Attachments Regulation Number: 21 CFR 882.4310 Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, And Their Accessories Regulatory Class: Class II Product Code: HBE, ERL, HBB, DZI Dated: March 11, 2019 Received: March 12, 2019
Dear Paminder Khurmi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Matthew Krueger, M.S.E. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190172
Device Name
Stryker Elite and Heavy Duty (HD) Attachments
### Indications for Use (Describe)
The Elite and Heavy Duty Attachments are intended to be used with the Stryker Consolidated Operating Room Equipment (CORE) console and electric and pneumatic motors. When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are intended to cut bone cement in the following manner: drilling, reaming, decorticating, shaping, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose and Throat (ENT)/Otology/Neurotology/Otothinolaryngology; Craniofacial; Dental; and Endoscopic applications.
The specific applications include Craniotomy, Laminotomy, Laminectomy, Minimally Invasive Surgery (MIS) Spine, Expanded Endonasal Approach (EEA)/Anterior Skull Base/Endoscopic/Transnasal/Transphenoidal, and Orthopedic Spine.
When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are intended to cut teeth in the following manner: sectioning, splitting, fragmenting, extracting, removing, drilling, and reaming. When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are also usable in the preparation for the placement of screws, metal, wires, pins, and other fixation devices, or the cutting of screws, metal, wires, pins, and other fixation devices in the following manner: sectioning, smoothing or shaping of metal, and removing/rounding sharp edges.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
4100 E. Milham Avenue Kalamazoo, MI 49001 t: 269 323 7700 f: 269 324 5412 www.stryker.com
# Section 5. 510(k)Summary
This section provides a summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.
#### SUBMITTER I.
| 510(k) Owner: | Stryker Instruments<br>4100 E. Milham Avenue<br>Kalamazoo, Michigan 49001 USA<br>Ph: +1-269-323-7700<br>Fax: +1-269-324-5412 |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Submitter/ Contact<br>Person: | Paminder Khurmi<br>Sr. Regulatory Affairs Specialist<br>Phone: +1-269-389-2264<br>Fax: +1-269-324-5412<br>Paminder.Khurmi@Stryker.com |
| FDA Establishment<br>Registration No. | 1811755 |
| Date Submitted: | March 14, 2019 |
#### II. DEVICE
| Trade Name: | Stryker® Elite and Heavy Duty (HD) Attachments |
|----------------------|--------------------------------------------------------------------------------------------------------|
| Common Name: | Surgical Drill Handpieces |
| Classification name: | Drills, Burs, Trephines & Accessories (Simple, Powered)<br>(21 CFR 882.4310, Product code HBE) |
| | Drill, Surgical ENT (Electric or Pneumatic) including Handpiece<br>(21 CFR 874.4250, Product code ERL) |
| | Pneumatic cranial drill motor<br>(21 CFR 882.4370, Product code HBB) |
| | Drill, Bone, Powered |
{4}------------------------------------------------
### 4100 E. Milham Avenue Kalamazoo, MI 49001 t: 269 323 7700 f: 269 324 5412 www.stryker.com
| | (21 CFR 872.4120, Product code DZI) |
|-----------------|-------------------------------------|
| Classification: | II |
#### III. PREDICATE DEVICE
| . 510 (K) Number | Product code | Trade name | Manufacturer |
|------------------|-----------------------|------------------------------------------------------|---------------------|
| K143320 | HBE, HBB, ERL,<br>DZI | Stryker® Elite and<br>Heavy Duty (HD)<br>Attachments | Stryker Instruments |
#### IV. SUBJECT DEVICE DESCRIPTION
The Stryker Elite and HD Attachments are used within a system consisting of a variety of devices, including a console, powered motors, and cutting accessories. The attachments connect to the motors and the cutting accessories to complete the system for physician use. The Stryker Elite and HD Attachments are offered for prescription use only. The Elite and HD Attachments are intended to serve as interfaces between powered motors and cutting accessories for the purposes of:
- Cutting bone, bone cement, and teeth;
- Placing or cutting screws, metal, wires, pins, and other fixation devices; and
- Providing a location for the user to hold and grip the device system. ●
The Elite and HD Attachments are provided in straight and angled configurations. The subject Elite Attachments are offered in the following lengths: 7cm, 12cm, 17cm, and 20cm. The Stryker HD Attachments are offered in the following lengths: 9cm and 14cm. The Elite and HD Attachments are powered by, and compatible with, the Stryker electric and pneumatic motors. The attachments are also used with the cutting accessories (burs). All the attachments demonstrate a rotary mode of action by transmitting torque.
The Elite and HD Attachments are made of stainless steel (SST). The attachments display a color band on the outer surface. The color bands serve to enhance the distinction of attachment and cutting accessory compatibility.
The purpose of this submission is to gain clearance for colorant band modification for the Subject Device Attachments color band that exceeds the threshold as per FDA Guidance, "Deciding When to Submit a 510(k) for a Change to an Existing Device".
{5}------------------------------------------------
4100 E. Milham Avenue Kalamazoo, MI 49001 t: 269 323 7700 f: 269 324 5412 www.stryker.com
#### INDICATIONS FOR USE V.
# ARI E 1. COMPADISON OF INDICATIONS FOR USE
The Subject Device Indications for Use remains identical to the Predicate Device. The modification described to the Subject Device in this 510(k) does not change the Indication for Use.
{6}------------------------------------------------
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
The Subject Device is compared to the Predicate Device for substantial equivalence of technological characteristics based on the modification described in this submission.
## Technological Characteristics
The Color Band location, Attachment to Cutting Accessory Locking Mechanism, and Attachment to Motor Locking Mechanism are identical when comparing the Subject and Predicate Device and, other than the modification described in this submission, all other technological characteristics remain the same between the Subject and Predicate Devices.
The technological characteristics that are not the same between the Subject and Predicate Devices are the color band material and colorant. However, the Subject Device color band material and colorant are similar to those cleared for the Predicate Device, and the different characteristics do not raise different questions of safety and effectiveness.
The change in color band material and colorant on the device does not change the Principle of Operation of the Subject Device compared to the Predicate Device.
The Principle of Operation remains: The Subject Device Attachments are combined with a power source, motor, and cutting accessory to achieve their function; the main function of the Subject Device Attachments is to provide a balanced location for the surgeon to hold and grip the device system, and to transmits torque from the motor to a cutting accessory.
### VII. PERFORMANCE DATA
The following testing were conducted, and the data was provided in support of the substantial equivalence determination:
### Performance Bench Testing
Verification testing was performed on the Subject Device as dictated by the results of the Risk Analysis, and no new questions of safety and effectiveness were raised. The Subject Device met all pre-defined acceptance criteria. The results of the tests support the substantial equivalence of the Subject Device to the Predicate Device. Testing data from the following was provided in support of the substantial equivalence determination.
- Colorfastness test
- Durability Test .
{7}------------------------------------------------
4100 E. Milham Avenue Kalamazoo, MI 49001 t: 269 323 7700 f: 269 324 5412 www.stryker.com
### Biocompatibility Testing
A biocompatibility evaluation was performed following the recommendations of ISO 10993-1: 2018 and FDA Guidance (Use of International Standard ISO- 10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" June 2016) as appropriate for limited exposure (< 24 hours) externally communicating, tissue/bone/dentin devices. The following assessments were completed to further assess the potential risk of the subject device material and colorant modification impacting color additive release.
- Chemical Characterization: ICP-MS
- . Toxicological Risk Assessment
The Biocompatibility testing was adopted from K143320
# Animal Testing
Animal testing was not required as a basis for substantial equivalence.
## Clinical Testing
Clinical testing was not required as a basis for substantial equivalence.
## VIII. CONCLUSIONS
The subject device, in comparison with the legally marketed predicate, has the same intended use, indications for use, operating principles, energy source, and functional outputs. Performance testing and risk analysis demonstrate that the device is as safe and effective as the predicate device and does not raise different significant questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.