The Cranial Drill Bits and accessories are intended to be used for the skull for neurological procedures, such as brain biopsy, brain contacting electrode and electrode accessory device placement.
Device Story
Cranial drill bits and accessories create access holes in the skull for neurological procedures; used for brain biopsy, electrode placement, and accessory device placement. System includes twist drill bits (16-30 cm length, 2.4-3.2 mm diameter), drill stops, drill stop wrenches, drill sleeve guides, and guide blocks. Used intraoperatively by surgeons; drill sleeve guides and guide blocks interface with stereotactic frames to minimize bit wobble and ensure precise placement. Device provides mechanical access; does not involve electronic processing or AI. Benefits include precise, controlled skull penetration for neurosurgical access.
Clinical Evidence
Bench testing only. Evaluations confirmed drilling efficiency (drilling through simulated skull in < 1 minute), drill bit run-out (wobble) performance with and without guide accessories, and mechanical compatibility with stereotactic frames. All tests passed.
Indicated for patients undergoing neurological procedures requiring skull access, including brain biopsy, brain contacting electrode placement, and electrode accessory device placement.
Regulatory Classification
Identification
Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure with three faces in profile, connected by flowing lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 7, 2017
Ad-tech Medical Instrument Corporation % Gary Syring Principal Consultant Ouality & Regulatory Associates, LLC 800 Levanger Lane Stoughton, Wisconsin 53589
Re: K170442
Trade/Device Name: Cranial Drill Bits and Accessories Regulation Number: 21 CFR 882.4310 Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, and Their Accessories Regulatory Class: Class II Product Code: HBE, HBG Dated: August 7, 2017 Received: August 10, 2017
Dear Mr. Syring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Michael J. Hoffmann -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170442
Device Name Cranial Drill Bits and Accessories
Indications for Use (Describe)
The Cranial Drill Bits and accessories are intended to be used for the skull for neurological procedures, such as brain biopsy, brain contacting electrode and electrode accessory device placement.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary - K170442 5.0
This summary is provided to support the 510(k) pre-market notification for the Cranial Drill Bits and accessories.
| Company Name: | Ad-Tech Medical Instrument Corporation<br>1901 William Street<br>Racine, WI 53404<br>Phone: (262) 634-1555 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact: | Lisa Theama, Chief Operating Officer |
| Date Summary Prepared: | August 3, 2017 |
| Trade Name: | Cranial Drill Bits and Accessories |
| Common Name: | Cranial Drill Bits |
| Classification Name: | 21 CFR 882.4310 Powered simple cranial drills, burrs, trephines, and<br>their accessories, Product Code: HBE, Class II<br>21 CFR 882.4300 Manual cranial drills, burrs, trephines, and their<br>accessories, Product Code: HBG, Class II |
| Predicate Devices: | K062842 Twist Drills (Bits), Primary Predicate<br>K122456 MRI Cranial Drill<br>K160129 MRII Cranial Drill and Accessories |
#### 5.1 Product Description
The device under review is a family of cranial drill bits and accessories. These cranial drill bits and accessories are applied to create a hole through the skull in support of neurological procedures.
Cranial drill bits are twist drill bits used for drilling holes in the skull, supporting access to the brain for brain biopsy, brain contacting electrode placement, electrode accessory placement such as Anchor Bolts or other needs as determined by the user. The cranial drill bit variations are 16 cm to 30 cm length and 2.4 mm to 3.2 mm outer diameter.
Accessories to the cranial drill bits include:
- Drill Stop: A collar placed on the outer diameter of the drill bit to provide a depth placement ● indicator.
- Drill Stop Wrench: The Drill Stop Wrench attaches the Drill Stop to the cranial drill bit.
- . Drill Sleeve Guide: The Drill Sleeve Guide is an optional accessory, consisting of a hollow tube of stainless steel that the cranial drill bit passes through to minimize the wobble of the drill bit as it is passed to the skull. The Drill Sleeve Guide is placed within a stereotactic frame.
- . Guide Block: A Guide Block is an optional accessory that adapts to a stereotactic frame to hold the Drill Sleeve Guide.
{4}------------------------------------------------
#### 5.2 Indications for Use
The Cranial Drill Bits and accessories are intended to be used for drilling holes in the skull for neurological procedures, such as brain biopsy, brain contacting electrode and electrode accessory device placement.
#### 5.3 Summary of Technological Characteristics
The following table provides a side-by-side comparison the Cranial Drill Bits and accessories to the predicate devices applied to support this pre-market notification.
| Table 5.3-1: Substantial Equivalence Technical Characteristics | | | | |
|----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Feature | Cranial Drill Bits and<br>Accessories<br>(Under Review) | Twist Drills<br>(Primary Predicate<br>K062842) | MRI Cranial Drill<br>(Predicate K122456,<br>K160129) | Comment |
| Intended use | The Cranial Drill Bit<br>and Accessories are<br>used with drills for<br>creating a hole in the<br>skull to provide access<br>or placement of<br>accessory components. | The Twist Drill Bit and<br>Accessories are used<br>with drills for creating a<br>hole in bone to provide<br>access or placement of<br>hardware. | The Cranial Drill Bit and<br>Accessories are used<br>with drills for creating a<br>hole in the skull to<br>provide access. | Same |
| Indications<br>for Use | The Cranial Drill Bits<br>and accessories are<br>intended to be used for<br>drilling holes in the<br>skull for neurological<br>procedures, such as<br>brain biopsy, brain<br>contacting electrode<br>and electrode accessory<br>device placement. | The Lorenz Twist Drills<br>are intended to be used<br>for drilling holes in large<br>and small bone during<br>orthopedic, spinal,<br>neurosurgical, medial<br>sternotomy, and oral and<br>maxillofacial procedures. | The MRII Cranial Drill<br>and accessories is<br>intended to provide<br>access through the skull<br>for ventriculostomy or<br>other neurological<br>procedures, such as<br>biopsy or catheter<br>placement, in or near an<br>MR scanner of 3T<br>maximum field strength.<br>The MRI Cranial Drill<br>and accessories are<br>intended to be used only<br>when the scanner is not<br>performing a scan. The<br>MRII Cranial Drill is<br>intended for single use<br>only. | The drill bits<br>create an<br>access hole in<br>the skull to<br>support<br>neurological<br>procedures. |
| Cranial drill<br>bit diameter | 2.4 mm to 3.2 mm | 0.7 mm to 2.4 mm | 2.0 mm to 3.2 mm | Equivalent |
| Cranial Drill<br>Bit | Yes | Yes | Yes | Equivalent |
| Accessories | Drill Stop<br>Drill Stop Wrench<br>Drill Sleeve Guide<br>Guide Block | Drill Sleeve Guide | Lancet<br>Depth Stop, Guard<br>(Drill Stop)<br>Ruler | The<br>accessories<br>are optional<br>and may be<br>used to<br>support<br>stereotactic<br>use of the<br>cranial drill<br>bit. |
| Provided<br>Sterile | Yes<br>(Option: Drill Bit, Drill<br>Stop, Drill Stop<br>Wrench) | No | Yes | Equivalent |
| Table 5.3-1: Substantial Equivalence Technical Characteristics | | | | |
| Feature | Cranial Drill Bits and<br>Accessories<br>(Under Review) | Twist Drills<br>(Primary Predicate<br>K062842) | MRI Cranial Drill<br>(Predicate K122456,<br>K160129) | Comment |
| Single patient<br>use,<br>Disposable | Yes<br>(Drill Bit, Drill Stop,<br>Drill Stop Wrench) | Yes | Yes | Equivalent |
| Reusable | Accessory Guide Block<br>and Sleeve Guide | No applicable | No applicable | Qualification<br>of user<br>applied<br>reusable<br>processes |
| Environment<br>of Use | Intraoperative | Intraoperative | Intraoperative | Same |
| Duration of<br>Use | < 1 Minute | < 1 Minutes | < 1 Minutes | Same |
| Patient<br>contact<br>material | Stainless Steel | Stainless Steel | Stainless Steel | Same |
| Drill<br>Diameters | 2.4 mm to 3.2 mm | 0.7 mm to 2.4 mm | 2.0 mm to 3.2 mm,<br>3.2mm to 6.0 mm | Equivalent |
| Sterilization<br>Method<br>(when<br>supplied<br>sterile) | Ethylene Oxide process | Not applicable,<br>Not supplied sterile | Ethylene Oxide process | Same |
{5}------------------------------------------------
#### 5.4 Performance Tests to Demonstrate Substantial Equivalency
To establish the technical equivalency of the cranial drill bits and accessories, evaluations were conducted to confirm compliance with performance requirements, including:
| Test | Summary of Requirement | Result |
|--------------------------------------------------------------------------------------------------|------------------------------------------------------------------|--------|
| Drilling efficiency | Drill through simulated skull material in < 1 min. | Pass |
| Drill bit run-out ("wobble")<br>Test with Guide Block, Drill<br>Sleeve Guide and Drill Stop. | Comparison of drilled hole diameter to the desired<br>hole size. | Pass |
| Drill bit run-out ("wobble")<br>Test with out Guide Block, Drill<br>Sleeve Guide and Drill Stop. | Comparison of drilled hole diameter to the desired<br>hole size. | Pass |
| Mechanical compatibility | Guide Block compatible with stereotactic frame | Pass |
#### ર્ રંડ Conclusion
The cranial drill bits and accessories meet performance requirements equivalent to the predicates. The intended use and technology of the cranial drill bits and accessories are the same as the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.