The Carevature DReal™ is intended to be used with high-speed compatible electric and pneumatic motors. When used with these motors, it is intended to cut bone by drilling, reaming, decorticating, shaping, dissecting, shaving and smoothing for neurosurgical and spinal applications. Specific applications include laminectomy/laminotomy and craniotomy/craniectomy/ skull base cuts < 1 cm3.
Device Story
Handheld, single-use bone-removal device; features rotating cutter with protective shield (hood) and curved distal shaft (45° or 75° angles). Connects to high-speed electric or pneumatic motors via ISO 3964 'micro' adaptor; requires external irrigation. Operated by surgeons in open surgical procedures. Device acts as interface between motor and bone; rotating cutter performs bone removal. Protective hood shields surrounding tissue. Output is physical bone modification; assists in neurosurgical/spinal procedures. Benefits include improved access and visibility via curved tip and tissue protection via hooded design.
Clinical Evidence
Bench testing only. Included biocompatibility (ISO 10993), sterilization validation (ISO 11137), shelf life, and performance testing (simulated use, temperature, slippage, vibration, fatigue). Animal feasibility study (1 pig) and cadaver feasibility study (1 human) confirmed safety and performance for intended spinal applications. No human clinical trial data provided.
Technological Characteristics
Handheld, single-use, powered bone-cutting tool. Materials: Steralloy (USP-6 plastic) handle, medical-grade stainless steel shaft and cutter. Powered by external electric/pneumatic motors (10,000-40,000 rpm). Features: 45°/75° curved distal tip, partially-hooded burr, irrigation port. Sterilization: Gamma irradiation. Connectivity: Mechanical interface to motor via ISO 3964 adaptor.
Indications for Use
Indicated for neurosurgical and spinal applications requiring bone cutting (drilling, reaming, decorticating, shaping, dissecting, shaving, smoothing), specifically laminectomy/laminotomy and craniotomy/craniectomy/skull base cuts < 1 cm3. For use with high-speed electric or pneumatic motors.
Regulatory Classification
Identification
Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.
Predicate Devices
Stryker MIS Attachments and Cutting Accessories (K143540)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. In the center of the seal is an abstract image of a bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W 066-G609 Silver Spring, MD 20993-0002
March 24, 2016
Carevature Medical, Ltd. Ms. Orly Maor Company Consultant 42nd Tvuot Haarets Street Tel Aviv, Israel 6954648
Re: K153519
Trade/Device Name: DReal™ Regulation Number: 21 CFR 882.4310 Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, and Their Accessories Regulatory Class: Class II Product Code: HBE Dated: February 9, 2016 Received: February 11, 2016
Dear Ms. Orly Maor:
This letter corrects our substantially equivalent letter of March 16, 2016.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S/Δ
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K153519
Device Name DReal™M
Indications for Use (Describe)
The Carevature DReal™ is intended to be used with high-speed compatible electric and pneumatic motors. When used with these motors, it is intended to cut bone by drilling, reaming, decorticating, shaping, dissecting, shaving and smoothing for neurosurgical and spinal applications. Specific applications include laminectomy/laminotomy and craniotomy/craniectomy/ skull base cuts < 1 cm3.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Traditional Premarket Notification Submission – 510(k) Summary DReal™ 510(k) Number K153519
### I. SUBMITTER
Carevature Medical Ltd. 42 Tvuot Haarets St. Tel Aviv, Israel 6954648 Tel: +972-8-6757073 Fax: +972-8-6757073
#### Contact Person
Orly Maor Company consultant ra@carevature.com Tel: +972-9-7453607 Fax: +972-8-6757073 42nd Tvuot Haarets St. Tel Aviv, Israel 6954648
Date Prepared: February 26, 2016
### II. DEVICE
Name of Device: DReal™ Common or Usual Name: DReal™ Classification Name: drills, burrs, trephines & accessories (simple, powered) CFR: 21 CFR §882.4310 Regulatory Class: II Product Code: HBE
### III. PREDICATE DEVICE
The predicate device is the Stryker Corporation Stryker MIS Attachments and Cutting Accessories cleared under K143540 (product code HBE, Regulation No. 21 CFR \$882.4310).
### IV. DEVICE DESCRIPTION
The DReal™ is a handheld bone-removal device based on a rotating cutter, with a protective shield (hood) partially covering the cutter; following its indications for use, it is designed to cut bone < 1 cm3.
The DReal™ comprises a handpiece with a plastic grip ("handle") and a central shaft that curves at its distal end ("tip"). The curved tip is equipped with an active, drill-like
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "carevature MEDICAL". The word "carevature" is in a sans-serif font, with the "c" in a darker shade of blue than the rest of the word. Below "carevature" is the word "MEDICAL" in a smaller, sans-serif font. There are two curved lines above and below the word "carevature", both in shades of blue.
cutter that is protected by an extension of the central shaft as a protective shield. The DReal™ has two models, with different angles of curvature at the distal end: Short 45° and Long 75°.
Located in the proximal end of the handle, is the motor adaptor (the motor is not part of the supplied product).
The DReal™ principles of operation are the same as any other bone cutting tool, as well as the predicate device; it uses electrical or pneumatic power to rotate a cutting element that is approximated to the bone requiring cutting.
The DReal™ has an irrigation port on its handle that connects to irrigation source. Irrigation should be running at all times while the DReal™ is active. In case of irrigation malfunction, irrigation can be added externally as with any high-speed, rotating power device, including the predicate.
The DReal™ can be powered by an electric or pneumatic motors using a 'micro' adaptor as per ISO 3964; however, it can rotate only clockwise (right-hand, or Forward).
The following motor system (Class II, Product Code HBE) was validated for use with the DRealTM.
- Stryker Consolidated Operating Room Equipment (CORE) -K112593 ●
## V. INDICATIONS FOR USE
The Carevature DReal™ is intended to be used with high-speed compatible electric and pneumatic motors. When used with these motors, it is intended to cut bone by drilling, reaming, decorticating, shaping, dissecting, shaving and smoothing for neurosurgical and applications. Specific applications include laminectomy/laminotomy and craniotomy/craniectomy/ skull base cuts < 1cm .
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
A substantial equivalence table, which summarizes the similarities and differences between the DReal™ and the predicate device, is provided below.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "carevature MEDICAL". The word "carevature" is in a sans-serif font, with the word "MEDICAL" in a smaller font size below it. There are two curved lines above and below the text, creating a circular shape around the logo. The logo is in a light blue color.
| | Carevature Medical DReal™ | Stryker MIS Attachments<br>and Cutting Accessories | SE<br>JUSTIFICATION |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K153519 | K143540 | — |
| Manufacturer | Carevature Medical Ltd. | Stryker Corporation | — |
| Product Code | HBE | HBE, ERL | Same |
| CFR | §882.4310 | §882.4310, §874.4240 | Same |
| Intended Use | The Carevature DReal™ is<br>intended to be used with high-<br>speed compatible electric and<br>pneumatic motors | The MIS Attachments and<br>Cutting Accessories are<br>intended to be used with the<br>Stryker<br>Consolidated Operating Room<br>Equipment (CORE Console<br>and electric and pneumatic<br>motors | Same-<br>Substantially<br>equivalent |
| | When used with these motors,<br>it is intended to cut bone by<br>drilling, reaming, decorticating,<br>shaping, dissecting, shaving<br>and smoothing for<br>neurosurgical and spinal<br>applications. | When used with these motors,<br>the MIS Attachments and<br>Cutting Accessories are<br>intended to cut bone in the<br>following manner: drilling,<br>reaming, decorticating,<br>shaping, dissecting, shaving,<br>and smoothing for the<br>following medical applications:<br>Neuro; Spine; Ear, Nose, and<br>Throat<br>(ENT)/Otorhinolaryngology;<br>and Endoscopic applications. | Different<br>The DReal™ has<br>limited<br>indications than<br>the Stryker MIS<br>attachments. The<br>specific<br>indications are a<br>subset of already<br>cleared<br>indications for<br>the predicate<br>device. |
| | Specific applications include<br>laminectomy/laminotomy and<br>craniotomy/craniectomy/ skull<br>base cuts < 1cm3. | Specific applications include<br>Craniotomy/Craniectomy,<br>Laminotomy/Laminectomy,<br>Minimally Invasive Surgery<br>(MIS) Spine, Expanded<br>Endonasal Approach<br>(EEA)/Anterior Skull Base/<br>Endoscopic/Transnasal/Transph<br>enoidal, and Orthopedic Spine.<br>These devices are also usable in<br>the preparation for the<br>placement of screws, metal,<br>wires, pins, and other fixation<br>devices. | |
| | Carevature Medical DReal™ | Stryker MIS Attachments<br>and Cutting Accessories | SE<br>JUSTIFICATION |
| Patient<br>Population | General<br>Prescription medical device | General<br>Prescription medical device | Same |
| Function | • The subject Carevature<br>DReal™ models act as an<br>interface between the high speed<br>motors and the cutting<br>accessories<br>• Carevature DReal™ models<br>are intended as a location for the<br>user to hold and grip the device<br>system | • The predicate Stryker MIS<br>Attachments act as an interface<br>between the high speed motors<br>and the cutting accessories<br>• The predicate MIS<br>attachments are intended as a<br>location for the user to hold<br>and grip the device system.<br>MIS Attachments and Cutting<br>Accessories are intended to cut<br>bone and used in the preparation<br>for the placement of screws,<br>metal, wires, pins, and other<br>fixation devices | Same |
| Contraindications | None known | None known | Same |
| Type of operation | Open surgical procedures | Open surgical procedures | Same |
| Conditions for<br>Use | DReal™– Single Use | Attachments – Reusable<br>Cutting accessories – Single use | Same<br>Carevature<br>DReal™ devices<br>are unified<br>attachments and<br>cutting<br>accessories<br>therefore, a single<br>use |
| Sterilization | DReal™– supplied sterile,<br>gamma irradiated | Attachments– sterilized by the<br>user<br>Cutting accessories – supplied<br>sterile, gamma irradiated | Similar<br>Carevature<br>DReal™ devices<br>are unified<br>attachments and<br>cutting<br>accessories<br>therefore,<br>sterilized by the<br>manufacturer |
| | Carevature Medical DReal™ | Stryker MIS Attachments<br>and Cutting Accessories | SE<br>JUSTIFICATION |
| Main Materials | Commonly used medical grade<br>metals and plastics:<br>The handle is made of<br>Steralloy, a USP-6 plastic<br>material The central shaft is made<br>of medical grade stainless<br>steel The cutter is made of<br>medical grade stainless<br>steel | Commonly used medical grade<br>metals and plastics | Same<br>Both the subject<br>and the predicate<br>are comprised of<br>similar materials.<br><br>Different<br>The DReal™ has<br>no color strip;<br>instead, the<br>various models<br>are designated on<br>the single-use<br>outer and inner-<br>packaging. |
| Packaging<br>Configuration | Carevature DReal™ products -<br>double sterile sealed pouches<br>and corresponding cardboard<br>box. | Attachments - Retention insert in<br>a rigid carton<br><br>MIS Cutting Accessories –<br>sealed chevron style pouch<br>sterile barrier system | Similar<br>Carevature<br>DReal™ models<br>are unified<br>attachments and<br>cutting accessories<br>therefore, packed<br>in sterile sealed<br>way |
| Available Style | Curved, Angled (45° and 75°),<br>partially-hooded burr | Straight, Curved, Angled | Similar<br>The curve and<br>angle of the tip<br>are intended to<br>improve its access<br>and visibility<br><br>Different<br>• The DReal™<br>possesses a<br>sharper<br>angulation near<br>the distal end<br>• The DReal™<br>possesses a<br>partially-hooded<br>burr to provide<br>protection for<br>surrounding<br>tissue |
| | Carevature Medical DRealTM | Stryker MIS Attachments<br>and Cutting Accessories | SE<br>JUSTIFICATION |
| Driving motor | The Carevature DRealTM<br>models were validated to be<br>used with the Stryker<br>Consolidated Operating Room<br>Equipment (CORE Console<br>and electric motors) | The MIS Attachments and<br>Cutting Accessories are<br>intended to be used with the<br>Stryker Consolidated Operating<br>Room Equipment (CORE<br>Console and electric or<br>pneumatic motors) | Same |
| Head Style<br>Offering | Match Head | Round, Diamond, Match Head | Similar |
| Diameter/Head<br>Size | 2.5 mm | 1.5 mm - 5.0 mm | Similar |
| Length | Shaft length 11cm | Two lengths 13 and 16 cm | Similar |
| Motor power<br>supply | Electric and Pneumatic | Electric and Pneumatic | Same |
| Speed | 10,000-40,000 rpm | 5,000-75,000 rpm | Similar |
| Pneumatic<br>Pressure<br>Recommendation | 120 psi (pounds per square inch) | 120 psi (pounds per square inch) | Same |
| Source of<br>Activation | Footswitch | Handswitch and Footswitch | Similar |
| Performance | 1. Shelf life testing<br>2. Temperature Testing<br>3. Slippage testing<br>4. Simulated distribution testing<br>transportation validation<br>(Chatter testing)<br>5. Simulated use testing<br>6. Fatigue test<br>7. Motor compatibility<br>8. Excessive vibrations | 1. Shelf life testing<br>2. Temperature Testing<br>3. Slippage testing<br>4. Chatter testing<br>5. Whip testing | Same<br>Whip test was<br>deemed<br>unnecessary due<br>to the product<br>design. Whipping<br>effect is<br>inherently limited<br>to an acceptable<br>value. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for "carevature MEDICAL". The word "carevature" is in a larger, bold font, and the word "MEDICAL" is in a smaller font below it. There are two curved lines above and below the text, creating a circular effect.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Carevature Medical. The logo consists of the word "carevature" in a sans-serif font, with the word "MEDICAL" in a smaller font below it. There are two curved lines above and below the word "carevature", which are blue.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for "carevature MEDICAL". The word "carevature" is in a sans-serif font, with the word "MEDICAL" in a smaller font below it. There are two curved lines above and below the word "carevature", which are blue in color. The logo is simple and modern.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for "carevature MEDICAL". The word "carevature" is in a sans-serif font, with the word "MEDICAL" in smaller letters below it. There are two curved lines above and below the word "carevature", which are blue in color. The logo is simple and modern.
## VII. PERFORMANCE DATA
The following performance data was provided in support of the substantial equivalence determination:
## Biocompatibility testing
An evaluation of biocompatibility was performed in compliance with ISO 10993-1. The biocompatibility tests are listed in the table below:
| Test | Results | Conclusions |
|--------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| GLP Cytotoxicity per<br>ISO 10993-5 | Less than grade 2 (mild reactivity) | Non-cytotoxic |
| ISO- intracutaneous in<br>Rabbits (irritation) per<br>ISO 10993-10 | The difference between each test article<br>extract overall mean score and<br>corresponding control was 0.0 and 0.1<br>for test articles extract | Not irritant |
| Sensitization per ISO<br>10993-10 | The test article showed no evidence of<br>causing delayed dermal contact in the<br>guinea pig | Not considered a<br>sensitizer |
| Systemic toxicity per<br>ISO 10993-11 | No mortality or evidence of systemic<br>toxicity from the extracts injected into<br>mice | Nontoxic |
| Systemic toxicity USP<br>pyrogen Study per ISO<br>10993-11 | The total rise of rabbit temperatures<br>was within acceptable USP limits | Nonpyrogenic |
| LAL test per NAMSA<br>LAL SOP | Less than 0.06 EU/ml | Acceptable |
All tests were successfully completed.
## Sterilization, Packaging and Shelf Life Testing
Sterilization validation testing was performed to demonstrate compliance with ISO 11137-1. In addition, shelf life and packaging testing were performed to support the labeled shelf life. All tests, including packaging integrity and simulated use (performance), were performed after simulated distribution testing (transportation validation) and successfully completed.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for "carevature MEDICAL". The word "carevature" is in a sans-serif font, with the word "MEDICAL" in smaller letters below it. There are two curved lines above and below the text, creating a circular shape around the logo. The colors are shades of blue.
# Performance Testing
Performance testing included the following:
| Test | Test Method Summary | Results |
|----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Simulated use testing | To demonstrate the performance of the DRealTM for its intended use: activation of the DRealTM on<br>a hard-tissue model to show efficacy (material<br>removal), durability and integrity for the removal<br>of material needed with the device' indications in<br>clinical settings. | The device<br>passed the test.<br>The acceptance<br>criteria were met. |
| Temperature Testing | To demonstrate the compliance of the DRealTM to<br>safety requirements derived from tissue necrosis | The device<br>passed the test.<br>The acceptance<br>criteria were met. |
| Slippage testing | To demonstrate the compatibility of the DRealTM<br>to the specified motors | The device<br>passed the test.<br>The acceptance<br>criteria were met. |
| Simulated<br>distribution testing<br>transportation<br>validation (Chatter<br>testing) | To demonstrate the compatibility of the DRealTM<br>to ASTM D4169-09 and ASTM D4332-13 | The device<br>passed the test.<br>The acceptance<br>criteria were met. |
| Fatigue test | To demonstrate the durability of the DRealTM for<br>its intended use.<br>Method: Activation of the DRealTM on a hard-<br>tissue (bone) to show durability and integrity<br>needed with the device' indications in clinical<br>settings. | The device<br>passed the test.<br>The acceptance<br>criteria were met. |
| Motor compatibility | To demonstrate the compatibility of the DRealTM<br>to the specified motors | The device<br>passed the test.<br>The acceptance<br>criteria were met.<br>The motors that<br>are approved<br>appears on the<br>device IFU |
| Excessive Vibration | To demonstrate the compatibility of the DRealTM<br>to the specified motor without malfunction | The device<br>passed the test.<br>The acceptance<br>criteria were met. |
| Animal Feasibility<br>Test | To investigate the safety and performance of the<br>DRealTM for its intended use in the spine. One<br>live adult (7 months) pig was used. Intra-<br>operative performance, gross bone volume<br>reduction, and histopathological sections were | The device<br>passed the test.<br>The acceptance<br>criteria were met. |
| Test | Test Method Summary | Results |
| Cadaver Feasibility<br>Test | To investigate the performance of the DReal™ for its intended use in the spine. One skeletally<br>mature human cadaver was used. Gross bone<br>volume reduction was observed. | The device<br>passed the test.<br>The acceptance<br>criteria were met. |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for "carevature MEDICAL". The word "carevature" is in a sans-serif font, with the "c" in a darker shade of gray than the rest of the word. Below "carevature" is the word "MEDICAL" in a smaller, sans-serif font. Above and below the words is a curved line that is blue on the ends and fades to white in the middle.
Conclusions:
The tests met the predefined acceptance criteria and passed.
## VIII. CONCLUSIONS
The DReal™ was determined to be substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.