Stryker Elite Attachments, Stryker Heavy Duty (HD) Attachments
Applicant
Stryker Corp.
Product Code
HBE · Neurology
Decision Date
Apr 23, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Elite and Heavy Duty Attachments are intended to be used with the Stryker Consolidated Operating Room Equipment (CORE) console and electric and pneumatic motors. When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are intended to cut bone cement in the following manner: drilling, reaming, decorticating, shaping, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose and Throat (ENT)/Otology/Neurotology/Otothinolaryngology; Craniofacial and Maxillofacial; Dental; and Endoscopic applications. The specific applications include Craniectomy, Laminotomy/Laminectomy, Minimally Invasive Surgery (MIS) Spine. Expanded Endonasal (EEA)/Anterior Skull Base/Endoscopic/Transnasal/Transphenoidal, and Orthopedic Spine. When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are intended to cut teeth in the following manner: sectioning, splitting, fragmenting, extracting, removing, drilling, and reaming. When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are also usable in the preparation for the placement of screws, metal, wires, pins, and other fixation devices, or the cutting of screws, metal, wires, pins, and other fixation devices in the following manner: sectioning, smoothing or shaping of metal, and removing/rounding sharp edges.
Device Story
Stryker Elite and HD Attachments serve as mechanical interfaces between powered surgical motors (electric/pneumatic) and cutting accessories (burs). Attachments transmit torque from motor to bur to cut bone, bone cement, teeth, or fixation hardware (screws/wires). Used in OR settings by surgeons; attachments provide grip surface for manual control. Features include stainless steel construction, color-coded bands for compatibility identification, and varied lengths (7-20cm). Reusable; requires end-user steam sterilization. Design modifications include sleeker profiles and grooved/knurled textures for improved balance and grip. Output is mechanical cutting action; clinical benefit is precise bone/tissue removal or hardware manipulation during complex surgical procedures.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility (ISO 10993-1) confirmed non-cytotoxic, non-sensitizing, non-toxic, and non-pyrogenic. Functional testing verified 1-year life, locking mechanism reliability, surface temperature limits, bur whip/chatter performance, and package integrity (ASTM D4169).
Technological Characteristics
Materials: Stainless steel. Principle: Rotary torque transmission. Power: Electric/pneumatic motors. Form factor: Straight/angled attachments, 7-20cm lengths. Connectivity: Mechanical interface to CORE console/motors. Sterilization: Steam (moist heat), SAL 10^-6. Standards: ISO 10993-1 (biocompatibility), ASTM D4169 (packaging).
Indications for Use
Indicated for patients requiring bone, bone cement, or tooth cutting (drilling, reaming, decorticating, shaping, smoothing, sectioning, splitting, fragmenting, extracting) or fixation device preparation/cutting in neuro, spine, ENT/otology, craniofacial, maxillofacial, dental, and endoscopic procedures. Includes specific applications: craniectomy, laminotomy/laminectomy, MIS spine, EEA/anterior skull base, and orthopedic spine.
Regulatory Classification
Identification
Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.
Predicate Devices
Stryker Consolidated Operating Room Equipment (CORE) System (K112593)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized eagle or bird in flight, composed of three overlapping profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 23, 2015
Stryker Instruments Ms. Brittney M. Larsen, RAC Senior Regulatory Affairs Representative 4100 E. Milham Ave. Kalamazoo, Michigan 49001
Re: K143320 Trade/Device Name: Stryker Elite and Heavy Duty (HD) Attachments Regulation Number: 21 CFR 882.4310 Regulation Name: Powered Simple Cranial Drills, Burrs, Trephines, and Their Accessories Regulatory Class: Class II Product Code: HBE, HBB, ERL, DZI Dated: March 20, 2015 Received: March 24, 2015
Dear Ms. Larsen,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143320
Device Name Stryker Elite and Heavy Duty (HD) Attachments
#### Indications for Use (Describe)
The Elite and Heavy Duty Attachments are intended to be used with the Stryker Consolidated Operating Room Equipment (CORE) console and electric and pneumatic motors. When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are intended to cut bone cement in the following manner: drilling, reaming, decorticating, shaping, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose and Throat (ENT)/Otology/Neurotology/Otothinolaryngology; Craniofacial and Maxillofacial; Dental; and Endoscopic applications.
The specific applications include Craniectomy, Laminotomy/Laminectomy, Minimally Invasive Surgery (MIS) Spine. Expanded Endonasal (EEA)/Anterior Skull Base/Endoscopic/Transnasal/Transphenoidal, and Orthopedic Spine.
When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are intended to cut teeth in the following manner: sectioning, splitting, fragmenting, extracting, removing, drilling, and reaming.
When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are also usable in the preparation for the placement of screws, metal, wires, pins, and other fixation devices, or the cutting of screws, metal, wires, pins, and other fixation devices in the following manner: sectioning, smoothing or shaping of metal, and removing/rounding sharp edges.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| Contact Details | | | |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|------------------------|
| 510(k) Owner | Stryker Instruments<br>4100 E. Milham Avenue<br>Kalamazoo, Michigan 49001 USA<br>Ph: +1-269-323-5293<br>Fax: +1-269-324-5412 | | |
| FDA Establishment<br>Registration No. | 1811755 | | |
| Contact Person | Brittney M. Larsen, RAC<br>Ph: +1-269-389-5293<br>Fax: +1-269-389-5412<br>Brittney.Larsen@Stryker.com | | |
| Date Submitted | April 22, 2015 | | |
| Device Name | | | |
| Trade Name | Stryker® Elite and Heavy Duty (HD) Attachments | | |
| Common Name | Surgical Drill Handpieces | | |
| Classification | II | | |
| Primary<br>Classification Name | Drills, Burs, Trephines & Accessories (Simple, Powered)<br>(21 CFR 882.4310, Product code HBE) | | |
| Secondary<br>Classification Name | Drill, Surgical ENT (Electric or Pneumatic) including Handpiece<br>(21 CFR 874.4250, Product code ERL)<br><br>Pneumatic cranial drill motor<br>(21 CFR 882.4370, Product code HBB)<br><br>Drill, Bone, Powered<br>(21 CFR 872.4120, Product code DZI) | | |
| Legally Marketed Predicate Device(s) | | | |
| 510(k) Number | Product Code | Trade Name | Manufacturer |
| Primary Predicate | | | |
| K112593 | ERL (primary)<br>DZI, DZJ, HBE<br>(secondary) | Stryker® Consolidated<br>Operating Room Equipment<br>(CORE) System | Stryker<br>Instruments |
| Secondary Predicate | | | |
| K041754 | HBB (primary)<br>ERL, HBE<br>(secondary) | Stryker Maestro Pneumatic<br>System | Stryker<br>Instruments |
{5}------------------------------------------------
stryker®
# Purpose of this Traditional 510(k) Premarket Notification
Stryker submits this Traditional 510(k), for the Stryker® Elite and HD Attachments, to request clearance for an expansion of indications and addition of color bands to the predicate devices. The predicate devices are currently cleared for a variety of general indications for use. The subject changes include adding the following specific procedures (hereafter referred to as medical applications): Craniectomy, Laminotomy/Laminectomy, Minimally Invasive Surgery (MIS) Spine, Expanded Endonasal Approach (EEA)/Anterior Skull Base/Transnasal/Transphenoidal, and Orthopedic Spine. This expansion of indications does not change the intended use or the fundamental scientific technology of the predicate devices. The subject modifications of this submission also include the addition of color bands to enhance the distinction of attachment and cutting accessory compatibility.
### Device Description and Intended Use
The Stryker® Elite and HD Attachments are used within a system consisting of a variety of devices, including a console, powered motors, and cutting accessories. The attachments connect to the motors and the cutting accessories to complete the system for physician use. The Stryker® Elite and HD Attachments are offered for prescription use only. The Elite and HD Attachments are intended to serve as interfaces between powered motors and cutting accessories for the purposes of:
- Cutting bone, bone cement, and teeth; ●
- Placing or cutting screws, metal, wires, pins, and other fixation devices; and ●
- Providing a location for the user to hold and grip the device system. ●
The Elite and HD Attachments are provided in straight and angled configurations. The subject Elite Attachments are offered in the following lengths: 7cm, 12cm, 14cm, 17cm, and 20cm. The Stryker® HD Attachments are offered in the following lengths: 9cm and 14cm.
The Elite and HD Attachments are powered by, and compatible with, the Stryker electric and pneumatic motors. The attachments are also used with the cutting accessories (burs). All the attachments demonstrate a rotary mode of action by transmitting torque.
The Elite and HD Attachments are made of stainless steel (SST). The attachments display a color band on the outer surface. The color bands serve to enhance the distinction of attachment and cutting accessory compatibility. The Elite and HD Attachment outer profiles feature a sleeker design. The Elite and HD Attachment outer surface textures include laterally grooved and continuous knurling textures for improved user grip.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase, and there is a registered trademark symbol to the right of the "r". The word is likely a logo for the Stryker Corporation, a medical technology company.
The Elite and HD Attachments are reusable, provided non-sterile, and end user sterilized. The attachments may be cleaned by manual or mechanical (automated) cleaning procedures. The attachment end user sterilization method is moist heat (steam) and the sterility assurance level (SAL) is 10-6. The Elite and HD Attachments have an expected life of one (1) year. The attachments are individually packaged in a sealed retention insert. The attachment labels contain a label specified part description, quantity, lot number, and contact information. There are no known contraindications for the Elite and HD Attachments.
# Indications for Use
The Elite and Heavy Duty Attachments are intended to be used with the Stryker Consolidated Operating Room Equipment (CORE®) console and electric and pneumatic motors. When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are intended to cut bone and bone cement in the following manner: drilling, reaming, decorticating, shaping, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose and Throat (ENT)/Otology/Neurotology/Otorhinolaryngology; Cramiofacial and Maxillofacial; Dental; and Endoscopic applications.
specific applications include Craniotomy/Craniectomy, Laminotomy/ The Laminectomy, Minimally Invasive Surgery (MIS) Spine, Expanded Endonasal Approach (EEA)/Anterior Skull Base/Endoscopic/Transnasal/Transphenoidal, and Orthopedic Spine.
When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are intended to cut teeth in the following manner: sectioning, segmenting, splitting, fragmenting, extracting, removing, drilling, and reaming.
When used with these motors, the Elite and Heavy Duty Attachments and Cutting Accessories are also usable in the preparation for the placement of screws, metal, wires, pins, and other fixation devices, or the cutting of screws, metal, wires, pins, and other fixation devices in the following manner: sectioning, deburring, smoothing or shaping of metal, and removing/rounding sharp edges.
{7}------------------------------------------------
stryker®
#### Table 6-1. Comparison to predicate devices
| Description | Stryker® CORE System<br>(Primary Predicate - K112593) | Stryker Maestro Pneumatic<br>System<br>(Secondary Predicate -<br>K041754) | Stryker® Elite and HD Attachments<br>(Subject) | Explanation<br>of Differences |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | Class II | Class II | Class II | Identical |
| Primary<br>Product Code | ERL<br>Drill, Surgical, ENT (Electric or<br>Pneumatic) including Handpiece | HBB<br>Pneumatic cranial drill motor | HBE<br>Drills, Burs, Trephines & Accessories (Simple,<br>Powered) | Equivalent |
| Primary<br>Regulation | 21 CFR 874.4250<br>Ear, nose, and throat electric or<br>pneumatic surgical drill | 21 CFR 882.4370<br>Pneumatic cranial drill motor | 21 CFR 882.4310<br>Powered simple cranial drills, burrs, trephines, and<br>their accessories. | Equivalent |
| Product<br>Types | Straight and angled attachments | Straight and angled attachments | Straight and angled attachments | Identical |
| Conditions<br>for Use | Reusable | Reusable | Reusable | Identical |
| Mechanism of<br>Action | Powered by electric motors | Powered by pneumatic motors | Powered by electric and pneumatic motors | Equivalent |
| Mode of<br>Action | Rotary (transmits torque) | Rotary (transmits torque) | Rotary (transmits torque) | Identical |
| Expected Life | 1 year | 1 year | 1 year | Identical |
| Type of Use | Prescription use only | Prescription use only | Prescription use only | Identical |
| Intended Use | To serve as an interface between powered motors and cutting<br>accessories for the purposes of:<br>• Cutting bone, bone cement, and teeth;<br>• Placing or cutting screws, metal, wires, pins, and other<br>fixation devices;<br>• Providing a location for the user to hold and grip device<br>system. | To serve as an interface between powered motors and cutting accessories for the purposes of:<br>• Cutting bone, bone cement, and teeth;<br>• Placing or cutting screws, metal, wires, pins, and other fixation devices;<br>• Providing a location for the user to hold and grip device system. | To serve as an interface between powered motors and cutting accessories for the purposes of:<br>• Cutting bone, bone cement, and teeth;<br>• Placing or cutting screws, metal, wires,<br>pins, and other fixation devices:<br>• Providing a location for the user to hold and grip device system. | Identical |
| Descript-<br>ion | Stryker® CORE<br>System<br>(Primary Predicate<br>- K112593) | Stryker Maestro<br>Pneumatic System<br>(Secondary<br>Predicate -<br>K041754) | Stryker® Elite and HD Attachments<br>(Subject) | Explanation of<br>Differences |
| Indications<br>for Use | The Stryker®<br>Consolidated<br>Operating Room<br>Equipment (CORE)<br>System is intended<br>for use in the<br>cutting, drilling,<br>reaming,<br>decorticating,<br>shaping, and<br>smoothing of bone,<br>bone cement and<br>teeth in a variety of<br>surgical procedures,<br>including but not<br>limited to, dental,<br>ENT (ear, nose,<br>throat), neuro, spine,<br>and endoscopic<br>applications. It is<br>also usable in the<br>placement or cutting<br>of screws, metal,<br>wires, pins, and<br>other fixation<br>devices. | The Stryker Maestro<br>Pneumatic system is<br>a pneumatically<br>operated surgical<br>instrument system.<br>The pneumatic<br>motor provides<br>power to operate<br>removable rotating<br>surgical cutting<br>tools and their<br>accessories intended<br>for use in<br>neurosurgery,<br>including<br>craniotomy and<br>spinal surgery; as<br>well as Ear, Nose<br>and Throat (ENT),<br>orthopedic, and<br>general surgical<br>applications<br>including<br>maxillofacial,<br>craniofacial and<br>sternotomy<br>surgeries. | The Elite and Heavy Duty Attachments are intended to be used with<br>the Stryker Consolidated Operating Room Equipment (CORE®)<br>console and electric and pneumatic motors. When used with these<br>motors, the Elite and Heavy Duty Attachments and Cutting<br>Accessories are intended to cut bone and bone cement in the<br>following manner: drilling, reaming, decorticating, shaping, and<br>smoothing for the following medical applications: Neuro; Spine; Ear,<br>Nose and Throat (ENT)/Otology/ Neurotology/Otorhinolaryngology:<br>Craniofacial and Maxillofacial; Dental; and Endoscopic applications.<br>The specific applications include Craniotomy/Craniectomy,<br>Laminotomy/Laminectomy, Minimally Invasive Surgery (MIS)<br>Spine, Expanded Endonasal Approach (EEA)/Anterior Skull<br>Base/Endoscopic/Transnasal/Transphenoidal, and Orthopedic Spine.<br>When used with these motors, the Elite and Heavy Duty Attachments<br>and Cutting Accessories are intended to cut teeth in the following<br>manner: sectioning, segmenting, splitting, fragmenting, extracting,<br>removing, drilling, and reaming.<br>When used with these motors, the Elite and Heavy Duty Attachments<br>and Cutting Accessories are also usable in the preparation for the<br>placement of screws, metal, wires, pins, and other fixation devices, or<br>the cutting of screws, metal, wires, pins, and other fixation devices in<br>the following manner: sectioning, deburring, smoothing or shaping of<br>metal, and removing/rounding sharp edges. | The subject devices add the<br>following indications for<br>use to the existing, cleared<br>indications: Craniectomy,<br>Laminotomy/Laminectomy,<br>Minimally Invasive<br>Surgery (MIS) Spine,<br>Expanded Endonasal<br>Approach (EEA)/Anterior<br>Skull Base/Transnasal/<br>Transphenoidal, and<br>Orthopedic Spine. The<br>subject indications add<br>verbiage that defines<br>cutting for each medium in<br>the variety of surgical<br>procedures listed. The<br>purpose of the changes<br>made to the indications for<br>use is to list the procedures<br>within each general<br>indication as per current<br>standard of care as known<br>by the medical community.<br>Verification and validation<br>testing conducted has<br>demonstrated that the<br>subject devices have a<br>similar safety and<br>effectiveness profile as the<br>legally marketed predicate<br>devices |
| Description | Stryker® CORE System<br>(Primary Predicate - K112593); and<br>Stryker Maestro Pneumatic System<br>(Secondary Predicate - K041754) | Stryker® Elite and HD<br>Attachments<br>(Subject) | Explanation of Differences | |
| Size Information | | | | |
| Straight and Angled<br>Attachment Length<br>Offerings | 7cm, 9cm, 12cm, 14cm, 17cm (straight<br>only), 20cm | 7cm, 9cm, 12cm, 14cm, 17cm,<br>20cm | Added the Elite 14cm Hooded and the Elite 17cm Angled<br>Attachment offerings for marketing of a more competitive product<br>offering. Both of the added attachments are within the cleared<br>length range. The 14cm hooded attachment includes a built in<br>hood. Verification and validation testing conducted<br>has<br>demonstrated that the subject devices have a similar safety and<br>effectiveness profile as the legally marketed predicate devices. | |
| Attachment Rear<br>Housing Outer<br>Diameters | 0.677in (nominal) | 0.677in (nominal) | Identical…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.