K130755 · Adeor Medical AG · HBE · Jun 27, 2014 · Neurology
Device Facts
Record ID
K130755
Device Name
HICUT HIGHSPEED INSTRUMENT
Applicant
Adeor Medical AG
Product Code
HBE · Neurology
Decision Date
Jun 27, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HiCut® Highspeed Instruments are intended to be single-used with a compatible surgical drill system for surgical bone work.
Device Story
HiCut™ Highspeed Instrument is a sterile, single-use surgical bur; designed for attachment to pneumatic or electric surgical motor drill systems. Device features various tip designs and lengths for cutting, shaping, or removing bone and soft tissue. Used by surgeons in clinical settings (OR) across multiple specialties including orthopedics, neurosurgery, and ENT. Operates as a mechanical cutting tool driven by external motor systems; provides physical tissue modification to facilitate surgical procedures. Benefits include precise bone/tissue removal during complex surgeries.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993), sterility validation (ISO 11137), packaging integrity (ISO 11607), and mechanical parameter comparison of geometrical conditions and rear-end design to ensure secure fit in drill chucks.
Technological Characteristics
Materials: surgical steel, optional diamond coating. Energy source: external pneumatic or electric motor drill. Form factor: single-use burs with various tip designs/lengths. Sterilization: radiation (ISO 11137). Connectivity: mechanical interface for drill chucks.
Indications for Use
Indicated for trephination, incision, cutting, removal, shaping, and sawing of soft and hard tissue, bone, and biomaterials in orthopedic, neurological, spinal, ENT, plastic, maxillofacial, arthroscopic, sternotomy, and general surgery.
Regulatory Classification
Identification
Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.
{0}------------------------------------------------
JUN 2 7 2014
Page 1 of 3
K130755
## Traditional 510(k) Summary
| Submitter: | adeor Medical AG<br>Kirchplatz 1<br>82049 Pullach<br>Germany |
|-------------------------|--------------------------------------------------------------|
| Contact Person: | Fabio von Zeppelin<br>COO |
| | Phone: +49 (0)89-744 42 398<br>Fax: +49 (0)89-744 24 809 |
| Preparation<br>Date: | May 9th, 2014 |
| Trade Name: | HiCut™ Highspeed Instrument |
| Common Name: | Bur |
| Classification<br>Name: | Neurological surgical devices |
#### Device Description:
The HiCut™ Highspeed Instrument is a sterile packed, single-use cutting device with numerous tip designs and various lengths. The rear end is designed to fit in various commonly marketed surgical motor drill systems. The tip designs are similar to the commonly used tip designs.
The HiCut™ Highspeed Instrument is classified Class II (USA) (Class IIa in EU) per CFR, Part 882.4310. The HiCut™ Highspeed Instruments are simple powered burs to be inserted in pneumatic or electric powered motor drill devices.
#### Indications for Use:
HiCut™ Highspeed Instruments are surgical burs indicated for trephination, incision, cutting, removal, shaping, sawing of soft and hard tissue, bone and biomaterials in Orthopedic Surgery, Neurological and Spinal Surgery, ENT Surgery, Plastic and Maxillofacial Surgery, Arthroscopy, Sternotomy and General Surgery,
#### Intended Use:
The HiCut® Highspeed Instruments are intended to be single-used with a compatible surgical drill system for surgical bone work.
{1}------------------------------------------------
## Predicate Devices:
| K113476 | Drills, Burrs, Trephines &<br>Accessories (Simple, Powered)<br>(HBE) | The Anspach<br>Effort, Inc. |
|--------------------|----------------------------------------------------------------------|-----------------------------------------------------|
| K922299<br>K013091 | Drills, Burrs, Trephines &<br>Accessories (Simple, Powered)<br>(HBE) | Von Zeppelin<br>Chirurgische<br>Instrumente<br>GmbH |
| K081475 | Drills, Burrs, Trephines &<br>Accessories (Simple, Powered)<br>(HBE) | Medtronic<br>Xomed, Inc. |
## Substantial Equivalence:
The HiCut™ Highspeed Instrument is substantially equivalent to the predicate devices. The equivalence is based on the device similarity to the predicate devices in indications of use, functionality and technological characteristics. The predicate devices K113476 and K081475 and the subject device are marketed in the European Union as substantially equivalent.
## Technological Characteristics Comparison:
The technological characteristics of the subject device are based on the same HiCut™ Highspeed Instrument technology as the predicate devices. The material of both the HiCut™ Highspeed Instruments and of the predicate devices is surgical steel and some have a layer of diamonds. Similar to the other legally marketed burs used for surgical bone trephination (predicate devices) the Hicut™ devices are sterilized, individually packed devices and are used in conjunction with a surgical motor drive either pneumatic or electric.
#### Performance data testing:
This submissions includes the following non-clinical performance testing:
- 1) Biocompatibility: evaluated in compliance with ISO 10993.
Conclusion: Based on the selected materials and the provided evaluation, sufficient biocompatibility can be assumed.
- 2) Sterility: The method of radiation complies with the norms ISO 11137-1. 2 and 3. Conclusion: The sterilization process is validated and ensures the sterility of the device throughout the shelf life and transport.
{2}------------------------------------------------
K130755 3 Page 3 of
The following sterility and shelf life testing is included in the submission:
## Procedure/Criteria
Validation and Revalidation of the Radiation Sterilization Method
Establishment of the sterilization dose
Sterility assurance level (SAL)
Testing to ensure that packaging maintains device sterility (tested: integrity of the sealed seam according to ISO 11607) and is not adversely affected by aging (10-year of artificial aging test in accordance with ASTM F 1980) or transport (drop and vibration test based on ISO13355 and ASTM D 5276-98). The compliance with these criteria is ensured for at least the duration of shelf life.
- 3) Parameter comparison: Substantial equivalence to the predicate devices is based on the comparison of geometrical conditions and their critical values. Samples for each system of both the subject and predicate devices were compared. As a result, it can be stated that the subject devices have an equivalent rear end design, which ensures a secure fit into the corresponding drill chucks. Further qualification criteria of the subject device such as materials, design and mechanical constitution exceed the same results as the predicate devices. Thus, the substantial equivalence and therefore, the safety and effectiveness of the subject device can be established.
Considering the conclusions of the non-clinical tests, it can be stated that the subject device is as safe, as effective, and performs as well as the predicate devices.
#### Conclusions
Based upon the comparison of their functionality, intended use and technological characteristics to the predicate devices and the conclusions drawn from the non-clinical performance testing, the Highspeed Instruments proved to be substantially equivalent. The subject device described in this submission performs as intended and raises no new safety or effectiveness issues.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of human services, featuring three curved lines.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### June 27, 2014
Adeor Medical AG Mr. Fabio von Zeppelin Chief Operating Officer Kirchplatz 1 Pullach, 82049 Germany
Re: K130755
Trade/Device Name: HiCut™ Highspeed Instrument Regulation Number: 21 CFR 882.4310 Regulation Name: Powered simple cranial drills, burrs, trephines, and their accessories Regulatory Class: Class II Product Code: HBE Dated: May 9, 2014 Received: May 27, 2014
Dear Mr. Zeppelin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Mr. Fabio von Zeppelin
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Binita S. Ashar -S. 2014.06.27 10:50:49 -04'00'
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K130755
Device Name HiCutTM Highspeed Instrument
#### Indications for Use (Describe)
HiCut™ Highspeed Instruments are surgical burs indicated for trephination, incision, cutting, removal, shaping, sawing of soft and hard tissue, bone, and biomaterials in Orthopedic Surgery, Neurological and Spinal Surgery, Plastic and Maxillofacial Surgery, Arthroscopy, Sternotomy and General Surgery.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED,
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Joshua C. N lipper -S
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.