TRITON ELECTRIC HIGH TORQUE HANDPIECE AND SOFTWARE MODULE
Applicant
Medtronic Surgical Technologies
Product Code
HBE · Neurology
Decision Date
Aug 27, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IPC System with the TRITON Electric High-Torque Handpiece is indicated for the incision / cutting, removal, drilling, and sawing of soft and hard tissue and biomaterials in Neurosurgical (Cranial, Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures.
Device Story
TRITON Electric High Torque Handpiece is a line extension for the Integrated Power Console (IPC) System. Device functions as an electric-powered surgical instrument; motor rotates to drive interchangeable attachments (saws, drills, drivers) for bone/tissue manipulation. Operated by surgeons in clinical settings (OR). IPC console provides power and control; software module enables recognition and control of handpiece position, torque, and direction. Output is mechanical rotation of attachments. Benefits include precise tissue/bone removal and screw/wire placement during spinal, cranial, and orthopedic procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Laboratory bench testing demonstrated performance characteristics substantially equivalent to predicate devices.
Technological Characteristics
Electric-powered handpiece; connects to IPC console. Features interchangeable attachments for drilling, sawing, and driving. Software-controlled torque, position, and direction. Duty cycle: 20s on/20s off; max torque 3.1 Nm (with attachment). Thermal safety: max surface temperature <51°C (<1 min) and <48°C (<10 min).
Indications for Use
Indicated for incision, cutting, removal, drilling, and sawing of soft/hard tissue, bone, and biomaterials in neurosurgical (cranial, craniofacial), orthopedic, arthroscopic, spinal, sternotomy, and general surgical procedures.
Regulatory Classification
Identification
Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.
Predicate Devices
XPS 4000 System, Midas Rex Legend EHS System, Integrated Power Console (IPC) (K081475)
XPS 4000 System, Midas Rex Legend EHS System, Integrated Power Console (IPC) (K081277)
{0}------------------------------------------------
K12/264
# 5. 510(k) Summary
# AUG 2 7 2012
This summary of safety and effectiveness is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| Address of Manufacturer: | Medtronic Powered Surgical Solutions<br>4620 North Beach Street<br>Fort Worth, TX 76137<br>(817) 788-6400 Phone |
|--------------------------|-----------------------------------------------------------------------------------------------------------------|
|--------------------------|-----------------------------------------------------------------------------------------------------------------|
(817) 788-6222 Facsimile
Establishment Registration Number: 1625507
Contact Person:
Date:
Trade or Proprietary Name:
Jeffrey Henderson VP, Quality, Clinical & RA
August 10, 2012
TRITON Electric High Torque Handpiece and Software Module
IPC System with the TRITON High Torque Handpiece
Common usual or Classification Name:
21 CFR 882.4360, Product Code HBC, Class II - Electric cranial drill motor
21 CFR 882.4310, Product Code HBE, Class II - Powered simple cranial drills, burrs, trephines, and their accessories
21 CFR 888.1100, Product Code HRX, Class II - Arthroscope
21CFR 878.4820, Product Code HWE, Class I - Surgical instrument motors and accessories/attachments
### Description:
The purpose of this submission is to add a new handpiece and an accompanying software module as a line extension to the current Integrated Power Console IPC® System. The handpiece is powered by the previously cleared integrated Powered Console. In order to operate the TRITON Electric Handpiece, the IPC will be equipped with the appropriate software module to drive the handpiece. The handpiece is used for sawing, drilling, driving and placing screws, wires and pins when used in conjunction with various attachments previously cleared for use during spinal surgery.
{1}------------------------------------------------
The TRITON Electric High Torque Handpiece has the identical attachment connecting interfacing components as the handpiece cleared under K870157. As a result, the TRITON Electric High Torque Handpiece utilizes the same Attachments and Blades as those cleared under K870157.
# Indications for Use:
The TRITON Electric High-Torque Handpiece is indicated for the incision / cutting, removal, drilling, and sawing of soft and hard tissue and bone, and biomaterials in Neurosurgical (Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures.
# Predicate Device Identification:
The Medtronic TRITON Electric High Torque Handpiece and Software Module is substantially equivalent to the following predicate devices:
| Device | Manufacturer | 510(k) Number | Clearance Date |
|------------------------------------------------------------------------------------|-----------------------|---------------|----------------|
| XPS 4000 System, Midas Rex<br>Legend EHS System, Integrated<br>Power Console (IPC) | Medtronic Xomed, Inc. | K081475 | 10/17/2008 |
| XPS 4000 System, Midas Rex<br>Legend EHS System, Integrated<br>Power Console (IPC) | Medtronic Xomed, Inc. | K081277 | 09/05/2008 |
| IPC® POWEREASETM System<br>510(k) K111520 | Medtronic Xomed, Inc. | K111520 | 10/26/2011 |
#### Table 1: Predicate Devices for the IPC System with TRITON High Torque Handpiece
### Comparison to Predicate Device:
The TRITON Electric High Torque Handpiece and Software Module are similar in device design, function, intended use and fundamental scientific technology to the previously cleared devices listed above.
Table 2: Comparison of Subject Device to Predicate Device
| Feature | Proposed IPC TM System w/ TRITON<br>High Torque Handpiece | Predicate Devices |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Integrated Powered Console System w/<br>TRITON Handpiece<br>The IPC® system is indicated for the<br>incision / cutting, removal, drilling, and<br>sawing of soft and hard tissue and<br>bone, and biomaterials in<br>Neurosurgical (Cranial, Craniofacial)<br>Orthopedic, Arthroscopic, Spinal,<br>Sternotomy, and General surgical<br>procedures. The IPC® provides power<br>to various drills and saws to drive | Integrated Powered Console - K081475<br>The IPC® system is indicated for the<br>incision / cutting, removal, drilling, and<br>sawing of soft and hard tissue and bone,<br>and biomaterials in Neurosurgical<br>(Cranial, Craniofacial) Orthopedic,<br>Arthroscopic, Spinal, Sternotomy, and<br>General surgical procedures. The IPC®<br>provides power to various drills and saws<br>to drive attachments and tools during<br>surgical procedures. |
| Feature | Proposed IPC™ System w/ TRITON<br>High Torque Handpiece | Predicate Devices |
| | attachments and tools during surgical<br>procedures. | |
| Indication for<br>Use | Integrated Powered Console w/<br>TRITON Handpiece<br>The IPC® is indicated for the incision /<br>cutting, removal, drilling, and sawing of<br>soft and hard tissue and bone, and<br>biomaterials in Neurosurgical (Cranial,<br>Craniofacial) Orthopedic, Arthroscopic,<br>Spinal, Sternotomy, and General<br>surgical procedures | Integrated Powered Console - K081475<br>The IPC® is indicated for the incision /<br>cutting, removal, drilling, and sawing of<br>soft and hard tissue and bone, and<br>biomaterials in Neurosurgical (Cranial,<br>Craniofacial) Orthopedic, Arthroscopic,<br>Spinal, Sternotomy, and General surgical<br>procedures |
| Operating<br>Principle | Integrated Powered Console w/<br>TRITON Handpiece<br>The IPC® System is an electric<br>powered instrument system consisting<br>of multiple handpieces which are<br>powered by the IPC® console. The<br>handpieces are intended to drive<br>various types of interchangeable<br>attachments and instruments used in<br>various surgical procedures. When<br>active, the motor of the handpiece<br>rotates to drive the connected<br>attachment. | Integrated Powered Console - K081475<br>The IPC® System is an electric powered<br>instrument system consisting of multiple<br>handpieces which are powered by the<br>IPC® console. The handpieces are<br>intended to drive various types of<br>interchangeable attachments and<br>instruments used in various surgical<br>procedures. When active, the motor of the<br>handpiece rotates to drive the connected<br>attachment. |
| Software | Integrated Powered Console w/<br>TRITON Handpiece<br>The IPC software is designed to<br>recognize and control position, torque<br>and direction of the following<br>handpieces, POWEREASE Driver,<br>Stylus, Stylus Touch, EHS, Visao,<br>Skeeter and Indigo drills, the<br>Oscillating, Reciprocating and Sagittal<br>saws, and Triton driver and the M4,<br>SC1 microdebriders and the TRITON<br>High Torque Handpiece. | Integrated Powered Console - K081475<br>The IPC software is designed to<br>recognize and control position, torque and<br>direction of the following handpieces,<br>Stylus, Stylus Touch, EHS, Visao,<br>Skeeter and Indigo drills, the Oscillating,<br>Reciprocating and Sagittal saws, and<br>Triton driver and the M4 and SC1<br>microdebriders |
{2}------------------------------------------------
:
{3}------------------------------------------------
# Handpiece Comparison
| Metric | IPC® with TRITON Electric High Torque<br>Handpiece | | Predicate Device |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Image: Free Run | Image: With<br>Attachment | |
| Maximum<br>Applied Torque | 1.6 Nm | 3.1 Nm(1) | 7 Nm |
| Range of<br>Speed | 0 to max free run speed of<br>1250 - 1800 rpm | 0 - 250 rpm | 0 - 250 rpm |
| Range of Safe<br>Working<br>Temperatures | Max temperature through the entire operating<br>range at any external surface location on<br>handpiece shall be less than 51 °C for less than<br>1 minute and shall not exceed 48 °C for less<br>than 10 minutes | | 48°C max |
| Duty Cycle | Cycle Time: 20 seconds on maximum / 20<br>seconds off minimum<br>Maximum number of cycles before resting<br>handpiece: 6<br>Maximum number of cycles before resting<br>attachment: 3<br>Minimum rest period: 25 minutes | | Continuous Operation |
| Intended Use | The TRITON Electric High Torque Handpiece is<br>intended to remove hard and soft tissue, drill<br>pilot holes, drive screws, wires, and pins<br>during, but not limited to, spinal, cranial, and<br>small bone surgical procedures. The TRITON<br>Electric High Torque Handpiece is intended to<br>be used in both open and minimally invasive<br>surgical procedures. | | The IPC® POWEREASE™ System<br>is indicated for drilling, tapping, and<br>driving screws and working end<br>attachments during spinal surgery,<br>including open and minimally<br>invasive procedures. It is also used<br>in the placement of screws, or<br>cutting of screws, posts, and rods. |
ក្រុង
{4}------------------------------------------------
# Testing:
Laboratory bench testing conducted on the IPC System with the TRITON Electric High Torque Handpiece demonstrates substantially equivalent performance characteristics to the predicate devices currently on the market.
# Conclusion/Summary:
Based upon the laboratory bench test summaries, intended use, and the successful completion of design control activities; the IPC System with the TRITON Electric High Torque Handpiece and software module is shown to be substantially equivalent to currently marketed predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
AUG 27 2012
Medtronic Surgical Technologies c/o Mr. Jeffrey Henerson VP, Quality, Clinical & RA 4620 North Beach Street Fort Worth, TX 76137
Re: K121264
Trade/Device Name: TRITON Electric High Torque Handpiece and Software Module Regulation Number: 21 CFR 882.4360 Regulation Name: Electric Cranial Drill Motor Regulatory Class: Class II Product Code: HBC, HBE, HRX, HWE Dated: August 10, 2012 Received: August 13, 2012
Dear Mr. Henderson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2 - Mr. Jeffrey Henerson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
ர்ட
<http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm>.
Sincerely vours.
Kesia Alexander
Malvina B. Evdelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# 4. Indications for Use Statement
K121264 510(k) Number (if known):
Device Name: __ TRITON Electric High Torque Handpiece and Software Module
Indications for Use:
The IPC System with the TRITON Electric High-Torque Handpiece is indicated for the incision / cutting, removal, drilling, and sawing of soft and hard tissue and biomaterials in Neurosurgical (Cranial, Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ,
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K121264
Prescription Use (Per 21 CFR 801.109)
Medtronic TRITON Electric High-Torque Handpiece
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.