Wireless Hand Control, EG1A (WIRELESS-HC), Receiver for Wireless Hand Control, EG1A (RECEIVER-HC)

K220485 · The Anspach Effort, Inc. · HBC · May 19, 2022 · Neurology

Device Facts

Record IDK220485
Device NameWireless Hand Control, EG1A (WIRELESS-HC), Receiver for Wireless Hand Control, EG1A (RECEIVER-HC)
ApplicantThe Anspach Effort, Inc.
Product CodeHBC · Neurology
Decision DateMay 19, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4360
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The eG1 High Speed System is indicated for cutting and shaping bone including spine and cranium.

Device Story

The eG1 Wireless Hand Control System is an optional accessory for the eG1 High Speed System (K133604) that replaces the wired foot control pedal. It consists of a sterile, single-use Wireless Hand Control (WIRELESS-HC) that snaps onto the eG1 handpiece and a non-sterile, reusable Receiver (RECEIVER-HC) that plugs into the eG1 control console. The user applies pressure to a button on the WIRELESS-HC; this pressure is converted into a wireless radio signal (2.4 GHz) and transmitted to the receiver. The receiver transforms the signal into an electric command to modulate the eG1 handpiece motor speed. Used in surgical settings by clinicians, the device allows for hands-free or alternative control of surgical drill speed, potentially improving ergonomics and workflow during bone cutting and shaping procedures.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included design verification, software validation, electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), radio frequency compliance (FCC 47 CFR Part 15), and sterilization validation (ISO 11135).

Technological Characteristics

System includes a sterile, single-use wireless hand control (lithium coin battery powered) and a reusable receiver. Operates via 2.4 GHz radio frequency signal. Compliant with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and FCC 47 CFR Part 15. Sterilization of the hand control via Ethylene Oxide per ISO 11135.

Indications for Use

Indicated for patients requiring bone cutting and shaping, including spine and cranium procedures.

Regulatory Classification

Identification

An electric cranial drill motor is an electrically operated power source used with removable rotating surgical cutting tools or drill bits on a patient's skull.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. May 19, 2022 The Anspach Effort, Inc. % Marie Ferrier Senior Regulatory Affairs Program Lead Synthes GmbH Eimattstrasse 3 Oberdorf, 4436, Switzerland Re: K220485 Trade/Device Name: Wireless Hand Control, EG1A (WIRELESS-HC), Receiver for Wireless Hand Control, EG1A (RECEIVER-HC) Regulation Number: 21 CFR 882.4360 Regulation Name: Electric Cranial Drill Motor Regulatory Class: Class II Product Code: HBC, ERL Dated: April 20, 2022 Received: April 21, 2022 Dear Marie Ferrier: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {2}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K220485 #### Device Name Wireless Hand Control, EG1A (WIRELESS-HC); Receiver for Wireless Hand Control, EG1A (RECEIVER-HC) Indications for Use (Describe) The eG1 High Speed System is indicated for cutting and shaping bone including spine and cranium. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## The Anspach Effort, Inc. eG1 Wireless Hand Control System: Wireless Hand Control, EG1A (WIRELESS-HC), Receiver for Wireless Hand Control, EG1A (RECEIVER-HC) ## 510(k) Summary ### A. Device Information: | Category | Comments | |---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------| | Sponsor: | The Anspach Effort, Inc.<br>4500 Riverside Drive<br>Palm Beach Gardens FL 33410, United States<br>Tel: 1-561-627-1080 | | Correspondent Contact<br>Information: | Mrs. Marie Ferrier<br>Synthes GmbH<br>Eimattstrasse 3<br>Oberdorf 4436, Switzerland<br>Tel: +33(0)675765048 | | Device Common Name: | Wireless Hand Control System | | Device Regulation & Name: | 882.4360 – Electrical cranial drill motor<br>874.4250 – Ear, nose, and throat electric or pneumatic<br>surgical drill | | Classification & Product Code: | HBC (Class 2) – Motor, Drill, Electric<br>ERL (Class 2) – Drill, Surgical, Ent (Electric or<br>Pneumatic) Including Handpiece | | 510(k) Number: | K220485 | | Device Proprietary Name: | eG1 Wireless Hand Control System | ## Predicate Device Information: | Predicate Device: | Foot control pedals used with eG1 High<br>Speed System | |----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------| | Predicate Device Manufacturer: | The Anspach Effort, Inc. | | Predicate Device Common Name: | eG1 High Speed System | | Predicate Device Premarket Notification # | K133604 | | Predicate Device Regulation & Name: | 882.4360 - Electrical cranial drill motor<br>874.4250 - Ear, nose, and throat electric or<br>pneumatic surgical drill | | Predicate Device Classification<br>& Product Code: | HBC (Class 2) – Motor, Drill, Electric<br>ERL (Class 2) – Drill, Surgical, Ent<br>(Electric or Pneumatic) Including<br>Handpiece | {5}------------------------------------------------ ## B. Date Summary Prepared May 19th, 2022 ## C. Description of Device The eG1 Wireless Hand Control System is intended to be used as an optional accessory to the eG1 High Speed System (K133604) to provide an alternative method for controlling the eG1 Handpiece motor speed instead of using the existing wired foot control pedal. The eG1 Wireless Hand Control System consists of two articles: 1) The Wireless Hand Control, EG1A (article # WIRELESS-HC) is a sterile, single-use device that snaps on the eG1 Handpiece. 2) The Receiver for Wireless Hand Control, EG1A (article # RECEIVER-HC) is a non-sterile reusable device that plugs into the eG1 Control Console. The WIRELESS-HC sends information about the amount of pressure applied by the user on its button to the RECEIVER-HC wirelessly. The RECEIVER-HC uses this information to modulate the speed of the eG1 Handpiece motor. ## D. Indications for Use The eG1 High Speed System is indicated for cutting and shaping bone including spine and cranium. {6}------------------------------------------------ | Characteristic | Application Device:<br>eG1 Wireless Hand<br>Control System<br>K220485 | Predicate Device:<br>Foot control pedals used<br>with eG1 High Speed<br>System K133604 | Impact on Substantial<br>Equivalence | |---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Company | The Anspach Effort, Inc. | The Anspach Effort, Inc. | - | | Regulation<br>Number | 882.4360 – Electrical<br>cranial drill motor<br>874.4250 – Ear, nose, and<br>throat electric or<br>pneumatic surgical drill | 882.4360 – Electrical<br>cranial drill motor<br>874.4250 – Ear, nose, and<br>throat electric or<br>pneumatic surgical drill | Identical | | FDA Product<br>Code | HBC (Class 2) – Motor,<br>Drill, Electric<br>ERL (Class 2) – Drill,<br>Surgical, Ent (Electric or<br>Pneumatic) Including<br>Handpiece | HBC (Class 2) – Motor,<br>Drill, Electric<br>ERL (Class 2) – Drill,<br>Surgical, Ent (Electric or<br>Pneumatic) Including<br>Handpiece | Identical | | Intended Use<br>& Indications<br>for Use | Cutting and shaping bone<br>including spine and<br>cranium | Cutting and shaping bone<br>including spine and<br>cranium | Identical | | The device<br>controls/<br>function | Speed (increase/<br>decrease) of the<br>handpiece motor | Speed (increase/ decrease)<br>of the handpiece motor<br>Depending on model:<br>-rotation<br>-irrigation pump | Identical (the predicate<br>has additional functions) | | Principle of<br>operation/<br>Technology | The pressure on the<br>button is transferred from<br>the WIRELESS-HC to<br>RECEIVER-HC<br>wirelessly (radio signal)<br>and transformed into an<br>electric signal | The pressure on the pedal<br>is transformed into an<br>electric signal | The predicate device<br>does not contain any<br>wireless technology.<br>The device Wireless<br>Foot control used with<br>ELAN 4 Electro Motor<br>System (K203739) is<br>used as a reference<br>device for this<br>application as it shares<br>the same indications for<br>use, function and usage<br>of the wireless<br>technology to transmit a<br>signal in order to<br>regulate the speed of the<br>handpiece motor. The<br>proposed device and the<br>reference device operate<br>in the same frequency<br>band (2.4 GHz).<br>Software, Wireless and<br>Electromagnetic<br>Compatibility testing are | | | | | provided in this<br>application to<br>demonstrate the<br>Substantial Equivalence. | | Power tool<br>compatibility | eG1 High Speed System | eG1 High Speed System | Identical | | Interface with<br>the user (user<br>action) | Increase/ decrease the<br>pressure on the button to<br>increase/ decrease the<br>speed | Increase/ decrease the<br>pressure on the pedal to<br>increase/ decrease the<br>speed<br>Depending on model:<br>-Press the dedicated button<br>to change rotation<br>-Press the dedicated button<br>to activate irrigation pump | Identical (the predicate<br>has additional functions) | | Interface with<br>the console | Receiver plugs into the<br>eG1 consoles (foot pedal<br>connector port) | Cord plugs into the eG1<br>consoles (foot pedal<br>connector port) | Identical | | Power source | RECEIVER-HC: main<br>power through the<br>console<br>WIRELESS-HC: contains<br>a lithium coin battery | Main power through the<br>console | Electrical Safety testing<br>is provided in this<br>application to<br>demonstrate the<br>Substantial Equivalence. | | Patient<br>contact | No direct or indirect<br>patient contact | No direct or indirect<br>patient contact | Identical | | Sterilization/<br>Reprocessing | RECEIVER-HC: non<br>sterile, reusable (cleaning<br>only)<br>WIRELESS-HC: sterile,<br>single-use | Non sterile, reusable<br>(cleaning only) | The predicate device is<br>non sterile. Sterilization<br>Validation is provided<br>in this application to<br>demonstrate the<br>Substantial Equivalence. | # E. Comparison of the Technological Characteristics {7}------------------------------------------------ ## F. Summary of Supporting Data The proposed device eG1 Wireless Hand Control System has the same intended use, function, interface with the eG1 console and patient contact as the predicate device Foot control pedals used with eG1 High Speed System (K133604). The proposed device shows some differences regarding the power source, the technology (wireless) and sterilization. A design verification analysis was conducted to verify that the physical characteristics of the eG1 Wireless Hand Control System meet the design inputs related to the interfaces with the user, eG1 Handpiece and Console. Design verification testing was conducted to verify the functional aspects of the proposed device. The software of the proposed device was developed applying IEC 62304:2015 Medical device software - Software life cycle processes. Design verification analysis, Design verification testing and Software Validation were conducted on the proposed device firmware. The proposed device was developed, tested and certified according to IEC 60601-1:2012 / ANSI/AAMI ES60601-1:2012 Medical Electrical Equipment – General Requirements for {8}------------------------------------------------ ### Basic Safety and Essential Performance. The proposed device was developed, tested and certified according to IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests. The proposed device was found compliant with applicable US FCC regulations and standards 47 CFR Part 15 Subpart B and 47 CFR § 15.247. The radio emissions were tested according to IEC 62311:2007 Assessment of electronic and electrical equipment related to human exposure restrictions for electromagnetic fields (0 Hz to 300 GHz). The WIRELESS-HC is sterilized by Ethylene Oxide. The sterilization cycle was validated according to ISO 11135:2014 Sterilization of health-care products - Ethylene oxide -Requirements for the development, validation and routine control of a sterilization process for medical devices. ## G. Discussion of Performance Testing All acceptance criteria are met. The differences in technological characteristics do not raise new questions of safety or effectiveness based on results of risk-based verification and validation testing. ### H. Conclusion Following the 510(k) Decision-Making Flowchart, it is determined that the eG1 Wireless Hand Control System is substantially equivalent to the predicate Foot control pedals used with eG1 High Speed System (K133604) as there is no difference in the intended use and the differences in technological characteristics do not raise new questions of safety or effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...