IPC systems, Legend EHS handpieces and Legend Stylus Touch handpiece, Microsaw handpieces,Triton electric high torque handpiece, Attachments and surgical dissecting tools
Applicant
Medtronic Powered Surgical Solutions
Product Code
HBC · Neurology
Decision Date
May 15, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4360
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K170312 · May 15, 2017
IPC systems, Legend EHS handpieces and Legend Stylus Touch handpiece, Microsaw handpieces,Triton electric high torque handpiece, Attachments and surgical dissecting tools
Medtronic Powered Surgical Solutions
Published clinical literature
A comprehensive clinical literature review was conducted to support the safety and effectiveness of the Electric Drill System for expanded indications in various spinal surgical procedures (LM, PLIF, TLIF, LSD, ALIF, and DLIF).
Clinical Literature Review; Spinal Surgery; Expanded Indications
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Clinical Literature Review; Systematic literature review
Safe and effective use of the Electric Drill System for expanded indications
Indications for Use
The Electric Drill System is a electrically operated surgical instrument system. The electric motor provides power to operate removable rotating surgical cutting tools and their accessories intended for use in neurosurgery, including craniotomy and spinal surgery; as well as Ear Nose and Throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial and sternotomy surgeries. Additionally, the Electric Drill System is indicated for the incision / cutting, removal, drilling, and sawing of soft and hard tissue, bone, and biomaterials during open and minimally invasive spine procedures, which may incorporate application of various surgical techniques during the following lumbar spinal procedures: - Lumbar Microdiscectomy - Lumbar Stenosis Decompression - Posterior Lumbar Interbody Fusion (PLIF) - Transforaminal Lumbar Interbody Fusion (TLIF) - Anterior Lumbar Interbody Fusion (ALIF) - Direct Lateral Interbody Fusion (DLIF)
Device Story
Electric surgical instrument system; comprises Integrated Power Console (IPC), electric foot control, connection cables, irrigation/cooling sets, and various electric handpieces/motors. System drives interchangeable rotating surgical cutting tools and attachments. Used in OR settings by surgeons for neurosurgical, spinal, ENT, and orthopedic procedures. Provides power for incision, cutting, removal, drilling, and sawing of bone and soft tissue. IPC also functions as endoscope lens cleaning system. Surgeon operates via foot pedal; output is mechanical rotation of cutting tools. Benefits include precise tissue removal during complex spinal and cranial surgeries.
Clinical Evidence
No clinical trials. Evidence consists of a comprehensive clinical literature review supporting the safety and effectiveness of the system for the expanded spinal indications (LM, PLIF, TLIF, LSD, ALIF, DLIF) and cadaveric testing evaluating user acceptability and performance across these procedures.
Technological Characteristics
Electrically powered surgical drill system. Components include Integrated Power Console, electric handpieces, attachments, and surgical dissecting tools. Operating principle: electric energy supplied to handpiece to drive rotating cutting tools. Materials are similar to predicate devices. System includes irrigation/cooling control. Connectivity via cables to foot control and console.
Indications for Use
Indicated for patients undergoing neurosurgery (craniotomy, spinal), ENT, orthopedic, and general surgical procedures (maxillofacial, craniofacial, sternotomy). Specifically indicated for incision, cutting, removal, drilling, and sawing of soft/hard tissue, bone, and biomaterials during open and minimally invasive spine procedures, including lumbar microdiscectomy, lumbar stenosis decompression, PLIF, TLIF, ALIF, and DLIF.
Regulatory Classification
Identification
An electric cranial drill motor is an electrically operated power source used with removable rotating surgical cutting tools or drill bits on a patient's skull.
Predicate Devices
Electric Drill System (XPS 4000 System, Midas Rex Legend EHS System, IPC) (K081475)
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare and medicine.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Medtronic Powered Surgical Solutions % Xiaojian Sun Sr. Regulatory Affairs Specialist Medtronic Neurosurgery 125 Cremona Dr. Goleta, California 93117
Re: K170312
Trade/Device Name: IPC Systems, Legend EHS Handpieces and Legend Stylus Touch Handpiece, Microsaw Handpieces, Triton Electric High Torque Handpiece, Attachments and Surgical Dissecting Tools Regulation Number: 21 CFR 882.4360 Regulation Name: Electric Cranial Drill Motor Regulatory Class: Class II Product Code: HBC, HBE, HRX, HWE, EOJ, ERL, KFK Dated: January 31, 2017 Received: February 1, 2017
May 15, 2017
Dear Xiaojian Sun:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Michael J. Hoffmann -S
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170312
### Device Name
IPC Systems, Legend EHS Handpieces and Legend Stylus Touch Handpieces, Triton Electric High Torque Handpiece, Attachments and Surgical Dissecting Tools
## Indications for Use (Describe)
The Electric Drill System is a electrically operated surgical instrument system. The electric motor provides power to operate removable rotating surgical cutting tools and their accessories intended for use in neurosurgery, including craniotomy and spinal surgery; as well as Ear Nose and Throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial and sternotomy surgeries.
Additionally, the Electric Drill System is indicated for the incision / cutting, removal, drilling, and sawing of soft and hard tissue, bone, and biomaterials during open and minimally invasive spine procedures, which may incorporate application of various surgical techniques during the following lumbar spinal procedures:
- Lumbar Microdiscectomy
- Lumbar Stenosis Decompression
- Posterior Lumbar Interbody Fusion (PLIF)
- Transforaminal Lumbar Interbody Fusion (TLIF)
- Anterior Lumbar Interbody Fusion (ALIF)
- Direct Lateral Interbody Fusion (DLIF)
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <div> <span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SECTION 9: 510(k) SUMMARY
This summary is submitted in accordance with the requirements of 21CFR807.92.
#### 9.1 DATE PREPARED
January 31, 2017
#### 9.2 NAME AND ADDRESS OF MANUFACTURER
# Table 4: Manufacturer information
| Establishment Name | Establishment Registration Number |
|---------------------------------------------------------------------------------------------|-----------------------------------|
| Medtronic Powered Surgical Solutions<br>4620 North Beach Street<br>Fort Worth, TX 76137 USA | 1625507 |
| Medtronic Xomed, Inc.<br>6743 Southpoint Drive North<br>Jacksonville, FL 33216 USA | 1045254 |
#### 9.3 CONTACT PERSON
Xiaojian Sun Senior Regulatory Affairs Specialist 805.571.8758 Telephone: E-Mail: xiaojian.sun@medtronic.com
#### 9.4 PROPRIETARY NAME OF THE PROPOSED DEVICES
IPC Systems, Legend EHS Handpieces and Legend Stylus Touch Handpiece, Microsaw Handpieces, Triton Electric High Torque Handpiece, Attachments and Surgical Dissecting Tools
#### 9.5 COMMON/USUAL NAME
Surgical Drill Motors and Accessories
#### 9.6 DEVICE CLASSIFICATION NAME
# Table 5: FDA device classification information
| Description | FDA Code | Regulation Number |
|-----------------------------------------------------------------------------|----------|-------------------|
| Drills, burrs, trephines and accessories (simple, powered) | HBE | 21CFR882.4310 |
| Motor, drill, electric | HBC | 21CFR882.4360 |
| Arthroscope | HRX | 21CFR888.1100 |
| Instrument, surgical, orthopedic, ac-powered motor and accessory/attachment | HWE | 21CFR878.4820 |
| Bur, ear, nose and throat | EQJ | 21CFR874.4140 |
| Drill, surgical, ent (electric or pneumatic) including handpiece | ERL | 21CFR874.4250 |
| Saw, pneumatically powered | KFK | 21CFR878.4820 |
{4}------------------------------------------------
#### 9.7 PREDICATE DEVICE IDENTIFICATION
| SUBJECT SYSTEM DESCRIPTION | PREDICATE | |
|------------------------------------------|------------------------------------------------------------------------------|------------------|
| | SYSTEM DESCRIPTION | 510(k) NUMBER |
| Medtronic Electric Drill System | Electric Drill System [XPS 4000 System, Midas<br>Rex Legend EHS System, IPC) | K081475 |
| - IPC (Integrated Power Console) | - IPC (Integrated Power Console) | K081475 |
| - Legend EHS Handpieces | - Legend EHS Handpieces | K081475, K012457 |
| - Legend Stylus Touch Handpiece | - Legend Stylus Touch Handpiece | K081475, K012457 |
| - Microsaw Handpieces and Blades | - Microsaw Handpieces and Blades | K081475 |
| - Surgical Dissecting Tools | Surgical Dissecting Tools | K020069 |
| - Surgical Dissecting Tools (Curved Bur) | - Surgical Dissecting Tools (Curved Bur) | K072315 |
| - Legend Attachments | - Legend Attachments | K020069 |
| - Electric Foot Pedal and Y-Splitter | - Electric Foot Pedal and Y-Splitter | K081475 |
| - Irrigation Tubing Sets | - Irrigation Tubing Sets | K081475 |
| - System Accessories | - System Accessories | K081475 |
| Triton Electric High Torque Handpiece | Triton Electric High Torque Handpiece | K121264 |
| - Triton Electric High Torque Handpiece | - Triton Electric High Torque Handpiece | K121264 |
| - Surgical Dissecting Tools | Surgical Dissecting Tools | K121264, K870157 |
| - Attachments | - Attachments | K121264, K870157 |
| - System Accessories | - System Accessories | K121264, K870157 |
| Legend Electric High Speed Drill System | Legend Electric High Speed Drill System | K012457 |
| - Electric Handpiece | Electric Handpiece | K081475, K012457 |
| - Surgical Dissecting Tools | Surgical Dissecting Tools | K020069 |
| - Surgical Dissecting Tools (Curved Bur) | - Surgical Dissecting Tools (Curved Bur) | K072315 |
| - Legend Attachments | - Legend Attachments | K020069 |
Table 6: Subject Devices that are in the scope of this Submission
#### 9.8 DEVICE DESCRIPTION
The Electric Drill System consists of an electric Integrated Power Console, Electric Foot Control unit with Y-Splitter, Connection Cables, Irrigation/Cooling Sets including a remote Irrigation Control Unit, and various Electric Handpieces and/or electric motors and Attachments to drive various Surgical Dissecting Tools. The Integrated Power Console device of the Electric Drill System can also function as an endoscope lens cleaning system.
The device design, function, the intended use and the general operating principles, and conditions of use of the overall Medtronic Electric Drill System remain similar to those cleared under K012457, K081475, and K121264.
The surgical dissecting tools and attachments remain similar to those covered under K870157, K020069, and K072315.
#### 9.9 INDICATIONS FOR USE
The Electric Drill System is a electrically operated surgical instrument system. The electric motor provides power to operate removable rotating surgical cutting tools and their accessories intended for use in neurosurgery, including craniotomy and spinal surgery; as well as Ear Nose and Throat (ENT), orthopedic, and general surgical applications including maxillofacial, craniofacial and sternotomy surgeries.
Additionally, the Electric Drill System is indicated for the incision / cutting, removal, drilling, and sawing of soft and hard tissue, bone, and biomaterials during open and minimally invasive spine
{5}------------------------------------------------
procedures, which may incorporate application of various surgical techniques during the following lumbar spinal procedures:
- Lumbar Microdiscectomy O
- Lumbar Stenosis Decompression o
- Posterior Lumbar Interbody Fusion (PLIF) o
- Transforaminal Lumbar Interbody Fusion (TLIF) o
- Anterior Lumbar Interbody Fusion (ALIF) O
- Direct Lateral Interbody Fusion (DLIF) O
#### 9.10 COMPARISON OF INTENDED USE
The expanded indication of the subject device did not alter the intended use of the device, which is to remove soft and hard tissue, bone, and biomaterials during surgical procedures. The cadaveric study demonstrated that the addition of the proposed indications do not present any new issues of safety or effectiveness, and the systems perform as intended during surgical use, similar to the use during other various surgical procedures on currently cleared indications. Therefore it is concluded that the subject device has the same intended use as the predicate device.
#### 9.11 COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Electrically powered drill system, designed to remove soft and hard tissue, and bone, and biomaterials is the technological principle for both the subject and predicate systems.
The subject and predicate systems are based on the following same technological elements:
- A Electric Drill System Application: Designed to remove soft and hard tissue, bone, and biomaterials during various surgical applications.
- > Operating Principle: The electric energy is supplied to the Handpiece to provide power to operate interchangeable Surgical Dissecting Tools supported by Attachments and intended for use in various surgical procedures to remove soft and hard tissue, bone, and biomaterials.
In terms of the materials used in manufacturing of the patient contacting components of the subject Electric Drill System, the subject Attachments, Surgical Dissecting Tools, and irrigation tubing sets are similar to the predicates.
#### DISCUSSION OF THE PERFORMANCE TESTING 9.12
| TEST | DESCRIPTION | RESULTS |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clinical<br>Literature<br>Review | A comprehensive clinical literature<br>search was conducted and reviewed<br>regarding the safe and effective use<br>of the Electric Drill System for the<br>expanded indications for use, which<br>is the subject of this submission. | The review of supporting literature<br>supports the use of the Medtronic<br>Electric Drill System in LM, PLIF,<br>TLIF, LSD, ALIF, and DLIF spinal<br>surgical procedures, as well as any<br>variations of these procedures. |
{6}------------------------------------------------
| Cadaveric<br>Testing | Users evaluated the acceptability of<br>the subject electric drill system to its<br>intended use on a variety of<br>procedures using cadavers. | The Medtronic Electric Drill System<br>is acceptable for its intended use in<br>various surgical procedures. |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
#### CONCLUSION 9.13
The subject Medtronic Electric Drill Systems when compared to the predicate Drill Systems have same intended use, where the subject and the predicate drill systems are intended for use in various surgical procedures to remove soft and hard tissue, and biomaterials. Results of cadaveric testing have demonstrated that the addition of the proposed indications do not present any new issues of safety or effectiveness, and the systems perform as intended during surgical use, similar to the use during various other surgical procedures on currently cleared indications.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.