K040186 · The Anspach Effort, Inc. · HBB · Apr 15, 2004 · Neurology
Device Facts
Record ID
K040186
Device Name
ANSPACH MRI SAFE SURGICAL DRILL SYSTEM
Applicant
The Anspach Effort, Inc.
Product Code
HBB · Neurology
Decision Date
Apr 15, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4370
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MRI Safe Surgical Drill System (including pneumatic drill motor, foot control, nosepiece attachments and cutters), is indicated for intra-operative (non-imaging) use within a Magnetic Resonance Imaging (MRI) system rated at 0.5 Tesla or less, for the surgical cutting, shaping and removal of bone, including bones of the skull and spine.
Device Story
Pneumatic surgical drill system designed for use in MRI environments (0.5 Tesla or less). System components include motor, foot control, nosepiece attachments, and cutters. Materials selected for low magnetic susceptibility to minimize MRI image distortion and projectile risks. Operates at lower speeds (60-65k RPM) than standard non-MRI safe motors (80-85k RPM) to reduce vibration and whipping. Used by surgeons for bone dissection in skull and spine procedures. System includes a locking mechanism to prevent interchangeability with non-MRI safe components. Device must be moved to Zone 4 during imaging to prevent distortion. Healthcare providers use the drill for bone removal; clinical benefit includes the ability to perform surgical procedures in proximity to MRI equipment while mitigating magnetic field hazards.
Clinical Evidence
Bench testing only. No clinical data provided. Evaluation focused on magnetic susceptibility, heat generation, image distortion, and mechanical performance (speed/torque) compared to non-MRI safe predicate systems.
Technological Characteristics
Pneumatic-powered surgical drill system. Materials selected for low magnetic susceptibility. Includes motor, foot control, nosepiece attachments, and cutters. Features a locking mechanism to prevent interchangeability with non-MRI safe components. Operates at 60-65k RPM. Designed for use in MRI environments up to 0.5 Tesla.
Indications for Use
Indicated for intra-operative (non-imaging) use in patients requiring surgical cutting, shaping, or removal of bone, including skull and spine, within an MRI environment (0.5 Tesla or less).
Regulatory Classification
Identification
A pneumatic cranial drill motor is a pneumatically operated power source used with removable rotating surgical cutting tools or drill bits on a patient's skull.
Predicate Devices
Anspach MicroMax
Anspach 65K
Submission Summary (Full Text)
{0}------------------------------------------------
K040186
# Exhibit 10
#### Section 17
#### 510(k) SUMMARY
### Anspach MRi Safe Surgical Drill System
The MRI Safe Surgical Drill System is a class II medical device that includes a peumatic arill The MHT Sale Surgical Drill System is a class in mounts; and bone dissection tools. To make motor and fool control, a variety of noooplood comprising each system component have been replaced with materials known to have significantly lower or no magnetic susceptibility. replaced with Thatenals "Rhown" to "have" eighteaning ".
However, when imaging is being performed the drill system must be moved to at least Zone 4.
The MRI Safe Drill System foot control is not intended to be placed within the sterile field nor within MRI magnetic field (Zones 1-3). The MRI Safe motor does not require continuous oling within MHI magned (60-65k RPM) than non-MRI safe motors (80-85k RPM). Nosepiece and fulls at Siower speed (60 ook him) that herinn, wobble and whipping. They are a direct allaciments support the catter to millimize tean, non-sterile for reusable applications. MRI contact (with patient) deviol and are arounders with exception of materials and indications for use. Cutters are surgically invasive, direct contact (with patient) devices and are Indications for use. Oatters are bargloady invaters, arry additional warnings to advise user that "dulling" may occur faster than non-MRI safe cutters if overused or reused.
There are no known unacceptable risks/hazards associated with use of the MRI drill system, There are no nrown andoooplaabo risks identified (over non-MRI cutters) and addressed when used as indicated. Adultional note faction in surgeon control of the motor through design andror laboring melaces of magnetic fields on motor operation (speed/torque, uuring Surgical use, 2) Adverse etc.); 3) Drill, foot control, attachment or cutter becoming heat generation (over noating), notes, se in the MRI environment; 4) MRI image distortion; 5) a projectile dailing acc or while cutters; and 6) Unintended use of non-MRI safe system components with the MRI safe system.
MRI Safe system (including motor, foot control, nosepiece attachments and cutters), is indicated for intra-operative (non-imaging) use within a MRI system rated at 0.5 Tesla or less, indicated for intra operative (non removal of bone, including bones of the skull and spine. The for surgical brill System (including drill motor and foot control, nosepiece attachments Mill Care Curgiour Drill System (x presence within zones 1-3 of a MRI system while imaging is and cattle), to occomplished. Presence of the equipment in Zones 1-3 during imaging could cause unacceptable image distortion.
Cleaning, sterilization and storage conditions are specified on product package inserts and Orculing, blenneaton U.S. European Union and in other global market requirements.
The MRI safe drill system is substantially equivalent in form, design to other Anspach The Mill of Claim Systems including MicroMax and the original 65K, their nosepiece attachments pricumedio and by of of a materials, indications for use and a slightly modified locking mechanism, that prevents interchangeability between MRI safe and non-MRI safe system components.
______________________________________________________________________________________________________________________________________________________________________________
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Public Health Service
APR 1 5 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. William G. Conety Director, Regulatory Affairs and Quality Assurance Anspach Companies 4500 Riverside Drive Palm Beach Gardens, Florida 33410
Re: K040186
Trade/Device Name: MRI Safe Surgical Drill System Regulation Number: 21 CFR 883.4370; 21 CFR 882.4310 Regulation Name: Pneumatic drill, Powered simple cranial drills, burrs, trephines and their accessories Regulatory Class: II Product Code: HBB, HBE Dated: January 23, 2004
Received: January 28, 2004
Dear Mr. Conety:
We have reviewed your Section 510(k) promarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendnents, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Mr. William G. Conety
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to procced to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
(celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
K040186 510(k) Number (if known):
Device Name:
"MRI Safe Surgical Drill System"
Indications For Use:
The MRI Safe Surgical Drill System (including pneumatic drill motor, foot contributing a Masaction The Mirri Oale Outlers), is indicated for intra-operative (non-imaging) use within a Magnetic allaciments and contens), 15 indicated for the open of the surgical cutting, shaping and removal of bone, including bones of the skull and spine.
Prescription Use:
(Part 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use. (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Muriam C Provost
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number_________________________________________________________________________________________________________________________________________________________________
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.