SPINE APPLICATIONS FOR THE MAYFIELD/ACCISS OPERATING ARM AND OPTICAL ACCISS SYSTEMS MODULE, MAYFIELD/ACCISS SPINE RING A
K991267 · Ohio Medical Instrument Co., Inc. · HAW · Jul 12, 1999 · Neurology
Device Facts
Record ID
K991267
Device Name
SPINE APPLICATIONS FOR THE MAYFIELD/ACCISS OPERATING ARM AND OPTICAL ACCISS SYSTEMS MODULE, MAYFIELD/ACCISS SPINE RING A
Applicant
Ohio Medical Instrument Co., Inc.
Product Code
HAW · Neurology
Decision Date
Jul 12, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The intended use of the MAYFIELD®/ ACCISS™ Image-guided Stereotactic Workstation [K955397] (hereafter referred to as the Arm System), and the MAYFIELD® (Optical ACCISS™ Image-guided Stereotactic Workstation (K982244) (hereafter referred to as the Optical System) is unchanged by Spine Applications for the MAYFIELD® ACCISS" Operating Arm and Optical ACCISS " Systems Module [4-12-A-1137] (hereafter referred to as Spine Applications). However, the Systems' indications for use have been expanded to include spinal surgery procedures. The Arm and Optical Systems are intended for use as devices that use diagnostic images of the patient, acquired specifically to assist the surgeon with presurgical planning, to provide orientation and reference information during intra-operative procedures.
Device Story
Computer-based image-guided stereotactic surgery system for surgical theatre use. Inputs: preoperative CT or MRI patient scans. Operation: surgeon correlates images to patient anatomy via anatomical landmarks; system tracks probe orientation/tip position relative to image data. Arm System uses position-sensing articulated arm; Optical System uses 3-camera sensor assembly tracking infrared LEDs on probes. Output: real-time visualization of probe orientation on monitor screen overlaid on CT/MRI data. Spine Applications module allows cropping of spinal column images. Spine Ring assembly provides rigid link between patient and system. Surgeon uses output to guide surgical instruments during procedures. Benefits: improved surgical planning and intra-operative localization accuracy.
Clinical Evidence
Clinical data obtained under Institutional Review Board (IRB) oversight. Surgeons evaluated the system for preoperative planning and intra-operative accuracy. Results compared favorably to established cranial surgery performance metrics for the predicate systems.
Technological Characteristics
System components: workstation, software, probe assembly (articulated arm or optical tracking), and Spine Ring. Materials: Anodized Aluminum 6061, Stainless steel (ASTM F899, passivated QQ-P-35 Type 2). Sensing: Optically encoded joints (Arm) or infrared LED tracking (Optical). Connectivity: Standalone workstation. Sterilization: Steam sterilization (132°C, high vacuum) for Spine Ring/Halo. Software: Graphical User Interface-based navigation.
Indications for Use
Indicated for guidance and localization in open craniotomies and spinal surgeries. Applicable to any medical condition where stereotactic surgery is safe and effective, and where a rigid anatomical structure (skull, long bone, or vertebra) can be identified relative to a CT or MR-based anatomical model.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
JUL 12 1999
1.
### II 510(k) SUMMARY AND CERTIFICATION
K 99/267
#### Summary of Safety and Effectiveness
#### Spine Applications for the MAYFIELD®/ ACCISS™ Operating Arm and Optical ACCISS™ Systems
Pursuant to Section 513(i) of the Federal Food, Drug and Cosmetics Act, as amended by the Safe Medical Devices Act [SMDA] of 1990.
#### Summary Preparation Date: April 9, 1999.
| General Information:<br>Classification Name: | Stereotactic Instrumentation |
|----------------------------------------------|---------------------------------------------------------------------------------------------|
| Common/Usual Name: | Computer-Based Image-guided Stereotactic Surgery<br>Planning System - Spine Applications |
| Proprietary Name: | Spine Applications for the MAYFIELD®/ ACCISS™<br>Operating Arm and Optical ACCISS™ Systems. |
#### Applicant's Name and Address:
Kenneth B. Miller, Regulatory Affairs and Quality Assurance Director Ohio Medical Instrument Company, Inc. 4900 Charlemar Drive Cincinnati, Ohio 45227
#### 2. Name of predicate device(s):
MAYFIELD®/ ACCISS" Image-guided Stereotactic Workstation (K955397), MAYFIELD®/Optical ACCISS™ Image-guided Stereotactic Workstation [OASYS"] (K982244), BUDDE® Halo Retractor System [K830332]
#### 3. Classification:
Neurosurgical stereotactic instruments and accessories are Class II (21CFR 882.4800 and 882.4560).
#### 4. Performance Standards:
No applicable performance standards have been established by FDA under section 514 of the Food, Drug and Cosmetic Act.
{1}------------------------------------------------
# II 510(k) SUMMARY AND CERTIFICATION (continued)
#### 5. Intended Use and Device Description:
Intended Use: The intended use of the MAYFIELD®/ ACCISS™ Image-guided Stereotactic Workstation [K955397] (hereafter referred to as the Arm System), and the MAYFIELD® (Optical ACCISS™ Image-guided Stereotactic Workstation (K982244) (hereafter referred to as the Optical System) is unchanged by Spine Applications for the MAYFIELD® ACCISS" Operating Arm and Optical ACCISS " Systems Module [4-12-A-1137] (hereafter referred to as Spine Applications). However, the Systems' indications for use have been expanded to include spinal surgery procedures.
The Arm and Optical Systems are intended for use as devices that use diagnostic images of the patient, acquired specifically to assist the surgeon with presurgical planning, to provide orientation and reference information during intra-operative procedures.
The MAYFIELD® ACCISS Image-guided Stereotactic Workstation and the MAYFIELD® Optical ACCISS™ Image-guided Stereotactic Workstation (OASYS") are indicated for use in guidance and localization in open craniotomies and in spinal surgeries. The ACCISS™ Systems are indicated for any medical condition in which the use of stereotactic surgery may be considered safe and effective, and where a reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model of the anatomy.
Device Description: Spine Applications is software that allows the user to select ("crop") images of the portion of the spine under consideration from images of the entire spinal column. Also included is an accessory, the MAYFIELD® ACCISS" Spine Ring" Assembly (hereafter referred to as Spine Ring). The indications for use of the Arm and Optical Systems have been expanded to include use in spinal surgeries.
With Spine Applications, the Arm and Optical Systems can be used to identify spinal anatomical landmarks on the preoperative patient scans and also on the patient's spine to achieve registration with the imaged data set. After the correlation, the indication of the probe orientation appears on the monitor screen and moves through the CT or MRI data in correct relation to the probe as manipulated by the surgeon. This is the same process as that used in cranial surgery with the Systems.
With all image-guided surgery systems there must be a fixed positional relationship or link between the patient and the system. For cranial surgery, this link to the patient is achieved through the use of the Halo that is attached to the MAYFIELD® Skull Clamp A2000 [K932807] (hereafter referred to as the Skull Clamp). The Arm or the DRF is attached to the Halo, thereby achieving a firm and rigid relationship to the patient.
{2}------------------------------------------------
# II 510(k) SUMMARY AND CERTIFICATION (continued)
#### 5. Intended Use and Device Description: (continued)
For spinal procedures and use of Spine Applications, a similar "ring", the Spine Ring, may be used, if the surgeon elects. The Spine Ring is actually a segment of the Halo that can be connected to any standard spine retractor system. Attachment of the Spine Ring to the retractor that is attached to the patient achieves the same firm and rigid relationship to the patient as that of the Halo. The Spine Ring is sterilized via steam sterilization methods just as the Halo.
Spine Applications, in conjunction with the Arm and Optical Systems, is a computerbased system designed for use in the surgical theatre. The Arm System uses a position-sensing articulated Arm with a probe that acts as a localizing device. The Optical System uses a Sensor Assembly positioned above/beside the surgical site that tracks the position and orientation of digitizing probes and other instruments.
To use either System, the computer is loaded with the patient's computed tomography (CT) or magnetic resonance image (MRI) data and these images are correlated to the patient by physically matching points of anatomical landmarks. The System computer interfaces with CT or MR images and provides probe orientation/ probe-tip position. The indication of the probe orientation and probe-tip position appears on the monitor screen and moves through the CT or MRI data in correct relation to the probe as manipulated by the surgeon.
The Arm System and the Optical System, in conjunction with Spine Applications (the underlined items], consists of the following items:
#### A. Workstation
- B. Software
- · Includes MAYFIELD® ACCISS Spine Applications Module [4-12-A-1137]
- C. Probe Assembly [articulated arm or optical tracking system] · Includes MAYFIELD® ACCISS" Spine Ring" Assembly [4-12-C-1123]
- D. Optical Tracking Accessories
- E. Training Accessories
- F. Supported Modalities
- CT
- MRI
#### G. Options:
Tektronix Phaser II SDX Color Printer Overland Data 5612 9-Track Tape Drive
{3}------------------------------------------------
# II 510(k) SUMMARY AND CERTIFICATION (continued)
#### 6. Summary of Substantial Equivalence:
Spine Applications is substantially equivalent to the following devices:
- · The MAYFIELD® ACCISS™ Image-guided Stereotactic Workstation [K955397],
- · The MAYFIELD®Optical ACCISS™ Image-guided Stereotactic Workstation (OASYS") [K982244].
- · The BUDDE® Halo Retractor System [K830332]
#### Indications:
- · The Arm System and the Optical System, in conjunction with Spine Applications, are used in the identical manner as the predicate devices. Both Systems are used for localization as well as correlation of preoperative CT or MRI data with patient anatomy to assist in planning and/or performance of surgery. Spine Applications provides the surgeon with image-guided technology, using the identical algorithms currently used for cranial image-quidance, plus the ability to select or "crop" the images of the specific vertebral bodies involved.
- · The Spine Ring is used in an identical manner as the predicate device to provide a rigid base for the Arm System or Optical System components and accessories.
#### Design:
- · The Arm System, in conjunction with Spine Applications, utilizes a mechanical articulated Arm that acts as a 3-0 digitizer, which interfaces to a computer graphics workstation that displays reformatted CT or MR images in a variety of configurations. The Optical System, in conjunction with Spine Applications, utilizes a 3-camera Sensor Assembly, a DRF with 3 LEDs, and a probe handle with 3 LEDs that acts as a 3-D digitizer. The Sensor Assembly interfaces to a computer graphics workstation ' that displays reformatted CT or MR images in a variety of configurations. Both Systems utilize the identical equipment and algorithms for image-guided cranial surgery planning and for image-quided spinal surgery planning. The spinal surgery planning algorithms of both Systems include the module for spinal vertebral body selection {"cropping"}.
#### Materials:
- · Materials used to manufacture Spine Applications and the Spine Ring are the same as those used for the predicate devices with the exceptions noted in Design, above.
#### Manufacturing:
- · Manufacturing processes for Spine Applications and the Spine Ring are the same as those used for the predicate devices. with exceptions noted in Design, above.
#### Specifications:
- · Specifications of Spine Applications and the Spine Ring are the same as those of the predicate devices, with the exceptions noted in Design, above.
{4}------------------------------------------------
# 11 510(k) SUMMARY AND CERTIFICATION (continued)
#### 6. Summary of Substantial Equivalence: (continued)
#### Performance:
- · The Arm System and the Optical System, both with Spine Applications, were used by surgeons under authority of the Institutional Review Boards for the hospitals involved. The Spine Applications performed to the surgeons' satisfaction with regard to preoperative planning and accuracy. The clinical data for Spine Applications compared favorably with the data for cranial surgery performed using the Systems.
#### 7. Packaging:
The Arm and Optical Systems, Spine Applications, and the Spine Ring are supplied in industry standard medical-grade packaging suitable for electronic equipment and surgical instruments. The shippers and cartons are of suitable design to ensure product identification and protection from damage during shipping and storage.
#### 8. Sterilization:
The Arm and Optical Systems, Spine Applications, and the Spine Ring are supplied NON-STERILE. These non-sterile devices are packaged in "clean" condition, and must be removed from their shipping and packing materials, then appropriately cleaned before use as described in the instruction manuals.
A sterilization process, following AAMI TIR No. 12-1994, has been validated for the Spine Ring and the Halo. The validated sterilization cycle for the Halo and the Spine Ring is:
- · Steam sterilization at 132 degrees C, under high vacuum [2.0 psia] for 3 minutes [STS Test No.: 98-3359, STS Study No. GMP-1998-0082, on file at OMI].
The Instruction Manuals for Spine Applications, the Optical System, and the Arm System provide instructions on sterile use and reprocessing of components that require sterilization by the user.
#### 9. Conclusion:
The Intended Use, Indications for Use, Design, Materials, Manufacturing, Specifications, Performance, Packaging and Sterilization of Spine Applications and the Spine Ring do not raise any new unresolved issues relating to safety and effectiveness. Ohio Medical Instrument Company, Inc. therefore considers Spine Applications and the Spine Ring as substantially equivalent to the predicate devices.
{5}------------------------------------------------
.
# 11 510(k) SUMMARY AND CERTIFICATION (continued)
### 10. Comparison Table:
| FEATURE | MAYFIELD ®/ ACCISS™ Operating Arm<br>and Optical ACCISS™ Systems<br>[K955397, K982244] | Spine Applications for the<br>MAYFIELD ®/ ACCISS™ Operating Arm<br>and Optical ACCISS™ Systems |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Arm System and the Optical System<br>are intended for use as devices that, by<br>the use of diagnostic images of the<br>patient, acquired specifically to assist the<br>surgeon with presurgical planning.<br>provide orientation and reference<br>information during intra-operative<br>procedures. | • Same. |
| Indications for<br>Use | The Arm System and the Optical System<br>are indicated for: | The Arm System and Optical System,<br>with Spine Applications, are indicated for: |
| | • Guidance and localization in open<br>craniotomies, and for surgeries that are<br>traditionally performed with a<br>stereotactic apparatus, such as<br>biopsies, thalamotomies, and electrode<br>implants. The Systems may also be<br>used to review medical images in a<br>neurosurgical context. | • Use in guidance and localization in<br>open craniotomies and in spinal<br>surgeries. The ACCISS Systems are<br>indicated for any medical condition in<br>which the use of stereotactic surgery<br>may be considered safe and effective,<br>and where a reference to a rigid<br>anatomical structure, such as a skull, a<br>long bone, or vertebra, can be<br>identified relative to a CT or MR based<br>model of the anatomy. |
| Design | • Arm or Optical tracking for<br>3-D digitizing. | • Same. |
| | • Interfaces to computer graphics<br>workstation. | • Same. |
| | • Displays reformatted CT or MR images<br>in variety of configurations. | • Same. |
| | • Algorithms for image-guided surgery<br>planning. | • Same. |
| | • Software to show body part undergoing<br>surgery - head. | • Same. |
| | | • Software to show body part undergoing<br>surgery - spine. |
{6}------------------------------------------------
# III 510(k) SUMMARY AND CERTIFICATION (continued)
# 10. Comparison Table: (continued)
| FEATURE | MAYFIELD®/ ACCISS" Operating Arm<br>and Optical ACCISS" Systems<br>[K955397, K982244] | Spine Applications for the<br>MAYFIELD®/ ACCISS™ Operating Arm<br>and Optical ACCISS" Systems |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Type of Detector | • Optically encoded signals from joints of<br>articulated arm. | • Same. |
| | • A 3-camera Sensor Assembly detects<br>infrared signals emitted from LEDs on<br>hand-held probe handle. The Sensor<br>Assembly is mounted on a dedicated,<br>mobile stand. | • Same. |
| Registration<br>Technique | • Scanned Fiducials | • Same. |
| | • Anatomical Fiducials | • Same. |
| Operating<br>Software | • Anatomical Fiducials<br>• Structure: MS-DOS, with two major<br>functions: Image Importation, and<br>Surgical Applications [navigation].<br>Uses a Graphical User Interface to<br>facilitate interaction with user. | • Same. Modified Graphical User<br>Interface with similar functionality. |
| | • Image Manipulation: Unreformatted,<br>reformatted, 3-D surface rendering. | • Same. |
| Materials | • Aluminum 6061 [anodized] for tracking<br>probe [K853627 for use of anodized<br>aluminum]. | • Same |
| | • Stainless steel ASTM F899, passivated<br>QQ-P-35 Type 2 for sterile probe-tips. | • Same |
| Accessories | • BUDDE® Halo Retractor System<br>[K830332] | • Same.<br>• MAYFIELD®/ ACCISS™ Spine Ring™<br>Assembly [4-12-C-1123] |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 12 1999
Mr. Kenneth B. Miller Director, Regulatory Affairs and Quality Assurance Ohio Medical Instrument Co., Inc. 4900 Charlemar Drive Cincinnati, Ohio 45227
- K991267 Re:
Trade Name: MA YFIELD® ACCISS™ Operating Arm and Optical ACCISS™ Systems Regulatory Class: II Product Code: HAW Dated: April 9, 1999 Received: April 13, 1999
Dear Mr. Miller:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{8}------------------------------------------------
Page 2 – Mr. Kenneth B. Miller
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
Page 1 of ﻠﺴ
510(k) Number (if known):
K991267
Device Name: Spine Applications for the MAYFIELD®/ACCISS™ Operating Arm and Optical ACCISS™ Systems
Indications For Use:
The MAYFIELD®/ ACCISS™ Image-guided Stereotactic Workstation and the MAYFIELD®/Optical ACCISS™ Image-guided Stereotactic Workstation (OASYS"), are indicated for use in guidance and localization in open craniotomies and in spinal indicated for use in galadino and indicated for any medical condition in which the use of stereotactic surgery may be considered to be safe and effective, and where a reference to a rigid anatomical structure, such as a skull, a long bone, or vertebra, can be identified relative to a CT or MR based model of the anatomy.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General Restorative Devices K991267
510(k) Number
Prescription Use X
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format - 1 - 2 - 96)
Page A-7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.