← Product Code [HAW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW) · K990860

# VIEWPOINT MICROSCOPE OPTION (K990860)

_Philips Medical Systems (Cleveland), Inc. · HAW · Mar 30, 1999 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW/K990860

## Device Facts

- **Applicant:** Philips Medical Systems (Cleveland), Inc.
- **Product Code:** [HAW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW.md)
- **Decision Date:** Mar 30, 1999
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The ViewPoint is intended for use as a device which uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intra-operative procedures.

## Device Story

ViewPoint Microscope Option is an accessory for the ViewPoint image-guided surgery system; enables optical tracking of a surgical microscope. System uses diagnostic patient images to assist physicians with presurgical planning and intra-operative orientation. Microscope focal point serves as the tracked tool tip; allows user to view cross-sectional images corresponding to anatomy seen through the microscope. Used in OR by surgeons; provides real-time spatial reference during intracranial and spinal procedures. System utilizes infrared signals from diodes on tools, detected by a position sensor assembly. Output is visual overlay of patient anatomy on surgical view; aids in surgical navigation and decision-making.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Optical tracking system using infrared signals emitted from diodes on tools; detected by a position sensor assembly (two optical detectors). Operates in a silo-shaped digitizer volume (0.5m radius hemisphere/cylinder). Software runs in a UNIX environment with a graphical user interface. Supports MPR and surface rendering. Connectivity via system-integrated tracking hardware.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- ViewPoint 3.0 Software ([K970604](/device/K970604.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

3/30/99

K990860

## SUMMARY OF SAFETY AND EFFECTIVENESS (As required by 21 CFR 807.92)

#### General Information 1.

| Classification:             | Class II<br>Image Assisted Surgery Device                                                                                                                                                                                                 |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name:          | Image Assisted Surgery Device Option                                                                                                                                                                                                      |
| Proprietary Name:           | ViewPoint Microscope Option                                                                                                                                                                                                               |
| Establishment Registration: | Picker International, Inc.<br>World Headquarters<br>595 Miner Road<br>Highland Heights, Ohio 44143<br>Contact: Elaine K. Keeler, Ph.D.<br>Phone Number: (440) 473-3000<br>FDA Owner Number: #1580240<br>FDA Registration Number: #1525965 |
| Performance Standards:      | No applicable performance standards have been<br>issued under section 514 of the Food, Drug and<br>Cosmetic Act.                                                                                                                          |

#### 2. Intended Uses

The ViewPoint Microscope Option does not change the existing intended use and indications for use for the ViewPoint as defined below.

The ViewPoint is intended for use as a device which uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intra-operative procedures.

The ViewPoint is indicated for use in:

- Intra-cranial surgical procedures involving space occupying lesions or malformations (including soft tissue, vascular and osseous)
- Spinal surgical procedures involving spinal stabilization, neural decompression, or resection of spinal neoplasms.

| PICKER INTERNATIONAL, INC. | (MICRO) | 03/11/99 | I-1 |
|----------------------------|---------|----------|-----|
|----------------------------|---------|----------|-----|

{1}------------------------------------------------

#### Device Description 3.

The Microscope Option for the ViewPoint system allows for optical tracking of a surgical microscope. The tip of the tool in this case is the focal point of the Microscope instead of the physical end of the instrument. This option gives the user the ability to see crosssectional images that correspond with the anatomy viewed through the microscope.

#### Safety and Effectiveness 4.

The ViewPoint system with the Microscope Option is substantially equivalent to the ViewPoint system described in the 510(k) submission K970604. The following chart has been compiled to demonstrate this substantial equivalency.

| Parameter               | ViewPoint Microscope Option                                                                               | Predicate Device - View Point 3.0<br>Software (K970604)                                                                                                                                                                              |
|-------------------------|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Tools and Accessories   | Same tools as previous<br>submission with the additional<br>capability of using a surgical<br>microscope. | Probe with various length tips, drill<br>guide, tracking device (head and<br>spine), and phantoms. Each trackable<br>tool has a minimum of four IREDs.                                                                               |
| Average Tool Accuracy   | Same.                                                                                                     | 2.0 - 5.0 mm                                                                                                                                                                                                                         |
| Type of Detector        | Same.                                                                                                     | Infrared signals emitted from diodes<br>on a tool are detected by a Position<br>Sensor Assembly with two optical<br>detectors. The assembly is either on a<br>mobile pedestal, mounted to the OR<br>table or mounted to the ceiling. |
| Active Digitizer Volume | Same.                                                                                                     | Silo shape comprised of 0.5m radius<br>hemisphere and a cylinder with 0.5m<br>radius and 0.5m length. Detector<br>Positioning Feature added to guide<br>user in finding center of active<br>digitizer volume.                        |
| Registration Technique  | Same.                                                                                                     | Scanned Fiducials and Anatomical<br>Fiducials.                                                                                                                                                                                       |

## Substantial Equivalence Chart

PICKER INTERNATIONAL, INC.

(MICRO) 03/11/99

{2}------------------------------------------------

| Parameter                       | ViewPoint Microscope Option                                                                        | Predicate Device - ViewPoint 3.0<br>Software (K970604)                                                                                                                                                                                                                                                                   |
|---------------------------------|----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Operating Software<br>Structure | Same with some additions to the<br>Surgery Application to<br>accommodate the Microscope<br>Option. | UNIX environment with three major<br>processes: Import, Surgery<br>Application and Foot Switch. Uses a<br>Graphical User Interface to facilitate<br>interaction with user.                                                                                                                                               |
| Image Manipulation              | Same.                                                                                              | MPR and surface rendering.                                                                                                                                                                                                                                                                                               |
| Other Features                  | Same.                                                                                              | Detector Positioning Feature.                                                                                                                                                                                                                                                                                            |
| Intended Use                    | Same.                                                                                              | The ViewPoint is intended for use as a<br>device which uses diagnostic images of<br>the patient acquired specifically to<br>assist the physician with presurgical<br>planning and to provide orientation<br>and reference information during intra-<br>operative procedures.                                             |
| Indications for Use             | Same.                                                                                              | The ViewPoint is indicated for use in:<br>- Intra-cranial surgical procedures<br>involving space occupying lesions<br>or malformations (including soft<br>tissue, vascular and osseous)<br>- Spinal surgical procedures<br>involving spinal stabilization,<br>neural decompression, or resection<br>of spinal neoplasms. |

PICKER INTERNATIONAL, INC.

(MICRO) 03/11/99

I - 3

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized symbol featuring three abstract human profiles facing to the right, with flowing lines above them.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 3 0 1999

Elaine K. Keeler, Ph.D. Manager, Clinical Science Picker International, Inc. 595 Miner Road Highland Heights, Ohio 44143

K990860 Re: Trade Name: ViewPoint Microscope Option Regulatory Class: II Product Code: HAW Dated: March 12, 1999 Received: March 16, 1999

Dear Dr. Keeler:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 - Elaine K. Keeler, Ph.D.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.

Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

# 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

ViewPoint Microscope Option Device Name:

### Indications for Use:

The ViewPoint Microscope Option does not change the existing intended use and indications for use for the ViewPoint as defined below.

The ViewPoint is intended for use as a device which uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intra-operative procedures.

The ViewPoint is indicated for use in:

- Intra-cranial surgical procedures involving space occupying lesions or malformations (including soft tissue, vascular and osseous)
- cluding soll though varial and observinal stabilization, neural decompression, or resection of spinal neoplasms.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |                                          |                                                  |
|--------------------------------------------------------|------------------------------------------|--------------------------------------------------|
| (Division Sign-Off)                                    |                                          |                                                  |
| Division of General Restorative Devices                |                                          |                                                  |
| 510(k) Number                                          | K990860                                  |                                                  |
| Prescription Use<br>(Per 21 CFR 801.109)               | <div style="text-align:center;">OR</div> | Over-The-Counter Use<br>(Optional Format 1-2-96) |

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW/K990860](https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW/K990860)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
