← Product Code [HAW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW) · K983764

# VIEWPOINT TOOLS- STERRAD (K983764)

_Philips Medical Systems (Cleveland), Inc. · HAW · Jul 8, 1999 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW/K983764

## Device Facts

- **Applicant:** Philips Medical Systems (Cleveland), Inc.
- **Product Code:** [HAW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW.md)
- **Decision Date:** Jul 8, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The ViewPoint is intended for use as a device that uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intra-operative procedures.

## Device Story

ViewPoint Tools - Sterrad is an image-assisted surgery system providing intra-operative orientation and reference using pre-acquired diagnostic patient images. The system utilizes trackable surgical instruments, including a Y-probe, guide blocks, and a trackable awl, equipped with infrared emitting diodes (IREDs). These tools are tracked by the system to provide real-time trajectory and orientation information to the surgeon. The device is used in clinical settings by physicians for presurgical planning and intra-operative guidance during intracranial and spinal procedures. The primary modification in this submission is the validation of the Sterrad 100 sterilization process for these tools, ensuring that sterilization does not compromise the accuracy or function of the tracking components. The system assists the physician in navigating surgical sites, potentially improving procedural precision.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing validating that the Sterrad 100 sterilization process does not adversely affect the accuracy, material properties, or functional performance of the surgical tools compared to the predicate devices.

## Technological Characteristics

System consists of trackable surgical instruments (Y-probe, guide blocks, awl) utilizing infrared emitting diodes (IREDs) for spatial tracking. Materials include a combination of metal and non-metal components. The system is designed for use with diagnostic imaging data. Sterilization is validated for the Sterrad 100 system. Connectivity and software architecture details are not specified beyond the tracking functionality.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- [K963221](/device/K963221.md)
- [K970604](/device/K970604.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a sequence of handwritten characters. The sequence starts with the letter 'K', followed by the numbers 9, 8, 3, 7, 6, and 4. The characters are written in a dark, bold font, and they appear to be slightly slanted.

## SUMMARY OF SAFETY AND EFFECTIVENESS (As required by 21 CFR 807.92)

| 1. | General Information                                                                                                                                                                               |
|----|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|    | Classification: Class II<br>Image Assisted Surgery Device                                                                                                                                         |
|    | Common/Usual Name: Image Assisted Surgery Device Option                                                                                                                                           |
|    | Proprietary Name: ViewPoint Tools - Sterrad                                                                                                                                                       |
|    | Establishment Registration: Picker International, Inc.<br>World Headquarters<br>595 Miner Road<br>Highland Heights, Ohio 44143<br>FDA Owner Number: #1580240<br>FDA Registration Number: #1525965 |
|    | Performance Standards: No applicable performance standards have been<br>issued under section 514 of the Food, Drug and<br>Cosmetic Act.                                                           |

#### Intended Uses 2.

The ViewPoint is intended for use as a device that uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intra-operative procedures.

The ViewPoint is indicated for use in:

· Intra-cranial surgical procedures involving space occupying lesions or malformations (including soft tissue, vascular and osseous)

· Spinal surgical procedures involving spinal stabilization, neural decompression, or resection of spinal neoplasms.

#### 3. Device Description

Prior to use the ViewPoint tools must be sterilized. Testing has been completed to validate the use of the Sterrad 100 system for this process. The Guide Block (Non-Trackable and Trackable) and the Trackable Awl are three additional tools now available for the ViewPoint. The Guide Blocks guide and track the trajectory of a biopsy needle during a procedure. The Trackable Awl is a standard awl that has been adapted to include infrared emitting diodes so that the tip of the awl can be tracked similar to the standard ViewPoint Y-probe.

PICKER INTERNATIONAL, INC. (STERRAD) 10/22/98 I - 1

{1}------------------------------------------------

### 4. Safety and Effectiveness

The use of the Sterrad 100 system adequately sterilizes the ViewPoint tools for intraoperative procedures and does not affect the accuracy or function of the tools. The Guide Blocks and the Trackable Awl described in this submission are equivalent to the tools described in the 510(k) submissions K963221 and K970604. This equivalence is demonstrated in the following table.

| Parameter               | ViewPoint - Sterrad                                                                                                                                                                          | Predicate Devices                                                                                                                                                                                                                                       |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Tools                   | • Y-Probe with various tips<br>• Cable<br>• Head Tracker<br>• Spine Tracker<br>• Drill Guide (Non-Trackable and Trackable)<br>• Guide Block (Non-Trackable and Trackable)<br>• Trackable Awl | • Y-Probe with various tips<br>• Cable<br>• Head Tracker<br>• Spine Tracker<br>• Drill Guide (Non-Trackable and Trackable)                                                                                                                              |
| Material Considerations | Same.                                                                                                                                                                                        | Combination of metal and non-metal materials. IREDs on trackable tools are sensitive to heat.                                                                                                                                                           |
| Lumens                  | Same.                                                                                                                                                                                        | Trackable tools have a dead-end lumen in the LEMO Connector. The drill guides have a single-channel stainless steel lumen.                                                                                                                              |
| Use limits              | Same.                                                                                                                                                                                        | None.                                                                                                                                                                                                                                                   |
| Accuracy (Y-probe)      | Same.                                                                                                                                                                                        | Repeatability/Resolution: 1mm<br>Distance measurement: $\pm$ 0.75 mm<br>3D Localization: $\leq$ 1.57 mm<br>Fourth Fiducial Checkpoint: $<$ 5.0 mm                                                                                                       |
| Sterilization Technique | Sterrad                                                                                                                                                                                      | Ethylene Oxide                                                                                                                                                                                                                                          |
| Intended Use            | Same.                                                                                                                                                                                        | The ViewPoint is intended for use as a device that uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intra-operative procedures |

# Substantial Equivalence Table

PICKER INTERNATIONAL, INC. (STERRAD)

10/22/98

{2}------------------------------------------------

| Parameter           | ViewPoint - Sterrad | Predicate Devices                                                                                                                                                                                                                                                                                          |
|---------------------|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Same.               | The ViewPoint is indicated for use in:<br>• Intra-cranial surgical procedures<br>involving space occupying lesions or<br>malformations (including soft tissue,<br>vascular and osseous)<br>• Spinal surgical procedures involving<br>spinal stabilization, neural<br>decompression, or resection of spinal |

PICKER INTERNATIONAL, INC. (STERRAD)

10/22/98

、

I - 3

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a circular design with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract emblem resembling an eagle or bird in flight, composed of three stylized, curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 8 1999

Elaine K. Keeler, Ph.D. Manager, MR Clinical Science Picker International, Inc. 595 Miner Road Highland Heights, Ohio 44143

Re: K983764 Trade Name: ViewPoint Tools - Sterrad Regulatory Class: II Product Code: HAW Dated: April 8, 1999 Received: April 9, 1999

Dear Dr. Keeler:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 - Elaine K. Keeler, Ph.D.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

510(k) Number (if known): ___________________________________________________________________________________________________________________________________________________

983764

ViewPoint - Sterrad Sterilization Device Name:

Indications for Use:

The ViewPoint is intended for use as a device which uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intra-operative procedures.

The ViewPoint is indicated for use in:

- Intra-cranial surgical procedures involving space occupying lesions or malformations (including soft tissue, vascular and osseous)
· Spinal surgical procedures involving spinal stabilization, neural decompression, or resection of spinal neoplasms.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Pcsell

(Division Sign-Off)
Division of General Restorative Devices
510(k) Number. K983764

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use (Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW/K983764](https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW/K983764)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
