EZ-FIDUCIALS

K250505 · Phasor Health, LLC · HAW · Jul 9, 2025 · Neurology

Device Facts

Record IDK250505
Device NameEZ-FIDUCIALS
ApplicantPhasor Health, LLC
Product CodeHAW · Neurology
Decision DateJul 9, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

Phasor EZ-Fiducials is intended to provide fixed reference point(s) in patients requiring stereotactic surgery in conjunction with CT imaging with the included screws placed using the included EZ-Driver (electric screwdriver) or manual screwdriver exclusively for each.

Device Story

EZ-Fiducials provides fixed reference points for stereotactic neurosurgery; system includes four titanium screws, a manual screwdriver, and a single-use, battery-operated electric screwdriver (EZ-Driver). Device is used in a clinical setting by neurosurgeons. Screws are inserted into the skull to serve as fiducial markers for pre-surgical CT imaging. The EZ-Driver facilitates screw placement; manual screwdriver provides an alternative. Output is the physical placement of fiducials on the patient's skull, which are then visualized on CT scans to guide surgical planning and navigation. Benefits include providing stable, fixed reference points to improve accuracy in stereotactic procedures.

Clinical Evidence

Bench testing only. No clinical data. Validation included biocompatibility, packaging, sterility, and mechanical performance testing. Screw deflection was verified using a Coordinate Measuring Machine (CMM) in bone simulant. Compliance with ASTM F543-23 was confirmed for screws, and the electric screwdriver performance was validated.

Technological Characteristics

Titanium screws (ASTM F543-23 compliant); single-use, gamma-sterilized. EZ-Driver: electric, non-rechargeable lithium battery-powered, handheld, no software. Manual screwdriver: handheld. System is standalone, non-networked. Screws feature 2.9mm non-threaded shaft diameter, 1.98mm threaded shaft diameter.

Indications for Use

Indicated for patients requiring stereotactic surgery who need fixed reference points for CT imaging.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION July 9, 2025 Phasor Health, LLC Ray King Chief Executive Officer 8944 Kirby Drive Houston, Texas 77054 Re: K250505 Trade/Device Name: EZ-Fiducials Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: June 9, 2025 Received: June 9, 2025 Dear Ray King: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250505 - Ray King Page 2 Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K250505 - Ray King Page 3 Sincerely, Adam D. Pierce -S Digitally signed by Adam D. Pierce -S Date: 2025.07.09 09:43:44 -04'00' Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K250505 | | | Device Name EZ-Fiducials | | | Indications for Use (Describe) Phasor EZ-Fiducials is intended to provide fixed reference point(s) in patients requiring stereotactic surgery in conjunction with CT imaging with the included screws placed using the included EZ-Driver (electric screwdriver) or manual screwdriver exclusively for each. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} K250505, Page 1 of 3 510(k) Summary Phasor Health, LLC # EZ-FIDUCIALS™ A. Device Description | Category | Comments | | --- | --- | | Sponsor | Phasor Health, LLC, Ray King, 8944 Kirby Drive, Houston, TX 77054, U.S.A. (832) 982-1234 | | Correspondent Contact Information | Ray King 8944 Kirby Drive, Houston, TX 77054, U.S.A. (832) 982-1234 | | Device Common Name | subdural fluid drainage kit | | Device Regulation & Name | 21 CFR 882.4560 | | Classification & Product Code | Class II, HAW | | Device Proprietary Name | EZ-FIDUCIALS | Predicate Device(s) Information: | Predicate Device(s)/ Reference Devices(s) | Navigus Unibody Fiducial Marker System/ OsteoMed Pinnacle Driver | | --- | --- | | Predicate Device/ Reference device Manufacturer(s) | (1) Medtronic, Inc., (2) OsteoMed; respectively | | Predicate Device Common Name | (1) Navigus Unibody Fiducial Marker System; (2) Battery-operated drill/driver; respectively | | Predicate Device Notification # | (1) K033619; (2) K162542 | | Predicate Device Classification & Name | (1) 21 CFR 882.4560; (2) 21 CFR 882.4310; respectively | | Predicate Device Classification & Product Code: | Class II for each of (1) HAW & (2) HBE, respectively | B. Date Summary Prepared: July 8, 2025 # C. Description of Device Phasor EZ-FIDUCIALS™ provide fixed reference points during neurosurgical procedures. The device is composed of 3 items, all single-use, provided gamma-sterilized, and placed within the same primary packaging (sealed Tyvek tray within a shelf carton): (1) Screws: Each of the four provided titanium screws is identical, with specific dimensions of each EZ-Fiducials screw as follows: screwhead circular shape screwhead with square-shaped engagement, non-threaded shaft diameter 2.9mm and length 15mm, threaded shaft diameter (major threads) 1.98mm, thread length 5mm; no protective caps are provided, and the screws should exclusively be utilized in conjunction with the screwdrivers (manual or EZ-Driver) provided together in the same tray (and with no other screwdriver); (2) Manual Screwdriver: handheld, to be solely used with the screws provided, not intended for use with any other screws; (3) EZ-driver: (electric, handheld, non-rechargeable battery-operated screwdriver, without software, single-use for solely tightening or loosening the provided screws exclusively, not intended for use with any other screws). # D. Indications for Use Phasor EZ-Fiducials is intended to provide fixed reference point(s) in patients requiring stereotactic surgery in conjunction with CT imaging with the included screws placed using the included EZ-Driver (electric screwdriver) or manual screwdriver exclusively for each. # E. Comparison of Technological Characteristics | | Application Device | Predicate Device | Reference Device | Impact on Substantial Equivalence | | --- | --- | --- | --- | --- | | Company | Phasor Health, LLC | Medtronic, Inc. | Osteomed Pinnacle | N/A | {5} K250505, Page 2 of 3 | 510(k) | Pending | K033619 | K162542 | Comparable | | --- | --- | --- | --- | --- | | Regulation Number | 21 CFR 882.4560 | 21 CFR 882.4560 | 21 CFR 882.4310 | screws comparable to predicate device; accessory screwdriver similar to reference device | | Product Code | HAW | HAW | HBE | comparable for predicate device | | Device Class | II | II | II | Comparable | | Product Code | HAW (Class II) | HAW (Class II) | HBE (Class II) | comparable for predicate device | | Intended Use & Indications for Use | Intended Use: Phasor EZ FIDUCIALS™ are intended for use as fixed reference points for presurgery CT imaging for patients requiring stereotactic surgery. Indications for Use: Phasor EZ-Fiducials is intended to provide fixed reference point(s) in patients requiring stereotactic surgery in conjunction with CT imaging with the included screws placed using the included EZ-Driver (electric screwdriver) or manual screwdriver exclusively for each. | Intended Use: The Medtronic Unibody Models FM-4007, FM-4010, and FM-4013 Bone Fiducials are intended for use as fixed reference points for presurgery CT imaging for patients requiring stereotactic surgery. Indications for Use: Navigus Unibody Fiducial Marker System is intended to provide fixed reference point(s) in patients requiring stereotactic surgery in conjunction with CT imaging. | Intended Use (and Indications for Use): The PINNACLE Driver is intended for driving screws and for drilling into bone, in conjunction with cranial surgical procedures. | taken together, comparable to both devices | | Technology | | | | | | Screwdriver Power Source | Single-use Lithium non-rechargeable Batteries (for EZ-Driver); Manually operated for "Manual Screwdriver" included | Manual | Disposable, lithium, single use, sterile battery | comparable when considered together | | Performance Characteristics | screws into skull at threaded end | screws into skull at threaded end | N/A | comparable screws to predicate device | | Screw | screws into skull at threaded end | screws into skull at threaded end | N/A | comparable screws to predicate device | | Activation buttons | Forward/Reverse button for EZ-Driver; N/A for manual screwdriver which can be turned in forward or reverse direction | N/A for manual screwdriver which can be turned in forward or reverse direction | Forward/Reverse buttons for Driver | comparable to reference device | | Insertion material | bone | bone | bone | Same | | Screw Sizes/Insertion | 2.0mm | 1.6mm | 1.2mm, 1.6mm, 2.0mm | Comparable | | Biocompatibility | Yes | Yes | | | | Sterile | Yes | Yes | | | {6} K250505, Page 3 of 3 | Single-use | Yes | Yes | | | | --- | --- | --- | --- | --- | | Speed of Electric Driver | 30rpm | N/A | 120rpm | different, but ability to drive screw supported by performance testing | | Performance Testing Standard(s) Met for Screws | ASTM F543-23 | No standard referenced | No standard referenced | Met testing to noted standard for Application Device | ## F. Summary of Supporting Data EZ-FIDUCIALS™ has the same intended use, patient population, and classification as the predicate. Accessory screwdrivers are comparable to the reference device noted. The subject and predicate devices are based upon the following technological elements, when the predicate and reference devices are taken together: * Sterile packaged for single use, and disposable. * OsteoMed Pinnacle Driver operates on single-use battery power, as per the reference predicate (lithium-ion battery); * Bone screws purchase into the skull in similar manner. The following technical characteristics of the subject device differ from the predicate device: * Navigus Unibody bone fiducials are provided without an electric or manual screwdriver (although one would be needed) for insertion; * Navigus Unibody supplies "protective cap" and "screwdriver guide" (which are optional items, not required to the indications or intended use); * Osteomed Pinnacle driver is a high-speed driver with screwdriving and drilling capabilities, but otherwise comparable to EZ-Driver in that either can be used to drive screws (or remove them). Moreover, while Osteomed Pinnacle screwdriver can be used electrically or manually (in the still setting with power off to turn screws), Phasor EZ-Driver as part of EZ-Fiducials may be used similarly in power-on or power-off manual settings, but the EZ-Fiducials manual screwdriver is provided additionally to serve this same purpose of manual screw turning. * Navigus Unibody bone fiducials is sterilized using EO, whereas EZ-FIDUCIALS™ is sterilized with gamma with the manual screwdriver, electric screwdriver and bone fiducials/screws sealed together in a tray packaged in a shelf carton. The proposed device does not appear to differ from the predicates taken together in any manner which would negatively impact the safety or effectiveness of the device. ## G. Discussion of Performance Testing Various tests including biocompatibility, packaging, sterility, safety, and performance have been validated with justifications as comparable and/or supported by documentation with adequate performance. Specifically, testing for a deflection of bone screws using a Coordinate Measuring Machine (CMM) with firm purchase into bone simulant was established and verified by testing. Further testing was performed to ensure the EZ-Fiducials screws was compliant with the ASTM F543-23 standard, with the electric screwdriver also tested and performing adequately (per report "02670-019011-1" including Table 6). No clinical testing was needed or performed otherwise. ## H. Conclusion The predicate device was cleared based upon bench testing alone. The performance and design validation testing conducted on the EZ-FIDUCIALS™ device on the bench demonstrated that it performs equivalent to the stated predicate and reference device that are currently marketed individually for the same intended use. The proposed device, comprised of the elements noted above, demonstrates that it should perform as safely as effectively as the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...