K233155 · ClearPoint Neuro, Inc. · HAW · Jan 12, 2024 · Neurology
Device Facts
Record ID
K233155
Device Name
ClearPointer Optical Navigation Wand
Applicant
ClearPoint Neuro, Inc.
Product Code
HAW · Neurology
Decision Date
Jan 12, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The ClearPointer is intended to be used in conjunction with the SmartFrame OR System and a compatible stereotactic optical navigation system for patient registration and navigation. The ClearPointer is intended for use by a Neurosurgeon in a standard operating room environment to place and orient the SmartFrame OR Tower towards a target in the brain along a planned trajectory during stereotactic functional neurosurgical procedures.
Device Story
ClearPointer is an optical navigation wand used in stereotactic functional neurosurgery; functions as an optical tracker for the SmartFrame OR System. Input: optical signals captured by Medtronic StealthStation S8 navigation system. Device consists of 4 optical spheres in specific geometry; includes removable pointer attachment for image registration and entry point location. Operation: surgeon attaches device to SmartFrame OR tower or uses pointer attachment to align tower to brain target. Output: positional data for navigation system. Benefit: enables precise alignment of surgical trajectory to planned target. Used in OR by neurosurgeons.
Clinical Evidence
Bench testing only. Accuracy testing performed using simulated workflow with StealthStation S8 software. Instrument error measured at 0.11–0.30 mm; registration error measured at 0.2–0.6 mm. Design verification confirmed all physical, performance, and safety specifications met.
Technological Characteristics
Materials: Titanium Ti 6Al-4V, Silicone, PEEK, Nylon, 370HR composite, 20% glass-filled polycarbonate. Sensing: Passive optical tracking via 4 spheres. Connectivity: Compatible with Medtronic StealthStation S8. Form factor: Wand with removable pointer attachment and reference array bracket. Sterilization: EtO. Single-use.
Indications for Use
Indicated for use by neurosurgeons in standard OR environments for patient registration and navigation to orient the SmartFrame OR Tower toward a brain target along a planned trajectory during stereotactic functional neurosurgical procedures.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
NexProbe Image-Guided Probe (Navigus Passive Probe, Model NP-1000) (K040334)
Submission Summary (Full Text)
{0}------------------------------------------------
January 12, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
ClearPoint Neuro, Inc. Brennan Sullivan Regulatory Affairs Manager 120 S. Sierra Avenue, Suite 100 Solana Beach, California 92075
Re: K233155
Trade/Device Name: ClearPointer Optical Navigation Wand Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: December 8, 2023 Received: December 11, 2023
Dear Brennan Sullivan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Image /page/1/Picture/6 description: The image shows a digital signature. The name "Adam D. Pierce -S" is written on the left side of the image and also appears as the signer on the right side. The date of the signature is 2024.01.12, and the time is 08:21:19 -05'00'.
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233155
Device Name ClearPointer Optical Navigation Wand
Indications for Use (Describe)
The ClearPointer is intended to be used in conjunction with the SmartFrame OR System and a compatible stereotactic optical navigation system for patient registration and navigation.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) Summary
#### A. Device Information
| Category | Comments |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Sponsor: | ClearPoint Neuro, Inc.<br>6349 Paseo Del Lago<br>Carlsbad, CA 92011 |
| Correspondent Contact<br>Information: | Brennan Sullivan<br>120 S. Sierra Ave.<br>Solana Beach, CA 92075<br>617-678-1028<br>bsullivan@clearpointneuro.com |
| Device Common Name: | Neurological Stereotactic Instrument |
| Device Regulation & Name: | Neurological Stereotaxic Instrument, 21CFR<br>882.4560 |
| Classification & Product Code: | Class II<br>HAW |
| 510(k) Number: | K233155 |
| Device Proprietary Name: | ClearPointer Optical Navigation Wand |
### Primary Predicate Device Information:
| Predicate Device: | NexProbe Image-Guided Probe<br>(Navigus Passive Probe, Model NP-1000) |
|----------------------------------------------------|-----------------------------------------------------------------------|
| Predicate Device Manufacturer: | Medtronic, Inc. |
| Predicate Device Common Name: | Neurological Stereotactic Instrument |
| Predicate Device Premarket Notification # | K040334 |
| Predicate Device Classification & Name | Neurological Stereotaxic Instrument,<br>21CFR 882.4560 |
| Predicate Device Classification &<br>Product Code: | Class II<br>HAW |
### B. Date Summary Prepared
December 8, 2023
### C. Description of Device
The ClearPointer Optical Navigation Wand is an optical tracker intended to be used with the SmartFrame OR Stereotactic System and ClearPoint Bone Screw Fiducials, both of which are subject of pending 510(k) submission. The ClearPointer is used in conjunction with a Medtronic StealthStation Navigation System to align the SmartFrame OR tower to the desired
{5}------------------------------------------------
# ClearPoint Neuro, Inc. ClearPointer
target. It is composed of 4 optical spheres and utilizes a sphere positional geometry that is recognizable by the Medtronic StealthStation S8 Optical navigation System. It is equipped with a Pointer attachment to allow for image registration and entry point location. The Pointer Attachment can also be removed from the Optical Tracker component, so that the Tracker can be mounted to a SmartFrame OR. The SmartFrame OR Hardware Kit is provided with the following components:
- 1x ClearPointer™ ●
- 1x Pointer Attachment ●
- 1x Reference Array Bracket Arm ●
- 1x Torx Screwdriver ●
- 10x Reflective Spheres
### D. Indications for Use
The ClearPointer is intended to be used in conjunction with the SmartFrame OR System and a compatible stereotactic optical navigation system for patient registration and navigation.
### E. Comparison of Technological Characteristics
The ClearPointer is substantially equivalent to the predicate device NexProbe, subject of K040334. The devices are similar in design in that they consist of a consist of a reference frame in a similar geometry and are assembled with 4 optical tracking spheres. They can both be attached to a stereotactic frame for registration and alignment using the Medtronic StealthStation Software and they utilize similar workflows to achieve alignment.
The differences between the devices primarily relate to the difference in stereotactic frame with which they are compatible. The ClearPointer is intended to be used with the ClearPoint SmartFrame OR System and the NexProbe is intended to be used with NexFrame. Any differences between the ClearPointer and NexProbe do not present any additional risks of safety or efficacy.
| | ClearPointer | NexProbe® Image-Guided<br>Probe<br>(Navigus Passive Probe,<br>Model NP-1000) | Comparison<br>on |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Classification | 21CFR 882.4560 | 21CFR 882.4560 | Identical |
| Product Code | HAW | HAW | Identical |
| | NexProbe® Image-Guided | | |
| | | Probe | Comparis |
| | ClearPointer | (Navigus Passive Probe,<br>Model NP-1000) | on |
| Indications for<br>Use | The ClearPointer is intended to be<br>used in conjunction with the<br>SmartFrame OR System and a<br>compatible stereotactic optical<br>navigation system for patient<br>registration and navigation. | The Navigus Passive Probe is<br>intended to be used in<br>conjunction with the Navigus<br>family of trajectory guides and<br>the Medtronic StealthStation®<br>Treatment Guidance System for<br>patient registration and<br>navigation. The device is<br>provided sterile and for single<br>use | Identical |
| Intended Use | The ClearPointer is intended for<br>use by a Neurosurgeon in a<br>standard operating room<br>environment to place and orient the<br>SmartFrame OR Tower towards a<br>target in the brain along a planned<br>trajectory during stereotactic<br>functional neurosurgical<br>procedures. | The NexProbe Image-guided<br>Probe is intended for use by a<br>Neurosurgeon in a standard<br>operating room environment to<br>place and orient the Nexframe<br>Tower towards a target in the<br>brain along a planned trajectory<br>during stereotactic functional<br>neurosurgical procedures. | Identical |
| Materials | Titanium Ti 6Al-4V<br>Silicone<br>PEEK<br>Nylon<br>370HR composite<br>20% Glass-filled polycarbonate | Polymer material<br>Stainless Steel | Similar |
| Components | ClearPointer Optical Tracker<br>Pointer Attachment<br>Optical Tracker Reflective Spheres<br>Reference Array Bracket Arm<br>T8 Torx Screwdriver | NexProbe subassembly<br>11.5mm passive spheres | Similar |
| Compatibility | Medtronic StealthStation S8<br>O-Arm® Registration<br>PointMerge Registration | Medtronic StealthStation S8<br>O-Arm® Registration<br>PointMerge Registration | Identical |
| Single Use or<br>Reusable | Single-Use | Single-use | Identical |
| Sterile? | EtO Sterilized | Sterile | Identical |
Table 1: Side-by-side comparison of ClearPointer with Predicate Device
{6}------------------------------------------------
## ClearPoint Neuro, Inc. ClearPointer
### F. Bench Testing
ClearPoint Neuro has performed extensive testing to demonstrate that the ClearPointer is safe and effective for its intended use. The development of the ClearPointer was conducted in conformance with the company's design control procedures. Design inputs provided the requirements for the respective product specifications. Design Verification testing was performed to confirm that the ClearPointer meets all design specifications. These tests included
{7}------------------------------------------------
# ClearPoint Neuro, Inc. ClearPointer
verification of physical, performance, and safety requirements, as well as benchtop accuracy testing. Accuracy testing was performed by measuring instrument error and registration error of a simulated workflow with Stealthstation S8 software. The results are summarized in Table 2 below:
| Test | Error |
|--------------------|----------------|
| Instrument Error | 0.11 – 0.30 mm |
| Registration Error | 0.2 – 0.6 mm |
#### Table 2: ClearPointer Accuracy Results
Risk analysis was also performed with mitigation of all identified risks to acceptable levels.
### G. Conclusion
The subject ClearPointer and the predicate Medtronic NexProbe have identical indications for use and similar technological characteristics and principles of operation. ClearPoint Neuro has performed extensive bench testing to demonstrate that the differences between ClearPointer and the predicate device do not raise any risks of safety or efficacy. ClearPointer has been demonstrated to meet all test specifications and has been shown to be as safe, as effective, and to perform as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.