The V-TAG is intended to assist with stereotactic guidance, placement, and fixation for the operation of surgical instruments or devices during the planning and operation of neurological procedures performed in conjunction with preoperative and perioperative MR imaging. These procedures include laser coagulation, biopsies, catheter placement and electrode placement procedures.
Device Story
V-TAG is a single-use, skull-mounted, rigid trajectory array guide for stereotactic neurosurgery in adults. Device consists of a base (fixed to skull with 4 bone screws), a trajectory guide with 7 channels, handle, center adaptor, imaging cartridge (for gadolinium-based contrast), and reducer. Neurosurgeon aligns V-TAG to a pre-planned trajectory using an image-guided neuronavigational system; once aligned, V-TAG is locked to provide a stable guide for 14-gauge or 16-gauge instruments. MRI confirms alignment before intracranial instrument placement. Used in OR and MRI suites. Benefits include stable, multi-lumen trajectory guidance for precise instrument delivery during minimally invasive procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (cytotoxicity, intracutaneous toxicity, systemic toxicity, sensitization, pyrogenicity), MRI compatibility (ASTM 2503-13, F2052-14, F2213-06), and integrity testing (dimensional verification, stability, seal integrity, bond/joint strength) at zero time and 1-year accelerated aging.
Technological Characteristics
Single-use, sterile, MR-compatible stereotaxic instrument. Components: Base (4 bone screws), 7-channel Guide, Handle, Center Adaptor, Imaging Cartridge, Reducer. Materials: MR-safe components. Sterilization: 60Co Gamma Irradiation. Standards: ISO 10993 (biocompatibility), ASTM F2503-13, ASTM F2052-14, ASTM F2213-06 (MRI safety).
Indications for Use
Indicated for adult patients undergoing neurological procedures requiring stereotactic guidance, placement, and fixation of surgical instruments, including laser coagulation, biopsies, catheter placement, and electrode placement, performed with preoperative and perioperative MR imaging.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features an abstract design of a human figure. The FDA logo is on the right and consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 23, 2018
Voyager Therapeutics, Inc. Lynn Bayless Director, Regulatory Affairs 75 Sidney Street Cambridge, Massachusetts 02139
Re: K180854
Trade/Device Name: Voyager Trajectory Array Guide (V-TAG) Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: June 20, 2018 Received: June 21, 2018
Dear Lynn Bayless:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Valerie A. Flournoy -S
For Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180854
Device Name Voyager Trajectory Array Guide (V-TAG)
### Indications for Use (Describe)
The V-TAG is intended to assist with stereotactic guidance, placement, and fixation for the operation of surgical instruments or devices during the planning and operation of neurological procedures performed in conjunction with preoperative and perioperative MR imaging. These procedures include laser coagulation, biopsies, catheter placement and electrode placement procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) SUMMARY 3
SUBMITTER:
CONTACT PERSON: DATE PREPARED: DEVICE TRADE NAME: COMMON NAME: CLASSIFICATION NAME: CFR CITATION: PRODUCT CODE: PREDICATE DEVICE:
Voyager Therapeutics, Inc 75 Sidney Street Cambridge, MA 02139 USA (857) 259-5340 Lynn E. Bayless, MS, RAC June 12, 2018 V-TAGTM Stereotaxic instrument Stereotaxic instrument 882.4560 HAW Monteris UFO (K090240)
#### 3.1 Device Description
The V-TAG™ device is a single-use, skull-mounted, rigid trajectory array guide used in stereotactic surgical procedures in adults. The V-TAG can be manipulated to provide a wide range of surgical trajectories into the head. An image-guided neuronavigational system loaded with a stereotactic surgical plan is used to align the V-TAG to the planned trajectory; once aligned, the V-TAG is securely locked to provide a stable instrument guide. After positioning of the V-TAG, magnetic resonance imaging (MRI) is used to confirm alignment with the planned trajectory. Intracranial placement of a neurosurgical device or surgical instrument using the V-TAG device is only to be performed after MRI confirmation of the trajectory.
#### 3.2 Indication for Use
The V-TAG is intended to assist with stereotactic guidance, placement, and fixation for the operation of surgical instruments or devices during the planning and operation of neurological procedures performed in conjunction with preoperative and perioperative MR imaging. These procedures include laser coagulation, biopsies, catheter placement and electrode placement procedures.
{4}------------------------------------------------
#### 3.3 Technological Characteristics
The V-TAG device is a single use, MR compatible, sterile medical device consisting of a Base that is temporarily attached to the skull with four bone screws during the surgical procedure , a trajectory Guide containing seven channels to guide the insertion of 14-gauge neurosurgical devices or instruments, a Handle to assist with alignment using standard stereotactic surgical procedures, a Center Adaptor to identify the center trajectory of the Guide, an Imaging Cartridge to be filled with an FDA approved diluted gadolinium-based contrast agent (not included, no patient contact), and a Reducer for use with 16-gauge neurosurgical devices or instruments. Table 1 demonstrates substantial equivalence of the V-TAG to the predicate device.
## Table 4: Substantial Equivalence: V-TAG to the Predicate Device (Monteris Medical-UFO™)
| Parameter | Predicate Device:<br>Monteris Medical UFO™ | 510(k) Device:<br>V-TAG™ |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | K090240 | - |
| Product | Monteris Medical UFO | Voyager Trajectory Array Guide |
| Manufacturer | Monteris Medical, Inc. | Voyager Therapeutics, Inc. |
| Indication for use | The Monteris Medical UFO is<br>intended to provide stereotactic<br>guidance, placement and fixation<br>for the operation of instruments or<br>devices during the planning and<br>operation of neurological<br>procedures performed in<br>conjunction with preoperative and<br>(or) perioperative MR or CT<br>imaging. These procedures include<br>laser coagulation, biopsies, catheter<br>placement and electrode<br>procedures. | The V-TAG is intended to assist<br>with stereotactic guidance,<br>placement, and fixation for the<br>operation of surgical instruments<br>or devices during the planning and<br>operation of neurological<br>procedures performed in<br>conjunction with preoperative and<br>perioperative MR imaging. These<br>procedures include laser<br>coagulation, biopsies, catheter<br>placement, and electrode<br>placement procedures. |
| Principal Operator | Neurosurgeon | Neurosurgeon |
| Use Location | MRI Suite and OR | MRI Suite and OR |
| Operating Principle | Stereotactic guiding & fixation<br>device | Stereotactic guiding & fixation<br>device |
| Design Principle | Ball and socket joint on a 3-leg<br>support using 3 bone screws and 9<br>spikes | Ball and socket joint fixed to skull<br>using 4 bone screws |
{5}------------------------------------------------
| Parameter | Predicate Device:<br>Monteris Medical UFO™ | 510(k) Device:<br>V-TAG™ |
|--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| “Frameless”<br>stereotaxy | Yes | Yes |
| MRI Compatible | Yes | Yes |
| Single use disposable | Yes | Yes |
| Facilitates selection<br>of trajectory | Yes | Yes |
| Provides alignment<br>for neurosurgical<br>device insertion | Yes | Yes |
| Facilitates temporary<br>fixation of<br>neurosurgical device | Yes | Yes |
| Verification of<br>trajectory alignment<br>by MRI | Yes | Yes |
| Range of motion | 15 degrees angular (pre-drilled<br>holes) or 58 degrees angular (no<br>pre-drilled entry hole), 360 degrees<br>rotation | 45 degrees angular, 360 degrees<br>rotation |
| Burr hole size | < 14 mm | < 14 mm |
| Bilateral option | Yes | Yes |
| Multi-lumen spacing | Single lumen | 7 lumens, 2.5 mm spacing |
| Sterilization Method | Ethylene oxide | 60Co Gamma Irradiation |
#### 3.4 Non-Clinical Test Results
The non-clinical tests performed were prospectively identified to show substantial equivalence to the predicate device.
All testing was conducted on the subject device.
All biocompatibility testing was in compliance with the FDA GLP regulations (21 CFR 58).
The device passed all tests that were conducted indicating substantial equivlence.
Thus, the non-clinical performance data demonstrate that the V-TAG performs as indicated in the Instructions for Use, and is substantially equivalent and performs as well as the predicate device that is marketed for the same intended use.
The tests conducted, method and overall conclusion is provided in Table 2.
{6}------------------------------------------------
# Table 5: Summary Nonclinical Test Results
| Test | Test Method | Conclusion |
|------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility: | | |
| Cytotoxicity:<br>1X MEM Elution<br>(Product and packaging<br>performed according to USP) | ISO 10993-5-Part 5: Tests for<br>In Vitro Cytotoxicity<br><br>In vitro mammalian cell culture<br>test | Non-cytotoxic.<br><br>No evidence of cell lysis or<br>toxicity.<br><br>Grade <2 (mild reactivity) |
| Intracutaneous Toxicity | UNI EN ISO 10993-10-Part<br>10: Tests for Irritation and Skin<br>Sensitization<br><br>0.9 % Sodium Chloride<br>Solution: Sesame Oil, NF<br><br>Intracutaneous injection in<br>rabbits. | No biological response elicited.<br><br>No adverse observations noted<br>throughout the duration of the<br>test.<br><br>Overall mean score was 0.0 for<br>both the sodium chloride and<br>sesame oil test article extracts. |
| Systemic Toxicity: ISO Acute<br>Systemic Injection Test | UNI EN ISO 10993-11-Part<br>11: Tests for Systemic Toxicity<br><br>0.9 % Sodium Chloride<br>Solution: Sesame Oil, NF | Study requirements met.<br><br>No biological response elicited.<br><br>No mortality nor evidence of<br>systemic toxicity. |
| Sensitization | UNI EN ISO 10993-10- Part<br>10: Tests for Irritation and Skin<br>Sensitization<br>ASTM F 720-81 (2002)<br><br>0.9% Sodium Chloride<br>Solution:<br>Sesame Oil, NF | Non-sensitizing.<br><br>Test article extracts showed no<br>evidence of causing delayed<br>dermal contact sensitization in<br>the guinea pig. |
| Material Mediated<br>Pyrogenicity | US, General Chapter <151>,<br>Pyrogen Test 0.9% Sodium<br>Chloride Solution | Test article deemed non-<br>pyrogenic.<br><br>No animal experienced a<br>temperature rise 0.5°C or more<br>above its baseline. |
| Test | Test Method | Conclusion |
| Bench Performance Testing: | | |
| | | |
| MRI Compatibility | ASTM 2503-13<br>ASTM F2052-14<br>ASTM F2213-06<br>Testing was not performed on<br>Center Adaptor or Handle as<br>they are not intended for use in<br>a MR environment. | V-TAG Base, Guide, Imaging<br>Cartridge, and Reducer are MR<br>safe. |
| Integrity Testing- Zero time and 1 year accelerated: | | |
| | | |
| Integrity Testing- Zero Time<br>and 1 year accelerated | Evaluated dimensional<br>verification, device stability,<br>seal integrity of the Image<br>Cartridge, and bond and<br>joint strength for the Image<br>Cartridge, Reducer Handle and<br>tip. | Predetermined test<br>specifications were met for all<br>tests at both time points. |
{7}------------------------------------------------
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.