Synergy Cranial

K150216 · Medtronic Navigation, Inc. · HAW · Jun 11, 2015 · Neurology

Device Facts

Record IDK150216
Device NameSynergy Cranial
ApplicantMedtronic Navigation, Inc.
Product CodeHAW · Neurology
Decision DateJun 11, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2
AttributesPediatric

Indications for Use

The StealthStation System, with Synergy® Cranial software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical procedures. The system is indicated for any medical condition in which reference to a rigid anatomical structure can be identified relative to images of the anatomy. This can include, but is not limited to, the following cranial procedures: Cranial Biopsies Tumor Resections Craniotomies/Craniectomies Skull Base Procedures Transsphenoidal Procedures Thalamotomies/Pallidotomies Pituitary Tumor Removal CSF Leak Repair Pediatric Catheter Shunt Placement General Catheter Shunt Placement

Device Story

StealthStation System with Synergy Cranial software is an image-guided surgery (IGS) system; assists surgeons in locating anatomical structures during neurosurgery. Inputs include diagnostic/intraoperative images (CT, MR, X-ray, Nuclear Medicine) and tracked surgical instruments. System uses optical (infra-red) or electromagnetic localization to track instrument position relative to patient anatomy. Software processes data to provide 3D visualization, registration, and surgical planning. Output displayed on screen for surgeon; provides real-time navigation, trajectory guidance, and probe-eye views. Used in OR by surgeons; aids in precise instrument placement. Benefits include improved surgical accuracy and guidance for complex cranial procedures.

Clinical Evidence

No clinical data. Bench testing only. Performance validated using anatomically representative phantoms under worst-case configurations. Demonstrated 3D positional accuracy (mean error ≤ 2.0 mm) and trajectory angle accuracy (mean error ≤ 2.0 degrees) using CT and T1-weighted MR images. Software verification and validation performed per SDLC requirements.

Technological Characteristics

Stereotaxic instrument; optical (infra-red) and electromagnetic localization. Imaging modalities: X-ray, MR, Nuclear Medicine. Connectivity: interfaces with surgical microscopes (Zeiss, Leica), ultrasound (Aloka, Sonosite), and Medtronic O-arm. Software: C++ based, GUI-driven. Accuracy: mean positional error ≤ 2mm, trajectory error ≤ 2 degrees.

Indications for Use

Indicated for patients undergoing open or percutaneous neurosurgical procedures where rigid anatomical structures can be referenced to diagnostic images, including biopsies, tumor resections, craniotomies, skull base/transsphenoidal procedures, thalamotomies/pallidotomies, pituitary tumor removal, CSF leak repair, and shunt placements.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract design featuring three stylized human profiles facing to the right, with flowing lines beneath them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 11, 2015 Medtronic Navigation, Inc. Ms. Kaye E. Waite Senior Regulatory Specialist 826 Coal Creek Circle Louisville, Colorado 80027 Re: K150216 Trade/Device Name: StealthStation System with Synergy Cranial Software Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: May 11, 2015 Received: May 12, 2015 Dear Ms. Waite: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {2}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Carlos L. Pena -S Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K150216 ### Device Name StealthStation System with Synergy Cranial Software ### Indications for Use (Describe) The StealthStation System, with Synergy® Cranial software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical procedures. The system is indicated for any medical condition in which reference to a rigid anatomical structure can be identified relative to images of the anatomy. This can include, but is not limited to, the following cranial procedures: - Cranial Biopsies - Tumor Resections - Craniotomies/Craniectomies - Skull Base Procedures - Transsphenoidal Procedures - Thalamotomies/Pallidotomies - Pituitary Tumor Removal - CSF Leak Repair - Pediatric Catheter Shunt Placement - General Catheter Shunt Placement | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------| | <div> <span> × Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary # June 10, 2015 - Medtronic Navigation, Inc. l. Company: 826 Coal Creek Circle Louisville, Colorado 80027 USA Telephone Number: 720-890-3200 Fax Number: 720-890-3500 - Kaye E. Waite Contact: Senior Requlatory Affairs Specialist Telephone Number: 720-890-3200 Fax Number: 720-890-3500 - ll. Proprietary Trade Name: StealthStation System with Synergy Cranial Software - .............................................................................................................................................................................. Classification Name: Stereotaxic Instrument (21 CFR 882.4560) - Classification: Class II, Stereotaxic Instrument IV. - V. Product Code: HAW #### VI. Product Description The StealthStation System, with Synergy Cranial v2.2.7 software helps guide surqeons during cranial surgical procedures such as biopsies, tumor resections, and shunt placements. The Synergy Cranial v2.2.7 software works in conjunction with an Image Guided System (IGS) which consists of clinical software, surqical instruments, a referencing system and platform/computer hardware. Image quidance, also called navigation, tracks the position of instruments in relation to the surgical anatomy and identifies this position on diagnostic or intraoperative images of the patient. Synergy Cranial v2.2.7 software functionality is described in terms of its feature sets which are categorized as imaging modalities, registration, planning, interfaces with medical devices, and views. Feature sets include functionality that contributes to clinical decision making and are necessary to achieve system performance. #### VII. Indications for Use The StealthStation System, with Synergy Cranial software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical procedures. The system is indicated for any medical condition in which reference to a rigid anatomical structure can be identified relative to images of the anatomy. {5}------------------------------------------------ This can include, but is not limited to, the following cranial procedures: - Cranial Biopsies - Tumor Resections - Craniotomies/Craniectomies - Skull Base Procedures - Transsphenoidal Procedures - Thalamotomies/Pallidotomies - Pituitary Tumor Removal - CSF Leak Repair - Pediatric Catheter Shunt Placement - General Catheter Shunt Placement #### VIII. Summary of the Technological Characteristics | Item | Subject Device (Synergy Cranial) | Predicate Device (Mach Cranial)<br>StealthStation System Update -<br>K050438 | |-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | The StealthStation® System, with Synergy®<br>Cranial software is designed as an aid for<br>precisely locating anatomical structures in<br>either open or percutaneous neurosurgical<br>procedures. | The StealthStation® System is intended as an<br>aid for precisely locating anatomical structures<br>in either open or percutaneous procedures. | | Indications for<br>Use | The StealthStation System, with Synergy<br>Cranial software, is intended as an aid for<br>precisely locating anatomical structures in<br>either open or percutaneous neurosurgical<br>procedures. The system is indicated for any<br>medical condition in which reference to a rigid<br>anatomical structure can be identified relative<br>to images of the anatomy.<br>This can include, but is not limited to, the<br>following cranial procedures:<br>- Cranial Biopsies<br>- Tumor Resections<br>- Craniotomies/Craniectomies<br>- Skull Base Procedures<br>- Transsphenoidal Procedures<br>- Thalamotomies/Pallidotomies<br>- Pituitary Tumor Removal<br>- CSF Leak Repair<br>- Pediatric Catheter Shunt Placement<br>- General Catheter Shunt Placement | The StealthStation® System is intended as an<br>aid for precisely locating anatomical structures<br>in either open or percutaneous procedures.<br>The StealthStation® System is indicated for<br>any medical condition in which the use of<br>stereotactic surgery may be appropriate, and<br>where reference to a rigid anatomical<br>structure, such as the skull, a long bone, or<br>vertebra, can be identified relative to a CT or<br>MR based model, fluoroscopy images, or<br>digitized landmarks of the anatomy.<br>For the optical-based and EM-based system,<br>example procedures include, but are not<br>limited to:<br>- Cranial Biopsies<br>- Tumor Resections<br>- Craniotomies/Craniectomies<br>- Skull Base Procedures<br>- Transsphenoidal Procedures<br>- Thalamotomies/Pallidotomies<br>- Pituitary Tumor Removal<br>- CSF Leak Repair<br>- Pediatric Catheter Shunt Placement<br>- General Catheter Shunt Placement | | System<br>Accuracy<br>Requirement | The System has demonstrated accuracy with | The System has demonstrated accuracy with | | | a mean positional error of 2mm and mean | a mean positional error of 2mm and mean | | | trajectory error of 2 degrees | trajectory error of 2 degrees | {6}------------------------------------------------ | Imaging Modalities | X-Ray based,<br>MR based<br>Nuclear Medicine based | |---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Registration Features | Exam-to-Exam Registration<br>Patient Registration | | Planning Features | Plan Entry and Target Selection<br>3D Model Building<br>Advanced Visualization | | Medical Device Interfaces | Microscope Navigation: Zeiss, Leica<br>Ultrasound Navigation: Aloka and Sonosite<br>Medtronic O-arm | | View (Display) Features | Ultrasound Video In, Ultrasound Overlay, 3D,<br>Anatomic Orthogonal, Trajectory 1 and 2,<br>Target Guidance, 2D Trajectory Guidance,<br>Probes Eye, Look Ahead, Microscope<br>Injection, Video Input | | Software Interface (GUI) | Blue style with chronological next/back task<br>flow at the top of the screen. Image controls<br>on the left. Planning information on the right. | | Programming Language | C++ | | Localization Technology | Optical (infra-red)<br>Electromagnetic | ### IX. Identification of Legally Marketing Device (Predicate Device) StealthStation System Update (K050438) #### Discussion of the Performance Testing X. The following table summarizes the testing conducted on the StealthStation System with Synergy Cranial v2.2.7 software: # Description Under representative worst-case configuration, the StealthStation® System with Synergy Cranial Software, has demonstrated performance in 3D positional accuracy with a mean error ≤ 2.0 mm and in trajectory angle accuracy with a mean error ≤ 2.0 degrees. This performance was determined using an anatomically representative phantom and utilizing a subset of system components and features that represent the worst-case combination of all potem components. The test configuration included CT images with slice spacing and thickness of 1.0 mm, and T1-weighted MR images with slice spacing and thickness of 1.5 mm. Software verification and validation testing for each requirement specification. System integration performance testing for cranial surgical procedures using anatomical phantoms. The following table summarizes the quality assurance measures that were applied during development of the software component of the system: {7}------------------------------------------------ ## Description Software Development Life Cycle Software Risk Assessment ## Software Configuration Management and Version Control Design verification and validation was performed using the StealthStation System with Synerqy Cranial v2.2.7 software in laboratory and simulated use settings. The results support the safety of the device and demonstrate that the software should perform as intended in the specified use conditions.. Clinical testing was not considered necessary prior to release as this is not new technology. #### XI. Conclusions The non-clinical data support the safety of the device and the software verification and validation demonstrate that the StealthStation System with Synergy Cranial v2.2.7 software should perform as intended in the specified use conditions. The non-clinical data demonstrate that the StealthStation System with Synergy Cranial 2.2.7 software performs comparably to the predicate device for the same intended use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...