← Product Code [HAW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW) · K141833

# STYLET, 23CM, SKULL-MOUNT PATIENT TRACKER, NON-INVASIVE PATIENT TRACKER, TRACER POINTER, TOUCH-N-GO POINTER, NAVIGATION (K141833)

_Medtronic Navigation, Inc. · HAW · Mar 25, 2015 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW/K141833

## Device Facts

- **Applicant:** Medtronic Navigation, Inc.
- **Product Code:** [HAW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW.md)
- **Decision Date:** Mar 25, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Pediatric

## Indications for Use

The Non-Invasive Patient Tracker is indicated for use as a stereotactic reference frame to track the position of patient anatomy during navigated cranial and ENT procedures in adult and pediatric patients. It is indicated for use with Medtronic computer-assisted surgery systems using electromagnetic navigation and cranial and ENT software. The Medtronic computer assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT-based or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.

## Device Story

AxiEM™ instruments (Stylet, Tracer Pointer, Touch-n-Go Pointer, Navigation Pointer) and trackers (Skull-Mount, Non-Invasive) are single-use, sterile devices for stereotactic surgery. Devices incorporate internal tracking sensors detecting low-energy magnetic fields generated by a mobile emitter. Navigation system software processes these signals to display instrument tip location relative to patient image planes (CT/MR/fluoroscopy). Used in ORs by surgeons for cranial/ENT procedures (e.g., tumor resection, catheter placement, craniotomy). Output provides real-time visualization of instrument position, aiding precise anatomical localization. Benefits include improved surgical accuracy and guidance during complex procedures.

## Clinical Evidence

Bench testing only. Performance validated via system accuracy testing (mean positional error <2mm, trajectory error <2 degrees), skin shift tests, chemical resistance, electrical safety (IEC 60601-1/1-2), MRI compatibility (1.5T/3T), and accelerated life testing. Biocompatibility confirmed via cytotoxicity, sensitization, irritation, systemic toxicity, and pyrogenicity testing per ISO 10993 and USP standards.

## Technological Characteristics

Electromagnetic tracking instruments/trackers. Materials: 302/303/304 Stainless Steel, PET, ABS plastic (Lustran 348), Pellethane 2363, silicone, titanium alloy. Sensing: Internal magnetic field detection. Connectivity: Compatible with Medtronic electromagnetic navigation systems. Sterilization: ETO, single-use. Form factor: Handheld pointers, stylets, and skull-mounted/non-invasive trackers.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- StealthStation® GoldenEye™ Micro-Magnetic Tracking System Option ([K001284](/device/K001284.md))
- Catheter Introducer for the StealthStation® System ([K022126](/device/K022126.md))
- StealthStation® System Update ([K050438](/device/K050438.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of a human figure, represented by three overlapping profiles facing to the right. The profiles are rendered in black and are arranged to create a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 25, 2015

Medtronic Navigation, Inc. Mr. Christopher Perman Principal Regulatory Affairs Specialist 826 Coal Creek Circle Louisville, Colorado 80027

Re: K141833

> Trade/Device Name: Stylet, 23cm, Skull-Mount Patient Tracker, Non-Invasive Patient Tracker, Tracer Pointer, Touch-n-Go Pointer, Navigation Pointer Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: February 20, 2015 Received: February 23, 2015

Dear Mr. Perman,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carlos L. Pena -S 同公

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known) K141833

Device Name Non-Invasive Patient Tracker

#### Indications for Use (Describe)

The Non-Invasive Patient Tracker is indicated for use as a stereotactic reference frame to track the position of patient anatomy during navigated cranial and ENT procedures in adult and pediatric patients. It is indicated for use with Medtronic computer-assisted surgery systems using electromagnetic navigation and cranial and ENT software.

The Medtronic computer assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT-based or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                                                                    |                                                                                                                     |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif; font-size: 10pt;"></span> <span style="font-family: Arial, sans-serif; font-size: 10pt;">☑</span> <span style="font-family: Arial, sans-serif; font-size: 10pt;"> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-family: Arial, sans-serif; font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{3}------------------------------------------------

510(k) Number (if known) K141833

Device Name Skull-Mount Patient Tracker

#### Indications for Use (Describe)

The Skull-Mount Patient Tracker is indicated for use as a stereotactic reference frame to track the position of patient anatomy during navigated cranial and ENT procedures. It is indicated for use with Medtronic computer-assisted surgery systems using electromagnetic navigation and cranial and ENT software.

The Medtronic computer assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT-based or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.

Type of Use (*Select one or both, as applicable*)> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{4}------------------------------------------------

510(k) Number (if known) K141833

Device Name Stylet, 23 cm

#### Indications for Use (Describe)

The Stylet is indicated for use in precisely locating anatomical structures during navigated cranial procedures. It is indicated for use with Medtronic computer-assisted surgery systems using electromagnetic navigation and cranial software.

The Medtronic computer assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT-based model, fluoroscopy images, or digitized landmarks of the anatomy.

The Stylet is indicated for use in the following procedures:

- · General ventricular catheter placement
- · Pediatric ventricular catheter placement
- · Tumor resections
- · Skull base procedures
- · Craniotomies/craniectomies
- · Transsphenoidal procedures

| Type of Use (Select one or both, as applicable)                                                    |  |
|----------------------------------------------------------------------------------------------------|--|
| <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)  </div> |  |
| <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)  </div>                 |  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{5}------------------------------------------------

510(k) Number (if known) K141833

Device Name Tracer Pointer

#### Indications for Use (Describe)

The Tracer® Pointer is indicated for use in registering patient anatomy and in precisely locating anatomical structures during navigated cranial procedures in adult and pediatric patients. It is indicated for use with Medtronic computerassisted surgery systems using electromagnetic navigation and cranial software.

The Medtronic computer assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT-based or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <div> </div> <div> </div> |
|----------------------------------------------|---------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <div> </div> <div> </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{6}------------------------------------------------

510(k) Number (if known) K141833

Device Name Touch-n-Go Pointer

#### Indications for Use (Describe)

The Touch-n-Go Pointer is indicated for use in registent anatomy during navigated cranial procedures in adult and pediatric patients. It is indicated for use with Medtronic computer-assisted surgery systems using electromagnetic navigation and cranial software.

The Medtronic computer assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT-based or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{7}------------------------------------------------

510(k) Number (if known) K141833

Device Name Navigation Pointer

#### Indications for Use (Describe)

The Navigation Pointer is indicated for use in precisely locating anatomical structures during navigated cranial procedures. It is indicated for use with Medironic computer-assisted surgery systems using electromagnetic navigation and cranial software.

The Medtronic computer assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT-based model, fluoroscopy images, or digitized landmarks of the anatomy.

The Navigation Pointer is indicated for use in the following procedures:

- · General tumor resections
- Pediatric tumor resections
- · Skull base procedures
- · Craniotomies/craniectomies
- · Transsphenoidal procedures

| Type of Use (Select one or both, as applicable)                               |                                                                              |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span>    Prescription Use (Part 21 CFR 801 Subpart D)   </div> | <div> <span> </span>    Over-The-Counter Use (21 CFR 801 Subpart C)   </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{8}------------------------------------------------

## 510(k) Summary

#### I. SUBMITTER

Medtronic Navigation. Inc. 826 Coal Creek Circle Louisville, Colorado 80027 USA

Telephone: 720-890-3200 Fax: 720-890-3500

Contact Person: Christopher Perman Date Prepared: March 24, 2015

#### II. DEVICE

Name of Device: Stylet, 23cm, Tracer® Pointer, Touch-n-Go Pointer, Navigation Pointer, Skull-Mount Patient Tracker, Non-Invasive Patient Tracker Common Name: Stereotaxic Instrument Classification Name: Stereotaxic Instrument (21 CFR 882.4560) Regulatory Class: Class II (21 CFR 882.4560) Product Code: HAW

#### III. PREDICATE DEVICE

StealthStation® GoldenEye™ Micro-Magnetic Tracking System Option, K001284 Catheter Introducer for the StealthStation® System, K022126 StealthStation® System Update, K050438

No reference devices were used in this submission.

#### IV. DEVICE DESCRIPTION

The AxiEM™ Instruments (Stylet, 23cm, Tracer® Pointer, Touch-n-Go Pointer, Navigation Pointer) and Trackers (Skull-Mount Patient Tracker. Non-Invasive Patient Tracker) are singleuse, sterile devices compatible with AxiEM™-enabled Medtronic computer assisted surgery systems. The instruments and trackers are electromagnetically navigated devices used for the purpose of localizing patient anatomy, patient registration, and tracking patient location during stereotactic surgery procedures.

Each device incorporates a tracking system within the instrument or tracker. The mobile emitter of the navigation system generates a low-energy magnetic field that is detected by the tracking system within the device. The navigation system software displays the location of the instrument's tip within multiple patient image planes and other anatomical renderings.

{9}------------------------------------------------

#### V. INDICATIONS FOR USE

## Stylet, 23cm

The Stylet is indicated for use in precisely locating anatomical structures during navigated cranial procedures. It is indicated for use with Medtronic computer-assisted surgery systems using electromagnetic navigation and cranial software.

The Medtronic computer assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT-based or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.

The Stylet is indicated for use in the following procedures:

- General ventricular catheter placement ●
- Pediatric ventricular catheter placement ●
- Tumor resections
- Skull base procedures ●
- Craniotomies/craniectomies ●
- Transsphenoidal procedures

## Tracer® Pointer

The Tracer® Pointer is indicated for use in registering patient anatomy and in precisely locating anatomical structures during navigated cranial procedures in adult and pediatric patients. It is indicated for use with Medtronic computer-assisted surgery systems using electromagnetic navigation and cranial software.

The Medtronic computer assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT-based or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.

## Touch-n-Go Pointer

The Touch-n-Go Pointer is indicated for use in registering patient anatomy during navigated cranial procedures in adult and pediatric patients. It is indicated for use with Medtronic computer-assisted surgery systems using electromagnetic navigation and cranial software.

The Medtronic computer assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT-based or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.

{10}------------------------------------------------

## Navigation Pointer

The Navigation Pointer is indicated for use in precisely locating anatomical structures during navigated cranial procedures. It is indicated for use with Medtronic computer-assisted surgery systems using electromagnetic navigation and cranial software.

The Medtronic computer assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT-based or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.

The Navigation Pointer is indicated for use in the following procedures:

- General tumor resections
- Pediatric tumor resections ●
- Skull base procedures
- Craniotomies/craniectomies ●
- Transsphenoidal procedures ●

## Skull-Mount Patient Tracker

The Skull-Mount Patient Tracker is indicated for use as a stereotactic reference frame to track the position of patient anatomy during navigated cranial and ENT procedures. It is indicated for use with Medtronic computer-assisted surgery systems using electromagnetic navigation and cranial and ENT software.

The Medtronic computer assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT-based or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.

## Non-Invasive Patient Tracker

The Non-Invasive Patient Tracker is indicated for use as a stereotactic reference frame to track the position of patient anatomy during navigated cranial and ENT procedures in adult and pediatric patients. It is indicated for use with Medtronic computer-assisted surgery systems using electromagnetic navigation and cranial and ENT software.

The Medtronic computer assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT-based or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy.

{11}------------------------------------------------

#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE VI.

## Stylet, 23cm

| Item                   | Subject Device<br>(Stylet, 23cm)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Stylet is indicated for use in<br>precisely locating anatomical structures<br>during navigated cranial procedures. It<br>is indicated for use with Medtronic<br>computer-assisted surgery systems using<br>electromagnetic navigation and cranial<br>software.<br><br>The Medtronic computer assisted surgery<br>system and its associated applications are<br>intended as an aid for precisely locating<br>anatomical structures in either open or<br>percutaneous procedures. Their use is<br>indicated for any medical condition in<br>which the use of stereotactic surgery may<br>be appropriate, and where reference to a<br>rigid anatomical structure, such as the<br>skull, can be identified relative to a CT- based or MR-based model, fluoroscopy<br>images, or digitized landmarks of the<br>anatomy. | StealthStation® GoldenEye™ Micro-Magnetic Tracking System Option<br>(K001284)<br><br>The StealthStation® System is intended as an aid for precisely locating anatomical<br>structures in either open or percutaneous procedures. The StealthStation® System is<br>indicated for any medical condition in which the use of stereotactic surgery may be<br>appropriate, and where reference to a rigid anatomical structure, such as the skull, a long<br>bone, or vertebra, can be identified relative to a CT or MR based model or fluoroscopy<br>images of the anatomy.<br><br>Example procedures include, but are not limited to:<br>Cranial Procedures:<br>Cranial biopsies<br>Tumor resections<br>Craniotomies/Craniectomies<br>Skull base procedures<br>Thalamotomies/Pallidotomies<br>ENT Procedures:<br>Transphenoidal procedures |
|                        | The Stylet is indicated for use in the<br>following procedures:<br>• General ventricular catheter<br>placement<br>• Pediatric ventricular catheter<br>placement<br>• Tumor resections<br>• Skull base procedures<br>• Craniotomies/craniectomies<br>• Transsphenoidal procedures                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Catheter Introducer for the StealthStation® System<br>(K022126)<br><br>The StealthStation® System is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The StealthStation® System is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to a CT or MR based model or fluoroscopy images of the anatomy.<br><br>Example procedures include, but are not limited to:<br>Cranial Procedures:<br>Cranial Biopsies<br>Placement<br>Tumor Resections                                                                                                                                                       |

{12}------------------------------------------------

| Craniotomies/ Craniectomies<br>Skull Base procedures<br>Thalamotomies/Pallidotomies<br>Pituitary Tumor Removal<br>CSF Leak Repair<br>Pediatric Catheter Shunt Placement<br>General Catheter Shunt Placement<br>ENT Procedures:<br>Transphenoidal Procedures                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| StealthStation® System Update<br>(K050438)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| The StealthStation® System is intended as an aid for precisely locating anatomical<br>structures in either open or percutaneous procedures. The StealthStation® System is<br>indicated for any medical condition in which the use of stereotactic surgery may be<br>appropriate, and where reference to a rigid anatomical structure, such as the skull, a long<br>bone, or vertebra, can be identified relative to a CT based or MR-based model, fluoroscopy<br>images, or digitized landmarks of the anatomy.<br>For the optical-based and EM-based system, example procedures include, but are not<br>limited to:<br>Cranial Procedures:<br>Cranial Biopsies<br>Tumor Resections<br>Craniotomies/ Craniectomies<br>Skull Base procedures<br>Thalamotomies/Pallidotomies<br>Pituitary Tumor Removal<br>CSF Leak Repair<br>Pediatric Catheter Shunt Placement<br>General Catheter Shunt Placement |
| ENT Procedures:<br>Transphenoidal Procedures                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |

{13}------------------------------------------------

| Establishment<br>of<br>Stereotactic<br>Coordinates<br>(Tracking<br>Method) | Electromagnetic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | StealthStation® GoldenEye™ Micro-Magnetic Tracking System Option<br>(K001284)<br>Electromagnetic<br>Catheter Introducer for the StealthStation® System<br>(K022126)<br>Electromagnetic or Optical<br>StealthStation® System Update<br>(K050438)<br>Electromagnetic or Optical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| System<br>Accuracy                                                         | The system has demonstrated accuracy<br>with a mean positional error of <2mm<br>and mean trajectory error of <2 degrees.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | The system has demonstrated accuracy with a mean positional error of <2mm and mean<br>trajectory error of <2 degrees.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Materials                                                                  | -302 Spring Steel<br>-PET<br>-Loctite Adhesive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Catheter Introducer for the StealthStation® System<br>(K022126)<br>-302 Spring Steel<br>-PET<br>-Loctite Adhesive                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Sterility                                                                  | ETO Sterile/ Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Catheter Introducer for the StealthStation® System<br>(K022126)<br>ETO Sterile/ Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Item                                                                       | Subject Device<br>(Tracer Pointer)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Predicate Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Indications<br>for Use                                                     | The Tracer® Pointer is indicated for use in<br>registering patient anatomy and in precisely<br>locating anatomical structures during navigated<br>cranial procedures in adult and pediatric patients.<br>It is indicated for use with Medtronic computer-<br>assisted surgery systems using electromagnetic<br>navigation and cranial software.<br><br>The Medtronic computer assisted surgery system<br>and its associated applications are intended as an<br>aid for precisely locating anatomical structures in<br>either open or percutaneous procedures. Their use<br>is indicated for any medical condition in which the<br>use of stereotactic surgery may be appropriate, and<br>where reference to a rigid anatomical structure,<br>such as the skull, can be identified relative to a CT-<br>based or MR-based model, fluoroscopy images, or<br>digitized landmarks of the anatomy. | StealthStation® GoldenEye™ Micro-Magnetic Tracking System Option<br>(K001284)<br><br>The StealthStation® System is intended as an aid for precisely locating<br>anatomical structures in either open or percutaneous procedures. The<br>StealthStation® System is indicated for any medical condition in which the use<br>of stereotactic surgery may be appropriate, and where reference to a rigid<br>anatomical structure, such as the skull, a long bone, or vertebra, can be identified<br>relative to a CT or MR based model or fluoroscopy images of the anatomy.<br><br>StealthStation® System Update<br>(K050438)<br><br>The StealthStation® System is intended as an aid for precisely locating<br>anatomical structures in either open or percutaneous procedures. The<br>StealthStation® System is indicated for any medical condition in which the use<br>of stereotactic surgery may be appropriate, and where reference to a rigid<br>anatomical structure, such as the skull, a long bone, or vertebra, can be identified<br>relative to a CT based or MR-based model, fluoroscopy images, or digitized<br>landmarks of the anatomy. |
| Establishment<br>of<br>Stereotactic<br>Coordinates<br>(Tracking<br>Method) | Electromagnetic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | StealthStation® GoldenEye™ Micro-Magnetic Tracking System Option<br>(K001284)<br><br>Electromagnetic<br><br>StealthStation® System Update<br>(K050438)<br><br>Electromagnetic or Optical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| System<br>Accuracy                                                         | The system has demonstrated accuracy with a<br>mean positional error of <2mm and mean<br>trajectory error of <2 degrees.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | The system has demonstrated accuracy with a mean positional error of <2mm<br>and mean trajectory error of <2 degrees.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Materials                                                                  | -303 Stainless Steel<br>-Lustran 348 (ABS Plastic)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | StealthStation® GoldenEye™ Micro-Magnetic Tracking System Option<br>(K001284)<br><br>-Medical Grade SS<br>-Titanium<br>-Plastic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Sterility                                                                  | ETO Sterile/ Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | StealthStation® System Update<br>(K050438)<br>ETO Sterile/ Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

{14}------------------------------------------------

# Tracer Pointer

{15}------------------------------------------------

# Touch-n-Go Pointer

| Item                                                                       | Subject Device<br>(Touch-n-Go Pointer)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Predicate Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use                                                     | The Touch-n-Go Pointer is indicated for use in<br>registering patient anatomy during navigated cranial<br>procedures in adult and pediatric patients. It is<br>indicated for use with Medtronic computer-assisted<br>surgery systems using electromagnetic navigation<br>and cranial software.<br><br>The Medtronic computer assisted surgery system<br>and its associated applications are intended as an aid<br>for precisely locating anatomical structures in either<br>open or percutaneous procedures. Their use is<br>indicated for any medical condition in which the use<br>of stereotactic surgery may be appropriate, and<br>where reference to a rigid anatomical structure, such<br>as the skull, can be identified relative to a CT-based<br>or MR-based model, fluoroscopy images, or<br>digitized landmarks of the anatomy. | StealthStation® GoldenEye™ Micro-Magnetic Tracking System Option<br>(K001284)<br>The StealthStation® System is intended as an aid for precisely locating<br>anatomical structures in either open or percutaneous procedures. The<br>StealthStation® System is indicated for any medical condition in which the use<br>of stereotactic surgery may be appropriate, and where reference to a rigid<br>anatomical structure, such as the skull, a long bone, or vertebra, can be<br>identified relative to a CT or MR based model or fluoroscopy images of the<br>anatomy.<br><br>StealthStation® System Update<br>(K050438)<br>The StealthStation® System is intended as an aid for precisely locating<br>anatomical structures in either open or percutaneous procedures. The<br>StealthStation® System is indicated for any medical condition in which the use<br>of stereotactic surgery may be appropriate, and where reference to a rigid<br>anatomical structure, such as the skull, a long bone, or vertebra, can be<br>identified relative to a CT based or MR-based model, fluoroscopy images, or digitized landmarks of the anatomy. |
| Establishment<br>of<br>Stereotactic<br>Coordinates<br>(Tracking<br>Method) | Electromagnetic<br>The system has demonstrated accuracy with a mean<br>positional error of <2mm and mean trajectory error<br>of <2 degrees.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | StealthStation® GoldenEye™ Micro-Magnetic Tracking System Option<br>(K001284)<br>Electromagnetic<br><br>StealthStation® System Update<br>(K050438)<br>Electromagnetic or Optical<br><br>The system has demonstrated accuracy with a mean positional error of <2mm and mean trajectory error of <2 degrees.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |

{16}------------------------------------------------

| Materials | -303 Stainless Steel<br>-Lustran 348 (ABS Plastic) | <i>StealthStation® GoldenEye™ Micro-Magnetic Tracking System Option</i><br>(K001284) |
|-----------|----------------------------------------------------|--------------------------------------------------------------------------------------|
|           |                                                    | -Medical Grade SS<br>-Titanium                                                       |
|           |                                                    | -Plastic                                                                             |
| Sterility | ETO Sterile/ Single Use                            | <i>StealthStation® System Update</i><br>(K050438)                                    |
|           |                                                    | ETO Sterile/ Single Use                                                              |

# Navigation Pointer

| Item                                                                       | Subject Device<br>(Navigation Pointer)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Item                                                                       | Subject Device<br>(Skull-Mount Patient Tracker)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Predicate Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use                                                     | The Navigation Pointer is indicated for<br>use in precisely locating anatomical<br>structures during navigated cranial<br>procedures. It is indicated for use with<br>Medtronic computer-assisted surgery<br>systems using electromagnetic<br>navigation and cranial software.<br><br>The Medtronic computer assisted<br>surgery system and its associated<br>applications are intended as an aid for<br>precisely locating anatomical structures<br>in either open or percutaneous<br>procedures. Their use is indicated for<br>any medical condition in which the use<br>of stereotactic surgery may be<br>appropriate, and where reference to a<br>rigid anatomical structure, such as the<br>skull, can be identified relative to a CT-<br>based or MR-based model, fluoroscopy<br>images, or digitized landmarks of the<br>anatomy.<br><br>The Navigation Pointer is indicated for<br>use in the following procedures:<br>• General tumor resections | <i>StealthStation® GoldenEye™ Micro-Magnetic Tracking System Option</i><br>(K001284)<br>The StealthStation® System is intended as an aid for precisely locating anatomical<br>structures in either open or percutaneous procedures. The StealthStation® System is<br>indicated for any medical condition in which the use of stereotactic surgery may be<br>appropriate, and where reference to a rigid anatomical structure, such as the skull, a long<br>bone, or vertebra, can be identified relative to a CT or MR based model or fluoroscopy<br>images of the anatomy.<br><br>Example procedures include, but are not limited to:<br>Cranial Procedures:<br>Cranial biopsies<br>Tumor resections<br>Craniotomies/Craniectomies<br>Skull base procedures<br>Thalamotomies/Pallidotomies<br><br>ENT Procedures:<br>Transphenoidal procedures<br><br><i>StealthStation® System Update</i><br>(K050438)<br>The StealthStation® System is intended as an aid for precisely locating anatomical<br>structures in either open or percutaneous procedures. The StealthStation® System is<br>indicated for any medical condition in which the use of stereotactic surgery may be<br>appropriate, and where reference to a rigid anatomical structure, such as the skull, a long | Indications<br>for Use                                                     | The Skull-Mount Patient Tracker is indicated for<br>use as a stereotactic reference frame to track the<br>position of patient anatomy during navigated<br>cranial and ENT procedures. It is indicated for use<br>with Medtronic computer-assisted surgery systems<br>using electromagnetic navigation and cranial and<br>ENT software.<br><br>The Medtronic computer assisted surgery system<br>and its associated applications are intended as an<br>aid for precisely locating anatomical structures in<br>either open or percutaneous procedures. Their use<br>is indicated for any medical condition in which the<br>use of stereotactic surgery may be appropriate, and<br>where reference to a rigid anatomical structure,<br>such as the skull, can be identified relative to a CT-<br>based or MR-based model, fluoroscopy images, or<br>digitized landmarks of the anatomy. | StealthStation® GoldenEye™ Micro-Magnetic Tracking System Option<br>(K001284)<br><br>The StealthStation® System is intended as an aid for precisely locating<br>anatomical structures in eit…

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW/K141833](https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW/K141833)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
