K130608 · Medtronic Navigation, Inc. · HAW · Aug 7, 2013 · Neurology
Device Facts
Record ID
K130608
Device Name
QUADCUT
Applicant
Medtronic Navigation, Inc.
Product Code
HAW · Neurology
Decision Date
Aug 7, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The XPS/IPC System is intended for the incision and removal of soft and hard tissue or bone in general otorhinolaryngology, head and neck, and otoneurological surgery. The Medtronic computer-assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT- or MR-based model, or digitized landmarks of the anatomy. The system and its associated applications should be used only as an adjunct for surgical guidance. They do not replace the surgeon's knowledge, expertise, or judgment.
Device Story
Quadcut is a sterile, single-use surgical blade accessory for the StraightShot M4 handpiece, part of Medtronic's XPS and IPC systems. Each blade features an integrated tracker for electromagnetic navigation. During ENT procedures, a mobile emitter generates a low-energy magnetic field to locate the tracker; the system's Fusion ENT software then displays the blade tip's position within patient image planes (CT/MR) or anatomical renderings. Used by surgeons in clinical settings, the device provides real-time visual guidance to assist in precise tissue/bone removal. It serves as an adjunct to surgical judgment, enhancing accuracy during open or percutaneous procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Testing focused on functional performance, compatibility with Medtronic Navigation systems and Fusion software, and navigational accuracy within simulated real-life use conditions.
Technological Characteristics
Materials: 304L stainless steel blade, ABS hub. Sensing: Electromagnetic tracking via integrated blade tracker and mobile emitter. Connectivity: Integrates with Medtronic XPS/IPC systems and Fusion ENT software. Sterilization: Ethylene Oxide. Form factor: Surgical blade accessory for M4 handpiece.
Indications for Use
Indicated for patients undergoing otorhinolaryngology, head and neck, or otoneurological surgery requiring incision/removal of soft/hard tissue or bone where stereotactic surgical guidance is appropriate and rigid anatomical structures (e.g., skull) can be referenced to CT/MR models or digitized landmarks.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
StealthStation Golden Eye Micro-Magnetic Tracking System Option (K001284)
Quadcut (non-navigated) (Exempt)
Submission Summary (Full Text)
{0}------------------------------------------------
# <130608
# 510(k) Summary
## February 27, 2012
- I. Company: Medtronic Navigation, Inc. 826 Coal Creek Circle Louisville. Colorado 80027 USA Telephone Number: 720-890-3200 Fax Number: 720-890-3500
- Contact: Sharon McDermott Senior Regulatory Affairs Specialist
### Proprietary Trade Name:
Quadcut®, 3.0 mm x 13 cm Quadcut®, 3.4 mm x 13 cm Quadcut®, 4.3 mm x 13 cm
- II. Common Name: Stereotaxic Instrument
- III. Classification Name: Stereotaxic Instrument (21 CFR 882.4560)
- IV. Classification: Class II (21 CFR 882.4560)
- V. Product Code: HAW
#### VI. Product Description:
The navigated Quadcut® is a sterile, single use surgical instrument intended for the incision and removal of soft and hard tissue or bone in general otothinolaryngology, head and neck and otoneurological surgery. Quadcut® is an accessory component of the StraightShot M4 handpiece, which is an integral part of the XPS and IPC Systems. The XPS and IPC systems provide power to drive the blades during ENT procedures.
Quadcut® is intended for attachment to the M4 hand piece for use in conjunction with Fusion ENT software on a Medtronic computer-assisted surgery system. Each blade has a tracker mounted on it to allow for navigation during the ENT surgical procedure. The system's mobile emitter generates a low-energy magnetic field to locate the tracker mounted on the blade. Then, the software displays the location of the blade's tip within multiple patient image planes and other anatomical renderings.
**AUG 07 2013**
{1}------------------------------------------------
#### VII. Indications for Use:
The XPS/IPC System is intended for the incision and removal of soft and hard tissue or bone in general otorhinolaryngology, head and neck, and otoneurological surgery.
The Medtronic computer-assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT - or MR-based model, or digitized landmarks of the anatomy.
The system and its associated applications should be used only as an adjunct for surgical guidance. They do not replace the surgeon's knowledge, expertise, or judgment.
#### VIII. Identification of Legally Marketing Devices (Predicate Devices)
| Description | 510(k) Number | Clearance<br>Date |
|---------------------------------------------------------------------|---------------|-------------------|
| StealthStation Golden Eye Micro-<br>Magnetic Tracking System Option | K001284 | 06/12/2000 |
| Quadcut® (non-navigated), Class I<br>(Product code: EQJ) | Exempt | NA |
{2}------------------------------------------------
| Item | Subject Devices | Predicate Devices |
|----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Quadcut Blades<br>Indications for Use | The XPS/IPC System is intended for the incision and removal<br>of soft and hard tissue or bone in general otorhinolaryngology,<br>head and neck, and otoneurological surgery.<br>The Medtronic computer-assisted surgery system and its<br>associated applications are intended as an aid for precisely<br>locating anatomical structures in either open or percutaneous<br>procedures. Their use is indicated for any medical condition in<br>which the use of stereotactic surgery may be appropriate, and<br>where reference to a rigid anatomical structure, such as the<br>skull, can be identified relative to a CT- or MR-based model,<br>or digitized landmarks of the anatomy.<br>The system and its associated applications should be used<br>only as an adjunct for surgical guidance. They do not replace<br>the surgeon's knowledge, expertise, or judgment. | Medtronic Xomed Quadcut<br>Blades (non-navigated)<br>Class I, 510(k) Exempt<br>Same intended use<br>StealthStation Golden Eye<br>Micro-magentic System.<br>K001284<br>Same intended use |
| Establishment of<br>Stereotactic<br>Coordinates<br>(Tracking Method) | Electromagnetic | StealthStation Golden Eye<br>Micro-magentic System.<br>K001284<br>Same |
| System Accuracy<br>Requirement | Benchtop and simulated environment:<br>95% confidence / 99.5% reliability, as dictated by risk analysis,<br>of $<=$ 3.00 mm. | StealthStation Golden Eye<br>Micro-magentic System.<br>K001284<br>Equivalent |
| Materials - blade | 304L stainless steel | Medtronic Xomed Quadcut<br>Blades (non-navigated)<br>Class 1, 510(k) Exempt<br>Same material |
| M4 Hub | ABS hub | Medtronic Xomed Quadcut<br>Blades (non-navigated)<br>Class I, 510(k) Exempt<br>Same materials |
| Sterilization Method | Ethylene Oxide sterilization | Medtronic Xomed Quadcut<br>Blades (non-navigated)<br>Class 1, 510(k) Exempt<br>Same sterilization method |
#### IX. Comparison of the Technological Characteristics:
The subject devices have the same intended use and technological characteristics as the predicate devices.
{3}------------------------------------------------
#### X. Discussion of the Performance Testing
Testing was completed to ensure the functionality and compatibility with the identified Medtronic products. Specifically, testing with Medtronic Navigation systems and Fusion software was conducted to ensure acceptable navigational accuracy. Test samples were subjected to simulated real-life use conditions during functional testing.
#### XI. Conclusions
.
The navigated Quadcut® has been shown through comparison and testing to be substantially equivalent to the identified predicate devices.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three tail feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002
August 7, 2013
Medtronic Navigation % Ms. Sharon McDermott Senior Regulatory Affairs Specialist 826 Coal Creek Circle Louisville, CO 80027
Re: K130608
Trade/Device Name: Quadcut Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: June 25, 2013 Received: June 25, 2013
Dear Ms. McDermott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Ms. Sharon McDermott
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalTroblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number: K130608
Device Name:
Quadcut®, 3.0 mm x 13 cm Quadcut®, 3.4 mm x 13 cm Quadcut®, 4.3 mm x 13 cm
Quadcut® is an accessory component of the StraightShot M4 handpiece, which is an integral part of the XPS and IPC Systems.
### Indications for Use:
The XPS/IPC System is intended for the incision and removal of soft and hard tissue or bone in general otorhinolaryngology, head and neck, and otoneurological surgery.
The Medtronic computer-assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to a CT- or MR-based model, or digitized landmarks of the anatomy.
The system and its associated applications should be used only as an adjunct for surgical guidance. They do not replace the surgeon's knowledge, expertise, or judgment.
| Prescription Use X | AND/OR | Over-The-Counter Use |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|------------------------|
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>(Part 21 CFR 801 Subpart D) | | (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Eric A. Mann -S
(Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.