← Product Code [HAW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW) · K122096

# KARL STORZ NAVIGATIONPANEL UNIT (NPU) SYSTEM (K122096)

_KARL STORZ Endoscopy-America, Inc. · HAW · Sep 27, 2012 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW/K122096

## Device Facts

- **Applicant:** KARL STORZ Endoscopy-America, Inc.
- **Product Code:** [HAW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW.md)
- **Decision Date:** Sep 27, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The Navigation Panel Unit System (NPU) is an intraoperative image guided localization system that links a freehand probe tracked by a passive marker sensor system to a virtual computer image space on a patient's preoperative diagnostic image data set. The system is intended to be used as a positioning aid for navigation in ENT surgery, including but not limited to endoscopic surgery. The NPU is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure can be identified relative to the radiological imaging-based model . of the anatomy. Surgery procedures include but are not limited to the following: transphenoidal procedures, maxillary antrostomies, ethmoidectomies, sphenoidectomies, sphenoid explorations, turbinate resections, frontal sinusotomies, intranasal procedures, intranasal tumor resections, otologic surgery, and ENT skull base surgery.

## Device Story

NPU System is an intraoperative image-guided localization system; functions as a positioning aid for ENT surgery. Inputs: preoperative diagnostic radiological image data; tracked freehand probe via passive marker sensor system. Operation: links tracked probe position to virtual computer image space; enables real-time visualization of instrument relative to patient anatomy. Used in clinical/OR settings by surgeons. Output: visual navigation display on computer monitor. Benefit: assists surgeon in precise anatomical localization during complex ENT/skull base procedures.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Multi-component system; passive marker sensor tracking; software-based planning and treatment navigation; utilizes preoperative radiological image data; intended for ENT/otologic surgical environments.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- BrainLAB AG: Kolibri Image Guided Surgery System ([K014256](/device/K014256.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K122096

Image /page/0/Picture/1 description: The image shows the logo for Karl Storz Endoscopy. The logo is in black and white and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "Karl Storz Endoscopy" in a smaller font.

#### 510(k) SUMMARY .

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document are accurate and complete to the best of KSEA's knowledge.

| Submitter:           | Karl Storz Endoscopy-America, Inc.<br>2151 E. Grand Avenue<br>El Segundo, CA 90245-5017<br>Phone: (424) 218-8381<br>Fax: (424) 218-8519 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:      | Leigh Spotten<br>Regulatory Affairs Manager<br>Email: leigh.spotten@karlstorz.com                                                       |
| Date Prepared:       | July 13, 2012                                                                                                                           |
| Device Trade Name:   | Karl Storz Navigation Panel Unit (NPU) System                                                                                           |
| Common Name:         | Image Guided Surgery System                                                                                                             |
| Classification Name: | Neurologic stereotaxic instrument                                                                                                       |
| Regulation Number:   | 21 CFR 882.4560                                                                                                                         |
| Product Code:        | HAW                                                                                                                                     |

## Predicate Device(s):

BrainLAB AG: Kolibri Image Guided Surgery System (K014256)

### Device Description:

The Karl Storz Navigation Panel Unit (NPU) System is a multi-component system to be used as a positioning aid for navigation in ENT surgery. The software comprises a planning and treatment program in which an instrument is navigated on the basis of pre-loaded radiological image data.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS) in the USA. The logo features a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

SEP 27 2012

Karl Storz Endoscopy-America, Inc. c/o Mr. Leigh Spotten Regulatory Affairs Manager 2151 E. Grand Ave. El Segundo, CA 90245-5017

Re: K122096

Trade/Device Name: Navigation Panel Unit System (NPU) Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: July 13, 2012 Received: July 16, 2012

Dear Mr. Spotten:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Rutherm

Malvina B. Eydelman, M.I Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

K122096

## Indication for use

## 510(k) Number (if known): K122096

Device Name: Navigation Panel Unit System (NPU)

Indications for use: The Navigation Panel Unit System (NPU) is an intraoperative image guided localization system that links a freehand probe tracked by a passive marker sensor system to a virtual computer image space on a patient's preoperative diagnostic image data set. The system is intended to be used as a positioning aid for navigation in ENT surgery, including but not limited to endoscopic surgery. The NPU is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure can be identified relative to the radiological imaging-based model . of the anatomy. Surgery procedures include but are not limited to the following: transphenoidal procedures, maxillary antrostomies, ethmoidectomies, sphenoidectomies, sphenoid explorations, turbinate resections, frontal sinusotomies, intranasal procedures, intranasal tumor resections, otologic surgery, and ENT skull base surgery.

Prescription Use __ V (21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

lethur Doucit

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K122096

| Prescription Use     |
|----------------------|
| (Per 21 CFR 801.109) |

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW/K122096](https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW/K122096)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
