K100314 · Medacta International S.A. · HAW · Mar 31, 2010 · Neurology
Device Facts
Record ID
K100314
Device Name
IMNS MEDACTA NAVIGATION SYSTEM,33.221000US
Applicant
Medacta International S.A.
Product Code
HAW · Neurology
Decision Date
Mar 31, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The iMNS Medacta Navigation System is intended to be used to support the surgeon during specific orthopedic surgical procedures by providing information on bone resections, instrument and implant positioning during joint replacement. The iMNS Medacta Navigation System provides computer assistance to the surgeon based on anatomical landmarks and other specific data obtained intra-operatively that are used to place surgical instruments. Examples of some surgical procedures include but are not limited to: Total Knee Replacement Minimally Invasive Total Knee Replacement
Device Story
iMNS Medacta Navigation System provides computer-aided surgical navigation for total knee replacement. System uses stereotaxic technology: passive markers attached to patient bones; infrared camera tracks spatial location of markers. System calculates anatomical landmarks (femur head, knee, ankle centers) intra-operatively. Data displayed on monitor in real-time. Surgeon uses system-navigated instruments to guide bone resections and implant positioning. System components: infrared camera acquisition system, computer with proprietary software, interface devices (keyboard, foot pedal, mouse), and manual reusable surgical instruments. Used in OR by surgeons. Provides real-time support to improve positioning accuracy of cutting guides and implants. Surgeon maintains control of all surgical decisions.
Clinical Evidence
Bench testing only. Design verification and validation were conducted per written protocols with pre-defined acceptance criteria. Testing included GUI verification, resection planning and cutting block positioning validation, single/multi-point directional acquisition validation, and hardware/software compatibility stability testing. Simulated user setting validation performed by a surgeon confirmed the system meets user needs.
Technological Characteristics
Stereotaxic optical tracking system using infrared cameras and passive markers. Components: infrared camera acquisition system, computer, monitor, interface devices (keyboard, foot pedal, mouse), and manual reusable surgical instruments. Software: GMK v4.2.2 application. Connectivity: Standalone system. Sterilization: Reusable instruments handled per standard orthopedic protocols.
Indications for Use
Indicated for patients undergoing orthopedic joint replacement surgery, specifically Total Knee Replacement or Minimally Invasive Total Knee Replacement, requiring intra-operative guidance for bone resection and implant positioning.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "K100314" at the top, followed by the word "Medacta" in a stylized font. Below "Medacta" is the word "International" on the left and a plus sign on the right. The text and symbols appear to be part of a logo.
MAR 8 1 2010
# Special 510(k) Summary
Manufacturer:
MEDACTA International SA Strada Regina CH6874 Castel San Pietro Switzerland Phone (+41) 91 696 60 60 FAX (+41) 91 696 60 66
Contact Person:
Heather Neelv Director RA/QA 4725 Calle Quetzal, Suite B Camarillo, CA 93012 USA Phone: 805-886-1069 Fax: 877-550-4749 e-mail: HNeely@medacta.us.com
Date Prepared:
January 29, 2010
#### DEVICE INFORMATION
Trade/Proprietary Name: iMNS Medacta Navigation System Common/Classification Name:
> 21 CFR 882.4560 Class II Device Product Code: HAW
K083872 iMNS Medacta Navigation System Predicate Device:
### Product Description:
The iMNS Medacta Navigation System is a device for computer aided navigation of surgical instruments used in total knee replacement surgery. The system works on the common principle of stereotaxic technology in which passive markers are securely mounted on the patient's bones and an infrared camera is used to monitor the spatial location of those markers. This information is used to locate the anatomical landmarks such as centers of rotation of the femur head, knee and ankle intraoperatively. These measurements are displayed on a computer screen in real time. The instruments are then outfitted with the passive markers to improve the positioning of the cutting guides. The information from the system with the "navigated" instruments assists the surgeon in conducting the bone resections and positioning of the orthopedic surgical implants. The surgeon maintains control of the surgery and makes any decisions required with
{1}------------------------------------------------
regard to bone resections and implant positioning but the iMNS Medacta Navigation System provides real time support and information throughout the surgery.
The iMNS Medacta Navigation System consists of the following kev components:
- · An acquisition system composed of two infrared cameras equipped . with infrared light emitting diodes (LED) to track the position of the passive markers.
- . A computer running the proprietary Medacta software and a monitor,
- . Interface devices of a keyboard, foot pedal and optional mouse to control the system, and
- Manual reusable surgical instruments. .
The software application called GMK v4.2.2 is designed to work with Medacta's GMK Total Knee System, cleared under K090988. The manual reusable surgical instruments include instruments specifically designed for navigated surgery and other standard surgical instruments needed to conduct total knee replacement.
#### Indications for Use:
The iMNS Medacta Navigation System is intended to be used to support the surgeon during specific orthopedic surgical procedures by providing information on bone resections, instrument and implant positioning during joint replacement.
The iMNS Medacta Navigation System provides computer assistance to the surgeon based on anatomical landmarks and other specific data obtained intra-operatively that are used to place surgical instruments.
Examples of some surgical procedures include but are not limited to:
Total Knee Replacement
Minimally Invasive Total Knee Replacement
#### Performance Testing
No performance standards applicable to this device have been adopted under Section 514 of the Food, Drug and Cosmetic Act. Performance testing of the iMNS Medacta Navigation System was conducted in accordance with design controls, international standards and FDA guidance documents.
## Comparison to Predicate Device
The indications for use of the modified system remain the same as the original 510(k) K083872.
The iMNS Medacta Navigation System v4.2.2 is identical to the predicate device, iMNS Medacta Navigation System v4.0 in the following aspects: Intended Use/Indications for Use, design features, operating principle, type of optical tracking system, graphical user interface, control switch, passive locators, navigated surgery process, sterility, and biocompatibility aspects.
{2}------------------------------------------------
The main difference between the iMNS Medacta Navigation System v4.2.2 and the predicate device is the software application. Previously there was only one software application, Evolis Global Femur First version 4.0 (K083872) and the second software application is GMK version 4.2.2 which is the subject of this 510(k).
The manual reusable surgical instruments have changed to those required by the non-navigated GMK Total Knee System. Additionally, there are separate micrometric positioners that are specific to GMK included with the instruments for use with navigated knee surgery and the iMNS Medacta Navigation System. These manual surgical instruments are made of the same type of materials as those in the original submission and handled the same as other reusable manual orthopedic instruments for non-navigated surgery.
The iMNS Medacta Navigation System was tested as part of design verification and validation to written protocols with pre-defined acceptance criteria. The testing met all acceptance criteria. This new software version has been completely verified and validated like an initial software release. The verification and validation activities are summarized in the below.
| Summary of Design Controls for Software Changes v 4.1.0 and v 4.2.1 | | |
|---------------------------------------------------------------------|------------------|---------------------------------------------|
| Protocol ID | Report ID | Title |
| IL 07.09.026 | EG08 and Annex 1 | GUI Verification |
| | EG13 and Annex 1 | |
| IL 07.09.028 | EG09 | Validation of resections planning and |
| | EG11 | cutting blocks positioning |
| IL 07.09.029 | EG12 | Single Point/ Multi Point/ Directional |
| | | Acquisition's Validation |
| IL 07.09.048 | EG10 | Functionality of the Navigation Software in |
| | | Deep Stress Cases |
| Summary of Design Controls for Software Changes v 4.2.2 | | |
|---------------------------------------------------------|-----------|------------------------------------------------------------|
| Protocol ID | Report ID | Title |
| IL07.09.026 | GMK00 | GUI Verification |
| IL07.09.027 | GMK06 | Mechanical axis and saggital plane<br>reconstruction |
| IL07.09.028 | GMK02 | Resection planning & cutting block<br>positioning |
| IL07.09.029 | GMK03 | Single point/multipoint directional<br>acquisition |
| IL07.09.030 | GMK05 | Single point acquisition |
| IL07.09.031 | GMK01 | Geometry of knee implants & ancillaries |
| IL07.09.092 | GMK04 | Validation of HW &SW compatibility &<br>relative stability |
Design validation was conducted on the iMNS Medacta Navigation System in a simulated user setting by a surgeon and demonstrated that the system meets user needs and intended uses.
{3}------------------------------------------------
There were no changes to the hardware and trolley with this device modification, there were no significant changes to the hardware system risk analysis or validation/verification.
#### Conclusion:
The data and information provided in this submission support the conclusion that the iMNS Medacta Navigation System (GMK v4.2.2) is substantially equivalent to its predicate device, iMNS Medacta Navigation System (Evolis v4.0) with respect to indications for use, operating principle, and technological characteristics.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name arranged in a circular pattern around the perimeter. In the center of the seal is a stylized depiction of a bird, which is the symbol of the HHS. The bird is shown with its wings spread, suggesting protection and care.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
MAR 3 1 2010
Medacta International SA % Medacta USA Ms. Heather Neely Director, RA/QA 4725 Calle Quetzal, Suite B Camarillo, California 93012
Re: K100314
Trade/Device Name: iMNS Medacta Navigation System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: HAW Dated: March 18, 2010 Received: March 22, 2010
Dear Ms. Neely:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
Page 2 - Ms. Heather Neely
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office _ of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark N Wilkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known):
Device Name: iMNS Medacta Navigation System
Klock314 Indications for Use:
The iMNS Medacta Navigation System is intended to be used to support the surgeon during specific orthopedic surgical procedures by providing information on bone resections, instrument and implant positioning during joint replacement.
The iMNS Medacta Navigation System provides computer assistance to the surgeon based on anatomical landmarks and other specific data obtained intra-operatively that are used to place surgical instruments.
Examples of some surgical procedures include but are not limited to: Total Knee Replacement
Minimally Invasive Total Knee Replacement
Over-The-Counter Use Prescription Use _ × AND/OR (Over-The-Counter Use __
AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R. Ogle for mxm
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K1000314
Page
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.