← Product Code [HAW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW) · K060556

# BRIGIT SURGICAL DEVICE (K060556)

_Zimmer, Inc. · HAW · Jul 31, 2006 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW/K060556

## Device Facts

- **Applicant:** Zimmer, Inc.
- **Product Code:** [HAW](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW.md)
- **Decision Date:** Jul 31, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

BRIGIT Surgical Device is a computer controlled electromechanical multi-jointed arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide to be used by the surgeon to manually guide orthopedic instruments. It is indicated for any orthopedic medical condition in which the use of stereotaxic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a long bone, can be identified relative to digitized landmarks of the anatomy. Example orthopedic procedures for these instruments include, but are not limited to: - Total Knee Arthroplasty (Primary and Revision) - Unicompartmental Knee Arthroplasty (Primary and Revision) - Minimally Invasive Knee Orthopedic Procedures - Total or Hemi-Hip Arthroplasty (Primary and Revision) - Minimally Invasive Hip Orthopedic Procedures - Tumor Resection and Bone/Joint Reconstruction - Stabilization or Repair of Pelvic/Femoral Fractures

## Device Story

BRIGIT Surgical Device is a computer-controlled, electromechanical, multi-jointed arm used in operating rooms. It assists surgeons by determining reference alignment axes relative to anatomical landmarks, planning instrument positions, and providing stable, accurate mechanical guidance for orthopedic instruments. The device does not contact the patient directly. In 'cooperative mode,' surgeons manually move the arm by grabbing the tip to collect anatomical landmarks. The system calculates guide positioning based on 3D surgical planning parameters and digitized anatomical landmarks. It provides rigid, steady support for tool guides, enhancing accuracy and reproducibility during procedures like arthroplasty, tumor resection, and fracture repair. The surgeon retains manual control of the instruments while the arm maintains the planned orientation.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance validation included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-2), software verification per FDA guidance, and device accuracy/repeatability testing.

## Technological Characteristics

Computer-controlled electromechanical multi-jointed arm. Features include cooperative mode for manual positioning and 3D spatial calculation engine. Connectivity/integration supports digitized landmark referencing. Electrical safety per IEC 60601-1; electromagnetic compatibility per IEC 60601-2. Software-controlled guidance system.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Neuromate Stereotactic system ([K963256](/device/K963256.md))
- Navitrack System - S&N Image Free Knee ([K043536](/device/K043536.md))
- Medtronic Stealth Station ([K050438](/device/K050438.md))
- Zimmer Ortho Guidance Systems - Knee Instruments ([K033011](/device/K033011.md))
- Zimmer Ortho Guidance Systems - Hip Instruments ([K033223](/device/K033223.md))
- Zimmer Computer Assisted Solutions - Electromagnetic and Imageless Knee Instrumentation ([K052425](/device/K052425.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

bos16

Page 1 of 3

### Summary of Safety and Effectiveness

JUL 31 2006

Submitter:

Contact Person:

Date:

Trade Name:

Common Name:

Classification Name and Reference:

Predicate Device:

Zimmer, Inc. P.O. Box 708 Warsaw, IN 46581-0708

Stephen H. McKelvey Manager, Corporate Regulatory Affairs Telephone: (574) 372-4944 Fax: (574) 372-4605

July 26, 2006

BRIGIT Surgical Device

Computer assisted surgical device

Stereotaxic instrument, 21 CFR § 882.4560

Neuromate Stereotactic system, manufactured by Integrated Surgical Systems, Inc., K963256, cleared May 9, 1997.

Navitrack System - S&N Image Free Knee, manufactured by Orthosoft, K043536, cleared January 14, 2005.

Medtronic Stealth Station, manufactured by Medtronic, K050438, cleared June 2, 2005.

Zimmer Ortho Guidance Systems - Knee Instruments, manufactured by Zimmer, Inc., K033011, cleared February 12, 2004.

Zimmer Ortho Guidance Systems - Hip Instruments, manufactured by Zimmer, Inc., K033223, cleared February 18, 2004.

Zimmer Computer Assisted Solutions -Electromagnetic and Imageless Knee Instrumentation, manufactured by Zimmer, Inc., K052425, cleared December 28, 2005.

006

{1}------------------------------------------------

## Device Description:

BRIGIT Surgical Device is a computer controlled electromechanical multi-jointed arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide to be used by the surgeon to manually guide orthopedic instruments.

BRIGIT Surgical Device assists the surgeon in:

- determining reference alignment axes in " relation to anatomical landmarks,
- planning the position of instruments relative to these axes,
- and providing a stable, accurate and reproducible mechanical guidance of surgical instruments in accordance with said planning, while not directly entering in contact with the patient.

BRIGIT Surgical Device can be shifted into a "cooperative mode" where the surgeon can manually move the arm anywhere in the operating field by simply grabbing the tip. Pinpoint collection of anatomical landmarks is carried out with BRIGIT Surgical Device and its pointing instrument in cooperative mode.

Adequate position of the guide is derived from three-dimensional calculations performed from desired surgical planning parameters and spatial positions of anatomical landmarks.

BRIGIT Surgical Device offers the rigidity essential for accurate and steady support of the guide.

BRIGIT Surgical Device is a computer controlled electromechanical multi-jointed arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide to be used by the surgeon to manually guide orthopedic instruments.

It is indicated for any orthopedic medical condition in which the use of stereotaxic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a long bone, can be

Intended Use:

{2}------------------------------------------------

060556

Pages of 3

identified relative to digitized landmarks of the anatomy.

Example orthopedic procedures for these instruments include, but are not limited to:

- · Total Knee Arthroplasty (Primary and Revision)
- · Unicompartmental Knee Arthoplasty (Primary and Revision)
- · Minimally Invasive Knee Orthopedic Procedures
- · Total or Hemi-Hip Arthroplasty (Primary and Revision)
- · Minimally Invasive Hip Orthopedic Procedures
- · Tumor Resection and Bone/Joint Reconstruction
- · Stabilization or Repair of Pelvic/Femoral Fractures

The technological characteristics, features, specifications, materials, mode of operation, and intended use of the BRIGIT Surgical Device are substantially equivalent to the predicate devices listed above.

The differences that exist between the devices do not raise new issues of safety or effectiveness regarding the BRIGIT Surgical Device.

Testing was carried out to assure compliance with recognized electrical safety standards: IEC 60601-1 standard for electrical safety and IEC 60601-2 standard for electromagnetic compatibility.

Tests were also carried out to satisfy the requirements of the FDA Guidance for the Content of Premarket Submisions for Software Contained in Medical Devices.

Device accuracy tests were performed to validate the accuracy and repeatability of the device.

Comparison to Predicate Device:

#### Performance Data

{3}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Zimmer, Inc. % Mr. Stephen H. McKelvey Manager, Corporate Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581-0708

Re: K060556

Trade/Device Name: BRIGIT Surgical Device Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: HA W Dated: July 6, 2006 Received: July 7. 2006

Dear Mr. McKelvey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

JUL 3 1 2006

Image /page/3/Picture/11 description: The image shows the logo for the U.S. Department of Health and Human Services (HHS). The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is composed of three parallel lines that curve and flow together. The logo is encircled by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA".

{4}------------------------------------------------

Page 2 – Mr. Stephen H. McKelvey

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Herbert Lemmerno

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

## 510(k) Number (if known): K060556

Device Name:

BRIGIT Surgical Device

### Indications for Use:

BRIGIT Surgical Device is a computer controlled electromechanical multi-jointed arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide to be used by the surgeon to manually guide orthopedic instruments.

It is indicated for any orthopedic medical condition in which the use of stereotaxic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a long bone, can be identified relative to digitized landmarks of the anatomy.

Example orthopedic procedures for these instruments include, but are not limited to:

- · Total Knee Arthroplasty (Primary and Revision)
- · Unicompartmental Knee Arthroplasty (Primary and Revision)
- · Minimally Invasive Knee Orthopedic Procedures
- · Total or Hemi-Hip Arthroplasty (Primary and Revision)
- · Minimally Invasive Hip Orthopedic Procedures
- · Tumor Resection and Bone/Joint Reconstruction
- · Stabilization or Repair of Pelvic/Femoral Fractures

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(Please do not write below this line - Continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Huler Lemur
(Division Sign-Off)

Division of General, Restorative, and Neurological Devices

Page 1 of 1

510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW/K060556](https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAW/K060556)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
