← Product Code [HAE](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAE) · K221818

# Rebellion; Phantom Multi-Bite Kerrison Rongeur (K221818)

_Morpheus AG · HAE · Oct 7, 2022 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAE/K221818

## Device Facts

- **Applicant:** Morpheus AG
- **Product Code:** [HAE](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAE.md)
- **Decision Date:** Oct 7, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.4840
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Rebellion is intended for cutting and removing bone, vertebral bodies and tissue in orthopedics, neurosurgery and spine surgery involving the skull or spinal column.

## Device Story

Manual bone punch (Kerrison rongeur) used in orthopedic, neuro, and spine surgery; operated by surgeons. Device features integrated Luer-Lock connection for NaCl 0.9% rinsing solution to prevent tube clogging; posterior tube connects to bone/tissue trap or standard suction system. Removes bone/tissue via manual biting action; suction facilitates debris removal. Single-use, sterile device; 160 mm length; 2-4 mm bite sizes; 90° up-biting angulation. Benefits include efficient bone removal and reduced clogging via integrated irrigation/suction.

## Clinical Evidence

No clinical performance testing submitted. Substantial equivalence is based on non-clinical performance data and comparison to the predicate device.

## Technological Characteristics

Manual rongeur; stainless steel (420, 304, 302) and polymers (PVC, ABS, PA6 GF60, TPE, fluoropolymer). Features integrated Luer-Lock for irrigation and suction port for bone/tissue trap. 160 mm shaft length; 14 mm jaw opening; 2-4 mm bite size; 90° up-biting. Sterile, single-use.

## Regulatory Identification

A manual rongeur is a manually operated instrument used for cutting or biting bone during surgery involving the skull or spinal column.

## Predicate Devices

- Rebellion; Phantom Multi-Bite Kerrison Rongeur ([K200768](/device/K200768.md))

## Reference Devices

- Fehling-punches ([K153243](/device/K153243.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 7, 2022

Morpheus AG Timo Rack, CEO Bahnhofstraße 18 Spaichingen, Baden Würrtemberg 78549 Germany

Re: K221818

Trade/Device Name: Rebellion; Phantom Multi-Bite Kerrison Rongeur Regulation Number: 21 CFR 882.4840 Regulation Name: Manual Rongeur Regulatory Class: Class II Product Code: HAE Dated: August 30, 2022 Received: September 8, 2022

Dear Timo Rack:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K221818

Device Name Rebellion; Phantom Multi-Bite Kerrison Rongeur

Indications for Use (Describe)

The Rebellion is intended for cutting and removing bone, vertebral bodies and tissue in orthopedics, neurosurgery and spine surgery involving the skull or spinal column.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Morpheus AG Rebellion 510(k) Premarket Notification

Image /page/3/Picture/2 description: The image is a logo for Morpheus AG, a company that specializes in inspiring medical technology. The logo features a stylized head with a pixelated effect on the right side, followed by the company name "MORPHEUS AG" in bold, sans-serif font. Below the company name, the tagline "INSPIRING MEDICAL TECHNOLOGY" is written in a smaller, sans-serif font.

#### DATE OF APPLICATION: 10/07/2022

APPLICANT:

Morpheus AG Bahnhofstrasse 18 78549 Spaichingen Germany Tel: + 49 (0) 7461 70382-0 E-Mail: info@morpheus.com.ag

CONTACT PERSON:

Timo Rack CEO Tel.: + 49 (0) 7461 70382-11 E-Mail: t.rack@morpheus.com.ag

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for Morpheus AG. The logo consists of a stylized head in profile on the left, with a series of squares emanating from the back of the head. To the right of the head is the text "MORPHEUS AG" in bold, followed by the text "INSPIRING MEDICAL TECHNOLOGY" in a smaller font.

#### 1 Device Name

| Trade Name:                 | Rebellion;<br>Phantom Multi-Bite Kerrison Rongeur |
|-----------------------------|---------------------------------------------------|
| Common Name:                | Kerrison Rongeur                                  |
| Device Classification Name: | Manual rongeur                                    |

#### Classification / Product Code 2

Rebellion can be classified according to following device name and product code:

| Device    | Regulation<br>Description | Regulation<br>Medical Specialty | Product Code | Regulation<br>Number | Device<br>Classification |
|-----------|---------------------------|---------------------------------|--------------|----------------------|--------------------------|
| Rebellion | Manual rongeur            | Neurology                       | HAE          | 21 CFR<br>882.4840   | II                       |

#### Predicate Device / Reference Device 3

| Subject Device | Predicate Device                                     | Reference Device | 510(k) number | 510(k) Holder                        |
|----------------|------------------------------------------------------|------------------|---------------|--------------------------------------|
| Rebellion      | Rebellion;<br>Phantom Multi-Bite<br>Kerrison Rongeur | --               | K200768       | Morpheus AG                          |
|                | --                                                   | Fehling-punches  | K153243       | Fehling Instruments<br>GmbH & Co. KG |

#### 4 Device Description

The Rebellion is a Bone Punch and it is available in the length 160 mm with tree different tip sizes (2, 3 and 3 mm). The Rebellion can be connected to a rinsing solution (NaCl 0,9%) via integrated Luer-Lock connection. The tubes are flushed with the rinsing solution, so a clogging of the tube is prevented. At posterior tube the Rebellion can be connected to the Morpheus Bone and Tissue Trap or to a standard suction system via sterile tubing.

Image /page/4/Picture/12 description: The image shows a tool that is labeled as "Figure 1: Rebellion". The tool has a long, white barrel with a black handle and trigger. There is also a gray cable coming out of the handle.

Figure 1: Rebellion

{5}------------------------------------------------

Morpheus AG Rebellion 510(k) Premarket Notification

Image /page/5/Picture/2 description: The image shows the logo for Morpheus AG, a company focused on inspiring medical technology. The logo features a stylized silhouette of a human head in profile, with a pixelated or fragmented design emanating from the back of the head. To the right of the head is the company name, "MORPHEUS AG," in bold, sans-serif font, with the tagline "INSPIRING MEDICAL TECHNOLOGY" underneath in a smaller font size.

#### Intended Use 5

The Rebellion is intended for cutting and removing bone, vertebral bodies and tissue in orthopedics, neurosurgery and spine surgery involving the skull or spinal column.

#### Technological Characteristics e

Morpheus Rebellion possesses similar technological characteristics as compared to the predicate device K200768 and reference device K153243. Different characteristics do not raise different questions of safety and effectiveness, and scientific methods were applied to evaluate different characteristics' effects on safety and effectiveness.

| Description            | Subject Device                                                                                                                                                               | Predicate Device<br>Morpheus AG<br>(Rebellion)                                                                                                                               | Reference Device<br>Fehling Instruments GmbH<br>& Co. KG<br>(Fehling-punches)                                                                                                 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)                 | --                                                                                                                                                                           | K200768                                                                                                                                                                      | K153243                                                                                                                                                                       |
| Product Code           | HAE                                                                                                                                                                          | HAE                                                                                                                                                                          | HAE                                                                                                                                                                           |
| Class                  | II                                                                                                                                                                           | II                                                                                                                                                                           | II                                                                                                                                                                            |
| Regulation #           | 882.4840                                                                                                                                                                     | 882.4840                                                                                                                                                                     | 882.4840                                                                                                                                                                      |
| Classification<br>Name | Manual, Rongeur                                                                                                                                                              | Manual, Rongeur                                                                                                                                                              | Manual, Rongeur                                                                                                                                                               |
| Indication for Use     | The Rebellion is indicated to cut<br>and remove bone, vertebral<br>body and tissue in the<br>orthopedic, neuro and spine<br>surgery involving the skull or<br>spinal column. | The Rebellion is indicated to<br>cut and remove bone,<br>vertebral body and tissue in<br>the orthopedic, neuro and<br>spine surgery involving the<br>skull or spinal column. | Fehling rongeurs (bone<br>punches) are manually<br>operated instruments<br>indicated for cutting or<br>biting bone during surgery<br>involving the skull or<br>spinal column. |
| Anatomical<br>location | Skull and spine                                                                                                                                                              | Skull and spine                                                                                                                                                              | Skull and spine                                                                                                                                                               |
| Sterility              | Sterile                                                                                                                                                                      | Sterile                                                                                                                                                                      | Non-Sterile                                                                                                                                                                   |
| Re-Use                 | No                                                                                                                                                                           | No                                                                                                                                                                           | Yes                                                                                                                                                                           |
| Material               | Stainless Steel: 420, 304 and<br>302<br>Polymer: PVC, ABS, PA6 GF60<br>(Grivory GV-6H); TPE<br>(TM5MED) and Fluoropolymer<br>(Altera MT1000A)                                | Stainless Steel: 420, 304 and<br>302<br>Polymer: PVC, ABS, PA6 GF60<br>(Grivory GV-6H); TPE<br>(TM5MED) and Fluoropolymer<br>(Altera MT1000A)                                | 420 and 304 Stainless<br>Steels;<br>Coating: CERAMO®<br>(TiAIN)<br>Polymer: Silicone                                                                                          |
| Patient<br>Contacting  | Stainless Steel: 420, 304<br>Polymer: Fluoropolymer (Altera<br>MT1000A)                                                                                                      | Stainless Steel: 420, 304<br>Polymer: Fluoropolymer<br>(Altera MT1000A)                                                                                                      | 420 and 304 Stainless<br>Steels; CERAMO® (TiAIN)                                                                                                                              |
| Design features        | Manual Rongeur                                                                                                                                                               | Manual Rongeur                                                                                                                                                               | Manual Rongeur                                                                                                                                                                |
| Bone Removal           | Suctional function                                                                                                                                                           | Suctional function                                                                                                                                                           | Manual                                                                                                                                                                        |
| Shaft-length           | 160 mm                                                                                                                                                                       | 200 - 250 mm                                                                                                                                                                 | 110 - 400 mm                                                                                                                                                                  |
| Jaw opening            | 14 mm                                                                                                                                                                        | 14 mm                                                                                                                                                                        | 9 - 19 mm                                                                                                                                                                     |
| Bite size              | 2 - 4 mm                                                                                                                                                                     | 2 - 4 mm                                                                                                                                                                     | 0.8 - 8 mm                                                                                                                                                                    |
| Description            | Subject Device                                                                                                                                                               | Predicate Device<br>Morpheus AG<br>(Rebellion)                                                                                                                               | Reference Device<br>Fehling Instruments GmbH<br>& Co. KG<br>(Fehling-punches)                                                                                                 |
| Cutting<br>angulation  | 90° up-biting                                                                                                                                                                | 90° up-biting                                                                                                                                                                | 40° and 90° up/down bi-<br>ting                                                                                                                                               |

## 6.1 Device Characteristics Table

{6}------------------------------------------------

## Morpheus AG Rebellion 510(k) Premarket Notification

Image /page/6/Picture/2 description: The image shows the logo for Morpheus AG, a company focused on inspiring medical technology. The logo features a stylized silhouette of a human head with a pixelated effect emanating from the back of the head. To the right of the head is the company name "MORPHEUS AG" in bold, uppercase letters, with the tagline "INSPIRING MEDICAL TECHNOLOGY" underneath in a smaller font.

#### Performance Data 7

# 7.1 Clinical performance testing

Clinical performance testing was not submitted in this 510(k).

# 7.2 Non-Clinical performance testing

The subject device does not represent a new worst-case when compared to the previously cleared Morpheus Rebellion Rongeur (K200768).

#### Conclusion 8

In accordance with 21 CFR Part 807 and based on the information provided in this premarket notification, Morpheus AG concludes that the Rebellion Rongeur is as safe and as effective for their intended use as the predicate device, the Rebellion Rongeur (K200768).

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAE/K221818](https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAE/K221818)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
