K152734 · Vitalitec Medizintechnik GmbH · HAE · Jun 18, 2016 · Neurology
Device Facts
Record ID
K152734
Device Name
Kerrison Rongeurs
Applicant
Vitalitec Medizintechnik GmbH
Product Code
HAE · Neurology
Decision Date
Jun 18, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4840
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Vitalitec Kerrison Rongeurs are manually operated reusable surgical instruments used for cutting or biting bone during surgery involving the skull or spinal column.
Device Story
Vitalitec Kerrison Rongeurs are manual, reusable surgical instruments used by surgeons in neurosurgical and orthopedic procedures. The device features a scoop-shaped cutting tip, a shaft, and a movable handle. Operation involves the surgeon manually activating the handle to move the jaws, which bite and cut bone tissue during skull or spinal surgery. The device is designed to be detachable for cleaning, with the upper shaft moving sideways to facilitate disassembly. It is supplied non-sterile and must be sterilized by the user before clinical use. The device provides a mechanical means for bone removal, assisting the surgeon in dissecting bone and tissue. Benefits include robust construction and reusability for multiple surgical interventions.
Clinical Evidence
No clinical data was used to determine substantial equivalence. Evidence is based on bench testing, including mechanical cutting performance testing comparing the subject device to the predicate using DIN 58298 standards, and adherence to recognized standards for sterilization (AAMI/ANSI ST81, ISO 17665-1), biocompatibility (ISO 10993-1), and risk management (ISO 14971).
Technological Characteristics
Materials: Stainless steel 1.4021 (420 SS) per DIN EN ISO 7153-1; Titanium-aluminium-nitride (TiAIN) surface coating. Dimensions: 1-6 mm bite sizes; 14-16 mm jaw openings; 40° and 90° cutting angles. Principle: Manual mechanical cutting/biting via scoop-shaped tip. Connectivity: None. Sterilization: User-sterilizable (moist heat).
Indications for Use
Indicated for patients undergoing neurosurgical or orthopedic procedures involving the skull or spinal column requiring the cutting or biting of bone.
Regulatory Classification
Identification
A manual rongeur is a manually operated instrument used for cutting or biting bone during surgery involving the skull or spinal column.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right, suggesting a connection to health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 18, 2016
Vitalitec Medizintechnik GmbH % Ms. Lynette Howard President Lyle Howard Corporation 106 East 5th Avenue Mount Dora, Florida 32757
Re: K152734
Trade/Device Name: Vitalitec Kerrison Rongeurs Regulation Number: 21 CFR 882.4840 Regulation Name: Manual Rongeur Regulatory Class: Class II Product Code: HAE Dated: May 17, 2016 Received: May 20, 2016
Dear Ms. Howard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -\$=D/Δ
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152734
Device Name Vitalitec Kerrison Rongeurs
Indications for Use (Describe)
Vitalitec Kerrison Rongeurs are manually operated reusable surgical instruments used for cutting or biting bone during surgery involving the skull or spinal column.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
### Device Submitter
Vitalitec Medizintechnik GmbH Steigäcker 20 Balgheim (Baden-Wurttemberg) 78582 Germany
Phone: 908.788.4580 Fax: 352.383.8338
Contact Person: Lynette Howard Date Prepared: June 16, 2016
Establishment Registration Number: 8044057
### Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
Trade Name: Vitalitec Kerrison Rongeurs Common Name: Kerrison Rongeurs Classification Name: Class II, 21 CFR 882.4840, Product Code: HAE
#### Predicate Device [21 CFR 807.92(a)(3)]
For this submission, Vitalitec Medizintechnik GmbH has chosen the Integra LifeSciences Corporation device, IntegraTM Kerrison Rongeurs, cleared under 510(k) K092227.
The Integra LifeSciences Corporation rongeur is a class II device according to 21 CFR 882.4840. Product Code HAE.
#### Description of the Device [21 CFR 807.92(a)(4)]
The Vitalitec Medizintechnik GmbH Kerrison Rongeur is a class II device according to 21 CFR 882.4840, Product Code HAE.
The Vitalitec Medizintechnik GmbH Kerrison Rongeurs are reusable stainless steel instruments that are sterilizable and packaged non-sterile. Devices are available with the following features: 1-6 mm bite sizes; 14 mm - 16 mm jaw openings; 40° and 90° up/down cutting angles; regular and thin footplates; various handle and shaft styles; sliding shaft; detachable.
A rongeur is a robust constructed reusable surgical instrument with sharp, spoon shaped working ends, which used for cutting or biting bone during surgery involving the skull or spinal
{4}------------------------------------------------
column.In Neurosurgery a rongeur is used for grasping and dissecting of bone and tissue during neurosurgical and orthopedic procedures.
## Indications for Use
Vitalitec Kerrison Rongeurs are manually operated reusable surgical instruments used for cutting or biting bone during surgery involving the skull or spinal column.
## Technological Characteristics [21 CFR 807.92(a)(6)]
The subject device is similar in design, indications for use, and mechanical performance to the predicate device.
| Characteristic | Subject | Predicate |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | K152734 | K092227 |
| Class | II | II |
| Product Code | HAE | HAE |
| Regulation # | 882.4840 | 882.4840 |
| Classification<br>Name | Manual, Rongeur | Manual, Rongeur |
| Trade Name | Vitalitec Kerrison Rongeurs | Integra™ (Jarit®, Ruggles™, R&B<br>Redmond™, (Redmond™),<br>Miltex®, MeisterHand®) Kerrison<br>rongeurs |
| Indications for Use | Vitalitec Kerrison Rongeurs are<br>manually operated reusable<br>surgical instruments used for<br>cutting or biting bone during<br>surgery involving the skull or<br>spinal column. | Integra™ (Jarit®, Ruggles™, R&B<br>Redmond™, (Redmond™),<br>Miltex®, MeisterHand®) Kerrison<br>rongeurs are manually operated<br>instruments indicated for cutting or<br>biting bone during surgery involving<br>the skull or spinal column. |
| Materials - Patient<br>Contact -<br>Transient Use | Handle, Shaft, Jaw Parts:<br>Stainless steel 1.4021 (420 SS)<br>composition X20Cr13 according<br>to DIN EN ISO 7153-1<br>Surface Coating:<br>Titanium-aluminium-nitride<br>coating (TiAIN) | 420 Stainless steel<br>Surface Coatings:<br>Titanium-Diamond coat & Hard<br>coat™-nitride coating (TiAIN);<br>Titanium Nitride (TiN); Stealth coat<br>& Smooth-coat™ (Dicronite DL-5);<br>Ultra-coat™ (ZrN); PTFE<br>(polytetrafluoroethylene (Teflon®);<br>Silicon (Elastosil® LR3003/80 A,B) |
| Packaging | Shipped non-sterile | Shipped non-sterile |
| Type of Device | Reusable Surgical Instrument | Reusable Surgical Instrument |
{5}------------------------------------------------
| Characteristic | Subject | Predicate |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | K152734 | K092227 |
| Class | II | II |
| Product Code | HAE | HAE |
| Regulation # | 882.4840 | 882.4840 |
| Classification | Manual, Rongeur | Manual, Rongeur |
| Name | | |
| Sterility | Sterilizable | Sterilizable |
| Utility | Reusable | Reusable |
| Technological<br>Principles | Rongeurs (Kerrisons) are manual<br>surgical instruments constructed<br>with a scoop-shaped cutting tip<br>which also compresses and cuts<br>bone. | The only difference is that the upper<br>slide of the Integra moves upward,<br>where the Vitalitec upper shaft<br>moves sideways to detach for<br>cleaning. |
| Elements | Instrument body, movable<br>handle to activate jaws, scoop<br>shaped cutting tip. | Instrument body, movable handle to<br>activate jaws, scoop shaped cutting<br>tip. |
The Vitalitec Medizintechnik GmbH Kerrison Rongeur is substantially equivalent to the predicate device since the basic features and the intended uses are the same. The only difference in design between the predicate is that the upper slide of the Integra moves upward, where the Vitalitec upper shaft moves sideways to detach for cleaning. The predicate device is available with additional coatings that are not used by Vitalitec. The minor differences between the Vitalitec Medizintechnik GmbH Kerrison Rongeur and the predicate device raise no new issues of safety and effectiveness. These differences have no effect on the function or intended use of the device.
## Performance Data [21 CFR 807.92(b)(1)]
The Vitalitec Medizintechnik GmbH Kerrison Rongeurs have been successfully tested for its functions, performance, and safety as per FDA recognized standards including:
Recognition Number 14-295: AAMI / ANSI ST81:2004/(R)2010, sterilization of medical devices - information to be provided by the manufacturer for the processing of resterilizable medical devices. (Sterility)
Recognition Number 5-40: ISO 14971 Second edition 2007-03-01, medical devices application of risk management to medical devices. (General I (QS/RM))
Recognition Number 5-90: ISO 15223-1 Second Edition 2012-07-01, medical devices symbols to be used with medical device labels, labelling, and information to be supplied - part 1: general requirements. (General I (QS/RM))
{6}------------------------------------------------
Recognition Number 2-156: AAMI / ANSI / ISO 10993-1:2009/(R) 2013, biological evaluation of medical devices -- part 1: evaluation and testing within a risk management process. (Biocompatibility)
Recognition Number 8-344: ISO 7153-1 Second edition 1991-04-01. surgical instruments -metallic materials -- part 1: stainless steel [including: amendment 1 (1999)]. (Materials)
Recognition Number 14-333: ISO 17665-1 First edition 2006-08-15, sterilization of health care products - moist heat - part 1: requirements for the development, validation and routine control of a sterilization process for medical devices. (Sterility)
| Test | Test Method Summary | Results |
|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mechanical<br>Testing<br>Cutting<br>performance<br>testing/bite<br>force | Mechanical testing was performed to<br>compare functional requirements of<br>the Vitalitec Kerrison Rongeurs with<br>the predicate device. Cutting of<br>control carton 250g/m2 according to<br>DIN 58298 using both the predicate<br>and proposed Kerrison Rongeur. | 100% Inspection resulted in all samples of<br>both the predicate and the proposed Kerrison<br>Ronguers meeting the requirements of DIN<br>58298, resulting in the determination of<br>substantial equivalence. Test showed the<br>measured functional specifications of the<br>Vitalitec Kerrison rongeurs were equivalent<br>or better than comparable specifications of<br>the predicate device. |
## Reprocessing:
Vitalitec Medizintechnik GmbH Kerrison Rongeurs cleaning and disinfection validations studies were performed in accordance with FDA Draft Guidance : Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling (May 2, 2011) The subject device is shipped non-sterile and is intended to be sterilized by user.
## Clinical Data [21 CFR 807.92(b)(2)]
Clinical data was not used to determine substantial equivalence.
# Conclusion [21 CFR 807.92(b)(3)]
Substantial equivalence for this device is based on similarities in intended use, function (design and technology), performance, and operational principle to the predicate device. Integra Kerrison Rongeurs (K092227).
The Vitalitec Medizintechnik GmbH Kerrison Rongeurs intended use and indications for use is identical to the Integra Kerrison Rongeurs (K092227) for cranial and spine applications: Neurosurgery.
{7}------------------------------------------------
The material used for the Vitalitec Medizintechnik GmbH Kerrison Rongeur is the same material as used for the Integra predicate Rongeur. The only difference is that the Integra devices have additional coating materials that are not used by Vitalitec.
The design, operational principle and technology of cutting bone are identical to Integra predicate Rongeurs. The material is captured and removed the same as the predicate device. The mechanical cut of the Vitalitec Medizintechnik GmbH Kerrison Rongeur is identical to the predicate. The only technological difference is that the upper slide of the Integra moves upward, where the Vitalitec upper shaft moves sideways to detach for cleaning. This does not impact the intended use of the device.
Based on predicate analysis as well as non-clinical performance testing, Vitalitec Medizintechnik GmbH has demonstrated the subject device is identical to the predicate in performance based on intended use. The results of the testing confirmed that Vitalitec Medizintechnik GmbH Kerrison Rongeurs and Integra Kerrison Rongeurs performed as intended.
Due to the similarity of the intended use, indications of use, function, materials, performance, and operational principle to the predicate device, Vitalitec Medizintechnik GmbH believes that the Vitalitec Medizintechnik GmbH Kerrison Rongeur does not raise any new safety or effectiveness issues.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.