← Product Code [HAE](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAE) · K150428

# Integra (R) Jarit (R) Kerrison Rongeurs, Integra (R) Ruggles (TM)-Redmons (TM) Kerrison Rongeurs, Integra (R) Miltex (R) Kerrison Rongeurs, Integra (R) Meisterhand (R) Kerrison Rongeurs (K150428)

_Integra Lifesciences Corp. · HAE · Apr 16, 2015 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAE/K150428

## Device Facts

- **Applicant:** Integra Lifesciences Corp.
- **Product Code:** [HAE](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAE.md)
- **Decision Date:** Apr 16, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.4840
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

Integra® (Jarit®, Ruggles™-Redmond™, Miltex®, MeisterHand®) Kerrison Rongeurs are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column.

## Device Story

Integra Kerrison Rongeurs are reusable, manually operated stainless steel surgical instruments. Used by surgeons in clinical settings for cutting or biting bone during skull or spinal procedures. Devices feature various bite sizes, jaw openings, cutting angles, and shaft lengths; some models include Detach shafts. Instruments are provided non-sterile and are designed to be sterilized by the user. The device functions through mechanical force applied by the surgeon to the handle, which actuates the cutting tip. This submission specifically updates the Instructions for Use to allow sterilization in either assembled or disassembled configurations, supported by existing validation data.

## Clinical Evidence

Bench testing only. Validation of pre-vacuum steam sterilization per ANSI/AAMI ST79:2010 and ANSI/AAMI/ISO 14937:2009 (270°F/132°C, 4-minute exposure, 20-minute dry time) was performed and passed.

## Technological Characteristics

Manual, non-electrical, reusable surgical instrument. Materials: 420 stainless steel; surface coatings include Titanium Nitride (TiN), TiAlN, Dicronite DL-5, ZrN, PTFE, and Silicone (Elastosil® LR3003/80). Dimensions: 1-6 mm bite sizes, 9-15.5 mm jaw openings, 4.75-15" shaft lengths. Sterilization: Steam (autoclave).

## Regulatory Identification

A manual rongeur is a manually operated instrument used for cutting or biting bone during surgery involving the skull or spinal column.

## Predicate Devices

- Integra® Jarit®, Ruggles™ -Redmond™, Miltex®, and MeisterHand® Kerrison Rongeurs ([K092227](/device/K092227.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 16, 2015

Integra York PA, Inc. Ms. Stephanie Sheeslev Senior Manager, Regulatory Affairs 589 Davies Drive York, Pennsylvania 17402

Re: K150428

> Trade/Device Name: Integra® (Jarit®, Ruggles™-Redmond™, Miltex®, MeisterHand®) Kerrison Rongeurs Regulation Number: 21 CFR 882.4840 Regulation Name: Manual Rongeur Regulatory Class: Class II Product Code: HAE Dated: February 18, 2015 Received: February 19, 2015

Dear Ms. Sheesley,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Carlos L. Pena-S

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K150428

### Device Name

Integra® (Jarit®, Ruggles™-Redmond™, Miltex®, MeisterHand®) Kerrison Rongeurs

Indications for Use (Describe)

Integra® (Jarit®, Ruggles™-Redmond™, Miltex®, MeisterHand®) Kerrison Rongeurs are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column.

Type of Use (*Select one or both, as applicable*)|X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

#### I. SUBMITTER

Integra York PA, Inc. 589 Davies Drive York, PA 17402 USA

Phone: (717) 717-840-3522 Fax: (717) 840-9347

Contact Person: Stephanie Sheesley Date Prepared: April 15, 2015

#### II. DEVICE

Name of Device: Integra® (Jarit®, Ruggles™-Redmond™, Miltex®, MeisterHand®) Kerrison Rongeurs Common Name: Kerrison Rongeur Classification Name: Manual, Rongeur (21 CFR 882.4840) Regulatory Class: Class II Product Code: HAE

#### III. PREDICATE DEVICE

Integra® Jarit®, Ruggles™ -Redmond™, Miltex®, and MeisterHand® Kerrison Rongeurs, K092227

This predicate device has not been subject to a design-related recall.

#### DEVICE DESCRIPTION IV.

Integra® Kerrison Rongeurs are reusable stainless steel instruments that are sterilizable and packaged non-sterile. Devices are available with the following features: with or without proprietary surface treatments; 1-6 mm bite sizes; 9 - 15.5 mm jaw openings; 40° and 90° up/down cutting angles: regular and thin/low profile footplates: standard and eiector tips: 4.75 -15" shaft lengths; and various handle and shaft styles, including Detach®. Integra® Kerrison Rongeurs are distributed under the following brand names: Jarit®, Ruggles™ -Redmond™, Miltex®, and MeisterHand®.

#### V. INDICATIONS FOR USE

Integra® (Jarit®, Ruggles™-Redmond™, Miltex®, MeisterHand®) Kerrison Rongeurs are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column.

{4}------------------------------------------------

### VI. COMPARISON OF TECHNOLOGICAL CHRACTERISTICS WITH THE PREDICATE DEVICE

|                                    | Integra York PA, Inc.<br>Integra® Kerrison Rongeurs | Integra Medical Instrument Group<br>(now Integra York PA, Inc.)<br>Integra® Kerrison Rongeurs |  |
|------------------------------------|-----------------------------------------------------|-----------------------------------------------------------------------------------------------|--|
| 510(k) #                           | Subject of Submission                               | K092227                                                                                       |  |
| Class                              | II                                                  | II                                                                                            |  |
| Pro Code                           | HAE                                                 | HAE                                                                                           |  |
| Regulation #                       | 882.4840                                            | 882.4840                                                                                      |  |
| Classification                     | Manual, Rongeur                                     | Manual, Rongeur                                                                               |  |
| Indications for Use                | Kerrison Rongeurs are manually                      | Kerrison Rongeurs are manually                                                                |  |
|                                    | operated instruments indicated for                  | operated instruments indicated for                                                            |  |
|                                    | cutting or biting bone during surgery               | cutting or biting bone during surgery                                                         |  |
|                                    | involving the skull or spinal column.               | involving the skull or spinal column.                                                         |  |
| Contact Materials                  | 420 Stainless steel                                 | 420 Stainless steel                                                                           |  |
|                                    | Surface coatings: Titanium                          | Surface coatings: Titanium                                                                    |  |
|                                    | Nitride(TiN); Diamond Coat and Hard-                | Nitride(TiN); Diamond Coat and Hard-                                                          |  |
|                                    | Coat™ (TiAIN): Stealth Coat and                     | Coat™ (TiAIN); Stealth Coat and                                                               |  |
|                                    | Smooth-Coat™ (Dicronite DL-5);                      | Smooth-Coat™ (Dicronite DL-5);                                                                |  |
|                                    | Ultra-Coat™ (ZrN); PTFE                             | Ultra-Coat™ (ZrN); PTFE                                                                       |  |
|                                    | (polytetrafluoroethylene [Teflon®]);                | (polytetrafluoroethylene [Teflon®]);                                                          |  |
|                                    | Silicone (Elastosil® LR3003/80 A,B)                 | Silicone (Elastosil® LR3003/80 A,B)                                                           |  |
| Instructions for Use               | Sterilize in either the Assembled or                | Sterilize in Disassembled. Open                                                               |  |
|                                    | Disassembled, Open Position.                        | Position.                                                                                     |  |
|                                    | Successful Sterilization Validation                 |                                                                                               |  |
|                                    | supports Assembled and Open                         |                                                                                               |  |
|                                    | Configuration                                       |                                                                                               |  |
| Technological                      | Identical to Predicate                              | Manual, non-electrical, non-sterile,                                                          |  |
| Characteristics                    |                                                     | reusable                                                                                      |  |
| Malleability                       | Non-malleable                                       | Non-malleable                                                                                 |  |
| Sterility                          | Non-sterile                                         | Non-sterile                                                                                   |  |
| Sterilization                      | Sterilizable                                        | Sterilizable                                                                                  |  |
| Utility                            | Reusable                                            | Reusable                                                                                      |  |
| Energy Source                      | N/A (manual)                                        | N/A (manual)                                                                                  |  |
| Image of Example                   |                                                     |                                                                                               |  |
| Kerrison Rongeur -<br>Detach Shaft |                                                     |                                                                                               |  |
|                                    |                                                     |                                                                                               |  |
|                                    |                                                     |                                                                                               |  |

The below table provides a comparison of the subject devices and the predicate devices.

There are no differences in technology, materials, intended use, or design between the subject devices and the above predicates. The only change is with respect to the labeling. The Instructions For Use for the predicate device stated that the device was to be sterilized in the disassembled but open configuration. The Instructions for Use for the subject devices allows the devices to be sterilized in the either the disassembled, but open configuration. This change is supported by a successful sterilization validation.

{5}------------------------------------------------

#### VII. PERFORMANCE DATA

No performance standards have been promulgated under Section 514 of the Food. Drug and Cosmetic Act for these devices. No new testing was required to be performed in support of this labeling change. This Special 510(k) pre-market notification includes the sterilization validation submitted with the predicate device submission #K092227 as a reference only.

The below testing is provided as a reference to support the labeling change subject to this Special 510(k). This testing is not new and was previously included in the 510(k) submission #K092227 and was deemed acceptable to support the proposed labeling change per Pre-Submission #Q141112.

| Testing Performed                                                                                                                                                                                          | Results |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Pre-Vacuum (wrapped) Steam Sterilization Validation per ANSI/AAMI ST79:2010<br>& A1:2010 and ANSI/AAMI/ISO 14937:2009 at 270°F (132°C) with an Exposure<br>Time of 4 minutes and Drying Time of 20 minutes | Pass    |

No biocompatibility testing was performed as a result of this Special 510(k) as the subject devices are not under-going a change in materials - this submission is for a labeling change only.

## VIII. CONCLUSIONS

The labeling was the only change submitted under this Special 510(k) and supports the substantial equivalence to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAE/K150428](https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/HAE/K150428)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
