K202072 · Minnetronix Neuro, Inc. · GZT · Aug 26, 2020 · Neurology
Device Facts
Record ID
K202072
Device Name
MindsEye Expandable Port
Applicant
Minnetronix Neuro, Inc.
Product Code
GZT · Neurology
Decision Date
Aug 26, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4800
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Expandable Port is intended to provide access, allow visualization of the surgical field, and retraction of soft tissue during neurological cranial surgery. The device is indicated for use in surgery during which subcortical access is required.
Device Story
MindsEye Expandable Port is a single-use, minimally invasive intracranial device for subcortical access and soft tissue retraction. It consists of a port with a flexible, compliant sheath, an obturator, and a guide. The device is inserted in a collapsed state using the obturator for atraumatic access; once positioned, the port is radially expanded to create a working channel (6-21mm diameter). The transparent sheath allows visualization of cerebral structures via endoscope, microscope, or loupes. It is used in neurosurgical procedures, often with optical navigation systems, and is compatible with standard surgical tools like aspirators and electrosurgical instruments. The device is operated by neurosurgeons in an OR setting. It benefits patients by providing a customizable, atraumatic corridor to target surgical locations, reducing tissue trauma compared to static retractors.
Clinical Evidence
Bench testing only. Testing included biocompatibility (cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity), sterilization validation (SAL 10-6), packaging/shelf-life, and mechanical damage resistance testing. All tests passed.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 26, 2020
Minnetronix Neuro, Inc. % Prithul Bom Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K202072
Trade/Device Name: MindsEye Expandable Port Regulation Number: 21 CFR 882.4800 Regulation Name: Self-Retaining Retractor For Neurosurgery Regulatory Class: Class II Product Code: GZT Dated: July 24, 2020 Received: July 27, 2020
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K202072
Device Name MindsEye™ Expandable Port
Indications for Use (Describe)
The Expandable Port is intended to provide access, allow visualization of the surgical field, and retraction of soft tissue during neurological cranial surgery.
The device is indicated for use in surgery during which subcortical access is required.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
ln accordance with 21 CFR 807.92 and FDA Guidance Document The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] dated July 28, 2014, Minnetronix Neuro provides this 510(k) Summary.
5. 510(k) NUMBER: K202072
#### 5.1 SUBMITTER INFORMATION
| Company Name: | Minnetronix Neuro, Inc. |
|------------------------|-------------------------------------------|
| Company Address: | 1625 Energy Park Drive, St. Paul MN 55108 |
| Company Phone: | (651) 917-4060 |
| Company Facsimile: | (651) 917-4066 |
| Contact Person: | Matt Adams, VP / GM |
| Date Summary Prepared: | August 25, 2020 |
#### 5.2 DEVICE IDENTIFICATION
| Device Trade Name: | MindsEye™ Expandable Port |
|----------------------|-------------------------------------------|
| Common Name: | Neurological surgical port |
| Classification Name: | Self-retaining retractor for neurosurgery |
| | (21 CFR 882.4800, Product Code GZT) |
| Device Class: | Class II |
| Advisory Panel: | Neurological Devices Panel |
#### 5.3 IDENTIFICATION OF PREDICATE DEVICE
The predicate device for the MindsEye Expandable Port is the Nico® BrainPath® (K172433) with reference to the Vycor VBAS (K060973).
#### 5.4 DEVICE DESCRIPTION
The MindsEye Expandable Port is a single use, minimally invasive intracranial medical device designed to provide access and allow visualization to the surgical field while retracting subcortical soft tissues during neurological cranial surgery. Its design allows the user to place
{4}------------------------------------------------
the port while collapsed in its minimal form for atraumatic access to the subcortical space. After placement, the port can be expanded radially to create a working channel to the target surgical location. The user is able to expand the port to the desired working channel diameter appropriate for the surgical plan then collapse the port upon surgical completion for safe removal. The working channel diameter may be adjusted as needed by the user throughout surgery.
The MindsEye device is expected to be used in conjunction with other standard, commercially available intracranial medical devices, such as an aspirator, micro/endo/exoscope, irrigation, electrosurgical systems/tools and/or hemostatic agents. In addition, MindsEye is compatible with existing optical navigation systems for accurate target acquisition and may be attached to other commercial retraction systems according to user preference.
The MindsEye device is made of biocompatible, medical grade materials (metals and polymers) with no software or electrical components. The MindsEye device and its components are provided sterile and not reusable.
MindsEye is comprised of the following components and is available in three lengths (Table 1).
- Port with attached Sheath
- Obturator and Guide
- · Optional Support System consisting of support clips, post, and legs
### Table 1: MindsEye Expandable Port Sizes
| Model Number | Port Length (mm) |
|--------------|------------------|
| N-EXP1A-052 | 52 |
| N-EXP1A-062 | 62 |
| N-EXP1A-077 | 77 |
#### 5.5 INDICATIONS FOR USE
The Expandable Port is intended to provide access, allow visualization of the surgical field, and retraction of soft tissue during neurological cranial surgery.
The device is indicated for use in surgery during which subcortical access is required.
{5}------------------------------------------------
#### TECHNOLOGICAL CHARACTERISTICS COMPARISON 5.6
While there are expected technological differences due to the ability of MindsEve to be placed in collapsed form then expanded (dilatable working channel) as compared to the static size working channel offered by the predicate; the overall indication for use, principles of operation and placement as well as general size of the device are similar to the predicate and reference devices.
### Similar Characteristics:
Like the predicate and reference devices, the MindsEye port is placed using an obturator for atraumatic introduction. Placement can be performed using a commercial optical navigation system often utilized for surgical planning. The obturator is then removed leaving a working channel to allow for visualization and surgical access within the brain tissue. MindsEye can accommodate a minimum inner diameter (i.e. working channel) of 6mm to a maximum diameter of 21mm. The inner diameter range of MindsEye (6 - 21mm) is smaller then or within the diameter range of the predicate and references devices (11 – 28mm). Similarly, MindsEye is available in three port lengths (52 – 77mm) to accommodate various surgical locations which is also within the length range of the predicate and references devices (30 - 95mm). The transparent sheath allows users to see cerebral vasculature, structures and tissue adjacent to the sheath and accommodates the visualization preferences (i.e. use of an endoscope, microscope, loupes, etc.) of the user during surgery. Finally, MindsEye is compatible with typical adjunct surgical instruments such as larger retraction systems, aspirators, forceps, bipolar and other electrosurgical and hemostasis devices.
### Different Characteristics:
To allow for the radial dilation of the MindsEye port, it was designed with a flexible, compliant sheath and controlled mechanism to expand and contract as intended. This is the primary difference from the predicate and reference devices which have a hard, static size sheath. To achieve this, different materials of manufacture are used as compared to the predicate and reference devices, however comprehensive biocompatibility testing was performed to maintain the safety profile of the device. Characterization and performance testing were completed to provide assurance that the sheath of the MindsEye device is robust enough to withstand the normal conditions of brain surgery. A summary of the predicate and reference device comparison is provided in Table 2.
{6}------------------------------------------------
# Table 2: Predicate Comparison
| | Predicate NICO BrainPath | MindsEye Expandable Port | Reference Vycor VBAS |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| 510(k)<br>Number | K172433 | Not yet assigned | K060973 |
| Intended<br>Use | To provide for access and<br>allow for visualization of the<br>surgical field during brain<br>and spinal surgery. | Provide access and allow<br>visualization to the surgical<br>field and to retract soft tissue<br>during neurological cranial<br>surgery. | Same as predicate |
| Device Components | | | |
| Sheath<br>Lengths | L: 50, 60, 75, 95mm | L: 52, 62, 77mm | TC model:<br>L: 30, 50, 70, 90mm |
| Sheath<br>Inner<br>Diameter<br>Options | ID: 11, 13.5mm | ID Expandable: 6 - 21mm | TC model:<br>ID: 12, 17, 21, 28mm |
| Obturator<br>Tip Length | L: 8mm and 15mm | L: 9 mm | Unknown |
| | Principle of Primary Operation | | |
| | Consists of an "obturator-<br>like" component and a<br>"sheath-like" component<br>which are assembled,<br>inserted, and disassembled<br>to provide corridor access | Same as Predicate | Same |
| Operation and Placement | | | |
| | Handheld and can be<br>assisted by third-party<br>navigation (if desired) | Same as Predicate | Same |
| Shipping Configuration | | | |
| | Obturator and sheath<br>packaged and shipped<br>separately and paired during<br>surgical case | The obturator and sheath are<br>packaged and shipped<br>together ready for use | The port and introducer<br>are packaged<br>assembled ready for use |
| Reusable or Single Patient Use | | | |
| | Single patient use and<br>reusable | Single patient use only | Single patient use only |
| Method of Sterilization | | | |
| | Disposables: Gamma<br>Reusables:<br>Autoclave/hydrogen<br>peroxide gas plasma | EO<br>Reusables: None | Gamma<br>Reusables: None |
| | Predicate NICO BrainPath | MindsEye Expandable Port | Reference Vycor VBAS |
| Biocompatibility | Demonstrated based on<br>externally communicating<br>device in direct contact with<br>tissue/bone/dentin for a<br>limited duration | Same as Predicate | Same |
| Materials of Manufacture | Obturator: Aluminum<br>Sheath: COC | Sheath - Chronoprene 5A<br>Guide - ABS<br>Obturator – 303 SS<br>Support system – 303 SS | Unknown |
| Cross Sectional Analysis of Obturator / Sheath | Obturator / Sheath<br>combination has a circular<br>cross section | Obturator / sheath (port)<br>combination has a circular<br>cross section when placed<br>Sheath (port) has a hexagonal<br>cross section when dilated | Various depending on<br>model from circular to<br>elliptical |
| General Shape of Obturator Tip | Distal end of obturator has a<br>conical shape with a<br>rounded tip and no opening | Same as Predicate | Various depending on<br>the model |
| Third-Party Instrumentation | Obturator component<br>interfaces with third party<br>instruments | Same as Predicate | Same |
| Navigation Locking Mechanism | Included as part of the<br>obturator | Same as Predicate | None |
| Surface of Sheath | Inner diameter and<br>horizontal proximal surface<br>of the knurled ring are<br>textured | Inner and outer surface of<br>sheath is non-reflective,<br>transparent, and flexible | Port is transparent and<br>non-compliant |
| Proximal End of Sheath | Updated knurled ring with<br>concave geometry to reduce<br>optical glare, still includes<br>holes, slots and tabs. | The proximal end has a non-<br>reflective head and dial for port<br>expansion with tabs that can<br>be used to connect to the<br>support system | Unknown |
{7}------------------------------------------------
{8}------------------------------------------------
#### 5.7 SUMMARY OF TESTING
A series of non-clinical tests, biocompatibility, sterility, and shelf life testing were performed to demonstrate that MindsEye met the applicable design and performance requirements to support a determination of substantial equivalence related to the safety and effectiveness of the MindsEye device as compared to the predicate and reference devices.
| Non-Clinical Test | MindsEye Component | Result |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|-----------------------|
| Cytotoxicity: MEM elution | All potential patient<br>contacting components | Pass - non-cytotoxic |
| Sensitization: (Maximum) | All potential patient<br>contacting components | Pass - non-sensitizer |
| Irritation: (Intracutaneous Reactivity) | All potential patient<br>contacting components | Pass - non-irritant |
| Acute System Toxicity | All potential patient<br>contacting components | Pass |
| Material Mediated Pyrogenicity | All potential patient<br>contacting components | Pass |
| Packaging and Shelf Life<br>• shipping/distribution simulation,<br>• environmental conditions, aging, visual<br>packaging inspection,<br>• bubble and seal strength packaging testing,<br>• environmental and shipping simulation | All | Pass |
| Sterilization<br>• B&F testing,<br>• BI for SAL 10-6,<br>• Endotoxin testing | All | Pass |
| Specification Review | All | Pass |
| Damage Resistance | Sheath | Pass |
### Table 3: Non-Clinical Testing Summary
#### 5.8 CONCLUSION
A review of the risk assessment, biocompatibility and other non-clinical testing, in addition to demonstrated compliance with recognized standards, have established that the MindsEye Expandable Port does not raise different questions of safety or effectiveness for the described intended use when compared to the predicate and reference devices. Therefore, the results of these analyses provide reasonable assurance that the MindsEye Expandable Port has a similar safety and effectiveness profile as compared to the predicate and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.