← Product Code [GZT](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/GZT) · K172433

# NICO BrainPath (K172433)

_Nico Corporation · GZT · Sep 7, 2017 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/GZT/K172433

## Device Facts

- **Applicant:** Nico Corporation
- **Product Code:** [GZT](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/GZT.md)
- **Decision Date:** Sep 7, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.4800
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

To provide for access and allow for visualization of the surgical field during brain and spinal surgery. Indications may include subcortical access to diseases such as the following: - Primary/Secondary Brain Tumors - Vascular Abnormalities/Malformations - Intraventricular Tumors/Cysts

## Device Story

NICO BrainPath provides minimally invasive subcortical access to neurological tissues; creates atraumatic surgical corridor. Device consists of reusable aluminum obturators and single-use cyclic olefin copolymer (COC) sheaths. Surgeon assembles obturator and sheath in OR; inserts into brain; removes obturator to leave sheath in place, providing 11 mm or 13.5 mm surgical corridor for visualization and instrument access. Used in neurosurgery; compatible with third-party navigation and retractor arms. Modifications include concave proximal sheath ring to reduce optical glare and expanded size offerings (11 mm diameter; 95 mm length). Benefits include minimally invasive access to deep-seated brain pathologies, potentially reducing tissue trauma compared to traditional open surgery.

## Clinical Evidence

Bench testing only. Testing included cytotoxicity, sensitization, irritation, simulated use with third-party instruments, packaging/shelf-life validation, cleaning validation (bioburden/endotoxin/protein), and sterility validation (autoclave/gas plasma). All tests passed.

## Technological Characteristics

Materials: Aluminum (obturator), COC (sheath). Principle: Mechanical obturator-sheath assembly for surgical corridor creation. Dimensions: 11 mm and 13.5 mm inner diameters; 50-95 mm lengths. Connectivity: None. Sterilization: Gamma (single-use), Autoclave/Hydrogen Peroxide Gas Plasma (reusable).

## Regulatory Identification

A self-retaining retractor for neurosurgery is a self-locking device used to hold the edges of a wound open during neurosurgery.

## Predicate Devices

- NICO BrainPath ([K150378](/device/K150378.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 7, 2017

NICO Corporation Sean Spence Regulatory Affairs Manager 250 E 96th St. Suite 125 Indianapolis, Indiana 46240

Re: K172433

Trade/Device Name: NICO BrainPath Regulation Number: 21 CFR 882.4800 Regulation Name: Self-Retaining Retractor for Neurosurgery Regulatory Class: Class II Product Code: GZT Dated: August 10, 2017 Received: August 11, 2017

Dear Mr. Spence:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Michael J. Hoffmann -S

for

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K172433

Device Name NICO BrainPath

Indications for Use (Describe)

To provide for access and allow for visualization of the surgical field during brain and spinal surgery. Indications may include subcortical access to diseases such as the following:

- Primary/Secondary Brain Tumors
- · Vascular Abnormalities/Malformations

· Intraventricular Tumors/Cysts"

| Type of Use (Select one or both, as applicable)                                                           | <table><tr><td><div> <input checked="true" type="checkbox"/>       Recreational Use (Part 21 CFR 201.327(p))      </div></td></tr><tr><td><div> <input type="checkbox"/>       Sunscreen Use (21 CFR 201.327(c))      </div></td></tr></table> | <div> <input checked="true" type="checkbox"/>       Recreational Use (Part 21 CFR 201.327(p))      </div> | <div> <input type="checkbox"/>       Sunscreen Use (21 CFR 201.327(c))      </div> |
|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <div> <input checked="true" type="checkbox"/>       Recreational Use (Part 21 CFR 201.327(p))      </div> |                                                                                                                                                                                                                                                |                                                                                                           |                                                                                    |
| <div> <input type="checkbox"/>       Sunscreen Use (21 CFR 201.327(c))      </div>                        |                                                                                                                                                                                                                                                |                                                                                                           |                                                                                    |

> Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for NICO Corporation. The letters "NICO" are in a bold, sans-serif font, with the "O" being a filled circle in a golden color. Below the letters is the word "CORPORATION" in a smaller, sans-serif font.

# NICO BrainPath®

21 CFR §807.92 Date Prepared: 06 September 2017

510(k) Number: K172433

| Submitter/Manufacturer: | NICO Corporation<br>250 E. 96th Street, Suite 125<br>Indianapolis, IN 46240 |
|-------------------------|-----------------------------------------------------------------------------|
| Contact Person:         | Sean Spence, RAC<br>Regulatory Affairs Manager<br>Office: 317.660.7118      |
| Trade Name:             | NICO® BrainPath®                                                            |
| Common/Usual Name:      | Self-retaining retractor for neurosurgery                                   |
| Classification / Panel: | 21 CFR §882.4800 / Neurology                                                |
| Product Codes:          | GZT                                                                         |
| Predicate Device:       | K150378, NICO BrainPath                                                     |

#### Device Description

The NICO BrainPath and accessories are designed to provide minimally invasive access to neurological tissues. The design specifically supports the creation of an atraumatic surgical corridor to access the brain. These BrainPath devices are part of what is being called the BrainPath Approach™, which integrates the expanding neurosurgical armamentarium of trajectory planning and navigation, optics, corridor resection and biopsy, and tissue preservation. To date, the BrainPath technology has been used to successfully access primary and secondary brain tumors, vascular abnormalities or malformations, and intraventricular tumors and cysts.

The BrainPath consists of multiple-sized reusable and re-sterilizable obturators with coordinating single patient use disposable sheaths. The obturator and sheath are assembled in the operating room immediately prior to use. After placement, the obturator is removed leaving behind the sheath which provides a 13.5 mm or 11 mm surgical corridor. Table 1 outlines the available BrainPath configurations.

| Name                 | Sheath Inner-Diameter (mm) | Sheath Length (mm) | Obturator Tip (mm) |
|----------------------|----------------------------|--------------------|--------------------|
| 13.5 mm x 50 mm (ST) |                            | 50                 | 7.5                |
| 13.5 mm x 50 mm      |                            | 50                 |                    |
| 13.5 mm x 60 mm      | 13.5                       | 60                 |                    |
| 13.5 mm x 75 mm      |                            | 75                 |                    |
| 13.5 mm x 95 mm      |                            | 95                 | 15                 |
| 11 mm x 50 mm        |                            | 50                 |                    |
| 11 mm x 60 mm        | 11                         | 60                 |                    |
| 11 mm x 75 mm        |                            | 75                 |                    |

#### Table 1: BrainPath Configurations

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for NICO Corporation. The logo consists of the word "NICO" in a bold, sans-serif font, with the "O" being a circle filled with a golden color. Below the word "NICO" is the word "CORPORATION" in a smaller, sans-serif font.

The BrainPath accessories include a Manipulation Tool which is used for manipulating the position of the sheath after it has been placed, Shepherd's Hooks which are for attaching to various commercially available retractor arms, a 95 mm Stylet for use with the 95 mm obturator, and a Sterilization Tray for sterilization of the reusable components (i.e., obturators, manipulation tools, and 95 mm stylet).

| BrainPath Component          | Sterile or Non-Sterile | Single-Use or Reusable |
|------------------------------|------------------------|------------------------|
| Sheath (all sizes)           | Sterile                | Single-Use             |
| Obturators (all sizes)       | Non-Sterile            | Reusable               |
| Manipulation Tool            | Non-Sterile            | Reusable               |
| 95mm Stylet                  | Non-Sterile            | Reusable               |
| Sterilization Tray           | Non-Sterile            | Reusable               |
| Shepherd's Hooks (all types) | Sterile                | Single-Use             |

#### Table 2: NICO BrainPath Components Supplied

## Intended Use

To provide for access and allow for visualization of the surgical field during brain and spinal surgery. Indications may include subcortical access to diseases such as the following:

- Primary/Secondary brain tumors -
- Vascular abnormalities/malformations -
- -Intraventricular tumors/cysts

#### Comparison to Predicate

The subject BrainPath iteration is nearly identical to the predicate NICO BrainPath (K150378). Modifications include a concave proximal sheath ring geometry to reduce optical glare, and a product line extension to include a 13.5 mm x 95 mm and 11 mm set of offerings. The 95 mm length Obturator also necessitated the creation of a new 95 mm stylet accessory. Table 3 provides additional details on how the subject and predicate devices compare.

|                                                     | Predicate Device<br>NICO BrainPath                                                                                                                                                                                                                                                                         | Subject Device<br>NICO Brain Port                                                                               |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Device Name                                         | BrainPath                                                                                                                                                                                                                                                                                                  | Identical                                                                                                       |
| Intended Use                                        | To provide for access and allow for visualization of the surgical field during brain and spinal surgery.<br>Indications may include subcortical access to diseases such as the following:<br>- Primary/Secondary brain tumors<br>- Vascular abnormalities/malformations<br>- Intraventricular tumors/cysts | Identical                                                                                                       |
| Principle of<br>Operation                           | Consists of an "obturator-like" component and a “sheath-like” component which are assembled, inserted, and disassembled to provide corridor access                                                                                                                                                         | Identical                                                                                                       |
| Operation and<br>placement                          | Handheld and can be assisted by third-party navigation (if desired)                                                                                                                                                                                                                                        | Identical                                                                                                       |
|                                                     | Predicate Device<br>NICO BrainPath                                                                                                                                                                                                                                                                         | Subject Device<br>NICO Brain Port                                                                               |
| Shipping<br>Configuration                           | Obturator and sheath packaged and shipped<br>separately and paired during surgical case                                                                                                                                                                                                                    | Identical                                                                                                       |
| Reusable or Single<br>Patient Use                   | Single Patient Use and Reusable                                                                                                                                                                                                                                                                            | Identical                                                                                                       |
| Method of<br>Sterilization / SAL                    | Disposables: Gamma (sheath/hooks) / 10-6<br>Reusables: Autoclave/hydrogen peroxide gas plasma                                                                                                                                                                                                              | Identical                                                                                                       |
| Biocompatibility                                    | Demonstrated based on externally<br>communicating device in direct contact with<br>tissue/bone/dentin for a limited duration                                                                                                                                                                               | Identical                                                                                                       |
| Materials of<br>Manufacture                         | Obturator: Aluminum<br>Sheath: COC                                                                                                                                                                                                                                                                         | Identical                                                                                                       |
| Cross Sectional<br>analysis of<br>Obturator/ Sheath | Obturator/Sheath combination has a circular<br>cross section                                                                                                                                                                                                                                               | Identical                                                                                                       |
| Depth markings                                      | Incremental depth markings on both sheath and<br>obturator.                                                                                                                                                                                                                                                | Identical                                                                                                       |
| Sheath inner<br>diameter (mm)<br>options            | 13.5 mm                                                                                                                                                                                                                                                                                                    | 13.5 mm and 11 mm                                                                                               |
| Sheath Lengths                                      | 13.5 mm diameter: 50, 60, and 75 mm                                                                                                                                                                                                                                                                        | 13.5 mm dia.: 50, 60, 75, and 95 mm<br>11 mm dia.: 50, 60, 75 mm                                                |
| General Shape of<br>Obturator Tip                   | Distal end of obturator has a conical shape with a<br>rounded tip and no opening.                                                                                                                                                                                                                          | Identical                                                                                                       |
| Obturator Tip                                       | For the blue, standard tip obturators, the tip<br>extends 15 mm beyond the sheath. For the Gold-<br>ST, shallow-tip obturator, the tip extends 7.5 mm<br>beyond the sheath.                                                                                                                                | Identical, new offerings all have 15 mm tips                                                                    |
| Third-Party<br>instrumentation                      | Obturator component interfaces with third-party<br>instruments.                                                                                                                                                                                                                                            | Identical                                                                                                       |
| Surface of Sheath                                   | Inner diameter and horizontal proximal surface<br>of the knurled ring are textured.                                                                                                                                                                                                                        | Identical                                                                                                       |
| Proximal End of<br>Sheath                           | Knurled ring and proximal portion of the sheath<br>has holes, slots, and tabs.                                                                                                                                                                                                                             | Updated knurled ring with concave geometry<br>to reduce optical glare, still includes holes,<br>slots and tabs. |
| Manipulation Tool                                   | "Manipulation Tool” is textured to prevent<br>glare/shine from surgical lighting.                                                                                                                                                                                                                          | Identical                                                                                                       |
| Shepherd's Hooks                                    | Three Shepherd's Hooks available to facilitate<br>interfacing with common third-party retractors.                                                                                                                                                                                                          | Identical                                                                                                       |
| 95mm Stylet                                         | None                                                                                                                                                                                                                                                                                                       | For use with 13.5 mm x 95 mm Obturator as a                                                                     |

#### Table 3: Comparison Table

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for NICO Corporation. The word "NICO" is written in large, bold, black letters. The "O" in NICO is a golden color. Underneath the word "NICO" is the word "CORPORATION" written in smaller, black letters.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for NICO Corporation. The letters "N," "I," and "C" are in black, and the "O" is in gold. Underneath the letters, the word "CORPORATION" is written in smaller, black letters.

## Non-Clinical Testing

The following tests were confirmed or repeated to demonstrate that the subject device modifications met applicable design and performance requirements, and supports a determination of substantial equivalence:

#### Table 4: Non-Clinical Testing

| Testing                                                                                                                                                                                                                                                     | Device(s) Supported                                                  | Result/Conclusion |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|-------------------|
| Cytotoxicity - MEM Elution: 72 hour incubation                                                                                                                                                                                                              | All                                                                  | Non-cytotoxic     |
| Sensitization - Maximization (2 extracts)                                                                                                                                                                                                                   |                                                                      | Non-sensitizer    |
| Irritation - Intracutaneous Reactivity (2 extracts)                                                                                                                                                                                                         | All                                                                  | Non-irritant      |
| Simulated Use to demonstrate the BrainPath has the ability to<br>interface with third-party Instruments and meets design input<br>requirements                                                                                                              | All                                                                  | Pass              |
| Packaging & Shelf Life – shipping/distribution simulation,<br>environmental conditions, aging, visual packaging inspection,<br>bubble and seal strength packaging testing, and functional<br>testing following aging, environmental and shipping simulation | All                                                                  | Pass              |
| Specification Review                                                                                                                                                                                                                                        | All                                                                  | Pass              |
| Cleaning Validation (Reusable Devices) – Establishment of<br>cleaning validation per miles soil test using bioburden<br>endotoxin and protein testing                                                                                                       | Obturator, Manipulation<br>Tools, Stylet, and<br>Sterilization Tray  | Pass              |
| Sterility Validation (Reusable Devices) – Steam autoclaving,<br>IUSS, and hydrogen peroxide gas plasma                                                                                                                                                      | Obturators, Manipulation<br>Tools, Stylet, and<br>Sterilization Tray | Pass              |
| Sterility Validation (Single-Use) – B&F testing, VDmax for<br>SAL 10-6, along with routine Endotoxin testing                                                                                                                                                | BrainPath Sheath and<br>Shepherd's Hooks                             | Pass              |
| Sterilization Tray Drop Test                                                                                                                                                                                                                                | Obturators, Manipulation<br>Tools, Stylet, and<br>Sterilization Tray | Pass              |

### Conclusion

Risk assessment, biocompatibility, non-clinical testing, design validation, and compliance with recognized standards have demonstrated that the subject NICO BrainPath modifications do not raise different questions of safety or effectiveness when compared to the predicate. Therefore, the results of these analyses provide reasonable assurance that the NICO BrainPath has a similar safety and effectiveness profile as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/GZT/K172433](https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/GZT/K172433)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
