← Product Code [GXD](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/GXD) · K192491

# Coolief Radiofrequency Generator (CRG) System (K192491)

_Avanos Medical, Inc. · GXD · Feb 21, 2020 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/GXD/K192491

## Device Facts

- **Applicant:** Avanos Medical, Inc.
- **Product Code:** [GXD](/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/GXD.md)
- **Decision Date:** Feb 21, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4400
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

COOLIEF* Radiofrequency Generator (CRG) System is indicated for use to create lesions during neurological lesion procedures and for the coagulation and decompression of disc material to treat symptomatic patients with contained herniated discs. The CRG is to be used with only Avanos RF Probes and accessories, such as Avanos TransDiscal Probes, COOLIEF*/Standard RF Pain Management Probes, Cannulas, Introducers, and Fluid Delivery Introducers. Use of CRG System is limited to the indications cleared under these Avanos devices and accessories.

## Device Story

System uses controlled 460kHz radiofrequency energy to create lesions in neurological tissue or coagulate/decompress disc material. Comprises RF generator, quad pump unit (QPU), and therapy cables. QPU circulates fluid via peristaltic pump heads to cool RF probe tips. Operated in clinical settings (OR/outpatient surgery) by licensed healthcare professionals using fluoroscopic or ultrasound guidance. Touchscreen interface allows selection of six treatment modes (Standard, Bipolar, Cooled, TransDiscal Monopolar/Bipolar). System monitors impedance and temperature; provides automatic shut-off and audible alarms for safety. Output affects clinical decision-making by enabling targeted tissue ablation/coagulation. Benefits include controlled thermal treatment for pain management.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including hardware performance, flow-rate verification, mechanical testing, software V&V, temperature control, and lesion validation. Compliance with IEC 60601-1, 60601-1-2, 60601-2-2, 62366-1, and 62304 confirmed.

## Technological Characteristics

RF generator (80W max), 4-channel peristaltic pump, therapy cables (copper/nickel, silicone/thermoplastic elastomer). Energy: 460kHz +/- 1% quasi-sinusoidal. Connectivity: Standalone, non-networked. Sterilization: Non-sterile, non-sterilizable. Software: IEC 62304 compliant.

## Regulatory Identification

A radiofrequency lesion generator is a device used to produce lesions in the nervous system or other tissue by the direct application of radiofrequency currents to selected sites.

## Predicate Devices

- Baylis Pain Management Generator – TD ([K072478](/device/K072478.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 21, 2020

Avanos Medical, Inc. Thomas Kozma, Ph.D. Director, Regulatory Affairs 5405 Windward Parkway Alpharetta, Georgia 30004

Re: K192491

Trade/Device Name: COOLIEF* Radiofrequency Generator (CRG) System Regulation Number: 21 CFR 882.4400 Regulation Name: Radiofrequency Lesion Generator Regulatory Class: Class II Product Code: GXD Dated: January 20, 2020 Received: January 22, 2020

Dear Dr. Thomas Kozma:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Xiaolin Zheng, Ph.D., M.S. Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K192491

Device Name

COOLIEF* Radiofrequency Generator (CRG) System

#### Indications for Use (Describe)

COOLIEF* Radiofrequency Generator (CRG) System is indicated for use to create lesions during neurological lesion procedures and for the coagulation and decompression of disc material to treat symptomatic patients with contained herniated discs. The CRG is to be used with only Avanos RF Probes and accessories, such as Avanos TransDiscal Probes, COOLIEF*/Standard RF Pain Management Probes, Cannulas, Introducers, and Fluid Delivery Introducers. Use of CRG System is limited to the indications cleared under these Avanos devices and accessories.

| Type of Use (Select one or both, as applicable) |                                                   |
|-------------------------------------------------|---------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D)    | <span style="text-decoration: overline;">X</span> |
| Over-The-Counter Use (21 CFR 801 Subpart C)     |                                                   |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

## MANUFACTURER/SPONSOR NAME, ADDRESS, TELEPHONE, CONTACT INFORMATION

| Date Summary Prepared                   | February 18, 2020                                                          |
|-----------------------------------------|----------------------------------------------------------------------------|
| 510(k) Manufacturer/ Sponsor<br>Address | Avanos Medical, Inc.<br>5405 Windward Parkway<br>Alpharetta, GA  30004     |
| Contact Person                          | Thomas Kozma, PhD<br>Director, Regulatory Affairs<br>Phone: (470) 448-5681 |

#### CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME

| Trade Name            | COOLIEF* Radiofrequency Generator (CRG) System |
|-----------------------|------------------------------------------------|
| Common Name           | Lesion Radiofrequency Generator                |
| Classification Name   | Generator, Lesion, Radiofrequency              |
| Regulation Number     | 21 CFR § 882.4400                              |
| Product Code          | GXD                                            |
| Device Classification | II                                             |

### PREDICATE DEVICES

| Manufacturer Name                                                      | Predicate Device Trade Name           | 510(k)<br>Number |
|------------------------------------------------------------------------|---------------------------------------|------------------|
| Baylis Medical Company<br>(Currently Owned by Avanos<br>Medical, Inc.) | Baylis Pain Management Generator – TD | K072478          |

#### DEVICE DESCRIPTION

The COOLIEF* Radiofrequency Generator (CRG) System uses controlled radiofrequency energy to generate heat that is used to create lesions in neurological tissues or to coagulate and decompress disc material to treat symptomatic patients with contained herniated discs. The following components of the predicate Pain Management Generator-TD (K072478), which must be used together, have been modified and are the subject of this 510(k): a) COOLIEF* Radiofrequency Generator (CRG), b) COOLIEF* Quad Pump Unit (QPU), and c) COOLIEF* Therapy Cables.

- a) COOLIEF* Radiofrequency Generator (CRG):
CRG is an 80-watt Radio-Frequency (RF) Generator with four independent RF channels, allowing all RF channels to be activated individually or altogether. The system is controlled via a touchscreen interface display that includes user controls, alarm messages, and on-screen selection of lesioning modalities. Like the predicate Pain Management Generator device (K072478), the CRG can operate in six treatment modes that include: Standard RF, Bipolar RF, Cooled RF, TransDiscal Monopolar RF, and TransDiscal Bipolar RF. The CRG can only be used with the other subject devices components of the COOLIEF* Radiofrequency Generator System (i.e the QPU and Therapy Cables).

{4}------------------------------------------------

- b) COOLIEF* Quad Pump Unit (QPU):
COOLIEF* QPU pump heads circulate fluid within a closed-loop system, using Avanos-marketed COOLIEF* fluid tubing kits, to cool the tip region of the COOLIEF* RF Probes. The QPU has four peristaltic pump heads that rotate to drive the fluid flow. The COOLIEF* QPU is controlled by the CRG and can be used with only components of the COOLIEF* Radiofrequency Generator System.

- c) COOLIEF* Therapy Cables:
COOLIEF* Therapy Cables are provided in the following three configurations for use with only the CRG:

- 1. COOLIEF* Multi-RF Therapy Cable is used for Standard, Pulsed, or Biopolar RF treatment modes. The cable is re-usable and not intended to be in the sterile field.
- 2. COOLIEF* Multi-Cooled RF Therapy Cable is used for Cooled RF treatment mode. The cable is re-usable and not intended to be in the sterile field.
- 3. COOLIEF* TransDiscal* Y-Connector Therapy Cable is used for the TransDiscal Monopolar or TransDiscal Bipolar RF treatment modes. The cable is re-usable and not intended to be in the sterile field.

Each configuration of the COOLIEF* cable is compatible with only the CRG.

The subject device components of the COOLIEF* Radiofrequency Generator System can be used with the following additional COOLIEF* components that have been previously cleared: RF Probe Electrodes (i.e., Standard, TransDiscal, or Cooled) (K002389, K031951, and K163461), Cannula/Introducers (K972846, K031951, and K163236), Fluid Tubing Kit (K031951), and Split Grounding Pad (K140658).

## INDICATIONS FOR USE

COOLIEF* Radiofrequency Generator (CRG) System is indicated for use to create lesions during neurological lesion procedures and for the coagulation and decompression of disc material to treat symptomatic patients with contained herniated discs. The CRG is to be used with only Avanos RF Probes and accessories, such as Avanos TransDiscal Probes, COOLIEF*/Standard RF Pain Management Probes, Cannulas, Introducers, and Fluid Delivery Introducers. Use of CRG System is limited to the indications cleared under these Avanos devices and accessories.

## COMPARISON OF SUBJECT DEVICE TO PREDICATE DEVICE

Like the predicate devices, the Generator, Quad Pump, and Therapy Cable are used outside of the sterile field and are reusable. The COOLIEF* Radiofrequency Generator (CRG) System is used in a clinical setting, such as an outpatient surgery center or operating room in a hospital by licensed healthcare professionals with access to fluoroscopic or ultrasound imaging. The following table compares the general and technological characteristics of the subject CRG System and predicate device (i.e., Baylis Pain Management Generator - TD (K072478) that must be used together to achieve intended use).

| COOLIEF* Radiofrequency Generator (CRG) |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                       |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                         | Subject Device - CRG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate Device - PMG                                                                                                                                                                                                                                                                                                                                                                                                         | Comment                                                                                                                                                                                                                                                                                               |
| Device Name<br>(510(k))                 | COOLIEF* Radiofrequency<br>Generator (CRG) System (RF<br>Generator, Peristaltic Pump,<br>and Therapy Cables) (K192491)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Baylis Pain Management<br>Generator-TD (RF Generator)<br>(K072478)                                                                                                                                                                                                                                                                                                                                                             | RF Generator, Peristaltic Pump,<br>and Therapy Cables must be used<br>together to achieve intended use.                                                                                                                                                                                               |
| Regulation                              | 21 CFR § 882.4400                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 21 CFR § 882.4400                                                                                                                                                                                                                                                                                                                                                                                                              | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| Product Code                            | GXD - RF Lesion Generator                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | GXD - RF Lesion Generator                                                                                                                                                                                                                                                                                                                                                                                                      | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| COOLIEF* Radiofrequency Generator (CRG) |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                       |
|                                         | Subject Device - CRG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate Device - PMG                                                                                                                                                                                                                                                                                                                                                                                                         | Comment                                                                                                                                                                                                                                                                                               |
| Indications for<br>Use                  | COOLIEF* Radiofrequency<br>Generator (CRG) System is<br>indicated for use to create lesions<br>during neurological lesion<br>procedures and for the<br>coagulation and decompression of<br>disc material to treat symptomatic<br>patients with contained herniated<br>discs. The CRG is to be used with<br>only Avanos RF Probes and<br>accessories, such as Avanos<br>TransDiscal Probes,<br>COOLIEF*/Standard RF Pain<br>Management Probes, Cannulas,<br>Introducers, and Fluid Delivery<br>Introducers. Use of CRG System is<br>limited to the indications cleared<br>under these Avanos devices and<br>accessories. | Baylis Pain Management<br>Generator is indicated for use to<br>create lesions during neurological<br>lesion procedures, and for the<br>coagulation and decompression of<br>disc material to treat symptomatic<br>patient with contained herniated<br>discs. The Baylis PMG-TD is to be<br>used with separately approved<br>probes such as Baylis TransDiscal<br>Probe, Oratec Spinecath™ and<br>Baylis Pain Management Probes. | Same as Predicate Device, except<br>for:<br>Updated branding from Baylis to<br>Avanos COOLIEF and specified use<br>to only Avanos probes and<br>accessories, and indications<br>limited to these Avanos probes<br>and Accessories. This branding<br>change has no impact on<br>performance or safety. |
| Prescriptive<br>Status                  | Rx Only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Rx Only                                                                                                                                                                                                                                                                                                                                                                                                                        | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| Environment of<br>Use                   | Clinical setting with licensed<br>professional                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Clinical setting with licensed<br>professional                                                                                                                                                                                                                                                                                                                                                                                 | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| AC Power<br>Compatibility               | 100-120V, 220-240V                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 100-120V, 220-240V                                                                                                                                                                                                                                                                                                                                                                                                             | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| Energy Delivery                         | Radiofrequency at 460kHz +/- 1%,<br>Quasi-sinusoidal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Radiofrequency at 460kHz +/- 1%,<br>Quasi-sinusoidal                                                                                                                                                                                                                                                                                                                                                                           | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| Generator Max<br>Power                  | 80 Watts                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 50 Watts                                                                                                                                                                                                                                                                                                                                                                                                                       | Equivalent to Predicate Device<br>Subject device has greater max<br>power capacity, but max power<br>per probe is equivalent at 50 W<br>(see row below); therefore, no<br>negative impact on performance<br>or safety.                                                                                |
| Max Power<br>(Current) per<br>Probe     | 50 Watts (900mA)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 50 Watts (900mA)                                                                                                                                                                                                                                                                                                                                                                                                               | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| Stimulation<br>Frequency                | Sensory = 50 Hz<br>Motor = 2Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Sensory = 50 Hz<br>Motor = 2Hz                                                                                                                                                                                                                                                                                                                                                                                                 | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| Stimulation<br>Pulse Width              | 0.1, 0.2, 0.5, and 1.0 msec                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 0.1, 0.2, 0.5, and 1.0 msec                                                                                                                                                                                                                                                                                                                                                                                                    | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| Stimulation<br>Voltage                  | Voltage Mode: 0 to 10V                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Voltage Mode: 0 to 10V                                                                                                                                                                                                                                                                                                                                                                                                         | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| Multiple Lesions                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                            | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| Probe<br>Connections                    | 1 to 4                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 1 to 4                                                                                                                                                                                                                                                                                                                                                                                                                         | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| Accessory<br>Compatibility              | Compatible with Avanos-marketed<br>RF Probes, Cannula/Introducers,<br>Fluid Tubing Kits, and Split<br>Grounding Pad.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Compatible with Avanos-marketed<br>RF Probes, Cannula/Introducers,<br>Fluid Tubing Kits, and Split<br>Grounding Pad.                                                                                                                                                                                                                                                                                                           | Same as Predicate Device<br>(All COOLIEF* system components<br>are used together to achieve the<br>intended use. The subject CRG,<br>QPU, and Therapy Cables are not<br>compatible/interchangeable with<br>the predicate PMG, PPU, or<br>Therapy Cables.)                                             |
| Type of Control                         | Automatic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Manual, Automatic, Pneumatic<br>Foot Switch                                                                                                                                                                                                                                                                                                                                                                                    | Equivalent to Predicate<br>Subject device no longer includes<br>Manual or Foot Switch Controls,<br>which were rarely used, optional<br>features.                                                                                                                                                      |
| Overall<br>Temperature                  | 30°C - 95°C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 30°C - 95°C                                                                                                                                                                                                                                                                                                                                                                                                                    | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| COOLIEF* Radiofrequency Generator (CRG) |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                       |
|                                         | Subject Device - CRG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate Device - PMG                                                                                                                                                                                                                                                                                                                                                                                                         | Comment                                                                                                                                                                                                                                                                                               |
| Lesioning Times                         | 10 sec to 30 min                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 10 sec to 30 min                                                                                                                                                                                                                                                                                                                                                                                                               | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| Pulse RF Time                           | 10 to 900 seconds                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 10 to 900 seconds                                                                                                                                                                                                                                                                                                                                                                                                              | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| Type of User<br>Interface               | Capacitive touch screen graphical<br>user interface and rotary dial                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Manual soft keys and rotary dial                                                                                                                                                                                                                                                                                                                                                                                               | Equivalent to Predicate Device<br>Subject device includes touch<br>screen display                                                                                                                                                                                                                     |
| System Safety<br>Features               | Automatic Shut-Off includes:<br>• Out-of-range impedance and<br>temperature<br>• Over power, voltage, current<br>Output of errors:<br>• Fault indicator, fault code<br>display and description<br>Audible alarm                                                                                                                                                                                                                                                                                                                                                                                                         | Automatic Shut-Off includes:<br>• Out-of-range impedance and<br>temperature<br>• Over power, voltage, current<br>Output of errors:<br>• Fault indicator, fault code<br>display and description<br>Audible alarm                                                                                                                                                                                                                | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| Contact Quality<br>Monitoring           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | No                                                                                                                                                                                                                                                                                                                                                                                                                             | Subject Device circuitry and<br>associated alarms were added to<br>comply with IEC 60601-2-2.                                                                                                                                                                                                         |
| Printer Output                          | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | No                                                                                                                                                                                                                                                                                                                                                                                                                             | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| WiFi Capability                         | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | No                                                                                                                                                                                                                                                                                                                                                                                                                             | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| Internet<br>Connectivity                | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | No                                                                                                                                                                                                                                                                                                                                                                                                                             | Same as Predicate Device                                                                                                                                                                                                                                                                              |
| Sterility                               | Supplied non-sterile; Non-<br>sterilizable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Supplied non-sterile; Non-<br>sterilizable                                                                                                                                                                                                                                                                                                                                                                                     | Same as Predicate Device                                                                                                                                                                                                                                                                              |

{5}------------------------------------------------

{6}------------------------------------------------

| COOLIEF* Quad Pump Unit (QPU) |                                                               |                                                |                          |
|-------------------------------|---------------------------------------------------------------|------------------------------------------------|--------------------------|
|                               | Subject Device<br>COOLIEF* Radiofrequency<br>Generator System | Predicate Peristaltic Pump                     | Comment                  |
| Device Name<br>(510(k))       | COOLIEF* Quad Pump Unit,<br>QPU (K192491)                     | Baylis Peristaltic Pump Unit, PPU<br>(K072478) | Same as Predicate Device |
| Pump Flow Rate                | >18mL/min                                                     | >18mL/min                                      | Same as Predicate Device |
| Pump Power<br>Requirement     | 24VDC from RF Generator,<br>1000mA max                        | 24VDC from RF Generator,<br>1000mA max         | Same as Predicate Device |
| Sterility                     | Supplied non-sterile; Non-<br>sterilizable                    | Supplied non-sterile; Non-<br>sterilizable     | Same as Predicate Device |

| COOLIEF* Therapy Cables              |                                                                                                                                                                  |                                                                                                                                                                  |                          |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
|                                      | Subject Devices – MRF, MCRF, and TDP2                                                                                                                            | Predicate Devices – MRF, MCRF, and TDP2 (K072478)                                                                                                                | Comment                  |
| Device Name<br>(510(k))              | Avanos Accessory Therapy Cables:<br>COOLIEF* Multi-RF Therapy Cable, MRF COOLIEF* Multi-Cooled Therapy Cable, MCRF COOLIEF* TransDiscal* Y-Connector Cable, TDP2 | Baylis Accessory Therapy Cables:<br>COOLIEF* Multi-RF Therapy Cable, MRF COOLIEF* Multi-Cooled Therapy Cable, MCRF COOLIEF* TransDiscal* Y-Connector Cable, TDP2 |                          |
| Cable Length                         | 8 Ft                                                                                                                                                             | 8 Ft                                                                                                                                                             | Same as Predicate Device |
| Probe<br>Connections -<br>Distal End | Pinned and keyed to be compatible<br>with Avanos-marketed probes                                                                                                 | Pinned and keyed to be compatible<br>with Avanos-marketed probes                                                                                                 | Same as Predicate Device |
| Number of<br>Probe<br>Connections    | Two (TDP2)<br>Four (MRF & MCRF)                                                                                                                                  | Two (TDP2)<br>Four (MRF & MCRF)                                                                                                                                  | Same as Predicate Device |
| Key Cable<br>Materials               | Internal Materials:<br>Copper/Nickel<br>External Materials:<br>Silicone & Thermoplastic elastomer                                                                | Internal Materials:<br>Copper/Nickel<br>External Materials:<br>Silicone & Thermoplastic elastomer                                                                | Same as Predicate Device |
| Sterility                            | Supplied non-sterile; Non-<br>sterilizable                                                                                                                       | Supplied non-sterile; Non-sterilizable                                                                                                                           | Same as Predicate Device |

{7}------------------------------------------------

## Summary of Non-Clinical Testing (Performance Testing)

## Non-Clinical Performance Data

The following performance data were provided in support of the substantial equivalence determination for the subject COOLIEF* Radiofrequency Generator (CRG) System. Performance testing of the subject device was conducted to demonstrate that the modified device continued to meet performance specifications. Results of design verification activities did not raise any new or different issues of safety or effectiveness. The risk management process was used throughout the non-clinical verification activities in accordance with ISO 14971. The following verification tests were conducted.

| Bench Testing Performed                                                                                                                                                                                                      | Result |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| COOLIEF* RF Generator Hardware Performance                                                                                                                                                                                   | Pass   |
| Quad Pump Unit Flow-Rate Verification                                                                                                                                                                                        | Pass   |
| Therapy Cable Mechanical Testing                                                                                                                                                                                             | Pass   |
| Software Verification and Validation                                                                                                                                                                                         | Pass   |
| Temperature Control - Benchtop and Perfused Tissue                                                                                                                                                                           | Pass   |
| Bench-Top Lesion Validation                                                                                                                                                                                                  | Pass   |
| Transportation and Handling (Packaging)                                                                                                                                                                                      | Pass   |
| IEC 60601-1 (Medical Electrical Equipment – General Requirements<br>for basic safety and essential performance 2012, Edition 3.1, Class 1.                                                                                   | Pass   |
| IEC 60601-1-2: 2014 (Medical Electrical Equipment - General requirements for basic safety and<br>essential performance – Collateral standard: Electromagnetic compatibility – Requirements and<br>tests.                     | Pass   |
| IEC 60601-2-2: 2009 (Fifth Ed.)<br>Medical electrical equipment Part 2-2: Particular requirements for basic safety and essential<br>performance of high frequency surgical equipment and high frequency surgical accessories | Pass   |
| IEC 62366-1:2015 (Part 1: Application of Usability Engineering to Medical Devices.)                                                                                                                                          | Pass   |
| IEC 62304:2006 A1:2015 (Medical device software – software life cycle processes)                                                                                                                                             | Pass   |

## Clinical Performance Data

Clinical data was not applicable to these subject devices.

#### Conclusion

The differences between the subject COOLIEF* Radiofrequency Generator (CRG) System and predicate device Baylis Pain Management Generator-TD (K072478)) do not raise any new or different questions of safety or effectiveness. The subject COOLIEF* Radiofrequency Generator System is substantially equivalent to the predicate devices with respect to the indications for use, technology, material composition, which are not patient contacting, and performance.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/GXD/K192491](https://fda.innolitics.com/submissions/NE/subpart-e%E2%80%94neurological-surgical-devices/GXD/K192491)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
