K181864 · Baylis Medical Company, Inc. · GXD · Jan 2, 2019 · Neurology
Device Facts
Record ID
K181864
Device Name
Polaris RF Ablation System
Applicant
Baylis Medical Company, Inc.
Product Code
GXD · Neurology
Decision Date
Jan 2, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Polaris RF Ablation System is intended for the creation of radiofrequency lesions in nervous tissue.
Device Story
The Polaris RF Ablation System is a radiofrequency (RF) lesion generator used by physicians to create thermal lesions in nervous tissue for pain management. The system delivers controlled RF energy from a generator to target tissues via various RF probes (standard or cooled). For cooled procedures, a pump unit circulates water through the probe to minimize tissue charring and adhesion, allowing for larger, more consistent ablation volumes. The system also delivers low-frequency stimulation pulses for nerve localization. The generator features a touchscreen interface, real-time monitoring of temperature, impedance, power, and voltage, and automatic safety shut-offs for out-of-range parameters. The device is used in a clinical setting by physicians familiar with RF lesion techniques. The output (lesions) is used to treat patients by inducing cellular necrosis through thermal coagulation.
Clinical Evidence
Bench testing only. No clinical data. Performance was validated through mechanical stress testing (IEC 60601-2-2), electrical safety testing (IEC 60601-1), biocompatibility (ISO 10993-1), packaging integrity (ISO 11607), and comparative benchtop lesion validation in a soft tissue model to confirm ablation performance against predicates.
Technological Characteristics
System includes RF generator, peristaltic pump, connector hub, footswitch, and various RF probes/cannulae. Materials include stainless steel, polyimide, ABS, silicone, PVC, polycarbonate, and copolyester. Energy source: RF generator (465 kHz sinusoidal). Connectivity: Standalone system with touchscreen interface. Sterilization: Ethylene oxide (probes/cannulae) or steam (reusable probes). Software: Embedded firmware for generator control and monitoring.
Indications for Use
Indicated for the creation of radiofrequency lesions in nervous tissue.
Regulatory Classification
Identification
A radiofrequency lesion generator is a device used to produce lesions in the nervous system or other tissue by the direct application of radiofrequency currents to selected sites.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Baylis Medical Company Inc. May Tsai Team Leader, Regulatory Affairs 2775 Matheson Blvd. East Mississauga, L4W 4P7 Canada
January 2, 2019
### Re: K181864
Trade/Device Name: Polaris RF Ablation System Regulation Number: 21 CFR 882.4400 Regulation Name: Radiofrequency Lesion Generator Regulatory Class: Class II Product Code: GXD, GXI Dated: November 30, 2018 Received: December 3, 2018
Dear May Tsai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
$$\textbf{John Marker-S}_{\textbf{Date:2019.01.02}\textbf{16:14:56}}^{\textbf{Digitalby signed by John}}$$
For Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181864
Device Name Polaris RF Ablation System
Indications for Use (Describe)
The Polaris RF Ablation System is intended for the creation of radiofrequency lesions in nervous tissue.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 7. 510(k) Summary
#### 7.1 Submitter Information
- A. Company Name: Baylis Medical Company Inc.
- B. Company Address: 2775 Matheson Blvd. East Mississauga, Ontario L4W 4P7 Canada
- C. Company Phone: (905) 602-4875
- D. Company Facsimile: (905) 602-5671
- E. Contact Person: May Tsai, Requlatory & Scientific Affairs Team Leader
- F. Summary Prepared on: 11-Jul-2018
#### 7.2 Device Identification
- A. Device Trade Name:
- . Polaris RF Ablation System
- B. Device Common Name:
- Radiofrequency lesion generator ●
- . Radiofrequency lesion probe
- C. Classification Name:
- CFR 882.4400 Radiofrequency lesion generator ●
- CFR 882.4725 Radiofrequency lesion probe .
- D. Product Code:
- . GXD, GXI
- E. Device Class: Class II
{5}------------------------------------------------
#### 7.3 Identification of Predicate Device
## Table 7.1: Predicate Devices
| Predicate Device | Manufacturer | 510(k) |
|-----------------------------------------------------|-----------------|---------|
| Baylis Pain Management Generator-TD | | K072478 |
| Baylis TransDiscal System | | K062937 |
| Baylis Pain Management Cooled Probe | Halyard Health* | K053082 |
| Baylis Pain Management Probe and<br>Connector Cable | | K002389 |
| BMC RF Cannula | | K972846 |
*Note: Originally submitted by Baylis Medical Company Inc. as the manufacturer, but product line has since been acquired by Halyard Health.
#### 7.4 Indications for Use
The Polaris RF Ablation System is intended for the creation of radiofrequency lesions in nervous tissue.
#### 7.5 Device Description
The Polaris RF Ablation System includes the following components:
- 1. Polaris Radiofrequency Generator and Desk Stand
- 2. Polaris Pump, Pump Cable and Desk Stand
- 3. Polaris Standard Connector Hub / Polaris Cooled Connector Hub
- 4. Polaris Footswitch
- 5. Polaris Cooled RF Probe Kit:
- i. Polaris Cooled RF Ablation Probe
- ii. Polaris Tube Kit
- iii. Polaris Introducer
- 6. Polaris Single-Use RF Probe
- 7. Polaris Reusable RF Probe
- 8. Polaris RF Cannula
The Polaris RF Ablation System is intended for the creation of radiofrequency lesions in nervous tissue. The system is designed to deliver controlled RF energy from the Polaris RF Generator to target tissues via the Polaris RF Probes for standard RF procedures or Polaris Cooled RF Probes for cooled RF
{6}------------------------------------------------
procedures. RF energy is applied based on the configured RF generator settings to create the desired lesions in the target tissue. The generator also delivers low frequency stimulation pulses during procedures.
During standard RF procedures, the Polaris RF Probe is used with a compatible Polaris RF Cannula to enable RF energy delivery via the active electrode to the target tissues. During cooled RF procedures with Polaris Cooled RF Probes, the system integrates a cooling mechanism by internally circulating water through the probes using the Polaris Pump Unit. This cooling results in minimal charring of the surrounding tissue and also prevents tissue from adhering to the electrodes. This allows for larger and more consistent ablation volumes to be achieved.
#### 7.6 Comparison to Predicate Device
The intended use of the subject Polaris RF Ablation System is the same as the predicate devices. The indication for use of the Polaris RF Ablation System is the same, or a subset of, the indications for use of the predicate devices. The subject and predicate devices share the same fundamental scientific technology, including principles of operation and mechanism of action (Tables 7.2 and 7.3). Differences in design and technological characteristics between the proposed and predicate devices do not raise any new types of safety and effectiveness questions. Verification and validation test results provide reasonable assurance of the substantial equivalence of the Polaris RF Ablation System compared to the predicate system.
{7}------------------------------------------------
# Table 7.2: Comparison of Subject and Predicate Devices – Overall System
| | PREDICATE DEVICES | | | | | SUBJECT<br>DEVICE | Identical /<br>Substantially<br>Equivalent<br>(SE) |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| Device Name | Baylis Pain<br>Management<br>Generator-TD | TransDiscal (TD)<br>System | BMC RF<br>Cannula | BMC PM Probe<br>and Connector<br>Cable | Baylis PM<br>Cooled Probe | Polaris RF<br>Ablation System | N/A |
| 510(k) # | K072478 | K062937 | K972846 | K002389 | K053082 | K181864 | N/A |
| Manufacturer | Halyard Health* | | | | | Baylis Medical<br>Company Inc. | N/A |
| Class | II | II | II | II | II | II | YES/YES |
| Product Code,<br>Regulation | GEI, 882.4400<br>GXD, 882.4400 | GXI, 882.4725 | GXI, 882.4725 | GXI, 882.4725 | GXI, 882.4725 | GXD, 882.4400<br>GXI, 882.4725 | YES/YES |
| Indications for<br>Use | Baylis Pain<br>Management<br>Generator - TD;<br>Model PMG-115-<br>TD (For<br>Domestic Use)<br>and Model PMG-<br>230-TD (For<br>International<br>Use) is indicated<br>for use to create<br>lesions during<br>neurological<br>lesion<br>procedures, and<br>for the<br>coagulation and<br>decompression<br>of disc material<br>to treat<br>symptomatic<br>patients with | The Baylis TD<br>system, used in<br>combination with<br>the Baylis PM<br>Generator, is<br>indicated for the<br>creation of RF<br>heat lesions in<br>nervous tissue<br>including that<br>which is situated<br>in the<br>intervertebral<br>disc material. | The BMC RF<br>Cannula is<br>intended for<br>use in RF heat<br>lesion<br>procedures for<br>relief of pain | To create RF<br>lesions in<br>nervous tissue. | The Baylis PM<br>Cooled Probe<br>will be used in<br>conjunction<br>with a RF<br>Generator to<br>create RF<br>lesions in<br>nervous tissue. | The Polaris RF<br>Ablation System<br>is intended for<br>the creation of<br>RF lesions in<br>nervous tissue. | NO/YES |
| | PREDICATE DEVICES | | | | SUBJECT<br>DEVICE | Identical /<br>Substantially<br>Equivalent<br>(SE) | |
| | contained<br>herniated discs.<br>The Baylis PMG-<br>TD is to be used<br>with separately<br>approved probes<br>such as Baylis<br>TransDiscal<br>Probe, Oratec<br>Spinecath and<br>Baylis Pain<br>Management<br>Probes. | | | | | | |
| Relevant<br>System<br>Components | RF Generator | Peristaltic Pump,<br>Connector Cable<br>Tube Kit<br>Introducer<br>Y-Connecting<br>Cable | RF Cannula | Standard RF<br>Probes (non-<br>cooled) | Cooled RF<br>Probe | RF Generator<br>and Desk Stand<br>Peristaltic Pump<br>and Desk Stand,<br>Pump Connector<br>Cable<br>Connector Hub<br>RF Cannula<br>Standard RF<br>Probes (non-<br>cooled)<br>Cooled RF Probe<br>kit (Cooled RF<br>Probe, Tube Kit,<br>Introducer) | NO/YES |
Traditional 510(k)
{8}------------------------------------------------
{9}------------------------------------------------
| | PREDICATE DEVICES | SUBJECT<br>DEVICE | Identical /<br>Substantially<br>Equivalent<br>(SE) |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| User | Physicians familiar with RF lesion techniques | Physicians<br>familiar with RF<br>lesion<br>techniques | YES/YES |
| Anatomical<br>Site of Use | Nervous tissue | Nervous tissue | YES/YES |
| Access Method | Percutaneous | Percutaneous | YES/YES |
| Energy<br>Type | Radiofrequency | Radiofrequency | YES/YES |
| Principle of<br>Operation | Operator controlled; RF delivered from RF generator to compatible RF probes | Operator<br>controlled; RF<br>delivered from<br>RF generator to<br>compatible RF<br>probes | YES/YES |
| Mechanism of<br>Action | Cellular necrosis through thermal coagulation | Cellular necrosis<br>through thermal<br>coagulation | YES/YES |
| System<br>Feedback<br>Mechanism | Temperature controlled | Temperature<br>controlled | YES/YES |
| Ability to Make<br>Multiple<br>Lesions | Yes | Yes | YES/YES |
| | PREDICATE DEVICE | SUBJECT DEVICE | Identical/<br>SE |
| | RF GENERATOR | | |
| Device Name | Baylis Pain Management<br>Generator-TD | Polaris Radiofrequency<br>Generator | N/A |
| 510(k) # | K072478 | K181864 | N/A |
| Available Generator Modes | Standard RF Mode<br>• Stimulation<br>(Sensory/Motor)<br>• Pulsed RF (Single<br>RF, Multi-RF,<br>Bipolar RF)<br>• RF lesion (Single<br>RF, Multi-RF,<br>Bipolar RF) | Standard RF Mode<br>• Stimulation<br>(Sensory/Motor)<br>• Pulsed RF (Single<br>RF, Multi-RF,<br>Bipolar RF)<br>• RF lesion (Single<br>RF, Multi-RF,<br>Bipolar RF) | YES/YES |
| | Cooled RF Mode<br>• Stimulation<br>(Sensory/Motor)<br>• RF lesion (Single<br>RF, Multi-RF,<br>Bipolar RF) | Cooled RF Mode<br>• Stimulation<br>(Sensory/Motor)<br>• RF lesion (Single<br>RF, Multi-RF,<br>Bipolar RF) | |
| Maximum Power Output | 50 Watts | 50 Watts | YES/YES |
| Operating Frequency<br>Waveform | 461 kHz Sinusoidal | 465 kHz Sinusoidal | NO/YES |
| Temperature Monitoring<br>Thermocouple (RF modes) | Yes | Yes | YES/YES |
| Display Parameters | Real time temperature,<br>impedance, power and<br>voltage (pulsed RF<br>mode) | Real time temperature,<br>impedance, power and<br>voltage (pulsed RF<br>mode) | YES/YES |
| Safety Features | • Automatic shut-off for:<br>o Out-of-range<br>impedance and<br>temperature<br>o Over power,<br>voltage, current | • Automatic shut-off for:<br>o Out-of-range<br>impedance and<br>temperature<br>o Over power,<br>voltage, current | YES/YES |
| | • Output of errors,<br>including:<br>o Fault indicator,<br>fault code display<br>and description<br>o Audible alarm | • Output of errors,<br>including :<br>o Alert icon, fault<br>code display and<br>description<br>o Audible alarm | |
| Touchscreen | No | Yes | NO/YES |
| Used with Desk Stand | No | Yes | NO/YES |
| Used with Optional<br>Footswitch | Yes | Yes | YES/YES |
| Environment | Supplied non-sterile;<br>Non-sterilisable | Supplied non-sterile;<br>Non-sterilisable | YES/YES |
{10}------------------------------------------------
## Table 7.3: Comparison of Subject and Predicate Devices – Relevant Components
{11}------------------------------------------------
| | | PREDICATE DEVICE | SUBJECT DEVICE | Identical/SE |
|-------------------------------------------------------------|---------------------------------------------------------|------------------------------------------------------------------|---------------------------------------------------|------------------|
| | | COOLED RF PROBE | | |
| Device Name | | PM Cooled Probe | Polaris Cooled RF<br>Ablation Probe | N/A |
| 510(k) # | | K053082 | K181864 | N/A |
| Probe Configuration | | Monopolar | Monopolar | YES/YES |
| Tissue Temperature<br>Measurement | Includes Thermocouple for | Yes | Yes | YES/YES |
| | Location of Thermocouple | Exposed at probe distal<br>tip | Embedded within probe<br>distal tip | NO/YES |
| Patient-<br>Contact | Shaft,<br>Thermocouple | Stainless steel | Stainless steel | YES/YES |
| Materials | Insulation | Polyimide | Polyimide | |
| Key Non- | Handle | Acetal | ABS | NO/YES |
| patient | Cable | Silicone | Silicone | |
| Contact<br>Materials | Tubing | Polyvinyl chloride | Polyvinyl chloride | |
| | Luer lock | Polycarbonate | Polycarbonate | |
| Key<br>Dimensions | Shaft<br>Diameter | 18 Gauge | 18 Gauge | NO/YES |
| | Active Tip<br>Lengths | 2, 4, 5.5, 6 mm | 4, 5.5 mm | |
| | Compatible<br>Introducer<br>Lengths | 50, 75, 100, 150 mm | 50, 100, 150 mm | |
| | Cable | 48" (length);<br>3.80 mm (OD) | 48" (length);<br>3.80 mm (OD) | |
| | Tubing | 48" (length);<br>3.18 mm (OD) | 48" (length);<br>3.18 mm (OD) | |
| Environment | | Provided sterile;<br>Single use | Provided sterile;<br>Single use | YES/YES |
| Sterilization Method:<br>Sterility Assurance Level<br>(SAL) | | Ethylene oxide;<br>10-6 | Ethylene oxide;<br>10-6 | YES/YES |
| | | TUBE KIT | | |
| Device Name | | TransDiscal Tube Kit | Polaris Tube Kit | N/A |
| 510(k) # | | K062937 | K181864 | N/A |
| Key non-<br>patient<br>Contact<br>Materials | Tubing<br>Burette,<br>Female Luer<br>Cap, Luer<br>Locks | Tygon<br>Polycarbonate | Tygon<br>Polycarbonate | NO/YES |
| | Crimp | N/A | Stainless steel | |
| Key | Tubing Length | 140" | 140" | YES/YES |
| Dimensions | Burette<br>Capacity | 70 mL | 70 mL | |
| Environment | | Provided sterile;<br>Single use | Provided sterile;<br>Single use | YES/YES |
| Sterilization Method;<br>(SAL) | Sterility Assurance Level | Ethylene oxide;<br>10-6 | Ethylene oxide;<br>10-6 | YES/YES |
| | | PREDICATE DEVICE | SUBJECT DEVICE | Identical/<br>SE |
| | | INTRODUCER | | |
| Device Name | | TransDiscal Introducer | Polaris Introducer | N/A |
| 510(k) # | | K062937 | K181864 | N/A |
| Comprises Cannula and Stylet | | Yes | Yes | YES/YES |
| Depth Markers | | Yes | Yes | YES/YES |
| Patient-<br>contact<br>Materials | Shaft<br>Introducer, Stylet | Stainless steel | Stainless steel | NO/YES |
| | Insulation | Polyimide | Polyimide | |
| | Hub | Unknown | Copolyester | |
| Key<br>Dimensions | Shaft Outer<br>Diameter | 17 Gauge | 17 Gauge | NO/YES |
| | Usable<br>Length(s) | 50, 75, 100, 150 mm | 50, 100, 150mm | |
| Environment | | Provided sterile;<br>Single use | Provided sterile;<br>Single use | YES/YES |
| Sterilization Method;<br>Sterility Assurance Level<br>(SAL) | | Ethylene oxide;<br>10-6 | Ethylene oxide;<br>10-6 | YES/YES |
| | | SINGLE-USE RF PROBE | | |
| Device Name | | Pain Management Probe | Polaris Single-Use RF<br>Probe | N/A |
| 510(k) # | | K002389 | K181864 | N/A |
| Configuration | | Monopolar | Monopolar | YES/YES |
| Used with Compatible<br>Cannula | | Yes | Yes | YES/YES |
| Thermocouple for Tissue<br>Temperature Measurement | | Yes | Yes | YES/YES |
| Patient-<br>contact<br>Materials | Shaft, Stylet | Stainless steel | Stainless steel | YES/YES |
| Key<br>Dimensions | Shaft<br>Diameter | 24, 28, 31 Ga | 28 Ga | NO/YES |
| | Compatible<br>Cannula<br>Length(s) | 50, 54, 60, 100, 145,<br>200 mm | 50, 100, and 150 mm | |
| Environment | | Provided non-sterile;<br>Re-sterilisable (steam) | Provided sterile;<br>Single use | NO/YES |
| Sterilization Method;<br>Sterility Assurance Level<br>(SAL) | | N/A | Ethylene oxide;<br>10-6 | NO/YES |
| REUSABLE RF PROBE | | | | |
| Device Name | | Pain Management Probe | Polaris Reusable RF<br>Probe | N/A |
| 510(k) # | | K002389 | K181864 | N/A |
| Configuration | | Monopolar | Monopolar | YES/YES |
| Used with Cannula | | Yes | Yes | YES/YES |
| Thermocouple for Tissue<br>Temperature Measurement | | Yes | Yes | YES/YES |
| | | PREDICATE DEVICE | SUBJECT DEVICE | Identical/<br>SE |
| Patient-<br>contact<br>Materials | Shaft, Stylet | Stainless steel or nitinol | Stainless steel or nitinol | YES/YES |
| Key<br>Dimensions | Shaft<br>Diameter | 24, 28, 31 Ga | 28 Ga | NO/YES |
| | Compatible<br>Cannula<br>Length(s) | 50, 54, 60, 100, 145,<br>200 mm | 50, 100, and 150 mm | NO/YES |
| Environment | | Provided non-sterile;<br>Re-sterilisable (steam) | Provided non-sterile;<br>Re-sterilisable (steam) | YES/YES |
| | | RF CANNULA | | |
| Device Name | | BMC RF Cannula | Polaris RF Cannula | N/A |
| 510(k) # | | K972846 | K181864 | N/A |
| Patient-<br>contact<br>Materials | Shaft, Stylet | Stainless steel | Stainless steel | YES/YES |
| | Shaft<br>Insulation | Polyethylene<br>terephthalate | Polyethylene<br>terephthalate | YES/YES |
| | Hub | Copolyester | Copolyester | YES/YES |
| Key<br>Dimensions | Shaft<br>Diameter | 16, 18, 20, 21, 22 Ga | 16, 18, 20, 22 Ga | NO/YES |
| | Usable<br>Length | 50, 54, 60, 100, 145,<br>200 mm | 50, 100, 150 mm | NO/YES |
| | Bare Active<br>Tip | 2, 4, 5, 10, 15 mm | 5, 10 mm | NO/YES |
| Tip Configurations | | Straight sharp<br>Straight blunt<br>Curved sharp<br>Curved blunt | Straight sharp<br>Curved sharp<br>Curved blunt | NO/YES |
| Environment | | Provided sterile;<br>Single use | Provided sterile;<br>Single use | YES/YES |
| Sterilization Method;<br>Sterility Assurance Level<br>(SAL) | | Ethylene oxide;<br>10-6 | Ethylene oxide;<br>10-6 | YES/YES |
| | | PUMP UNIT | | |
| Device Name | | TransDiscal Pump Unit | Polaris Pump Unit | N/A |
| 510(k) # | | K062937 | K181864 | N/A |
| Number of Pump Heads | | Two | Two | YES/YES |
| Pump Controlled by<br>Generator | | Yes | Yes | YES/YES |
| Used with Tube Kit(s) | | Yes | Yes | YES/YES |
| Safety Switch | | Yes | Yes | YES/YES |
| Used with Desk Stand | | No | Yes | NO/YES |
| Environment | | Supplied non-sterile;<br>Non-sterilisable | Supplied non-sterile;<br>Non-sterilisable | YES/YES |
| CONNECTOR HUB | | | | |
| Device Name | | TransDiscal Y-<br>Connecting Cable | Polaris Connector Hub | N/A |
| 510(k) # | | K062937 | K181864 | N/A |
| Number of Probe<br>Connections Possible | | Two | Four | NO/YES |
| | | PREDICATE DEVICE | SUBJECT DEVICE | Identical/<br>SE |
| Connector | Probe | 5-pin female | 7-pin female | NO/YES |
| | Generator | 19-pin male | 33-pin HG4 male | NO/YES |
| Key Cable Materials | | Silicone, Thermoplastic elastomer | Thermoplastic elastomer, injection molded plastic | NO/YES |
| Cable Length | | 9 feet | 8 feet | NO/YES |
| Environment | | Supplied non-sterile; Non-sterilisable | Supplied non-sterile; Non-sterilisable | YES/YES |
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
#### 7.7 Performance Testing
Performance testing was completed to demonstrate substantial equivalence of the Polaris RF Ablation System and relevant system components to the identified predicate devices in Table 7.1. The system components were subjected to the verification and validation testing listed in Table 7.4.
# Table 7.4: Performance Testing of Subject Device
| Tests | Test Method Summary | Results | | |
|------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Mechanical | Polaris Re-Usable RF Probe | | | |
| Mechanical | The following tests were conducted<br>to verify whether the subject device<br>was capable of withstanding the<br>relevant mechanical stresses without<br>failure:<br>• Temperature Test<br>• Pull Test<br>• Cleaning and Sterilization<br>Reuse Test<br>• Mechanical Reuse Test<br>• Mechanical Test (Flexion)<br>• Continuity Test<br>• Functional Test | All samples passed the<br>acceptance criteria. Results<br>confirmed compliance of the<br>device with IEC 60601-2-<br>2:2009 and support<br>substantial equivalence with<br>the predicate device. | | |
| Electrical | Electrical verification was conducted<br>through comparison with the Polaris<br>Standard Single-Use Probe to verify<br>whether the device was capable of<br>withstanding the relevant electrical<br>stresses without failure. | The device was verified to<br>meet the requirements of IEC<br>60601-1:2005+A1:2012 and<br>IEC 60601-2-2:2009. Results<br>support substantial<br>equivalence with the predicate<br>device. | | |
{15}------------------------------------------------
| Tests | Test Method Summary | Results |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | Biocompatibility of the subject device<br>was demonstrated through<br>comparisons to similar legally<br>marketed devices as per current ISO<br>10993-1 requirements and the FDA<br>guidance, "Use of International<br>Standard ISO 10993-1, "Biological<br>evaluation of medical devices - Part<br>1: Evaluation and testing within a<br>risk management process". | Results confirmed compliance<br>of the device with the current<br>ISO 10993-1 requirements<br>and support substantial<br>equivalence with the predicate<br>device. |
| Packaging | Ship testing was performed to ensure<br>the integrity of the subject device<br>packaging through the rigors of<br>shipping and handling. | All packaging met relevant<br>requirements. Results support<br>substantial equivalence with<br>the predicate device. |
| Cleaning and Sterilization | The recommended reprocessing<br>methods provided in the Instructions<br>for use were validated to ensure the<br>recommended parameters and<br>method could achieve a Sterility<br>Assurance Level of 10-6. | The device was verified to<br>meet the requirements of<br>AAMI TIR12, AAMI TIR30 and<br>ANSI/AAMI ST81<br>requirements and the FDA<br>guidance, "Reprocessing<br>Medical Devices in Health Care<br>Settings: Validation Methods<br>and Labeling". Results support<br>substantial equivalence with<br>the predicate device. |
| Mechanical | <b>Polaris Single-Use RF Probe</b><br>The following tests were conducted<br>to verify whether the subject device<br>was capable of withstanding the<br>relevant mechanical stresses without<br>failure:<br>● Temperature Test<br>● Pull Test<br>● Durability Test<br>● Mechanical Test (Flexion)<br>● Continuity Test<br>● Functional Test | All samples passed the<br>acceptance criteria. Results<br>confirmed compliance of the<br>device with IEC 60601-2-<br>2:2009 and support<br>substantial equivalence with<br>the predicate device. |
| Tests | Test Method Summary | Results |
| Electrical | The following tests were conducted<br>to verify whether the subject device<br>was capable of withstanding the<br>relevant electrical stresses without<br>failure:<br>High Frequency Leakage<br>Current Test - Cable High Frequency Dielectric<br>Strength Test – Cable High Frequency Dielectric<br>Strength Test - Handle Mains Frequency Dielectric<br>Strength Test - Cable Mains Frequency Dielectric<br>Strength Test - Handle Mains Frequency Dielectric<br>Strength Test – Connector | All samples passed the<br>acceptance criteria. Results<br>confirmed compliance of the<br>device with IEC 60601-<br>1:2005+A1:2012 and IEC<br>60601-2-2:2009 and support<br>substantial equivalence with<br>the predicate device. |
| Biocompatibility | Biocompatibility of the subject device<br>was demonstrated through<br>comparisons to similar legally<br>marketed devices as per current ISO<br>10993-1 requirements and the FDA<br>guidance, "Use of International<br>Standard ISO 10993-1, "Biological<br>evaluation of medical devices - Part<br>1: Evaluation and testing within a<br>risk management process". | Results confirmed compliance<br>of the device with the current<br>ISO 10993-1 requirements<br>and support substantial<br>equivalence with the predicate<br>device. |
| Packaging | Ship testing was performed to ensure<br>the integrity of the subject device<br>packaging through the rigors of<br>shipping and handling. The seal<br>strength and sterile barrier integrity<br>was validated per ANSI/AAMI/ISO<br>11607-1 and 11607-2 over the shelf<br>life of the device. | The device packaging met<br>relevant requirements. Results<br>confirmed compliance of<br>device packaging with<br>ANSI/AAMI/ISO 11607-1 and<br>11607-2 and support<br>substantial equivalence with<br>the predicate device. |
| Tests | Test Method Summary | Results |
| | Polaris RF Cannula | |
| Mechanical | The following tests were conducted<br>to verify whether the subject device<br>was capable of withstanding the<br>relevant mechanical stresses without<br>failure:<br>Positive Pressure Liquid<br>Leakage Test Sub-atmospheric Pressure Air<br>Leakage Test Resistance to Separation Test Stress Cracking Test Hub Strength Test | All samples passed the<br>acceptance criteria. Results<br>confirmed the device met with<br>the mechanical requirements<br>of ISO 80369-7:2016, ISO<br>594-1:1986 and ISO<br>7864:2016 and support<br>substantial equivalence with<br>the predicate device. |
| Electrical | The following tests were conducted<br>to verify whether the subject device<br>was capable of withstanding the<br>relevant electrical stresses without<br>failure:<br>High Frequency Leakage<br>Current Test High Frequency Dielectric<br>Strength Test Mains Frequency Dielectric<br>Strength Test | All samples passed the<br>acceptance criteria. Results<br>confirmed compliance of the<br>device with IEC 60601-<br>1:2005+A1:2012 and IEC<br>60601-2-2:2009 and support<br>substantial equivalence with<br>the predicate device. |
| Biocompatibility | Biocompatibility of the subject device<br>was demonstrated through biological<br>testing as per current ISO 10993-1<br>requirements and the FDA guidance,<br>"Use of International Standard ISO<br>10993-1, "Biological evaluation of<br>medical devices - Part 1: Evaluation<br>and testing within a risk<br>management process". | Results confirmed compliance<br>of the device with the current<br>ISO 10993-1 requirements<br>and support substantial<br>equivalence with the predicate<br>device. |
| Packaging | Ship testing was performed to ensure<br>the integrity of the subject device<br>packaging through the rigors of<br>shipping and handling. The seal<br>strength and sterile barrier integrity<br>was validated per ANSI/AAMI/ISO<br>11607-1 and 11607-2 over the shelf<br>life of the device. | The device packaging met<br>relevant requirements. Results<br>confirmed compliance of<br>device packaging with<br>ANSI/AAMI/ISO 11607-1 and<br>11607-2 and support<br>substantial equivalence with<br>the predicate device. |
| Tests | Test Method Summary | Results |
| | Polaris Cooled RF Probe Kit | |
| Mechanical | The following tests were conducted to verify whether the Polaris Cooled RF Probe was capable of withstanding the relevant mechanical stresses without failure:<br>Leak/Pressure Test Pull Test Temperature Test Tip Compression Test Mechanical Test (Flexion) Continuity Tes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.