K063753 · Aci, Inc. · GXD · Jun 20, 2007 · Neurology
Device Facts
Record ID
K063753
Device Name
GFX NERVE ABLATION SYSTEM
Applicant
Aci, Inc.
Product Code
GXD · Neurology
Decision Date
Jun 20, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4400
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Lesion Probe and Generator system is intended to create radiofrequency (RF) heat lesions in nerve tissue. It is intended for use only by trained clinicians in a hospital or clinical setting.
Device Story
GFX Nerve Ablation System comprises RF generator (Model RFG-100) and single-use probe (Model PB-100T). System used by clinicians in hospital/clinical settings to locate target nerves via stimulation and subsequently inhibit nerve function by creating thermal lesions using RF energy. Generator monitors temperature, impedance, and power output to ensure safety during ablation. System provides minimally invasive alternative for nerve function inhibition.
Clinical Evidence
Bench testing only. Device integrity and functionality validated under routine manufacturing conditions. Compliance with ISO 10993-1 (biological evaluation) and AAMI/ISO 10993-7 (ethylene oxide residues) confirmed.
Technological Characteristics
Radiofrequency lesion generator system. Single-use probes. Energy source: RF. Safety features: temperature, impedance, and power output monitoring. Sterilization: Ethylene oxide per ISO 11135. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for creation of radiofrequency heat lesions in nerve tissue by trained clinicians in hospital or clinical settings.
Regulatory Classification
Identification
A radiofrequency lesion generator is a device used to produce lesions in the nervous system or other tissue by the direct application of radiofrequency currents to selected sites.
K081729 — MODIFICATION TO GFX NERVE ABLATION SYSTEM · Bioform Medical, Inc. · Aug 7, 2008
K050084 — COSMAN RF LESION GENERATOR, MODEL RFG-1A; COSMAN TEW TRIGEMINAL ELECTRODE KIT; COSMAN TIC TRIGEMINAL ELECTRODE KIT; · Cosman Medical, Inc. · Mar 17, 2005
K181864 — Polaris RF Ablation System · Baylis Medical Company, Inc. · Jan 2, 2019
K071745 — BAYLIS PAIN MANAGEMENT SINGLE-USE PROBE · Baylis Medical Co., Inc. · Jul 19, 2007
K052878 — NT-1000 · Neuro Therm, Inc. · Jan 23, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(K) SUMMARY 11063753
### DATE: 1/16/07
### SUBMITTER INFORMATION
| A. Company Name: | ACI Inc. | JUN 2 0 2002 |
|-----------------------|--------------------------------------------------------|--------------|
| B. Company Address: | 10940 Parker Road, Suite 513<br>Parker, Colorado 80134 | |
| C. Company Phone: | (303) 790-7109 | |
| D. Company Facsimile: | (303) 790-1029 | |
| E. Company Contact: | W. Michael Janssen<br>President / CTO | |
#### DEVICE IDENTIFICATION
| A. Device Trade Name: | Lesion Probe and Generator System |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| B. Device Common Name: | GFX Nerve Ablation System consisting of;<br>GFX Nerve Ablation Generator Model RFG-100 and<br>GFX Nerve Ablation Probe Model PB-100T |
| C. Classification Name: | Radiofrequency Lesion Generator |
| D. Device Class: | Class II |
| E. Device Code: | GXD (proposed) (21 CFR 882.4400) |
#### PREDICATE DEVICES
Trade Name: Radionics RFG-3CF 510(k) Number: K965182
Trade Name: Mercury Medical Disposable RF Cannula 510(k) Number: K000073
## DEVICE DESCRIPTION
## GFX Nerve Ablation System (generator and probe):
The GFX Nerve Ablation System provides a minimally invasive technique for creating a neural lesion inhibiting the function of the target nerve. The Generator and probe are used as a system to both stimulate the nerve for the purpose of locating the probe correctly and to create a neural lesion to inhibit nerve function through the application of RF energy. The GFX Nerve Ablation probes are single use devices supplied sterile to the customer.
Page 8 of 57
{1}------------------------------------------------
#### INTENDED USE
The Lesion Probe and Generator system is intended to create radiofrequency (RF) heat lesions in nerve tissue. It is intended for use only by trained clinicians in a hospital or clinical setting.
### COMPARISON TO PREDICATE DEVICES
The GFX Nerve Ablation System is substantially equivalent in the following technological ways to the identified predicate devices;
- Indications For Use -
- RF Energy Output Type
- o Frequency
- o Power
- o Wave Type (pseudo-sinusoidal)
- Treatment Time o
- Functions (stimulation and ablation)
- Safety Monitoring Systems
- Temperature o
- Impedance o
- o Power Output
- Probe Type (single use) -
## BIOCOMPATIBILITY, STERILIZATION, PACKAGING, AND BENCH TESTING
ACI Inc. conducts and maintains valid ethylene oxide sterilization processes in accordance with ISO 11135, Medical Devices - Validation and Routine Control of Ethylene Oxide Sterilization. Package integrity for the sterile probe is initially validated and, in addition, 100% of ACI Inc. devices undergo packaging visual verification prior to transfer to finished goods inventory.
Device integrity and functionality were qualified and/or validated using samples produced under routine manufacturing conditions. All Lesion Probes and Generators meet ACI Inc.'s in-house requirements, and requirements listed in ISO 10993-1, Biological Evaluation of Medical Devices and AAMI/ISO 10993-7, Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residues.
### CONCLUSION
ﺳ
The GFX Nerve Ablation System (Generator and Probe) are substantially equivalent in technological characteristics and intended use to the predicate devices identified in this summary.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized graphic of an eagle or bird-like figure, composed of three curved lines that suggest movement or flight.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
ACI Inc. % Intertek Testing Services Mr. Daniel W. Lehtonen Sr. Staff Engineer - Medical Devices 2307 East Aurora Road, Unit B7 Twinsburg, Ohio 44087
JUN 2 0 2007
Re: K063753
Trade/Device Name: GFX Nerve Ablation System Regulation Number: 21 CFR 882.4400 Regulation Name: Radiofrequency lesion generator Regulatory Class: II Product Code: GXD, GXI Dated: June 4, 2007 Received: June 5, 2007
Dear Mr. Anselmo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 – Mr. Clay Anselmo
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
s,
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications For Use
510(k) Number (if known): K063753
Device Name: GFX Nerve Ablation System
Indications for Use: The Lesion Probe and Generator system is intended to create radiofrequency (RF) heat lesions in nerve tissue. It is intended for use only by trained clinicians in a hospital or clinical setting.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Concurrence of CDRH, Office of Device Evalu
Over-The-Counter Use (21 CFR 801 Subpart C)
Division of General, Restorative,
Page 1 of
(Division Sign-Off)
510(k) Number
and Neurological Devices
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Page 5 of 57
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.