K161443 · Ekso Bionics, Inc. · PHL · Jul 19, 2016 · Physical Medicine
Device Facts
Record ID
K161443
Device Name
Ekso™ (version 1.1) and Ekso GT™ (version 1.2)
Applicant
Ekso Bionics, Inc.
Product Code
PHL · Physical Medicine
Decision Date
Jul 19, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3480
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ekso™ (version 1.1) and Ekso GT™ (version 1.2) are intended to perform ambulatory functions in rehabilitation institutions under the supervision of a trained physical therapist for the following population: • Individuals with hemiplegia due to stroke (upper extremity motor function of at least 4/5 in at least one arm) • Individuals with spinal cord injuries at levels T4 to L5 (upper extremity motor function of at least 4/5 in both arms) • Individuals with spinal cord injuries at levels of C7 to T3 (ASIA D with upper extremity motor function of at least 4/5 in both arms). The therapist must complete a training program prior to use of the device. The devices are not intended for sports or stair climbing.
Device Story
Ekso is a powered motorized orthosis; consists of fitted metal brace supporting legs, feet, and torso; worn via straps. Battery-powered motors drive hip and knee joints. Integrated torso contains computer and power supply. Operated by physical therapist using handheld interface to specify settings and initiate steps. Used in rehabilitation institutions with mobility aids (cane, crutch, or walker). Device assists patients with sit, stand, walk, and turn functions. Provides visual/auditory feedback to therapist. Failsafe: power failure locks knees and frees hips (passive brace mode). Benefits include enabling ambulatory function in clinical settings for patients with specific neurological impairments.
Clinical Evidence
Clinical evidence includes two prospective, non-randomized studies of stroke patients with hemiplegia. Study 1 (n=54) and Study 2 (n=8) evaluated safety and performance over 216 total walking sessions. 93% of patients in Study 1 and 88% in Study 2 safely performed ambulatory functions using only one arm for stability. No falls or adverse events reported. Data supports safety and effectiveness for the expanded indication.
Technological Characteristics
Powered motorized orthosis; metal brace with adjustable hip, leg, and foot components. Energy source: 48.1V rechargeable lithium-ion battery. Connectivity: handheld controller interface. Safety: IEC 60601-1:2005, IEC 60601-1-2:2007. Failsafe: passive knee lock/hip free mode. Weight: 50 lbs. Max speed: ~2 km/hr. Operating temp: 10° to 95°F.
Indications for Use
Indicated for adults >18 with hemiplegia due to stroke (UE motor function ≥4/5 in at least one arm), spinal cord injury (SCI) T4-L5 (UE motor function ≥4/5 in both arms), or SCI C7-T3 ASIA D (UE motor function ≥4/5 in both arms). Contraindicated for sports or stair climbing.
Regulatory Classification
Identification
A powered lower extremity exoskeleton is a prescription device that is composed of an external, powered, motorized orthosis that is placed over a person's paralyzed or weakened limbs for medical purposes.
Special Controls
In combination with the general controls of the FD&C Act, the Powered Exoskeleton is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Elements of the device materials that may contact the patient must be demonstrated to be biocompatible.
(2) Appropriate analysis/testing must validate electromagnetic compatibility/interference (EMC/EMI), electrical safety, thermal safety, mechanical safety, battery performance and safety, and wireless performance, if applicable.
(3) Appropriate software verification, validation, and hazard analysis must be performed.
(4) Design characteristics must ensure geometry and materials composition are consistent with intended use.
(5) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Performance testing must include:
(i) Mechanical bench testing (including durability testing) to demonstrate that the device will withstand forces, conditions, and environments encountered during use;
(ii) Simulated use testing (
*i.e.,* cyclic loading testing) to demonstrate performance of device commands and safeguard under worst case conditions and after durability testing;(iii) Verification and validation of manual override controls are necessary, if present;
(iv) The accuracy of device features and safeguards; and
(v) Device functionality in terms of flame retardant materials, liquid/particle ingress prevention, sensor and actuator performance, and motor performance.
(6) Clinical testing must demonstrate a reasonable assurance of safe and effective use and capture any adverse events observed during clinical use when used under the proposed conditions of use, which must include considerations for:
(i) Level of supervision necessary, and
(ii) Environment of use (
*e.g.,* indoors and/or outdoors) including obstacles and terrain representative of the intended use environment.(7) A training program must be included with sufficient educational elements so that upon completion of training program, the clinician, user, and companion can:
(i) Identify the safe environments for device use,
(ii) Use all safety features of device, and
(iii) Operate the device in simulated or actual use environments representative of indicated environments and use.
(8) Labeling for the Physician and User must include the following:
(i) Appropriate instructions, warning, cautions, limitations, and information related to the necessary safeguards of the device, including warning against activities and environments that may put the user at greater risk.
(ii) Specific instructions and the clinical training needed for the safe use of the device, which includes:
(A) Instructions on assembling the device in all available configurations;
(B) Instructions on fitting the patient;
(C) Instructions and explanations of all available programs and how to program the device;
(D) Instructions and explanation of all controls, input, and outputs;
(E) Instructions on all available modes or states of the device;
(F) Instructions on all safety features of the device; and
(G) Instructions for properly maintaining the device.
(iii) Information on the patient population for which the device has been demonstrated to have a reasonable assurance of safety and effectiveness.
(iv) Pertinent non-clinical testing information (
*e.g.,* EMC, battery longevity).(v) A detailed summary of the clinical testing including:
(A) Adverse events encountered under use conditions,
(B) Summary of study outcomes and endpoints, and
(C) Information pertinent to use of the device including the conditions under which the device was studied (
*e.g.,* level of supervision or assistance, and environment of use (*e.g.,* indoors and/or outdoors) including obstacles and terrain).
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three stylized human profiles facing to the right, with a design element below them that resembles flowing water or fabric.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 19, 2016
Ekso Bionics, Inc. % John J. Smith, MD, JD Regulatory Counsel Hogan Lovells U.S. L.L.P. 555 Thirteenth Street NW Washington, DC 20004
Re: K161443
Trade/Device Name: Ekso™ (version 1.1) and Ekso GTTM (version 1.2) Regulation Number: 21 CFR 890.3480 Regulation Name: Powered Lower Extremity Exoskeleton Regulatory Class: Class II Product Code: PHL Dated: Mav 25, 2016 Received: May 25, 2016
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -A
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161443
Device Name
EksoTM (version 1.1) and Ekso GTTM (version 1.2)
Indications for Use (Describe)
The Ekso™ (version 1.1) and Ekso GT™ (version 1.2) are intended to perform ambulatory functions in rehabilitation institutions under the supervision of a trained physical therapist for the following population:
· Individuals with hemiplegia due to stroke (upper extremity motor function of at least one arm)
- · Individuals with spinal cord injuries at levels T4 to L5 (upper extremity motor function of at least 4/5 in both arms)
· Individuals with spinal cord injuries at levels of C7 to T3 (ASIA D with upper extremity motor function of at least 4/5 in both arms).
The therapist must complete a training program prior to use of the devices are not intended for sports or star climbing.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary as required by 21 CFR 807.92(c)
| Device name | Ekso |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitters name &<br>contact info | Ekso Bionics® Inc.<br>1414 Harbour Way South<br>Suite 1201<br>Richmond, CA 94804 |
| | <b>Contact Details:</b><br>Thomas Looby<br>CEO<br>Tel: +1 937-838-0842<br>Email: tom@eksobionics.com<br>Tel: +1 (510) 984-1761<br>Fax: +1 (510) 927-2647 |
| Preparation Date | June 27, 2016 |
| Device Name &<br>Classification | Trade Name: Ekso™ (version 1.1) and Ekso GT™ (version 1.2)<br>Common Name: Exoskeleton<br>Classification Name: Powered Exoskeleton<br>Device Classification: Class II, 21 CFR 890.3480<br>Product Code: PHL |
| Indications for Use | The Ekso™ (version 1.1) and Ekso GT™ (version 1.2) are intended to perform ambulatory functions in rehabilitation institutions under the supervision of a trained physical therapist for the following population:<br>• Individuals with hemiplegia due to stroke (upper extremity motor function of at least 4/5 in at least one arm)<br>• Individuals with spinal cord injuries at levels T4 to L5 (upper extremity motor function of at least 4/5 in both arms)<br>• Individuals with spinal cord injuries at levels of C7 to T3 (ASIA D with upper extremity motor function of at least 4/5 in both arms).<br>The therapist must complete a training program prior to use of the device. The devices are not intended for sports or stair climbing. |
| Device Description | The Ekso is a powered motorized orthosis. It consists of a fitted metal brace that supports the legs, feet, and torso. It is worn via straps on the body, legs, and feet. Battery powered motors drive knee and hip joints. It has an integrated solid torso containing the computer and power supply. It has a hand-held user interface to specify settings and initiate steps. The Ekso is used with a cane, crutch, or walker. |
| Predicate Device | Manufacturer: Ekso Bionics<br>510(k): K143690<br>Date of Clearance: 4/1/2016 |
{5}------------------------------------------------
| Manufacturer | Ekso Bionics®, Inc. | Predicate Ekso<br>Bionics®, Inc. | Differences |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Ekso™ (version 1.1)<br>and Ekso GT™ (version<br>1.2) | Ekso™ (version 1.1)<br>and Ekso GT™ (version<br>1.2) | |
| 510(k) Number | TBD | K143690 | N/A |
| Product Code | PHL | PHL | Same |
| Regulation Number | 890.3480 | 890.3480 | Same |
| Regulation Name | Powered Exoskeleton | Powered Exoskeleton | Same |
| Indications for Use | The Ekso™ is intended<br>to perform ambulatory<br>functions in<br>rehabilitation<br>institutions under the<br>supervision of a trained<br>physical therapist for<br>the following<br>population:<br>Individuals with<br>hemiplegia due to<br>stroke (upper<br>extremity motor<br>function of at least 4/5<br>in at least one arm),<br>Individuals with spinal<br>cord injuries at levels<br>T4 to L5 (upper<br>extremity motor<br>function of at least 4/5<br>in both arms),<br>Individuals with spinal<br>cord injuries at levels<br>of C7 to T3 (ASIA D<br>with upper extremity<br>motor function of at<br>least 4/5 in both arms).<br>The therapist must<br>complete a training<br>program prior to use of<br>the device. The<br>devices are not<br>intended for sports or<br>stair climbing. | The Ekso™ is intended<br>to perform ambulatory<br>functions in<br>rehabilitation<br>institutions under the<br>supervision of a trained<br>physical therapist for<br>the following<br>population with upper<br>extremity motor<br>function of at least 4/5<br>in both arms:<br>Individuals with<br>hemiplegia due to<br>stroke, Individuals with<br>spinal cord injuries at<br>levels T4 to L5, and<br>Individuals with spinal<br>cord injuries at levels<br>of C7 to T3 (ASIA D).<br>The therapist must<br>complete a training<br>program prior to use of<br>the device. The Ekso<br>1.1 and 1.2 are not<br>intended for sports or<br>stair climbing. | Similar; the new<br>indication is clarified<br>to explicitly include<br>individuals with<br>hemiplegia due to<br>stroke, who have<br>upper extremity motor<br>function of at least 4/5<br>in only one arm.<br>Individuals with spinal<br>cord injury still require<br>upper extremity motor<br>function of at least 4/5<br>in both arms. |
| Body Coverage | Worn over legs and<br>upper body with rigid<br>torso | Worn over legs and<br>upper body with rigid<br>torso | Same |
| Manufacturer | Ekso Bionics®, Inc. | Predicate Ekso<br>Bionics®, Inc. | Differences |
| Trade Name | Ekso™ (version 1.1)<br>and Ekso GT™ (version<br>1.2) | Ekso™ (version 1.1)<br>and Ekso GT™ (version<br>1.2) | |
| Size of Components | Adjustable upper leg,<br>lower leg, and hip<br>width; control unit<br>integrated into the<br>torso | Adjustable upper leg,<br>lower leg, and hip<br>width; control unit<br>integrated into the<br>torso | Same |
| Mobility Aid | Walker, Crutches, Cane | Walker, Crutches, Cane | Same |
| Ability of User<br>Mobility | Sit, stand, walk, and<br>turn | Sit, stand, walk, and<br>turn | Same |
| Walking Speed | ~2 km/hr | ~2 km/hr | Same |
| Grade of Inclination | 1.15 deg | 1.15 deg | Same |
| Type of Surface | Smooth, cement,<br>carpet | Smooth, cement,<br>carpet | Same |
| Patient Population | Adults over age of 18<br>with hemiplegia due to<br>stroke, Spinal Cord<br>Injury (SCI) from T4 to<br>L5, and SCI from C7 to<br>L5 ASIA D | Adults over age of 18<br>with hemiplegia due to<br>stroke, Spinal Cord<br>Injury (SCI) from T4 to<br>L5, and SCI from C7 to<br>L5 ASIA D | Same |
| Height of Patient | ~62″ to 74″ (1.58 m to<br>1.88 m) | ~62″ to 74″ (1.58 m to<br>1.88 m) | Same |
| Weight of Patient | Up to 220 lbs (100kg) | Up to 220 lbs (100kg) | Same |
| Control Method | Handheld interface for<br>PT; weight shift to<br>initiate steps | Handheld interface for<br>PT; weight shift to<br>initiate steps | Same |
| Range of Motion | Hips: 135° flexion to<br>20° extension<br>Knees: 130° flexion to<br>0° extension<br>Ankles: 10° flexion to<br>10° extension | Hips: 135° flexion to<br>20° extension<br>Knees: 130° flexion to<br>0° extension<br>Ankles: 10° flexion to<br>10° extension | Same |
| Weight | 50 lbs (23 kg) | 50 lbs (23 kg) | Same |
| Rechargeable Battery | Rechargeable lithium<br>ion batteries 48.1V,<br>30A peak current, 1<br>hour of continuous<br>usage per charge | Rechargeable lithium<br>ion batteries 48.1V,<br>30A peak current, 1<br>hour of continuous<br>usage per charge | Same |
| Battery Charge Time | 1 hour | 1 hour | Same |
| Expected Useable Life | 4 years | 4 years | Same |
| Training Program | Yes | Yes | Same |
| Certification Program | Yes | Yes | Same |
| Manufacturer | Ekso Bionics®, Inc. | Predicate Ekso<br>Bionics®, Inc. | Differences |
| Trade Name | Ekso™ (version 1.1)<br>and Ekso GT™ (version<br>1.2) | Ekso™ (version 1.1)<br>and Ekso GT™ (version<br>1.2) | |
| User Feedback | Provides visual<br>feedback on the<br>handheld controller<br>and auditory feedback | Provides visual<br>feedback on the<br>handheld controller<br>and auditory feedback | Same |
| Fall Detection and<br>Mitigation | None | None | Same |
| Failsafe Feature | In event of power<br>failure- knees become<br>locked and hips free<br>(similar to typical<br>passive leg braces) | In event of power<br>failure- knees become<br>locked and hips free<br>(similar to typical<br>passive leg braces) | Same |
| Operating<br>Temperature | 10° to 95°F (-12° to -<br>35°C) | 10° to 95°F (-12° to -<br>35°C) | Same |
| Operating Humidity | Not available | Not available | Same |
| Electrical Safety<br>Testing | IEC 60601-1:2005 with<br>US deviations | IEC 60601-1:2005 with<br>US deviations | Same |
| Electromagnetic<br>Compatibility Testing | Passed IEC 60601-1-2:<br>2007 | Passed IEC 60601-1-2:<br>2007 | Same |
{6}------------------------------------------------
{7}------------------------------------------------
# Table 2 - Clinical Information
| Study | Description…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.