K182242 · Mespere Lifesciences, Inc. · QEM · Dec 26, 2018 · Cardiovascular
Device Facts
Record ID
K182242
Device Name
VO200 – NeurOs Cerebral Oximetry System
Applicant
Mespere Lifesciences, Inc.
Product Code
QEM · Cardiovascular
Decision Date
Dec 26, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
The VO200 – NeurOs Cerebral Oximetry is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation of blood in the brain tissue beneath the sensor. The prospective clinical value of data from the VO200 - NeurOs Cerebral Oximetry has not been demonstrated in disease states. The VO200 - NeurOs Cerebral Oximetry should not be used as the sole basis for diagnosis or therapy.
Device Story
The VO200 - NeurOs Cerebral Oximetry System is a non-invasive monitor for regional cerebral tissue oxygen saturation (StO2). It utilizes multi-distance diffuse reflectance spectroscopy via a sensor probe placed on the patient's forehead. The sensor contains LEDs (660nm-905nm) and photodetectors to measure light absorption of oxygenated and deoxygenated hemoglobin. The system includes a reusable sensor, disposable adhesive, and display software running on a tablet, laptop, or PC connected via USB. The device processes signals to calculate StO2, providing real-time, continuous trend data graphs to clinicians in hospitals or clinics. It assists healthcare providers in monitoring cerebral oxygenation trends, though it is not intended for diagnostic use or as the sole basis for therapy.
Clinical Evidence
Clinical validation study (STP-9200001) conducted at UCSF on 12 healthy adult volunteers (ages 21-49). Study compared VO200 StO2 readings against invasive jugular bulb and radial artery blood samples analyzed by co-oximeter during induced hypoxia (40-80% saturation range). Results: absolute accuracy ±4.9%, trending accuracy ±1.5%. No adverse events reported.
Technological Characteristics
Multi-distance diffuse reflectance spectroscopy; near-infrared LEDs (660-905nm) and photodetectors. Components: reusable sensor (medical grade ABS), disposable adhesive, USB-connected display software. Connectivity: USB to PC/tablet. Standards: EN ISO 13485, EN ISO 14971, EN 60601-1, EN 60601-1-2, EN 62304, EN ISO 10993.
Indications for Use
Indicated for adult patients as an adjunct trend monitor of regional hemoglobin oxygen saturation in cerebral tissue. Not for use as the sole basis for diagnosis or therapy.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
INVOS 5100C Cerebral/Somatic Oximeter System (K082327)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 26, 2018
Mespere LifeSciences Inc. Rebecca Li Manager of Quality Assurance and Regulatory Affairs 180 Frobisher Drive, Suite 1C Waterloo, N2V 2A2 Ca
Re: K182242
Trade/Device Name: VO200 - NeurOs Cerebral Oximetry System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: QEM Dated: August 10, 2018 Received: August 20, 2018
Dear Rebecca Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Krause =S -S -S -S -S -S -S -S -S -S -S -S -S -S -S -S -S -S -05' -05' -05' -05' -05' -05' -05' -05' -05' -05' -05' -05' 00'
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182242
Device Name
VO200 - NeurOs Cerebral Oximetry System
#### Indications for Use (Describe)
The VO200 - NeurOs Cerebral Oximetry is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation of blood in the brain tissue beneath the sensor. The prospective clinical value of data from the VO200 - NeurOs Cerebral Oximetry has not been demonstrated in disease states. The VO200 - NeurOs Cerebral Oximetry should not be used as the sole basis for diagnosis or therapy.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:10pt">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:10pt">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Appendix 5-1
Image /page/3/Picture/1 description: The image shows the logo for "Mespere Life Sciences". The logo features a large, stylized "M" in gray, with the rest of the company name spelled out in a similar font. Below "Mespere", the words "LIFE SCIENCES" are written in a smaller, blue font. To the left of the "M", there is a vertical blue bar.
180 Frobisher Drive, Suite 1C
Waterloo, ON,
T: 519-884-7575
F: 519-884-8810
K182242
www.mespere.com
# 510(k) SUMMARY
| Submitted by: | Mespere LifeSciences Inc.<br>180 Frobisher Drive, Suite 1C<br>Waterloo, Ontario, Canada<br>N2V 2A2<br>Phone: 519-884-7575<br>Fax: 519-884-8810 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact: | Rebecca S. Li, Manager of Quality Assurance and Regulatory Affairs |
| Date Summary<br>prepared: | 8/30/2018 |
| Trade Name: | VO200 - NeurOs Cerebral Oximetry System |
| Common Name: | NeurOs Cerebral Oximetry |
| Classification<br>name: | Oximeter, Tissue Saturation (21CFR 870.2700) |
| Classification: | Class II Medical Device |
| Product Code: | QEM |
| Predicate Device: | K082327<br>INVOS 5100C Cerebral/Somatic Oximeter System<br>Product Code: QEM<br>Regulation Number: 870.2700<br>This predicate has not been subject to a design related recall |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Mespere Life Sciences. Next to the logo is the text "VO200 – NeurOs Cerebral Oximetry System - Traditional 510(k) Notification". The logo is a stylized "M" with the words "LIFE SCIENCES" below it.
## Intended Use/Indications For Use:
The VO200 – NeurOs Cerebral Oximetry is intended for use as an adjunct trend monitor of regional hemoglobin oxygen saturation of blood in the brain tissue beneath the sensor. The prospective clinical value of data from the VO200 - NeurOs Cerebral Oximetry has not been demonstrated in disease states. The VO200 - NeurOs Cerebral Oximetry should not be used as the sole basis for diagnosis or therapy.
### Device Description:
The Mesepre VO 200 - NeurOs Cerebral Oximetry System is based on the same optical operating principles of multi-distance diffuse reflectance spectroscopy as the predicate device. It measures oxygenated and de-oxygenated hemoglobin content underneath the sensor probe and allows clinicians to monitor the blood oxygenation level of cerebral tissue.
The VO 200 - NeurOs Cerebral Oximetry System consists of a reusable sensor, a single use disposable adhesive, and a display software. The sensor uses two wavelengths within the range from 660nm to 905nm to measure light absorption of tissue, with one wavelength is more sensitive to deoxygenated hemoglobin, the other is more sensitive to oxygenated hemoglobin. Regional cerebral tissue oxygen saturation is calculated through the ratio of tissue absorption to the two wavelengths and validated aqainst blood sample testing through the blood gas analyzer. Sensors are sized to provide desired depth and area coverage for the measurement.
The sensor consists of LEDs, photo detectors, pre-amplifier, analogue to digital convertor, and a CUP for controlling the light emission and detection sequence, auto power adjustment for LED, gain adjustment for photo detectors to ensure optimized signal to noise ratio, and data processing and calculations. It also has an isolator between the monitor and the sensor to ensure the electric safety of the sensor. The enclosure of the sensor is made of medical grade ABS.
The single use sensor adhesive is made of a soft foam with medical grade adhesives.
The Mespere Display Software can be loaded to a tablet, laptop, or stationary PC. The sensor is then connected through a standard USB connector. It displays the measurement through the computer user interface at a refresh rate of approximately one second. Screenshots of the measurement can be saved to a user specified file and can be exported through USB ports or the internet.
#### Technology Characteristics:
The VO200 - NeurOs Cerebral Oximetry System has the same technology characteristics as the predicate.
Similarities:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Mespere Life Sciences. The text next to the logo reads "VO200 - NeurOs Cerebral Oximetry System - Traditional 510(k) Notification". The logo is a stylized "M" with the words "LIFE SCIENCES" written below it.
- Both are based upon multi-distance, diffuse reflectance, near infrared spectroscopy . for analyzing oxygenated and deoxygenated hemoglobin content of blood within cerebral tissue.
- Both use near infrared light emitters (LED) and photo detectors. .
- Both uses two near infrared wavelengths ●
#### Differences:
- . NeurOs sensor probe are reusable, only the medical grade sensor adhesives are one time disposables. The predicate sensor probes are single-use disposables.
- . For the predicate sensor, analogue to digital conversion and pre-Amp is made through boxes connected to the probe through cables.
### Comparison Table:
| Characteristics | Mespere VO200 - NeurOs<br>Cerebral Oximetry System<br>(Subject Device) | INVOS 5100C<br>Cerebral/Somatic Oximeter<br>System<br>(Predicate) |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications For Use | NeurOs Cerebral Oximetry System<br>is intended for use as an adjunct<br>monitor of regional hemoglobin<br>oxygen saturation of blood in the<br>brain tissue beneath the sensor. | INVOS 5100C is intended for use as<br>an adjunct monitor of regional<br>hemoglobin oxygen saturation of<br>blood in the brain or in other tissue<br>beneath the sensor. |
| Target Population | For adult patients. | For adult, pediatric, and neonate<br>patients. |
| Anatomical Sites | Cerebral | Cerebral |
| Where used<br>(hospital, home,<br>ambulance, etc.) | Hospital, clinic and other health<br>care facilities | Hospital, clinic and other health care<br>facilities |
| Technology | near infrared, multi-distance diffuse<br>reflectance spectroscopy, light<br>emitters and photo detectors | near infrared, multi-distance diffuse<br>reflectance spectroscopy, light<br>emitters and photo detectors |
| Sterility | Non-sterile | Non-Sterile |
| Calibration | No need | No Need |
| Features/Method of<br>Use | Non-invasive | Non-Invasive |
| Monitored<br>parameters | Regional Oxygen Saturation | Regional Oxygen Saturation |
| Components | - Sensor<br>- Sensor cables<br>- display software<br>- adhesive patch | - Sensor<br>- Sensor cables<br>- Preamplifiers<br>- Monitor |
| Display | Real-time, continuous | Real-time, continuous |
| Risks | No compromise to patient care | No compromise to patient care |
| Optical | Yes<br>- LEDs and photodetectors | Yes<br>- LED and photodetector |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Mespere Life Sciences. The text next to the logo reads "VO200 – NeurOs Cerebral Oximetry System - Traditional 510(k) Notification". The logo is a stylized "M" in blue and gray.
| Characteristics | Mespere VO200 - NeurOs<br>Cerebral Oximetry System<br>(Subject Device) | INVOS 5100C<br>Cerebral/Somatic Oximeter<br>System<br>(Predicate) |
|-------------------|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Embedded Software | Yes | Yes |
| | Integrated software provides<br>analytics for measurements to be<br>displayed | Integrated software provides<br>analytics for measurements to be<br>displayed |
| Display | Display StO2 value (right and left)<br>and trend data graphs | Display rSO2 values (right and left)<br>and trend data graphs |
| Probe Attachment | Adhesive to skin | Adhesive to skin |
| Alarm | No | Yes |
| Device Type | Needs to be connected to a<br>computer | Standalone Device |
# Complied Standards:
The design of the Mespere VO200 - NeurOs Cerebral Oximetry system conforms to the following voluntary standards:
- EN ISO 13485:2016 ●
- EN ISO 14971:2012 ●
- EN 60601-1:2006/A1:2013 ●
- EN 60601-1-2:2007/AC:2010 ●
- EN 60601-1-6:2010 ●
- EN 62366:2008 ●
- EN ISO 14155:2011 ●
- EN 62304:2006/AC:2008
- EN 1041:2008 ●
- EN 980:2008
- EN 15223-1:2012 ●
- EN ISO 10993-1:2009 ●
- EN ISO 10993-5:2009 ●
- EN ISO 10993-10:2013 .
# Non-Clinical Tests:
The VO200 - NeurOs Cerebral Oximetry System was subjected to extensive safety, performance, and verification testing. Extensive benchtop tests verified the performance of the device to meet the design requirements in an ideal non-clinical environment. All tests were performed to ensure that the device complies with industry and safety standards.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Mespere Life Sciences, followed by the text "VO200 – NeurOs Cerebral Oximetry System - Traditional 510(k) Notification". The logo consists of the word "Mespere" in a combination of blue and gray, with the words "LIFE SCIENCES" in smaller font below. The text indicates that the VO200 is a cerebral oximetry system that has received traditional 510(k) notification.
## Clinical Tests:
Clinical tests were performed to verify and validate the clinical performance of the VO200 -NeurOs Cerebral Oximetry system and support the determination of substantial equivalence. The following clinical data was submitted:
Clinical Validation: The primary objective of this study was to validate the accuracy of the Mespere VO200 – NeurOs Cerebral Oximetry system for measuring the regional cerebral tissue blood oxygenation saturation (StO2), by comparing it to invasive blood samples drawn from the jugular bulb and radial artery, and compared by a co-oximeter. The subjects recruited for this study were healthy volunteers.
Methods: With IRB approval and written Informed Consent Forms, healthy volunteers were recruited. An ultrasound-quided small diameter internal jugular vein catheter was inserted and placed at the position of the subject's superior jugular bulb. A radial arterial line was inserted into a radial artery on the left arm. Two standard pulse oximeters were attached to the patient's fingers to help guide the hypoxia state plateau levels. The VO200 -NeurOs Cerebral Oximetry sensor was placed on the forehead. Blood gas analysis to determine oxyhemoglobin saturation was performed using a Co-Oximeter. Each subject had control data taken at the beginning of each experiment, with two control blood samples drawn while breathing room air. Hypoxia was induced to different levels of oxyhemoglobin saturation (between 70-100% SpO2) by having subjects breathe mixtures of nitrogen, room air, and carbon dioxide. Once a steady state level of hypoxia is achieved, a 1.0ml sample of blood are obtained simultaneously from the jugular line and radial arterial line. A second blood pair of samples, at the same steady-state saturation, is taken 30 seconds later. The VO200 - NeurOs Cerebral Oximetry samples were recorded simultaneously to the blood samples. The blood samples were immediately analyzed by the Co-Oximeter. Up to 27 paired samples were obtained on the plateaus across this span for each subject.
Results: The results showed the absolute accuracy of ± 4.9% and trending accuracy of ±1.5% (in range of 40%-80%), which is equivalent or better than the reported absolute accuracy of 5.0% and trending accuracy of ± 4.0 % of the predicate device.
No adverse events were reported.
| Parameter | Specification | Notes |
|----------------------------------------|---------------|-------|
| StO2 Measurement Range | 0 - 100% | |
| StO2 Validated Measurement Range | 40 - 80% | 1,2,3 |
| Trending Accuracy of StO2<br>(40%-80%) | $\pm$ 1.5% | 1,2,3 |
| Absolute Accuracy of StO2 | $\pm$ 4.9% | 1,2,3 |
Performance specifications
Appendix 5-1
{8}------------------------------------------------
# O200 – NeurOs Cerebral Oximetry System - Traditional 510(k) Notification
| (40%-80%) | |
|--------------------|----|
| Resolution of StO2 | 1% |
NOTES:
- 1 Validation was performed on a measurement range of 40% - 80%
- 2 Accuracy is based on a validation study comparison with venous jugular bulb and radial arterial blood samples and a Co-Oximeter on healthy volunteers. In the StO2 range of 40-80%, the trending accuracy was ± 1.5% and the absolute accuracy was ± 4.9% in the study STP-9200001 at the University of California San Francisco, UCSF (San Francisco, California, USA).
#### 3 Summary of validation study STP-9200001 at the University of California San Francisco, UCSF (San Francisco, California, USA)
#### Number of Study Sites:
Primary Clinical Site:
University of San Francisco California (UCSF)
Hypoxia Labs
513 Parnassus Ave.,
San Francisco, CA, 94143-0542
#### Subjects/data points:
The subjects recruited for this study consisted of healthy, non-smoking individuals of age 21-49 years' old who underwent hypoxia testing at the University of California at San Francisco (UCSF) located in San Francisco, California. No subjects was anemic (Hemoglobin ≤ gm·dl-1).
Twelve (12) healthy subjects were recruited for the study. A total of 284 paired venous and arterial blood sample readings, and 284 VO200 - NeurOs Cerebral Oximetry system readings were used to perform the comparison and statistical analysis.
#### Patient Demographics:
| Subject | Gender | Age | Skin Tone | Fitzpatrick Skin<br>Tone |
|------------|--------|-----|-----------------------|--------------------------|
| Subject #1 | Male | 26 | Olive | IV |
| Subject #2 | Female | 25 | Medium/light<br>olive | III |
| Subject #3 | Male | 26 | Medium/light | III |
Appendix 5-1
{9}------------------------------------------------
# VO200 – NeurOs Cerebral Oximetry System - Traditional 510(k) Notification
| | | | olive | |
|-------------|--------|----|-----------------------|-----|
| Subject #4 | Female | 33 | Fair | II |
| Subject #5 | Male | 24 | Brown | V |
| Subject #6 | Female | 32 | Medium/light<br>olive | III |
| Subject #7 | Male | 23 | Brown | V |
| Subject #8 | Female | 23 | Fair | II |
| Subject #9 | Male | 24 | Brown | V |
| Subject #10 | Male | 30 | Fair | II |
| Subject #11 | Male | 31 | Fair | II |
| Subject #12 | Male | 26 | Brown | V |
#### Agreement Analysis Results:
The calculated statistical results from the study are as follows:
| Trending Accuracy | ± 1.5% |
|-------------------|--------|
| Absolute Accuracy | ± 4.9% |
Bland Altman Plot of Absolute Measurement
Image /page/9/Figure/6 description: This scatter plot shows the difference between Mespere readings and blood gas readings in percentage on the y-axis versus blood gas readings in percentage on the x-axis. The plot includes data points representing the difference, along with three horizontal lines indicating the lower limit of agreement (LoA) at approximately -10, the upper LoA at around 10, and the mean difference near 0. The blood gas readings range from 40% to 85%, while the difference between readings varies from -15% to 15%.
Linear Regression Analysis of Absolute Measurement with 95% Confidence Interval of Slope and
Appendix 5-1
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image displays the logo for Mespere Life Sciences, followed by the text "VO200 - NeurOs Cerebral Oximetry System - Traditional 510(k) Notification". The logo features a stylized "M" in blue and gray, with the words "LIFE SCIENCES" in smaller font below. The text describes the VO200 NeurOs Cerebral Oximetry System and indicates that it is a Traditional 510(k) Notification.
### Intercept using Bootstrap Method
Slope = 0.7346
Intercept = 15.161
Image /page/10/Figure/4 description: The image is a scatter plot comparing Mespere readings to blood gas readings. The x-axis represents blood gas readings in percentage, ranging from 35 to 85. The y-axis represents Mespere readings in percentage, also ranging from 35 to 85. The plot includes data points representing Mespere readings, a line representing y=x, and a linear regression line for the Mespere readings.
Bland Altman of Absolute Trending Measurement
Image /page/10/Figure/6 description: This image is a scatter plot comparing Mespere Trend Readings to Blood Gas Trend Readings. The x-axis represents Blood Gas Trend Readings (%), ranging from approximately -15 to 25. The y-axis represents the difference between Mespere Trend Readings and Blood Gas Trend Readings (%), ranging from -9 to 5. The plot includes a cluster of data points, labeled as 'Difference', along with horizontal lines indicating 'Lower LoA' at approximately -3, 'Upper LoA' at approximately 3, and 'Mean' at approximately 0.
Linear Regression Analysis of Absolute Trending Measurement with 95% Confidence Interval of Slope and Intercept using Bootstrap Method
{11}------------------------------------------------
# VO200 – NeurOs Cerebral Oximetry System - Traditional 510(k) Notification
Image /page/11/Figure/1 description: This image is a scatter plot comparing Mespere Trend Readings (%) to Blood Gas Trend Readings (%). The x-axis represents Blood Gas Trend Readings (%), ranging from -15 to 30, while the y-axis represents Mespere Trend Readings (%), also ranging from -15 to 30. The plot includes a scatter of blue points representing Mespere Readings, an orange line representing y=x, and a dotted blue line representing a linear trendline with the equation y = 0.9497x + 0.0004.
# Conclusions:
The clinical data demonstrates that the VO200 – NeurOs Cerebral Oximetry system is substantially equivalent in performance to the predicate device when measuring regional cerebral tissue blood oxygenation saturation.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.