CMAP PRO

K113074 · Medical Technologies, Inc. · IKN · Feb 2, 2012 · Physical Medicine

Device Facts

Record IDK113074
Device NameCMAP PRO
ApplicantMedical Technologies, Inc.
Product CodeIKN · Physical Medicine
Decision DateFeb 2, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.1375
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

Surface electromyography with video recording, range of motion tracking, isometric functional testing and pinch and grip strength measuring.

Device Story

CMAP Pro is a battery-operated, stand-alone system for monitoring dynamic muscle function. It collects data via surface electromyography (sEMG) sensors, gyroscopic motion sensors, and a proprietary integrated functional capacity evaluation (FCE) sensor. The system acquires continuous analog signals, digitizes them at 2kHz, and transmits data wirelessly to an all-in-one PC. Software processes the signals to provide sEMG readings, motion tracking, and strength measurements, while simultaneously recording video of the subject. Used in clinical settings, the system allows healthcare providers to evaluate neuromuscular activity and functional capacity. The output provides objective data on muscle performance and range of motion, assisting clinicians in assessing patient functional status and guiding rehabilitation or treatment decisions.

Clinical Evidence

Bench testing only. Validation tests were performed to confirm the accuracy and repeatability of the range of motion (ROM) sensors, the FCE sensor, and the EMG leads. The device complies with IEC 60601-2-40 performance standards.

Technological Characteristics

Battery-operated, wireless system. Components: sEMG sensors, gyroscopic motion sensors, proprietary integrated FCE sensor. Signal processing: two-stage conditioning (external to POD and internal). Connectivity: ad hoc wireless network to all-in-one PC. Compliance: IEC 60601-2-40.

Indications for Use

Indicated for surface electromyography with video recording, range of motion tracking, isometric functional testing, and pinch and grip strength measuring in patients requiring neuromuscular monitoring.

Regulatory Classification

Identification

A diagnostic electromyograph is a device intended for medical purposes, such as to monitor and display the bioelectric signals produced by muscles, to stimulate peripheral nerves, and to monitor and display the electrical activity produced by nerves, for the diagnosis and prognosis of neuromuscular disease.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image contains a logo for Med-Tek Medical Technologies Unlimited. The logo consists of a circular graphic on the left and the text "MED-TEK" in bold, uppercase letters to the right. Below "MED-TEK" is the text "MEDICAL TECHNOLOGIES UNLIMITED" in a smaller font size. FEB - 2 2012 # 510(k) Summary for K113074 # Medical Technologies Unlimited, Inc. (MED-TEK) # Comprehensive Muscular Activity Profiler Pro (CMAP Pro™) ### 1. Submitter Information Submitter's Name and Address: Medical Technologies Unlimited, Inc. (Med-Tek) 2665 S. Bayshore Drive #502 Coconut Grove, FL 33133 Submitter's Telephone and fax numbers: Phone: 866-930-2627 Fax: 866-980-2627 Contact Person: Marco Vitiello, M.D. Medical Director Medical Technologies Unlimited, Inc. (Med-Tek) 2665 S. Bayshore Drive #502 Coconut Grove, FL 33133 Date this 510(k) summary was prepared: January 9, 2012. ### 2. General Device Information Comprehensive Muscular Activity Profiler Pro (CMAP Pro TM) Trade Name of Device: Diagnostic Electromyograph ( 21 CFR 890.1375) Classification Name: Predicate Device: Medical Technologies Unlimited Comprehensive Neuromuscular Profiler (CNMP) Predicate Device 510(k) number K031995 ### 3. Device Description: The CMAP Pro™ is a stand-alone, battery operated dynamic muscle function monitoring system, with a number of surface electromyography (sEMG) sensors connected to various parts of the subject's body for data collection. The data are then directly fed into a system for conditioning, acquiring, and transmitting sensor data. Analyzed signals include sEMG readings, motion detection, and muscle strength measurements. The system acquires continuous analog signals and then digitizes these signals by sampling at a rate of 2kHz. These data are then transferred wirelessly to an all-in-one PC for processing using software.. The system is capable of monitoring and video recording data from (SEMG) sensors {1}------------------------------------------------ connected to various muscle groups in the human body. During the acquisition of sEMG signals, the system will simultaneously acquire motion tracking of the body, isometric functional testing and pinch and grip strength measuring. Image /page/1/Figure/2 description: The image shows a diagram of a process flow. The process starts with a subject, then goes to sensors, then to signal conditioning, and finally to PC processing. Each step is represented by a square box with text inside, and the steps are connected by arrows. ### 4. Intended Use of Device: The CMAP Pro is indicated for surface electromyography with video recording, range of motion tracking, isometric functional testing and pinch and grip strength measuring. ### 5. Comparison to Predicate Device: The CMAP Pro and the predicate systems, our CNMP, both use a range of motion device in combination with functional capacity evaluation devices and surface EMG electrodes to obtain similar data. Instead of the electromagnetic field-based range-of-motion unit of the predicate device, the CMAP Pro uses gyroscopic sensors to detect motion. Instead of the 3 off-the-shelf functional capacity evaluation devices of the predicate system, the CMAP Pro uses a Med-Tek proprietary, integrated functional capacity evaluation (FCE) sensor. Both devices utilize surface EMG electrodes. The signal conditioning in the CMAP Pro takes place in two stages, one outside the POD and close to the sEMG sensor leads and the other inside the POD itself. In the predicate device, all signal conditioning took place inside the main unit. The CMAP Pro POD is smaller and thinner than the predicate, and it is battery powered rather than AC powered. The CMAP Pro transmits all data wirelessly via an ad hoc wireless network while the predicate device transmits data via USB 2.0. ### 6. Performance Testing: The CMAP Pro complies with IEC 60601-2-40 and all applicable performance tests in that standard. Additionally, validation tests were performed to validate the accuracy and repeatability of the range of motion (ROM) sensors, the FCE sensor, and the EMG leads. It was shown through the validation tests that the ROM sensors, FCE sensor, and EMG test leads captured data accurately and repeatably. In conclusion, the functionality of the new device is identical to the predicate device from a black box perspective. The underlying technology enhancements used to collect data using the range of motion (ROM) sensors, the FCE sensor, and the EMG leads are transparent to the user in terms of device functionality. Therefore the device is as safe, as effective, and performs the same as our predicate device and is therefore substantially equivalent. {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and science. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Medical Technologies Unlimited, Inc. % Underwriters Laboratories, Inc. Mr. Casey Conry 1285 Walt Whitman Road Melville, New York 11747 FEB -2 2012 Re: K113074 Trade/Device Name: CMAP Pro Regulation Number: 21 CFR 890.1375 Regulation Name: Diagnostic electromyograph Regulatory Class: II Product Code: IKN Dated: January 17, 2012 Received: January 18, 2012 Dear Mr. Conry: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ ## Page 2 - Mr. Casey Conry If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mark N. Melkerson Director Division of Surgical. Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Statement of Indications for Use Section 3: KII 3074 510(k) Number (if known): Device Name: _________________________________________________________________________________________________________________________________________________________________ Indications for Use: Surface electromyography with video recording, range of motion tracking, isometric functional testing and pinch and grip strength measuring. Prescription Use xi (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) . Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) (Division Sign of Surgical, Ofthopedic, and Restorative Devices 510(k) Number K113074 Page loof l 3.1 ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ 1 .. . . - -
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...