The Comby EEG Cap is intended for use in routine clinical settings where rapid placement of a large number of EEG electrodes is desired.
Device Story
Comby EEG Cap is an electrode positioning system for EEG recording; consists of high-quality profile rubber net, SEBS plastic electrode holders, and electrodes (sintered Ag/AgCl or gold-coated plates). Used in clinical settings by neurologists; cap secured via Velcro chin strap and chest belt. Electrodes placed per International 10-20 System; conductive gel/paste applied to reduce impedance between scalp and recessed electrodes. Wires exit cap to form cable connecting to EEG equipment via adapter or 25-pin D-sub connector. Device transmits electrophysiological signals; does not provide stimulation. Benefits include rapid electrode placement, consistent positioning, and patient comfort due to SEBS mounts. Shorter cable lengths minimize environmental artifacts.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against ANSI/AAMI EC12:2000 (R2015) for electrical safety and performance. Biocompatibility testing conducted per ISO 10993-1, ISO 10993-5 (cytotoxicity), and ISO 10993-10 (irritation and sensitization).
Technological Characteristics
Materials: Profile rubber net, SEBS (thermoplastic elastomer) electrode holders, Ag/AgCl or gold-coated electrodes. Form factor: Full head cap with 2-64 electrodes. Connectivity: Wired (25-pin D-sub or adapter). Standards: ANSI/AAMI EC12:2000 (R2015), ISO 10993-1/5/10. Passive device; no energy source.
Indications for Use
Indicated for use in routine clinical settings for rapid placement of EEG electrodes in adults and children to transmit electrophysiological signals to data collection devices.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
K240575 — WaveGuard Net EEG Cap (NA-245, NA-255, NA-261, NA-265, NA-271, NA-281) · Eemagine Medical Imaging Solutions GmbH · Nov 25, 2024
K223086 — Ceribell Instant EEG Headcap · Ceribell, Inc. · Mar 22, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 9, 2021
Pamel d.o.o % Stefan Bolleininger CEO be-On-Quality GmbH Bahnhofstrasse 85 Neunkirchen am Sand, D-91233 DE
Re: K211315
Trade/Device Name: Comby EEG Caps Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: October 13, 2021 Received: October 13, 2021
Dear Stefan Bolleininger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211315
Device Name Comby EEG Cap
Indications for Use (Describe)
The Comby EEG Cap is intended for use in routine clinical settings where rapid placement of large number of EEG electrodes is desired.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY OR 510(K) STATEMENT
### 5.1 Applicant
Date: 21-04-2021
Signature:
Name: Pamel d.o.o.
Mesekov put 16, 10000, Zagreb, Croatia Address:
| Contact Person in Germany: | Stefan Bolleininger |
|----------------------------|---------------------------------------|
| Telephone: | +49 9131 977986 – 50 |
| Fax: | +49 9131 977986 – 449 |
| E-Mail: | stefan.bolleininger@be-on-quality.com |
PAMEL
d.o.o.
proizvodnja, trgovina i usluge
Dubravica 41 f, Zagreb HR-10090
OIB: 14776088720
## 5.2 Trade Name : Comby EEG Cap
## 5.3 Common Name or Classification Name Disposable EEG Electrodes system
## 5.4 Establishment Registration Number
Not registered
# 5.5 Facility Address : Pamel d.o.o.
Mesekov put 16, 10000, Zagreb, Croatia
### 5.6 Device Classification
5.6.1 Classification: This is a class II device 5.6.2 Classification panel: Cutaneous electrodes
Product code: GXY
### 5.7 Regulation Number 21 CFR 882 1320
{4}------------------------------------------------
#### 5.8 Predicate Devices Descriptions
| Predicate device 1 | Predicate device 2 | Predicate device 3 |
|----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Electro-Cap System Electrode<br>Arrays (Predicate) 510(k)<br>#K112319 | Disposable EEG Electrodes<br>(MODEL: DL, E-CAP,<br>FLEX-CAP DEVICE<br>K200162 | Wave Guard Eemagine Medi-<br>cal (Predicate) 510(k)<br>#K110223 |
| Electro-Cap International Inc<br>011 W Lexington Rd, Eaton,<br>OH 45320, United States | Wuhan Greentek Pty Ltd.<br>Rm 03-2, Flr. 3, Dingye Bldg,<br>Phase, Wuhan CN 430074 | EEMAGINE MEDICAL IM-<br>AGING SOLUTIONS GMBH<br>8870 RAVELLO CT<br>Naples, FL 34114 |
#### 5.9 Proposed Device Description
The Comby EEG Cap is an electrode positioning system used to quickly place the standard 20 EEG electrodes or more in a uniform and consistent manner on the head and body in order to transmit electrophysiological signals from an individual to data collection devices. The Comby EEG Cap is made from high quality profile rubber net, electrode plastic holders made of SEBS and electrodes. Electrodes is made of two different conductors, sintered Ag/AgCl and gold coated plates. The EEG Comby Cap is fixed by Velcro tape holders which are fastened beneath the chin of the tested person. The cap holders is fixed to a belt, placed beneath the arms and around the chest of the tested person. The high quality profile rubber net holds the electrodes securely in position during the EEG recording. Wires are attached to each electrode and exit the cap to form a cable, which is used to connect the cap to the EEG equipment, either through an adapter cable, or directly to the equipment with the 25-pin D-sub connector. The electrical activity of the brain is transferred through the conductive Electro-Gel/paste to the recessed electrode and then to the EEG or computer equipment for evaluation. The electrodes on the standard caps are positioned in
{5}------------------------------------------------
accordance with the International 10-20 System of electrode placement. In addition, the Comby EEG Cap with as few as 2 to as many as 64 electrodes. Comby EEG are also produced based on customer specified electrode placements, with electrodes made of Ag/AgC1 material. Customized wiring and connectors offer additional options.
The Electro-Gel is used with external electrodes as the conductor between the scalp and the
(recessed) electrodes. It also reduces impedance (resistance to alternating current) between the
electrode surface and the skin "Electroconductive gel is a legally marketed product in US -
K111717" information "Company Name: Electra-Cap International, Inc. 1011 West Lexington
Road P.O. Box 87 Eaton, OH 45320 Contact: Amy Swallows Phone: 937-456-6099 Fax: 937-456-
7323 Trade Name: Electra-Gel Common Name: Electrode Gel.
#### 5.10 Indications for Use Statement
The Comby EEG Cap is intended for use in routine clinical settings where rapid placement of a number of EEG electrodes is desired.
#### 5.11 Required Components
The Comby EEG Cap needs to be used with conductive gel or paste (such as Electro-Gel), EEG device and computer equipment for evaluation.
#### 5.12 Bench Testing
The COMBY EEG CAP was tested for performance in order to verify compliance according to the following technical specifications and standards: ANSI/AAMI EC12:2000 (R2015) Disposable ECG Electrodes. Results indicate that the product complies with the applicable standard. The release test is also conducted on the product when it leaves the facility and the Performance release record is available in section 18. Compare to primary predicate devices specified and Comby EEG cap are same in, raw materials, physical features, and same manufacturing processes. The biocompatibility performance equiva- lence evidence of proposed electrode can be demonstrated according to ISO10993-1, ISO10993-5 and ISO10993-10. Clinical Data: Clinical testing is not required
Non clinical data:Non clinical tests performed include
Performance test :We performed electrical safety. The design of the Comby EEG cap is in conformance with performance in order to verify compliance according to the following technical specifications and standards: ANSI/ AAMI EC12:2000 (R2015) Disposable ECG Electrodes.
{6}------------------------------------------------
Styrene-ethylene-butylene-styrene, also known as SEBS, is an important thermoplastic elastomer (TPE) which behaves like rubber without undergoing vulcanization. Mounts made of SEBS provide an effective hold of the electrodes in the position without producing any pressure around the muscles. Muscle pressure artifacts are very common in EEG recording, and should be minimized to achieve the best recording result. SEBS electrode mounts are also more comfortable for the patient.
Comby EEG Caps have shorter cables when compared to the predicates. Shorter cables create less artifacts during the recoding form the environment. it is used on purpose to get better results as the recording can be made closer to patient without any environmental disturbance.
In COMBY EEEG CAP, maximum of 64 electrodes are required to provide coverage for the complete scalp.
The stated differences are made in the purpose of improving the EEG recording quality and patient comfort. Therefore COMBY EEG cap performs minimally as well or better that the predicates.
All the differences do not affect the safety and effectiveness of the subject device which is concluded after all the required testing, so there are no safety and effectiveness issues relating to the subject system.
Clinical testing is not required
We performed electrical safety. The design of the Comby EEG cap is in conformance with performance in order to verify compliance according to the following technical specifications and standards: ANSI/AAMI EC12:2000 (R2015) Disposable ECG Electrodes. Biocompatibility: 1. ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process 2. ISO 10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity 3. ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
Conclusion- Based on the comparison with predicate device, our Comby EEG cap has the same intended use, structure, and technologies, which is substantially equivalent to predicated device.
{7}------------------------------------------------
# 5.13 Summary table of comparison
# Predicate Comparison Chart -Table 1
| Feature | Subject device | Predicate device | Predicate device | Predicate device | Discussion |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | | K112319 | K200162 | K110223 | |
| Indica-<br>tion for<br>Use | The Comby<br>EEG is in-<br>tended for<br>use in rou-<br>tine clinical<br>settings<br>where rapid<br>placement<br>of a large<br>number of<br>EEG elec-<br>trodes is de-<br>sired | Electro-Cap<br>System Elec-<br>trode Arrays<br>The Electro-<br>Cap is intended<br>for use in rou-<br>tine clinical<br>set-tings where<br>rapid place-<br>ment of a large<br>number of EEG | Disposable<br>EEG Elec-<br>trodes<br>(MODEL: DL,<br>E-CAP, FLEX-<br>CAP) is in-<br>tended for use<br>in routine clini-<br>cal settings<br>where rapid<br>placement of a<br>number of EEG<br>electrodes<br>is<br>de-sired. | This is an EEG<br>electrode<br>set<br>intended for<br>use in routine<br>clinical settings<br>where rapid<br>placement of a<br>large number<br>of EEG<br>electrodes<br>is<br>de-sired | Same |
| Classifi-<br>cation | Class II | Class II | Class II | Class II | Same |
| Regula-<br>tion<br>number | 21CFR882.<br>1320 | 21CFR882.132<br>0 | 21CFR882.132<br>0 | 21CFR882.132<br>0 | Same |
| Regula-<br>tion | cutaneous<br>electrode | cutaneous elec-<br>trode | cutaneous elec-<br>trode | cutaneous elec-<br>trode | Same |
| Product<br>code | GXY, Elec-<br>trode, cuta-<br>neous | GXY, Elec-<br>trode, cutane-<br>ous | GXY, Elec-<br>trode, cutane-<br>ous | GXY, Elec-<br>trode, cutane-<br>ous | Same |
| Environ-<br>ment of<br>use | Electro-<br>physiologi-<br>cal | Electrophysio-<br>logical | Electrophysio-<br>logical | Electrophysio-<br>logical | Same |
| Intended<br>user | Neurolo-<br>gists | Neurologists | Neurologists | Neurologists | Same |
| Target<br>patient | Adults and<br>Children | Adults<br>and<br>Children | Adults<br>and<br>Children | Adults<br>and<br>Children | Same |
| Where<br>used | On the head | On the head | On the head | On the head | Same |
| Number<br>of con-<br>tacts | 2-64 | 2-256 | 2-128 | 1-256 | provide<br>better<br>recording and<br>patient comfort |
| Size<br>of<br>Caps | Various-<br>Babies<br>to<br>Large<br>27cm-62cm | Various-<br>Ba-<br>bies to Large<br>26cm-66cm | Various-<br>Ba-<br>bies to Large<br>26cm-66cm | Various-<br>Ba-<br>bies to Large<br>25cm-61cm | Same |
| Style<br>of<br>Caps | Full<br>head<br>cap | Full head cap | Full head cap | Full head cap | Same |
| Ear Slits | yes | yes | yes | yes | Same |
| Cap ma-<br>terial | High Qual-<br>ity Profile<br>Rubber | Spandex | Spandex | Elastic<br>Coolmax | equivalent |
| Electrode<br>Mounts | SEBS | Polyethylene | Silicone | Silicone | The electrode<br>mounts- base<br>is made of<br>comfort-able<br>SEBS plastic.<br>SEBS behaves<br>like rubber but<br>is artificial,<br>and the<br>electrode<br>holders have<br>rounded<br>corners so that<br>they are easy<br>to handle and<br>non-invasive<br>on the skin. |
| Cable<br>Length | 2-5 Feet | Various-<br>feet | 0.1m – 3.0m | Not Available | does not effect<br>safety |
| Type of<br>Cables | Standard<br>ribbon cable<br>and lead<br>wires | | Standard rib-<br>bon cable and<br>lead wires | Shielded Lead<br>Wires | Same |
| Elec-<br>trode<br>Metal | Ag/AgClAg<br>CL sintered<br>material<br>Gold plated<br>pure 99%<br>Silver | Pure Tin, Sil-<br>ver, Silver/Sil-<br>ver-Chloride,<br>Gold plated | 1. FLEX-CAP:<br>silver/silver<br>chloride ink<br>printed elec-<br>trodes on PET<br>(Polyethylene<br>terephthalate)<br>2 DL and E-<br>CAP silver /sil-<br>ver chloride<br>plated ABS<br>base | Silver/Sil-<br>ver/Chloride | Same as predi-<br>cate |
| Type<br>of<br>connect-<br>ors | D-Sub Con-<br>nectors 25-<br>in and 37-<br>pin, Touch<br>Proof Din<br>Sockets and<br>Special<br>Connectors<br>to<br>EEG<br>Equipment<br>& Comput-<br>ers | D-Sub<br>Con-<br>nectors, Touch<br>Proof<br>Din<br>Sockets<br>and<br>Special<br>Con-<br>nectors<br>to<br>Match EEG<br>Equipment<br>&<br>Computers | Touch-proof<br>safety socket<br>DIN42-802<br>(Φ=1.5mm) | D-Sub<br>Con-<br>nectors 25 and<br>35 pins. | Same<br>as<br>the<br>predicate<br>de-<br>vice |
| Biocom-<br>patibility | ISO 10993-<br>1,<br>ISO 10993-<br>5,<br>ISO 10993-<br>10 | None was Con-<br>ducted | ISO 10993-1,<br>ISO 10993-5,<br>ISO 10993-10 | ISO 10993-1,<br>ISO 10993-5,<br>ISO 10993-10 | The subject de-<br>vice has been<br>tested for bio-<br>compatibility<br>according to<br>the FDA guid-<br>ance, this dif-<br>ference does<br>not raise any<br>new safety or<br>effectiveness. |
| Perfor-<br>mance<br>require-<br>ments | Needs to<br>transmit<br>electrophys-<br>iological<br>signals from<br>an individ-<br>ual to data<br>collection<br>devices with<br>a maximum<br>impedance<br>of<br>5K/Ohms.<br>Does not<br>transmit<br>electrical<br>current, nor<br>are they in-<br>tended to be<br>used<br>for<br>stimulation. | Needs to trans-<br>mit electro-<br>physiological<br>signals from an<br>individual to<br>data collection<br>devices with a<br>maximum im-<br>pedance of<br>5K/Ohms.<br>Does not trans-<br>mit electrical<br>current, nor are<br>they intended<br>to be used for<br>stimulation. | 1. Re-<br>sistance <100<br>$ Ω $<br>2. AC im-<br>pedance <2<br>kΩ (at 10 Hz)<br>3. DC off-<br>set voltage<br><100 mV<br>4. Com-<br>bined offset in-<br>stability and in-<br>ternal noise:<br><150 μV<br>5. Bias<br>current toler-<br>ance <100 mV | Needs to trans-<br>mit electro-<br>physiological<br>signals from an<br>individual to<br>data collection<br>de-vices with a<br>maximum im-<br>pedance of<br>5K/Ohms.<br>Does not trans-<br>mit electrical<br>cur-rent, nor<br>are they in-<br>tended to be<br>used for stimu-<br>lation. | Performance of<br>the Comby<br>EEG cap is<br>tested as per<br>ANSI/AAMI<br>EC12:2000(R<br>2015) Disposa-<br>ble ECG Elec-<br>trodes |
{8}------------------------------------------------
{9}------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________
{10}------------------------------------------------
Pamel d.o.o.
{11}------------------------------------------------
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.