DYNATRON THERMOSTIM PROBE-DONE

K120835 · Dynatronics Corporation · GXY · Aug 20, 2012 · Neurology

Device Facts

Record IDK120835
Device NameDYNATRON THERMOSTIM PROBE-DONE
ApplicantDynatronics Corporation
Product CodeGXY · Neurology
Decision DateAug 20, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1320
Device ClassClass 2
AttributesTherapeutic

Indications for Use

A hand held cutaneous electrode to be used with Dynatronics Solaris devices to apply electrical stimulation and apply heat and cooling to the skin

Device Story

Dynatron ThermoStim probe is a passive, manual therapy accessory used with Dynatronics Solaris electrical stimulation devices. Device consists of a handle with a treatment head (flat or domed) and connections for lead wires and circulating water. It functions as an electrode for delivering electrical stimulation (Biphasic, Russian, HiVolt, IFC, Premodulated, Microcurrent) and as a thermal applicator (hot or cold) via circulating water from a Dynatron QUAD7. Operated by clinicians in a clinical setting to provide therapeutic benefits to targeted tissues. Thermal energy targets 110°F for heat and 37-50°F for cold. Healthcare providers use the probe to manually apply treatments; output affects clinical decision-making by enabling combined electrical and thermal therapy. Benefits include localized therapeutic stimulation and temperature modulation for patient treatment.

Clinical Evidence

Bench testing only. Waveform analysis confirmed no differences in electrical stimulation delivery compared to the predicate ultrasound probe. Dielectric strength and electromagnetic compatibility testing passed EN60601-1 and EN60601-1-2 standards. Thermal performance testing demonstrated consistent delivery of heat between 104°F and 112°F.

Technological Characteristics

Handheld cutaneous electrode with two treatment face options: 1.5" flat (11.4 cm²) and 1.1" dome (12.3 cm²). Passive accessory; energy source is external (Dynatronics Solaris for electrical, circulating water for thermal). Connectivity via lead wire and water circulation tubing. Tested for dielectric strength and EMC per EN60601-1 and EN60601-1-2.

Indications for Use

Indicated for patients requiring electrical stimulation and thermal therapy (heat or cold) applied to the skin via a handheld cutaneous electrode, used in conjunction with Dynatronics Solaris devices.

Regulatory Classification

Identification

A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Dynatronics. The logo features a stylized "D" shape to the left of the company name, "Dynatronics," which is written in a bold, italicized font. Below the logo, the address "7030 Park Centre Drive - Salt Lake City, Ut 84121" is printed, followed by the phone number "800-874-6251" and fax number "801-568-7711". # KI20835 PAGE I OF 3 AUG 2 0 2012 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS 16 August 2012 #### SUBMITTED BY: DYNATRONICS CORPORTATION 7030 Park Centre Drive Salt Lake City, UTAH 84121 Phone: (800) 874-6251; (801) 568-7000 Fax: (801) 568-7711 #### SUBMITTERS NAME: Douglas Sampson VP. Operations and R&D DYNATRONICS CORPORATION # 1. DEVICE NAME: Trade Name(s): Common Name: Classification: . Regulation Nos: Product Codes: Dynatron® ThermoStim™ Combination Electrical Stimulation and Thermal Therapy probe Class II 882.1320, 890.5720 GXY, ILO #### 2. PREDICATE DEVICES: Dynatron Solaris™ series Ultrasound probe in combination mode - K031329 (October 22, 2003) ArTek Spot® thermal probes - Exempt (hot/cold pack) from ThermoTek, Inc. #### 3. DESCRIPTION: The Dynatron ThermoStim probes are used to provide therapeutic electrical stimulation and thermal therapy treatments. The probes consist of a handle with a treatment head and connections for a lead wire and/or circulating water. The probes act as an electrode for therapeutic electrical current provided by a Dynatron Solaris device, or other similar Dynatronics electrical stimulation devices in the delivery of electrical stimulation therapy. The treatment head transfers thermal energy (hot or cold) from circulating water when connected to a Dynatron QUAD7™. The probes are a passive, manual therapy accessory. Treatment cycles are controlled through previously cleared Dynatronics electrical stimulation and OUAD7 devices. There are two versions of the probe face - a flat head with a surface area of 11.4 cm2 (1.5," diameter) and a domed head with a surface area of 12.3cm-(1.1" diameter). {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for Dynatronics. The logo features a stylized "D" shape to the left of the company name, which is written in a bold, sans-serif font. Below the logo, the address "7030 Park Centre Drive • Salt Lake City, Ut 84121" is printed. The image also includes a phone number, "800-874-6251", and a fax number, "801-568-7711". ### 4. INDICATIONS FOR USE: A hand held cutancous electrode to be used with Dynatronics Solaris devices to apply electrical stimulation and apply heat and cooling to the skin The Indications for Use stated herein are consistent with the cleared indications for the predicate device. ### 5. TECHNICAL ANALYSIS: The Dynatron ThermoStim probes generate therapeutic benefits as described in allowed claims through the delivery of electrical stimulation and thermal therapy to targeted tissues. The probes allow the practitioner to manually apply various electrical stimulation treatments (limited to modalities available from the FDA cleared Dynatronics Solaris series devices) as well as thermal therapy (hot or cold). The delivery of hot thermal energy has a target value of 110° Fahrenheit and the cold thermal energy has a selectable target value from 37 - 50° Fahrenheit. #### Performance Characteristics There are two treatment face options with the Dynatron ThermoStim probes. Both utilize the same handle and connections to therapy delivery devices. The treatment faces differ in that one is a 1.1" dome with 12.3 cm² surface area and the second is a 1.5" flat with 11.4 cm surface area. Both probes act as an accessory to provide the following treatment options: Biphasic, Russian, HiVolt, Interferential Current (IFC), Premodulated (IFC), and Microcurrent Thermal therapy - Hot = 110° Fahrenheit Cold = 37 - 50° Fahrenheit Non-clinical Bench Testing Wave forms were captured from a Dynatron Solaris device through the ThermoSTIM probe and the Ultrasound probe in combo mode. Analysis of the wave forms for delivery of electrical stimulation treatments show no differences between the two probes ThermoSTIM probes were tested for dielectric strength per EN60601-1 and for electromagnetic compatibility per EN60601-1-2. The probes pass the applicable test limits associated with the Solaris electrical stimulation device. Thermal data was collected using the ThermoSTIM probe in "hot" mode and delivering an electrical stimulation treatment. Treatments were delivered on two body parts per patient and using warm gel and room temperature gel. Results showed temperatures consistently between 104°F and 112°F. #### 6. SUBSTANTIAL EQUIVALENCE (SE) RATIONALE: The Dynatron ThermoStim probes share the same basic characteristics, features and intended uses as the predicate probes and, therefore, are substantially equivalent to the Dynatron Combination probe (applicable "K' number listed above) and the exempt ArTek Spot thermal probes. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for Dynatronics, with the company name in bold, stylized font. Below the logo is the address "7030 Park Centre Drive - Salt Lake City, Ut 84121". The logo features a stylized "D" shape to the left of the company name. 800-874-6251 · fax 801-568-77 K | 20835 PAGE 3 OF 3 # 7. SAFETY AND EFFECTIVENESS SUMMARY: There are no substantive differences between the product defined in this 510(k) submission and the predicate devices. This device is similar to the technology that is currently used in other similar medical devices. This device is subject to development and documentation as required in the Quality System Regulation, 21 CFT Part 820, under design/change control, and is subject to verification and validation to applicable standards / guidance documents. The sum of bench test results and design control outcomes allows us to conclude that the Dynatron ThermoSTIM probes are as safe and effective in delivering treatments when used as indicated in specific applications under a clinician's supervision / therapy program as the predicate probes. Signed: Dated: Douglas Sampson, VP, Operations and R&D DYNATRONICS CORPORATION {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, with stylized wings and body. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 AUG 20 2012 Dynatronics Corporation % Mr. Douglas Sampson Vice President 7030 Park Centre Drive Salt Lake City, Utah 84121 Re: K120835 Trade/Device Name: Dynatron ThermoStim probe Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous electrode Regulatory Class: Class II Product Code: GXY, ILO Dated: August 9, 2012 Received: August 10, 2012 Dear Mr. Sampson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ #### Page 2 - Mr. Douglas Sampson forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours Ron Dett Vsi Lira Dr. Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use Statement 510(k) Number (if known): K120835 Dynatron ThermoStim probe Device Name: #### Indications for Use: A hand held cutaneous electrode to be used with Dynatronics Solaris devices to apply electrical stimulation and apply heat and cooling to the skin Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-the-Counter Use (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Kinneth D. Ingram (Division Sign-Off) (Division Sign eigh eigh of Surgical, Orthopedic, ා Kestorative Devices 510(k) Number K120835 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...