DISPOSABLE ADHESIVE SURFACE ELECTRODE

K120494 · Spes Medica Srl · GXY · Jun 11, 2012 · Neurology

Device Facts

Record IDK120494
Device NameDISPOSABLE ADHESIVE SURFACE ELECTRODE
ApplicantSpes Medica Srl
Product CodeGXY · Neurology
Decision DateJun 11, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1320
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Disposable Adhesive Surface Electrodes are intended for use Spes Medica with electrodiagnostic or neurological monitoring equipment for the recording of electrophysiological activity and for electrical stimulation. The electrodes are non-sterile and for single patient use only.

Device Story

Disposable adhesive surface electrodes for electrodiagnostic/neurological monitoring and electrical stimulation. Device consists of three layers: cotton non-woven pad, conductive sensor layer, and solid hydro-gel adhesive layer for skin contact. Snap connector versions utilize Ag/AgCl pellets. Connectors include DIN 42802 touch-proof, 0.7mm pin, concentric bipolar socket, or 2mm socket. Used by clinicians in clinical settings to interface patient skin with monitoring/stimulation equipment. Facilitates signal acquisition for diagnostic purposes or delivery of electrical stimulation; benefits patient through non-invasive monitoring and therapeutic application.

Clinical Evidence

Bench testing only.

Technological Characteristics

Non-invasive cutaneous electrodes. Materials: cotton non-woven pad, conductive sensor layer (Ag/AgCl pellets in snap versions), solid hydro-gel adhesive. Connectivity: various touch-proof connectors (DIN 42802, 0.7mm pin, concentric bipolar, 2mm socket). Single patient use, non-sterile.

Indications for Use

Indicated for use with electrodiagnostic or neurological monitoring equipment for recording electrophysiological activity and electrical stimulation in patients requiring such monitoring or therapy.

Regulatory Classification

Identification

A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Spes Medica s.r.l. ्री 510(k)NOTIFICATION Disposable Adhesive Surface Electrodes Rev 0 K120494 # 510 (k) Summary of safety and effectiveness JUN 1 1 2012 # SUBMITTER INFORMATION - A. Company Name: - B. Company Address: - C. Company Phone: Company Fax: - D. Contact Person: Spes Medica s.r.l. Via Europa - Zona industriale Battipaglia (SA) - Italy 84091 + 39 0828 614191 + 39 0828 341788 Alfredo Spadavecchia Quality Assurance and Regulatory Affairs Spes Medica s.r.l. e-mail: a.spadavecchia@spesmedica.com E. Date Summary Prepared: February 14, 2012 #### DEVICE IDENTIFICATION | A. Device name: | Disposable Adhesive Surface Electrodes | |----------------------------|----------------------------------------| | B. Trade/Proprietary name: | Disposable Adhesive Surface Electrodes | | C. Classification name: | Cutaneous Electrode (21 CFR §882.1320) | | D. Product code: | GXY | ### LEGALLY MARKETED DEVICES (PREDICATE DEVICES) | Predicate device | 510(k) Holder/Applicant | 510(k) number | |-------------------------------------------|-------------------------------|---------------| | RLI Cutaneous Disposable Electrode | Rhythmlink International, LLC | K052188 | | Bio-logic Disposable Electrode | Bio-logic Systems Corp. | K941799 | | Sunspots Pre-gelled Surface<br>Electrodes | Axon System, Inc. | K062198 | {1}------------------------------------------------ 1 ### DEVICE DESCRIPTION Spes Medica Disposable Adhesive Surface Electrodes are single patient use, disposable devices. Electrodes are non-invasive as they are applied to the patient's skin using a selfadhesive solid-gel surface. The electrodes consist of 3 different layers of materials. The upper layer consists in a cotton non-woven pad, the second layer is the sensor and consists in a conductive layer. The lower layer, then the material in direct contact with the patient's skin consist in a solid Hydro-gel adhesive layer. Spes Medica Disposable Adhesive Surface Electrodes are supplied with leads or with snap connector. The electrodes which are attached to a lead wire terminate at the opposite end using different type of connectors such us: - DIN 42802 "touch proof" safety connectors; . - 0,7 mm "touch proof" pin; . - Concentric "touch proof" bipolar socket; . - 2 mm "touch proof" socket; . The snap connector electrode uses Ag/AgCl pellet(s) as the conductive element. The active surface of the pellet is comprised between the cotton non-woven pad and the Hydro-gel adhesive layer. The pellet is attached to a stainless steel male snap connector which is used as connection. #### INTENDED USE Disposable Adhesive Surface Electrodes are intended for use Spes Medica with electrodiagnostic or neurological monitoring equipment for the recording of electrophysiological activity and for electrical stimulation. The electrodes are non-sterile and for single patient use only. # TECNOLOGICAL COMPARISON Technologically. Spes Medica Disposable Adhesive Surface Electrodes are similar to the predicate devices. The electrodes are similar in intended use, design, materials, packaging and other technological characteristics to the predicate devices. No new technology or basic materials are used in these designs. #### CONCLUSION Spes Medica Disposable Adhesive Surface Electrodes are similar in intended use, design, materials, packaging and other technological characteristics to the predicate devices. After analyzing performance and safety testing, it is the conclusion of Spes Medica s.r.I. that the Disposable Adhesive Surface Electrodes are safe and effective as the predicate devices and introduce no new questions concerning safety and effectiveness. {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 JUN 1 1 2012 Spes Medica S.r.l. c/o Mr. Alfredo Spadavecchia Ouality Assurance and regulatory Affairs Via Europa, Zona Industriale Battipaglia, Italy 84091 Re: K120494 Trade/Device Name: Disposable Adhesive Surface Electrode Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: May 24, 2012 Received: May 29, 2012 Dear Mr. Spadavecchia: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ # Page 2 - Mr. Alfredo Spadavecchia Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Affoang for Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # INDICATIONS FOR USE K126494 510(k) Number (if known): Device Name: Disposable Adhesive Surface Electrodes Indications for Use: Spes Medica Disposable Adhesive Surface Electrodes are intended for use with electrodiagnostic or neurological monitoring equipment for the recording of electrophysiological activity and for electrical stimulation. The electrodes are non-sterile and for single patient use only. Prescription Use (21 CFR 801 Subpart C) AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) N.L.B. Nicholas Division Sign-Off) unvision of Ophthalmic, Neurological and Ear, Hose ...d Throat Devices 510(k) Number K120494
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...