ImPACT

K170209 · Impact Applications, Inc. · POM · Feb 23, 2017 · Neurology

Device Facts

Record IDK170209
Device NameImPACT
ApplicantImpact Applications, Inc.
Product CodePOM · Neurology
Decision DateFeb 23, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.1471
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

ImPACT is intended for use as a computer-based neurocognitive test battery to aid in the assessment and management of concussion. ImPACT is a neurocognitive test battery that provides healthcare professionals with objective measure of neurocognitive functioning as an assessment aid and in the management of concussion in individuals ages 12-59.

Device Story

ImPACT is a computer-based neurocognitive test battery; evaluates reaction time, memory, attention, spatial processing speed, and symptoms. Input: test-taker responses to stimuli presented on screen via standard input devices. Operation: software-based tool administered by healthcare professionals to gather objective performance data. Output: neurocognitive function metrics used by clinicians as part of a multidisciplinary approach for return-to-activity decisions. Context: used in clinical settings; facilitates objective assessment of concussion-related cognitive impairment. Benefit: provides standardized, medically accepted neurocognitive data to support clinical management.

Clinical Evidence

Bench testing and a volunteer study (n=not specified, age 19-22) compared the HTML5 version to the Flash predicate. Primary endpoint was reaction time performance. Results showed differences were statistically and clinically insignificant. Software verification and validation performed per IEC 62304; risk management per ISO 14971.

Technological Characteristics

Computer-based neurocognitive test battery; software application running on general-purpose computing platforms. Connectivity: networked/web-based. Software architecture: HTML5/CSS/JavaScript (migrated from Adobe ActionScript).

Indications for Use

Indicated for individuals ages 12-59 to provide healthcare professionals with objective measures of neurocognitive functioning to aid in the assessment and management of concussion.

Regulatory Classification

Identification

The computerized cognitive assessment aid for concussion is a prescription device that uses an individual's score(s) on a battery of cognitive tasks to provide an indication of the current level of cognitive function in response to concussion. The computerized cognitive assessment aid for concussion is used only as an assessment aid in the management of concussion to determine cognitive function for patients after a potential concussive event where other diagnostic tools are available and does not identify the presence or absence of concussion. It is not intended as a stand-alone diagnostic device.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Software, including any proprietary algorithm(s) used by the device to arrive at its interpretation of the patient's cognitive function, must be described in detail in the software requirements specification (SRS) and software design specification (SDS). Software verification, validation, and hazard analysis must be performed. (2) Clinical performance data must be provided that demonstrates how the device functions as an interpretation of the current level of cognitive function in an individual that has recently received an injury that causes concern about a possible concussion. The testing must: (i) Evaluate device output and clinical interpretation. (ii) Evaluate device test-retest reliability of the device output. (iii) Evaluate construct validity of the device cognitive assessments. (iv) Describe the construction of the normative database, which includes the following: (A) How the clinical workup was completed to establish a “normal” population, including the establishment of inclusion and exclusion criteria. (B) Statistical methods and model assumptions used. (3) The labeling must include: (i) A summary of any clinical testing conducted to demonstrate how the device functions as an interpretation of the current level of cognitive function in a patient that has recently received an injury that causes concern about a possible concussion. The summary of testing must include the following: (A) Device output and clinical interpretation. (B) Device test-retest reliability of the device output. (C) Construct validity of the device cognitive assessments. (D) A description of the normative database, which includes the following: ( *1* ) How the clinical workup was completed to establish a “normal” population, including the establishment of inclusion and exclusion criteria.( *2* ) How normal values will be reported to the user.( *3* ) Representative screen shots and reports that will be generated to provide the user results and normative data.( *4* ) Statistical methods and model assumptions used.( *5* ) Whether or not the normative database was adjusted due to differences in age and gender.(ii) A warning that the device should only be used by health care professionals who are trained in concussion management. (iii) A warning that the device does not identify the presence or absence of concussion or other clinical diagnoses. (iv) A warning that the device is not a stand-alone diagnostic. (v) Any instructions technicians must convey to patients regarding the administration of the test and collection of cognitive test data.

In combination with the general controls of the FD&C Act, the Computerized Cognitive Assessment Aid for Concussion is subject to the following special controls:

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Public Health Service Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of the human form. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 23, 2017 ImPACT Applications, Inc Michael Zagorski Director of Regulatory Affairs 9665 Granite Ridge Drive Suite 550 San Diego, California 92123 Re: K170209 Trade/Device Name: ImPACT Regulation Number: 21 CFR 882.1471 Regulation Name: Computerized Cognitive Assessment Aid For Concussion Regulatory Class: Class II Product Code: POM Dated: January 23, 2017 Received: January 24, 2017 Dear Mr. Zagorski: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. # Michael J. Hoffmann -S Carlos L. Peña. PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K170209 Device Name ImPACT #### Indications for Use (Describe) ImPACT is intended for use as a computer-based neurocognitive test battery to aid in the assessment and management of concussion. ImPACT is a neurocognitive test battery that provides healthcare professionals with objective measure of neurocognitive functioning as an assessment aid and in the management of concussion in individuals ages 12-59. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------| | <span style="font-size: 10pt;"> <span style="font-family: Arial;"> <span style="font-weight: bold;"> <span style="color: black;"> </span> </span> </span> </span> | | | <span style="font-size: 10pt;"> <span style="font-family: Arial;"> <span style="font-weight: bold;"> <span style="color: black;"> <input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </span> </span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a logo for ImPACT. The logo features a blue silhouette of a head with a brain inside. The word "ImPACT" is written in blue letters next to the head. The logo is surrounded by a blue border. # Section 5 - 510(k) Summary # Submission Date: January 23, 2016 # Submitter Information: | Company: | ImPACT Applications, Inc.<br>9665 Granite Ridge Drive, Suite 550<br>San Diego, CA 92123 | |----------|-----------------------------------------------------------------------------------------| |----------|-----------------------------------------------------------------------------------------| | Contract Person: | Michael Zagorski | |------------------|---------------------------------| | | Director of Regulatory Affairs | | | ImPACT Applications, Inc. | | | Tel: 412-567-8400 ext. 939 | | | Email: mzagorski@impacttest.com | #### Device Information: | Trade Name: | ImPACT®, ImPACT® Workplace | |------------------------|------------------------------------------------------| | Common Name: | Computerized Cognitive Test | | Classification Name: | Computerized cognitive assessment aid for concussion | | Device Classification: | Class II | | Product Code: | POM, 21 CFR 882.1471 | | Panel: | Neurology | | Predicate Device: | ImPACT, DEN150037 | | Reason for submission | Device Modifications | #### Indications for Use: ImPACT is intended for use as a computer-based neurocognitive test battery to aid in the assessment and management of concussion. lmPACT is a neurocognitive test battery that provides healthcare professionals with objective measure of neurocognitive functioning as an assessment aid and in the management of concussion in individuals ages 12-59. #### Device Description: ImPACT® (Immediate Post-Concussion Assessment and Cognitive Testing) is a computer-based neurocognitive test battery. {4}------------------------------------------------ ImPACT is a software-based tool that allows healthcare professionals to conduct a series of neurocognitive tests on individuals to gather basic data related to the neurocognitive functioning of the test subject. This computerized cognitive test battery evaluates and provides a healthcare professional with measures of various neurocognitive functions, including the reaction time, memory, attention, spatial processing speed and symptoms of an individual. ImPACT provides healthcare professionals with a set of well-developed and researched neurocognitive tasks that have been medically accepted as state-of-the-art best practices and is intended to be used as part of a multidisciplinary approach to making return to activity decisions. # Comparison to Predicate Device: ImPACT (HTML5 Version) is substantially equivalent to ImPACT (Flash version), manufactured by ImPACT Applications, Inc., and cleared under DEN150037. Both devices have intended use and indications for use as computerized neurocognitive tests to aid in the assessment of concussion. They are also similar in terms of technological characteristics as both are software applications using general purpose computing platform to electronically record objective performance measurements (speed and accuracy) as the test taker responds to stimuli presented on the screen via input devices. Further, all functionality of both devices is identical. The differences between the new device and the predicate are limited to a modification to the software code language used to develop the device. # Device Modifications The software of the predicate device, Adobe Flash Rich Internet Application (RIA) written in Adobe ActionScript programming language, was re-written using HTML5/CSS/JavaScript. The rewrite effort preserved all functions of the original version. The changes are preventive, to improve software maintainability, and adaptability to cope with changes in the software environment. There are no changes to the intended use, indication for use, including users, patient population, environment of use, or the conditions assessed. Further, there are no changes to the design of neurocognitive test battery; all tasks, stimuli, and captured information remain identical to the original version. # Summary of Performance Testing: Software verification and validation activities including code reviews, evaluations, analyses, traceability assessment, and manual testing were performed in accordance with IEC 62304 and other software standards to demonstrate device performance and functionality. All tests met the required acceptance criteria. Risk Management activities conducted in accordance with ISO 14971 assure all risk related to use of a computerized neurocognitive test, including use related risks are appropriately mitigated. Additionally, to validate the effect of the software change on the performance, two laboratory studies (one bench, one in volunteers age 19-22) were conducted to compare the results of the predicate device (Flash code) and the new device for reaction time. The studies showed that the differences in performance are statistically and clinically insignificant. #### Substantial Equivalence: The differences in design described above do not affectiveness of ImPACT for its intended use, functionality or performance, which was demonstrated through performance testing. Therefore, ImPACT (HTML5 version) is substantially equivalent to predicate ImPACT (Flash version).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...