DATEX-OHMEDA S/5 BIS MODULE, E-BIS AND ACCESSORIES
K052145 · GE Healthcare · OLW · Aug 23, 2005 · Neurology
Device Facts
Record ID
K052145
Device Name
DATEX-OHMEDA S/5 BIS MODULE, E-BIS AND ACCESSORIES
Applicant
GE Healthcare
Product Code
OLW · Neurology
Decision Date
Aug 23, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.1400
Device Class
Class 2
Intended Use
The Datex-Ohmeda S/5™ BIS Module, E-BIS and accessories are intended to be used with Datex-Ohmeda modular multiparameter monitors for monitoring neurophysiological status of hospitalized patients.
Device Story
The Datex-Ohmeda S/5 BIS Module (E-BIS) is a plug-in parameter module for Datex-Ohmeda modular monitors (e.g., S/5 Anesthesia/Critical Care Monitors). It acquires raw EEG signals from patients via patient interface cables and BIS sensors. The module processes these signals using a BIS Engine to calculate parameters: Bispectral Index (BIS, 0-100 scale for hypnosis level), Suppression Ratio (SR), Electromyograph (EMG) power, and Signal Quality Index (SQI). These parameters are displayed on the host monitor, where clinicians can view waveforms, trends, and set auditory/visual alarms. The device is used in clinical settings (OR, ICU) by qualified medical personnel to monitor anesthetic effects and neurophysiological status. It provides real-time data to assist in clinical decision-making regarding anesthesia administration.
Clinical Evidence
Bench testing only. The device underwent verification and validation of specifications against recognized standards including IEC 60601-1 (safety), IEC 60601-1-1 (medical electrical systems), IEC 60601-1-4 (programmable systems), IEC 60601-2-26 (electroencephalographs), and AAMI TIR No. 24 (physiologic waveform databases). No clinical trial data was required or provided.
Technological Characteristics
Single-width plug-in module for modular monitors. Uses EEG signal acquisition via BIS sensors. Parameters: BIS (0-100), Suppression Ratio (0-100%), EMG (70-110 Hz), SQI (0-100%). Connectivity: Proprietary module-to-monitor interface. Software: Embedded firmware (version 1.1). Complies with IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-2-26.
Indications for Use
Indicated for monitoring the state of the brain by data acquisition of electroencephalograph (EEG) signals of all hospitalized patients. The Bispectral index (BIS) may be used as an aid in monitoring the effects of certain anesthetic agents. Indicated for use by qualified medical personnel only.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
K013389 — DATEX-OHMEDA S/5 BIS MODULE, M-BIS AND ACCESSORIES · Datex-Ohmeda · Jan 10, 2002
K973596 — SPACELABS MEDICAL 90482 BISPECTRAL INDEX ANALYSIS (BIS) MODULE · Spacelabs Medical, Inc. · Feb 3, 1998
K061907 — MODIFICATION TO: DATEX-OHMEDA S/5 ENTROPY MODULE, E-ENTROPY AND ACCESSORIES · GE Healthcare · Mar 27, 2008
K012466 — BIS/EEG MODULE · General Electric Medical Systems Information Techn · Oct 1, 2001
K023459 — DATEX-OHMEDA S/5 ENTROPY MODULE, M-ENTROPY AND ACCESSORIES · Datex-Ohmeda · Jan 13, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
Page 1 of 4
AUG 2 3 2005
K 052145
# Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda S/5TM BIS Module, E-BIS and accessories
# GENERAL COMPANY INFORMATION as required by 807.92(a){1}
#### COMPANY NAME/ADDRESS/PHONE/FAX:
GE Healthcare 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344
## NAME OF CONTACT:
Mr. Joel Kent
#### DATE:
July 31, 2005
#### DEVICE NAME as required by 807.92(a)(2)
## TRADE NAME:
Datex-Ohmeda S/5TM BIS Module, E-BIS and accessories
## COMMON NAME:
EEG Measurement Module with BIS Index
#### CLASSIFICATION NAME:
The following Class II classification appears applicable:
Product Code Classification Name GWQ Electroencephalograph
CFR Section 882.1400
# NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)
The Datex-Ohmeda S/5™ BIS Module, E-BIS is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda M-BIS Module (K113389).
{1}------------------------------------------------
Page 2 of 4 -- August 5, 2005
# Declaration of Conformity to Design Controls/Fundamental Scientific Technology:
We have enclosed a copy of the Declaration of Conformity with Design Controls for the device.
There has been no change to the fundamental scientific technology from the predicate.
# Classification:
The following Class II classification appears applicable:
| Product Code | Classification Name | CFR Section |
|--------------|-----------------------|-------------|
| GWQ | Electroencephalograph | 882.1400 |
# Substantial Equivalence:
The Datex-Ohmeda S/5™ BIS Module, E-BIS is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda M-BIS Module (K013389).
# INTENDED USE /Indications for Use
The intended use and indications for use for the modified device are the same as the predicate. The Indication For Use form on a separate page is enclosed.
# Intended Use:
The Datex-Ohmeda S/5™ BIS Module, E-BIS and accessories are intended to be used with Datex-Ohmeda modular multiparameter monitoring neurophysiological status of hospitalized patients.
# Indications for use:
The Datex-Ohmeda S/5™ BIS Module, E-BIS and accessories are indicated for monitoring the state of the brain by data acquisition of electroencephalograph (EEG) signals of all hospitalized patients. The Bispectral index (BIS), a processed EEG variable, may be used as an aid in monitoring the effects of certain anesthetic agents. The device is indicated for use by qualified medical personnel only.
{2}------------------------------------------------
#### DEVICE DESCRIPTION as required by 807.92(a)(4)
The Datex-Ohmeda S/5™ BIS Module, E-BIS and accessories is a single-width plug-in parameter module for a Datex-Ohmeda modular monitoring system. The S/5 E-BIS module, M-BIS can be used with the following Datex-Ohmeda modular monitors: S/5 Anesthesia Monitor (AM), S/5 Compact Anesthesia Monitor (CAM), S/5 Critical Care Monitor (CCM) and S/5 Compact Critical Care Monitor (CCCM) with main software L-(C)ANE02(A)..00 or L-(C)ICU02(A)..00 or newer version. The Datex-Ohmeda S/5™ BIS Module, E-BIS and accessories is a parameter module for monitoring the state of the brain by data acquisition of EEG signals of all hospitalized patients. The BIS, a processed EEG variable, may be used as an aid in monitoring the effects of certain anesthetic agents. The Datex-Ohmeda S/5TM BIS Module, E-BIS and its predicate M-BIS (K013389) are used in conjunction with Aspect Medical Systems, Digital Signal Converter-Expanded Performance, DSC-XP (K011534), Patient Interface Cable; PIC (K011534), and BIS Sensor Plus (K994330), BIS Sensor XP (K002734) or BIS Sensor Pediatric (K001980). The raw EEG signal can be displayed from one of the two monitored channels. The waveform size, color and sweep speed can be adjusted.
Calculated parameters are:
- Bispectral Index,BIS(Range=0-100),continuous processed EEG parameter correlating to the patient's level of hypnosis, where 100=awake and 0=comatose.
- Suppression Ratio,SR,(Range=0-100%), the percentage of epochs in the past 63 seconds in which the EEG signal is considered suppressed.
- . Electromyograph, EMG, the absolute power (in decibels) in the frequency range 70-110 Hz
- Signal Quality Index, SQI (Range:0-100%), the percentage of good epochs in the last 60sec. that could be used to calculate the Bispectral Index and spectral variables.
All the calculated parameters can be selected on the display, and trended.
Alarms for E-BIS are taken care of by the host monitor and follow the user interface for alarms in Datex-Ohmeda \$/5 patient monitors. There are auditory and visual alarms and user adjustable limits for E-BIS. The default is OFF, because it doesn't provide information to be used for treatment or therapy. The BIS Engine at BIS module dictates error messages displayed at Datex-Ohmeda's host monitor's message fields and service page. These error messages are related to the BIS measurement, but follow Datex-Ohmeda user interface rules,
{3}------------------------------------------------
#### INTENDED USE as required by 807.92(a)(5)
#### Intended Use:
The Datex-Ohmeda S/5™ BIS Module, E-BIS and accessories are intended to be used with Datex-Ohmeda modular multiparameter monitors for monitoring neurophysiological status of hospitalized patients.
#### Indications for use:
The Datex-Ohmeda S/5TM BIS Module, E-BIS and accessories are indicated for monitoring the state of the brain by data acquisition of electroencephalograph (EEG) signals of all hospitalized patients. The Bispectral index (BIS), a processed EEG variable, may be used as an aid in monitoring the effects of certain anesthetic agents. The device is indicated for use by qualified medical personnel only.
# SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)
The Datex-Ohmeda S/5™ BIS Module, E-BIS is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda M-BIS Module (K013389).
The E-BIS module has the following similarities compared to the predicate M-BIS (K013389):
- identical intended use and indications for use .
- . identical fundamental scientific technology
- . identical electronic measurement board
- same module software (version changed from 1.0 to 1.1) .
- . same BIS Engine
- . use the same operating principle
- identical accessories .
- have the same user interface at the monitor and alarms (can be used with the same . monitor software)
- . the Customer and parameter specifications are the same
- have the same safety and effectiveness .
- . are manufactured using the same processes
The main differences between the new E-BIS and the predicate M-BIS (K013389) is primarily due to fact that the new E-BIS module has the following changes:
- . new color, shape, and size and thus differing mechanics
- The front panel and labeling have changed (added ESD symbol adjacent to BIS . sensor connector)
- . Minor module software change and monitor software correction
Based on the analysis and other documentation included in this 510(k) notification and attachments it is evident that the main features and indications for use of the Datex-Ohmeda S/5™ BIS Module, E-BIS are substantially equivalent to the predicate Datex-Ohmeda M-BIS Module (K013389).
{4}------------------------------------------------
# SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)
The Datex-Ohmeda S/5™ BIS Module, E-BIS has been assessed against the standards below. The device has been thoroughly tested through validation and verification of specifications.
- FDA/DCRND Reviewer Guidance for Premarket Notification Submissions, November 1993 .
- IEC 60601-1:1988 + Amdt. 1:1991 + Amdt. 2:1995 (Part 1: General requirements for safety) .
- EN 60601-1:1990+ A1:1993 + A13:1996 + A2:1995 (identical to IEC60601-1:1988 + Amdt. . 1:1991 + Amdt. 2:1995)
- CAN/CSA C22.2 No. 601.1-M90 + S1:1994 (Canadian deviations to IEC 60601-1:1988 + Amdt. . 1:1991) + S2:1998 (=IEC Amdt 2:1995)
- UL 2601-1, October 24, 1997 (U.S. deviations to IEC 60601-1:1988 + Amdt. 1:1991+ Amdt. . 2:1995)
- IEC 60601-1-1:1992 + Amdt.1:1995 (Safety requirements for medical electrical systems) .
- . EN 60601-1-1:1993 + A1:1996 (identical to IEC60601-1-1:1992 + Amdt.1:1995)
- AAMI ES1-1993 (Safe current limits for electromedical apparatus) .
- IEC 60601-1-4:2000 (Programmable electronic medical systems) .
- Electroencephalograph Devices Guidance for 510( k) Draft Document Version 1.0 November 3, . 1997
- . IEC 60601-2-26: 2002 Medical electrical equipment. Part 2: Particular requirements for the safety of electroencephalographs
- . ISO 14971 Ed. 1: Medical devices - Application of risk management to medical devices
- AAMI TIR No.24:1999 Acquisition and use of physiologic waveform databases for testing of . medical devices
- . IEC 60601-1-2.2001 (Electromagnetic compatibility - Requirements and tests) FDA/ODE Guidance for Content of Premarket Submission for Software Contained in Medical Devices. (May 11,2005)
- . FDA Performance standard, 21 CFR Part 898.12
## CONCLUSION:
The summary above shows that there are no new questions of safety and effectiveness for the Datex-Ohmeda S/5™ BIS Module, E-BIS compared to the legally marketed (predicate) Datex-Ohmeda M-BIS Module (K013389).
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 3 2005
Mr. Joel C. Kent Manager, Quality and Regulatory Affairs GE Healthcare 86 Pilgrim Road Needham, Massachusetts 02492
Re: K052145
Trade/Device Name: Datex-Ohmeda S/5" BIS Module, E-BIS and accessories Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: GWQ Dated: August 5, 2005 Received: August 8, 2005
Dear Mr. Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Barbara Briehud
for
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): _ 1 052145
Device Name: Datex-Ohmeda S/5™ BIS Module, E-BIS and accessories.
Indications for Use:
The Datex-Ohmeda S/5™ BIS Module, E-BIS and accessories are indicated for monitoring the state of the brain by data acquisition of electroencephalograph (EEG) signals of all hospitalized patients.
The Bispectral index (BIS), a processed EEG variable, may be used as an aid in monitoring the effects of certain anesthetic agents.
The device is indicated for use by qualified medical personnel only.
Prescription Use __ X Over-The-Counter Use _ AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Jadrian Lovelund for Melferson
Division of General. Restorative, ano Neurological Devices
**510(k) Number** K052145
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.