K212684 · Medeia, Inc. · OLU · Jan 28, 2023 · Neurology
Device Facts
Record ID
K212684
Device Name
BrainView QEEG Software
Applicant
Medeia, Inc.
Product Code
OLU · Neurology
Decision Date
Jan 28, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Intended Use
The BrainView QEEG Software Package is to be used by qualified medical or clinical professionals for the statistical evaluation of the human electroencephalogram (EEG).
Device Story
BrainView QEEG Software Package performs post-hoc statistical analysis of resting digital EEG data. Input consists of artifact-free, eyes-closed or eyes-open EEG recordings transferred from a host system. The software applies Fast-Fourier Transformation (FFT) to extract spectral power for delta, theta, alpha, and beta frequency bands. It derives over 4,000 measures, including absolute/relative power, asymmetry, and coherence for 19 monopolar and 171 bipolar derivations. These are compared against an age-regressed, Gaussian-validated normative database to generate Z-scores. Results are presented as statistical tables and topographical brain maps. Used in clinical settings by qualified medical personnel, the output serves as a supplementary tool for patient evaluation. It is contraindicated for use as a standalone diagnostic system; clinicians must review traditional EEG and selected epochs to avoid diagnostic errors. The device benefits patients by providing quantitative, age-matched normative comparisons to assist clinical decision-making.
Clinical Evidence
Bench-only validation using clinically acquired EEG waveforms from a small sample of subjects across the 4-85 age range. The study compared Z-scores for absolute power generated by the subject device against the predicate (K041263). Acceptance criteria required an R-squared factor of 0.8 or better. Results confirmed that the subject device produces results in agreement with the predicate across 23 age-grouped sets of Z-scores.
Technological Characteristics
Software-based QEEG analysis tool. Implements Fast-Fourier Transformation (FFT) for spectral analysis (1-40 Hz). Features include Z-score calculation and topographical mapping for 19 monopolar and 171 bipolar derivations. Normative database includes 2303 subjects (eyes-closed) and 1965 subjects (eyes-open) aged 4-85. Developed per IEC 62304:2015. Software level of concern: Moderate.
Indications for Use
Indicated for qualified medical or clinical professionals to perform post-hoc statistical evaluation of human EEG in patients aged 4-85 years.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
January 28, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Medeia, Inc. % Daniel Lehtonen Regulatory Consultant Compliance and Regulatory Services, LLC 3771 Southbrook Dr. Dayton, Ohio 45430
Re: K212684
Trade/Device Name: BrainView QEEG Software Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLU Dated: February 4, 2022 Received: February 8, 2022
Dear Daniel Lehtonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jay R. Gupta -S
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K212684
Device Name BrainView OEEG Software Package
Indications for Use (Describe)
The BrainView QEEG Software Package is to be used by qualified medical or clinical professionals for the statistical evaluation of the human electroencephalogram (EEG).
Type of Use (Select one or both, as applicable)
| <div> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------|
| <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Medeia. The logo has the word "Medeia" in bold, black letters. The first letter, "M", is colored with a gradient from green to blue. Below the word "Medeia" is the phrase "Designed for Clinicians" in blue letters.
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the Requirements of Safe Medical Device Systems Act 1990 and 21 CFR Sec. 807.92
510(k) Number:
K212684
#### a1 APPLICANT INFORMATION:
| Date Prepared: | 10 Dec 2021 |
|----------------|---------------------------------------------------------------|
| Name: | Medeia, Inc. |
| Address: | 7 W. Figueroa Street<br>Suite 300<br>Santa Barbara, CA, 93101 |
| Contact Person: | Slav Danev |
|-----------------|------------------|
| Phone Number: | +1 800 433 4609 |
| Fax Number: | +1 800 433 4609 |
| Email: | danev@medeia.com |
#### a2 NAME OF DEVICE:
| Trade Name: | BrainView QEEG Software Package |
|-----------------------|---------------------------------------------------------|
| Common Name: | Normalizing Quantitative Electroencephalograph Software |
| Classification Name: | Electroencephalograph; 21 CFR 882.1400 (OLU) |
| Classification Panel: | Neurology |
#### a3 PREDICATE DEVICES:
| Predicate Device: | K041263; NeuroGuide Analysis System (NAS) |
|-------------------|-------------------------------------------|
| Reference Device: | K171414; qEEG-Pro |
The FDA database for recalls was searched on 03 March 2021 during the preparation of the 510(k) submission and no recalls for the devices noted above were found.
#### STATEMENT OF INTENDED USE: a4
The BrainView QEEG Software Package is to be used by qualified medical or clinical professionals for the statistical evaluation of the human electroencephalogram (EEG).
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Medeia. The word "Medeia" is written in a combination of colors, with the "M" transitioning from green to blue, and the rest of the letters in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller, light blue font.
#### a5 DESCRIPTION OF THE DEVICE:
BrainView QEEG Software Package is a software program for the post-hoc statistical analysis of the human electroencephalogram (EEG). EEG recorded on a separate device (i.e., the host system) is transferred to the BrainView QEEG software package for display and user-review.
The device herein described consists of a set of tables that represent the reference means and standard deviations for representative samples. These tables are implemented as computer files that provide access to the exact tabular data resource for use by software that uses the tables as an information resource. The system requires that the user select reliable samples of artifact-free, eyes-closed or eyes open, resting digital EEG for purposes of analysis.
Analysis consists of the Fast-Fourier Transformation (FFT) of the data to extract the spectral power for each of the designated frequency bands (e.g. delta, theta, alpha, and beta), and frequency information from the EEG. The results of this analysis are then displayed in statistical tables and topographical brain maps of absolute and relative power asymmetry, and coherence for 19 monopolar and 171 selected bipolar derivations of the EEG.
In all over 4,000 measures are derived for comparison against carefully constructed and statistically controlled age-regressed, normative database in which the variables have been transformed and validated for their Gaussian distribution.
Each variable extracted by the analysis is compared to the database using parametric statistical procedures that express the differences between the patient and an appropriate age-matched reference group in the form of z-scores.
The BrainView QEEG Software Package is intended for prescription use by qualified medical personnel.
The device is intended for use by qualified medical personnel only and qualifies for exemption per 21 CFR 801 Subpart D Prescription devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Medeia. The logo has the word "Medeia" in bold, with the "M" in green and blue and the rest of the letters in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller font.
#### TECHNOLOGICAL CHARACTERISTIC COMPARISON: a6
| ltem | BrainView QEEG | NeuroGuide Analysis<br>System (NAS)<br>K041263 | qEEG-Pro<br>K171414 |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | Subject Device | Predicate Device | Reference Device |
| Indications for<br>Use | The BrainView QEEG<br>system is to be used by<br>qualified medical and<br>qualified clinical<br>professionals for the<br>post-hoc statistical<br>evaluation of the human<br>electroencephalogram<br>(EEG).<br>Rx-only | The NAS system is to be<br>used by qualified medical<br>and qualified clinical<br>professionals for the<br>post-hoc statistical<br>evaluation of the human<br>electroencephalogram<br>(EEG).<br>OTC | The qEEGpro system is to<br>be used by qualified<br>medical and qualified<br>clinical professionals for<br>the post-hoc statistical<br>evaluation of the human<br>electroencephalogram<br>(EEG).<br>Rx-only |
| EEG data<br>comparison<br>against normative<br>database | Yes; 2303 subjects<br>(eyes closed);<br>1965 subjects<br>(eyes open) | Yes; 625 samples | Yes; 1482 samples<br>(eyes closed);<br>1231 subjects<br>(eyes open) |
| EEG Spectral<br>Analysis | Yes; 4 frequency bands<br>(delta, theta, alpha, and<br>beta) | Yes; 4 frequency bands<br>(delta, theta, alpha, and<br>beta) | Yes; 4 frequency bands<br>(delta, theta, alpha, and<br>beta) |
| Age Range<br>Included in the<br>Normative<br>Database | 4-85 years | 2 months-82 years | 4-82 years |
| Product code | OLU | OLU | OLU |
| Classification | 882.1400 | 882.1400 | 882.1400 |
| Visual Display of<br>EEG | Yes | Yes | Yes |
| Software | Proprietary via DLL | Proprietary via DLL | Proprietary via DLL |
| Software<br>Features | Onscreen QEEG Z-Scores<br>and maps | Onscreen QEEG Z-Scores<br>and maps | Onscreen QEEG Z-Scores<br>and maps |
| Frequency Range | 1 - 40 Hz | 0.5 to 40 Hz | 1 - 40 Hz |
| Ratio of power | Yes | Yes | Yes |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Medeia. The logo has the word "Medeia" in bold, with the first letter "M" in green and blue, and the rest of the letters in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller font.
#### b1 NON-CLINICAL TESTING:
Non-clinical performance testing included delta, theta, alpha, and beta comparison of the subject and predicate device using a variety of simulated signals which were analyzed for frequency and power. These performance data demonstrated confirmation of pre-specified, objective evidence to specify that output requirements for the software have been fulfilled and met through static and dynamic analyses and code and document inspections. The software testing performance data further established that the software device's specifications consistently conform to the pre-specified user needs and the intended use. The algorithms and statistical methods used for data analysis were also evaluated through these tests. Therefore, the testing demonstrated the that the system accurately translates and presents EEGs from patients.
Potential adverse effects of the device are known if the BrainView QEEG software package is used as a standalone diagnostic system in the absence of other clinical data from more traditional means of patient evaluation. Relying only upon the use of a single index (such as relative power or the topological maps alone) without reviewing the traditional EEG, the epochs selected for analysis, or the complete set of statistical summary tables is also contraindicated and a source of potential error. Additional sources of error could arise from the inappropriate selection of EEG (selecting EEG epochs with artifacts, or by purposely selecting conditions for testing other than those specified). Additionally, it is possible that errors will occur through the purposeful falsification of symptoms in the patient history and patient age.
### Referenced Standards and Performance Testing:
The BrainView QEEG Software Package was developed using:
- IEC 62304:2015 [Edition 1.1] Medical device software — Software life cycle processes
## Software Verification and Validation Testing
Software verification and validation testing were conducted following the FDA guidance document for software contained in medical devices. The software was considered to be a "moderate" level of concern since a failure or latent flaw could indirectly result in a minor injury to the patient through incorrect or delayed information or through action of the operator.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Medeia. The word "Medeia" is written in a combination of colors, with the "M" in green and blue, and the rest of the letters in black. Below the word "Medeia" is the phrase "Designed for Clinicians" in a smaller font size.
#### b2 CLINICAL TESTING:
Clinical testing of the subject device included the use clinically acquired EEG waveforms from selected subjects who were used to validate performance of subject device database to that of the predicate K041263. These subjects were adequate to provide a range of values of the databases to verify performance since they are part of the adult and pediatric range of ages as well as the frequencies within the databases.
Acceptance criteria were defined as the BrainView QEEG software produces results sufficiently in agreement with the predicate device and that the R-squared factor shall be 0.8 or better. Additionally, the observed range of results obtained from the predicate device shall be used to verify that the BrainView QEEG produces results in agreement with the results obtained from the predicate device. The pre-defined acceptance criteria were met as 10-minute EEG recordings for eyes closed and eyes open of the subjects were de-artifacted and used to calculate z-scores for absolute power for the subject and predicate databases. Although the sample size was small, it was possible to validate results by computing values for all discrete ages ranging between 4 and 85, resulting in 23 age grouped sets of Z-scores for each subject's EEG sample which were compared with the predicate device's output and found to be similar.
#### b3 CONCLUSIONS:
The BrainView QEEG software has the same intended use as the predicate device, and it has the same manner of use and function, being a software-based database. Furthermore, it has similar requirements for training and expectations of intended users. The systems have equivalent performance in terms of data sampling and accuracy in the reference norms across age. Based on the device description, IFU, and performance testing, the BrainView QEEG software package is substantially equivalent to the predicate.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.