ControlSeal Electrode (ELSB)

K223605 · Coapt · GXY · Dec 30, 2022 · Neurology

Device Facts

Record IDK223605
Device NameControlSeal Electrode (ELSB)
ApplicantCoapt
Product CodeGXY · Neurology
Decision DateDec 30, 2022
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.1320
Device ClassClass 2

Indications for Use

The ControlSeal Electrodes are intended for non-invasive use with recording and monitoring equipment of Electromyography (EMG).

Device Story

ControlSeal Electrode is a non-invasive, reusable cutaneous electrode designed to record biopotential signals from the skin surface. It features a 316L stainless steel dome contacting surface and a standard threaded stud for connection to external EMG recording and monitoring equipment. The device functions as a passive conductor, creating an equipotential surface to detect electrical potentials. It is intended for use in clinical or diagnostic settings where EMG monitoring is required. The electrode is compatible with various EMG systems. By providing a stable interface for signal acquisition, it assists healthcare providers in obtaining EMG data for clinical decision-making and patient assessment.

Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing, including mechanical fit, voltage continuity, signal detection, mechanical strength, and sealing tests, as well as validation with compatible Coapt systems. The device conforms to IEEE 2010-2012 and ISO 10993-1 standards.

Technological Characteristics

Materials: 316L stainless steel (contacting surface), 18-8 stainless steel (screw), Buna-N rubber (O-ring). Sensing principle: Passive biopotential signal conduction. Form factor: Semi-spherical dome, 9.906 mm diameter. Connectivity: Standard threaded stud for connection to external leads. Biocompatibility: Evaluated per ISO 10993-1.

Indications for Use

Indicated for non-invasive recording and monitoring of Electromyography (EMG) signals in patients requiring such diagnostic or monitoring procedures.

Regulatory Classification

Identification

A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 30, 2022 Coapt Kevin Dwyer Regulatory Compliance & Legal Affairs Manager 303 W. Institute Pl. Suite 200 Chicago, Illinois 60610 Re: K223605 Trade/Device Name: ControlSeal Electrode (ELSB) Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: December 1, 2022 Received: December 2, 2022 Dear Kevin Dwyer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Heather L. Dean -S Heather Dean, PhD Assistant Director, Acute Injury Devices Team, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K223605 Device Name ControlSeal Electrode (ELSB) Indications for Use (Describe) The ControlSeal Electrodes are intended for non-invasive use with recording and monitoring equipment of Electromyography (EMG). # Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for COAPT. The letters "CO" are in orange, and the letters "APT" are in gray. Below the letters is the text "COMPLETE CONTROL" in gray. # Special 510(k) Summary - ControlSeal Electrode # 1. Submitter Information | Manufacturer: | Coapt, LLC<br>303 W. Institute Pl. Suite 200<br>Chicago, IL 60610 | |----------------------|-------------------------------------------------------------------| | Contact Person: | Kevin Dwyer<br>Regulatory Compliance & Legal Affairs<br>Manager | | Contact Information: | (844)262-7800 x716<br>kevin.dwyer@coaptengineering.com | | Date Prepared: | 12/01/2022 | ### 2. Device Identification | Trade/Proprietary Name: | ControlSeal Electrode | |-------------------------|--------------------------| | Common/Usual Name: | Cutaneous Electrode | | Classification Name: | Electrode, cutaneous | | Regulation Number: | 21 CFR §882.1320 | | Product Code: | GXY, Cutaneous electrode | | Device Class: | Class II | | Classification Panel: | Neurology | # 3. Legally Marketed Predicate Device | 510(k) Number | Device Name | |---------------|-----------------| | K190416 | Dome Electrodes | {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for COAPT Complete Control. The letters "COA" are in orange, and the letters "PT" are in gray. Below the letters is the text "COMPLETE CONTROL" in gray. # 4. Device Description The ControlSeal Electrode is an accessory designed for recording of biopotential signals. It is a non-invasive electrode that records biopotential signals from the surface of the skin. It has a standard mating thread that enables connection to electrical conductors of compatible devices. The ControlSeal Electrode contains the following components: - . 316L Stainless Steel ControlSeal Electrode Dome - Passivated 18-8 Stainless Steel Phillips Flat Head 2-56 Screw . - Stainless Steel Tooth Lock Washer . - . Protective Cap - Conducting Ring-Terminal . - Buna-N Rubber O-Ring . #### 5. Indication for Use Statement The ControlSeal Electrodes are intended for non-invasive use with recording and monitoring equipment of Electromyography (EMG). #### 6. Comparison to Predicate Devices The following table compares the ControlSeal Electrode to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials used, performance, and safety/performance testing. The subject device does not raise any new issues of safety or efficacy based on the similarities to the predicate device. | Manufacturer: | Coapt, LLC | Coapt, LLC | Device<br>Comparison | |--------------------------------|-----------------------|---------------------|----------------------| | Trade Name: | ControlSeal Electrode | Dome Electrode | | | 510(k) Number | TBD | K190416 | Not applicable | | Classification<br>Product Code | GXY | GXY | Same | | Regulation Number | 21 CFR 882.1320 | 21 CFR 882.1320 | Same | | Regulation Name | Cutaneous electrode | Cutaneous electrode | Same | {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for COAPT. The letters "CO" are in orange, while the letters "APT" are in gray. Below the letters is the phrase "COMPLETE CONTROL" in gray. | Indications for Use | The ControlSeal Electrodes<br>are intended for non-invasive<br>use with recording and<br>monitoring equipment of<br>Electromyography (EMG). | The Dome Electrodes are<br>intended for non-invasive use<br>with recording and<br>monitoring equipment of<br>Electromyography (EMG). | Same | |--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------| | Use/Field of<br>Application | The Coapt ControlSeal<br>Electrodes are compatible<br>with most of the EMG<br>equipment that are used in<br>the world. | The Coapt Dome Electrodes<br>are compatible with most of<br>the EMG equipment that are<br>used in the world. | Same | | Mechanism of<br>Action | Non-invasive transfer of<br>biopotential signals | Non-invasive transfer of<br>biopotential signals | Same | | Electrode<br>Dimensions | Diameter - 9.906 mm<br>Surface Area – 2.73 cm2 | Diameter - 9.525 mm<br>Surface Area – 2.80 cm2 | Similar | | Electrode Contact<br>Material | 316L Stainless Steel | 316L Stainless Steel | Same | | Electrode resistivity<br>as a % of typical<br>skin impedance | $3.57 x 10^{-12}\%$ | $3.57 x 10^{-12}\%$ | Same | | Electrode<br>Connection Thread | Common thread stud | Common thread stud | Similar.<br>Different<br>standard<br>thread sizes,<br>but achieve the<br>same purpose. | | Biocompatibility<br>Testing Passed | See ControlSeal ISO 10993<br>Biocompatibility Discussion | ISO 10993-5: Cytotoxicity<br>ISO 10993-10: Sensitization<br>ISO 10993-10: Irritation | Similar | Both the ControlSeal Electrode and the predicate, Dome Electrode are classified as product code: GXY, cutaneous electrodes under 21 CFR 882.1320. Both devices operate in the same manner – the non-invasive transfer of biopotential signals. The intended use of the predicate Dome Electrode is for non-invasive use with recording and monitoring equipment, (active and reference), of Electromyography (EMG). The intended use of the ControlSeal Electrode the same - for non-invasive use with recording and monitoring equipment of Electromyography (EMG). This does not raise any concerns of safety or effectiveness. The ControlSeal Electrode and predicate Dome Electrode are both recording the electrical potentials passively. Both the ControlSeal Electrode and Dome Electrode are able to detect these biopotential signals because they create an equipotential surface that conduct the electrical potentials. The use/field of application for the ControlSeal Electrode and predicate are the same. Both the ControlSeal Electrode and the predicate record biopotential signals from cutaneous locations. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for COAPT. The letters "COA" are in orange, and the letters "PT" are in gray. Below the letters is the phrase "COMPLETE CONTROL" in a smaller font. The ControlSeal Electrode and predicate device share similar physical dimensions and shape. They both have a diameter between 9 and 10mm. Both the ControlSeal Electrode and Dome Electrode share a semi-spherical shape for the contacting surface of the electrodes. The difference in surface area of the contacting surfaces of the ControlSeal Electrode and predicate do not cause any concerns of safety. The ControlSeal Electrode has also passed performance testing demonstrating the ability to detect quality EMG signals from cutaneous locations on a user. The ControlSeal Electrode contacting surface is made from 316L Stainless Steel. The material of the contacting surface for the predicate Dome Electrode is also made from 316L Stainless Steel. The goal in designing the ControlSeal Electrodes was to create a surface electrode for recording EMG signals that was reusable, biocompatible, and cost effective. The ControlSeal Electrodes underwent performance testing to assess the conductivity of EMG signals. The ControlSeal Electrodes passed all performance testing for EMG signal detection. Additionally, the predicate device, Dome Electrode was tested for conformance to the biocompatibility specifications of ISO 10993-5 for cytotoxicity, ISO 10993-10 for skin sensitization, and ISO 10993-10 for irritation. Following the Biocompatibility Risk Analysis, no additional testing is required to evaluate the biocompatibility of the ControlSeal Electrodes. The history of safe use of 316L Stainless Steel in medical devices and comparison to the previously cleared version of the device have resulted in no additional concerns for safety and effectiveness of the ControlSeal Electrode. The intended use, anatomical location, duration, size, material, passivation standard, of the ControlSeal Electrode is the same as the previously cleared device. Dome Electrode. Additionally, there have not been any adverse events in relation to biocompatibility reported for the previously cleared. Dome Electrodes. The 316L Stainless Steel was chosen as the material for the ControlSeal Electrode because it fits the intended application of a reusable, biocompatible, low-cost EMG surface electrode. Also, as discussed in the comparison to uncleared predicate electrodes, stainless steel is widely used in the prosthetic industry. Because the subject device, ControlSeal Electrode and predicate device, Dome Electrode are made from the same material, they would have the same Electrode resistivity as a % of typical skin impedance. A calculation using the conductivity of the materials, and the average impedance of the skin was performed by comparing the resistivity of gold and 316L Stainless Steel as a percentage of the typical impedance of the skin. The resistance of the skin is about 1000 ohms, making the resistivity of the electrodes 3.57 x 10-12%. As with the predicate, the ControlSeal Electrode is designed to operate with a wide range of compatible devices. The ControlSeal Electrode can be connected to compatible devices through a standard sized threaded stud. It is not permanently connected to any electrode leads. The predicate device, Dome Electrodes use a 4-40 standard sized thread, and the subject device. ControlSeal Electrode uses a 2-56 standard sized thread. The differences in electrode connection do not raise any concerns of safety or effectiveness of the ControlSeal Electrode, they are simply alternative means of connection to compatible devices. The method of connection of the ControlSeal Electrode does not have an effect on its ability to detect EMG {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for COAPT Complete Control. The word "COAPT" is written in large, bold letters, with the first two letters in orange and the last two in gray. Below the word "COAPT" is the phrase "COMPLETE CONTROL" in smaller, gray letters. The logo is simple and modern, and the use of color helps to make it stand out. signals. The ControlSeal Electrodes have passed performance testing for detection of EMG signals. # 7. Non-Clinical Performance Data As part of demonstrating safety and effectiveness of the ControlSeal Electrode. substantial equivalence to the predicate device, and in showing conformance to the recognized consensus standard IEEE 2010-2012 Recommended Practice for Neurofeedback Systems that are subject to this Special 510(k) submission, Coapt, LLC completed a number of non-clinical performance tests. The ControlSeal Electrode meets all the requirements for overall design, safety, and biocompatibility results, confirming that the design output meets the design inputs and specifications for the device. The ControlSeal Electrode passed all the testing in accordance with internal requirements, national standards, and international standards shown below to support safety and efficacy of the subject device: - IEEE 2010-2012 Recommended Practice for Neurofeedback Systems . - . ISO 10993-1: 2009/(R)2013 Biological evaluation of medical devices—Part 1: Evaluation and testing within a risk management process The ControlSeal Electrodes were also tested internally to ensure that it meets device specifications & requirements and operates as intended. The following validation testing was performed on the finished device: | Test Name | Result | |------------------------------------------------------------------------|--------| | Mechanical Fit Test | Pass | | Voltage Continuity Test | Pass | | Signal Detection Test | Pass | | Mechanical Strength Test | Pass | | Sealing Test | Pass | | Validation – Compatible Device – Coapt<br>COMPLETE CONTROL System Gen2 | Pass | | Validation – Compatible Device – Coapt<br>Evaluation Kit | Pass | # 8. Clinical Performance Data No human clinical testing was required to support the medical device as it was designed to conform to IEEE 2010-2012 Recommended Practice for Neurofeedback Systems. Cutaneous electrodes have been on the market for many years with proven safety and efficacy. The non-clinical testing detailed in {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the logo for COAPT Complete Control. The letters "COA" are in orange, while the letters "PT" are in gray. Below the letters is the text "COMPLETE CONTROL" in a smaller font. this submission supports the supports substantial safety and efficacy when used for its intended purposes. # 9. Statement of Substantial Equivalence to Predicate Device By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise additional questions regarding its safety and effectiveness as compared to the predicate device(s). The technological characteristics between the ControlSeal Electrode and predicate are remarkably similar and therefore substantially equivalent. The slight differences between devices do not raise new questions of safety and effectiveness as compared to the predicate as the ControlSeal Electrodes have received passing results for performance testing of mechanical fit, voltage continuity, signal detection, sealing, and validation testing with compatible devices. The ControlSeal Electrode, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...