LONG LIFE PADTM

K120516 · Omron Healthcare, Inc. · GXY · May 17, 2012 · Neurology

Device Facts

Record IDK120516
Device NameLONG LIFE PADTM
ApplicantOmron Healthcare, Inc.
Product CodeGXY · Neurology
Decision DateMay 17, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1320
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The LONG LIFE PAD™ is intended for use as a disposable, conductive adhesive interface between the patient's skin and the Electrical Stimulator. These pads should be used only with the Omron Electrotherapy Pain Relief Unit, (PM3030) and applied on normal, healthy, dry, clean skin of adults.

Device Story

Long Life Pad is a disposable, conductive adhesive electrode interface; functions as a bridge between patient skin and Omron Electrotherapy Pain Relief Unit (PM3030). Device consists of a laminated, flexible structure: polyester protective film, biocompatible hydrogel, carbon layer, and non-skin contact polyethylene lamination with a snap connection. Used in clinics, hospitals, or home environments by adult patients. Hydrogel provides adhesion, eliminating need for external securing materials. Device is restricted to use with the specified Omron stimulator; provides electrical stimulation path for pain relief. Benefits include ease of use and biocompatible interface for electrotherapy.

Clinical Evidence

Bench testing only. Testing included dispersion, adhesion, stability (storage/shelf life), and shipment testing. Biocompatibility testing performed in accordance with ISO 10993-1 and FDA guidance.

Technological Characteristics

Laminated flexible structure: polyester protective film, biocompatible hydrogel, carbon layer, non-skin contact polyethylene lamination, and snap connection. Single-patient, multiple-application use. Non-sterile. Connectivity via snap connection compatible only with Omron PM3030. Materials tested per ISO 10993-1.

Indications for Use

Indicated for use as a disposable, conductive adhesive interface between skin and electrical stimulator for adult patients. Contraindicated for patients with cardiac pacemakers, implanted defibrillators, or other implanted metallic or electronic devices. Must be applied to normal, healthy, dry, clean skin.

Regulatory Classification

Identification

A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K120516 # MAY 1 7 2012 # 510(k) Summary Page 1 of 5 May 9, 2012 | Omron Healthcare, Inc.<br>1925 West Field Court<br>Lake Forest, IL 60045 | Tel - 847-247-5625<br>Fax - 847-680-6911 | |--------------------------------------------------------------------------|------------------------------------------------------------------------------| | Official Contact: | Mirna DiPano- Director, Quality & Regulatory | | Proprietary or Trade Name: | LONG LIFE PADTM | | Common/Usual Name: | Cutaneous electrode | | Classification Name/Code: | GXY – electrode, cutaneous<br>CFR 882.1320<br>Class 2 | | Device: | Long Life PadTM | | Predicate Device: | Well Life Healthcare Limited – Well-Life Self<br>Adhesive Electrodes K082065 | #### Device Description: The LONG LIFE PAD™ is intended for use as a disposable, conductive adhesive interface between the patient's skin and the Electrical Stimulator. These pads should be used only with the Omron Electrotherapy Pain Relief Unit, (PM3030) and applied on normal, healthy, dry, clean skin of adults. The LONG LIFE PADs™ are non-sterile, disposable, laminated, flexible structures composed of the following three main construction layers from bottom to top: - . Polyester protective film, - . gel, - . carbon layer, - . non-skin contact polyethylene lamination layer - . snap connection The electrodes are designed for single-patient/multiple application use. Because of the Adhesive's nature of the biocompatible hydrogel, no securing materials are required to secure the device to the patient's skin. The pad patient contacting surface is a biocompatible hydrogel. The electrode has one connection point that is used to connect the stimulation device to the electrodes. This connection point is compatible only with the Omron PM3030. {1}------------------------------------------------ # 510(k) Summary Page 2 of 5 May 9, 2012 #### Indications for Use: The LONG LIFE PAD™ is intended for use as a disposable, conductive adhesive interface between the patient's skin and the Electrical Stimulator. These pads should be used only with the Omron Electrotherapy Pain Relief Unit, (PM3030) and applied on normal, healthy, dry, clean skin of adults. #### Environment of Use: Clinics, hospital and home use environments #### Contraindications Do not use these pads for stimulation if you have a cardiac pacemaker, implanted defibrillator, or other implanted metallic or electronic device. Such use could cause electric shock, burns, electrical interference, or death. #### Summary of substantial equivalence The LONG LIFE PAD™ was compared to the predicate Well-Life Self Adhesive Electrodes (K082065) as in the device comparison table below. {2}------------------------------------------------ | Device Comparison | | | | |----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------| | Indications for Use | Well-Life Pad<br>510(k) K082065<br>The Well Life Self Adhesive<br>Electrode/CM, FA, PU and SP<br>series are intended for use as a<br>disposable, conductive adhesive<br>interface between the patient's<br>skin and the Electrical<br>Stimulator. Well Life Self<br>Adhesive Electrode / AP, CM,<br>FA, PU and SP series are<br>designed and intended to be<br>used With marketed Electrical<br>Stimulators, i.e. TENS<br>(Transcutaneous Electrical<br>Nerve Stimulation), MENS<br>(Microcurrent Electrical Nerve<br>Stimulation), EMS (Electrical<br>Muscular Stimulation), IF<br>(Interferential) and PGS (Pulsed<br>Galvanic Stimulation) | May 9, 2012<br>Long Life Pad<br>The LONG LIFE PADTM is<br>intended for use as a disposable,<br>conductive adhesive interface<br>between the patient's skin and the<br>Electrical Stimulator. These pads<br>should be used only with the<br>Omron Electrotherapy Pain Relief<br>Unit, (PM3030) and applied on<br>normal, healthy, dry, clean skin of<br>adults. | Comparison<br>Similar indications. The Long<br>Life Pad is restricted to the<br>Omron PM3030 stimulator<br>(K110068) | | Product Code<br>Prescriptive or<br>OTC | GXY | GXY | Identical | | | Both | OTC | Similar | | Environment Of<br>use | Clinics, hospital and home use<br>environments | Clinics, hospital and home use<br>environments | Identical | | | Well-Life Pad<br>510(k) K082065 | Long Life Pad<br>510(k) TBD | Comparison | | Patient<br>Population | Adult | Adult | | | Conductive<br>Interface | Biocompatible hydrogel | Biocompatible hydrogel | Identical | | Construction &<br>Materials | Bottom to top:<br>• PET protective layer,<br>• gel,<br>• scrim layer,<br>• carbon layer,<br>• non-skin contact white<br>foam layer<br>• with snap connection | Bottom to top:<br>• Polyester protective film,<br>• gel,<br>• carbon layer,<br>• non-skin contact<br>polyethylene lamination<br>layer<br>• with snap connection | Identical<br><br>Construction layers and materials<br>are very similar. | 510(k) Summary {3}------------------------------------------------ # 510(k) Summary : 、一 {4}------------------------------------------------ # 510(k) Summary Page 5 of 5 May 9, 2012 #### Differences Between Other Legally Marketed Predicate Devices The Omron Long Life Pad is viewed as substantially equivalent to the predicate device because: The Long Life Pad uses the same principal of operation as the predicate. The Long Life Pad uses similar materials. The Long Life Pad has approximately the same surface area. The differences that exist between the devices are insignificant in the terms of safety or effectiveness. Indications - The Long Life Pad and the predicate (Well-Life Self Adhesive Electrodes K 082065) are intended for use as disposable, conductive adhesive interfaces between the patient's skin and an electrical stimulator. Prescriptive - The Long Life Pad TM is OTC as is the predicate. Design and Technology - The Long Life Pad "M has equivalent design and features as the predicate and has the identical technology to the predicate. Performance and Specifications - The Long Life Pad TM has equivalent specifications of performance as the predicate. Materials - The patient contacting materials of the electrode has been tested in accordance to ISO 10993-1 and FDA Guidance as was the predicate. Environment of Use - Same Patient Population - The Long Life Pads are restricted to adults #### Performance Testing - We have performed bench tests and found that the Long Life Pads met all requirements specifications and "Draft Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Cutaneous Electrode". The testing is comprised of - . Dispersion Test Protocol Report Summary - . Adhesion Test Protocol Report Summary - Stability Testing (Storage Conditions) . - Stability Test Protocol Report (Shelf life) . - Stability Shipment Test Report Summary . #### Conclusion - The LONG LIFE PAD™ is substantially equivalent to the predicate Well-Life Self Adhesive Electrodes (K082065) in indications for use, patient population, and environment for use, technology characteristics, specifications and performance. {5}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Omron HealthCare, Inc. c/o Mr. Paul E. Dryden President, Regulatory Consultant for Omron 24301 Woodsage Drive Bonita Springs, FL 34134 Re: K120516 Trade/Device Name: Long Life Pad™ Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: February 15, 2012 Received: February 21, 2012 Dear Mr. Dryden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. MAY 1 7 2012 {6}------------------------------------------------ # Page 2 - Mr. Paul E. Dryden Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number -(800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Kevin Norouzi for Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ # Indications for Use Statement Page 1 of 1 #### 510(k) Number: # LONG LIFE PAD™ K120516 Device Name: #### Indications for Use: The LONG LIFE PAD™ is intended for use as a disposable, conductive adhesive interface between the patient's skin and the Electrical Stimulator. These pads should be used only with the Omron Electrotherapy Pain Relief Unit, (PM3030) and applied on normal, healthy, dry, clean skin of adults. Prescription Use (Part 21 CFR 801 Subpart D) or ### Over-the-counter use X (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) # Concurrence of CDRH, Office of Device Evaluation (ODE) MzLB Nicholas (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number K120516
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...