THERATRODE

K112312 · Phoenix Medical Devices, Llp · GXY · Jan 25, 2012 · Neurology

Device Facts

Record IDK112312
Device NameTHERATRODE
ApplicantPhoenix Medical Devices, Llp
Product CodeGXY · Neurology
Decision DateJan 25, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1320
Device ClassClass 2

Indications for Use

To conduct electrical stimulation from commercially available nerve stimulation devices to the patient's skin. Single patient use - re-usable. Self adhering and re-positionable. Over the counter use.

Device Story

Theratrode cutaneous electrodes serve as interface between nerve stimulation devices and patient skin. Device consists of four-layer assembly: patient-contacting hydrogel (conductive/adhesive), carbon dispersion middle layer (current distribution), non-conductive top layer (spun lace, polyethylene, or polypropylene foam), and lead wire with .080" (2mm) female connector. Connector mates with standard nerve stimulation devices. Intended for single-patient, multiple-use, over-the-counter application. Electrodes facilitate electrical current transmission to patient. No active electronic components; passive interface device.

Clinical Evidence

Bench testing only. Comparison of electrical conductivity (inverse of impedance) between subject and predicate devices demonstrated equivalent performance. Hydrogel material passed biocompatibility testing for skin irritation, cytotoxicity, and delayed contact sensitization.

Technological Characteristics

Layered assembly: hydrogel patient-contacting layer, carbon dispersion middle layer, non-conductive top layer (spun lace, polyethylene, or polypropylene foam). Passive device. .080" (2mm) female connector compliant with IEC60601-1 Sub clause 56 3(c). Non-sterile.

Indications for Use

Indicated for patients requiring electrical stimulation via nerve stimulation devices. For single-patient, re-usable, self-adhering, re-positionable, over-the-counter use.

Regulatory Classification

Identification

A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K112312 # Attachment H Image /page/0/Picture/2 description: The image shows the logo for Phoenix Medical Devices. The logo features a stylized phoenix bird with its wings spread above the company name. The text "PHOENIX MEDICAL DEVICES" is written in a bold, sans-serif font, with "PHOENIX" on the top line and "MEDICAL DEVICES" on the bottom line. JAN 2 5 2012 # 510(k) Summary for the Theratrode Family of Cutaneous Electrodes #### 1. Sponsor Phoenix Medical Devices, LLC 2458 Alton Parkway Irvine, Ca 92606 Registration Number: 3004620982 Contact Person: Jim Klett Telephone: (800) 689-9892 (949) 266-8126 Fax: Date Prepared: February 26, 2011 #### 2. Device Name Theratrode Brand Cutaneous Electrodes comprised of the following model numbers and descriptions: T1 = 2" (5 0.5mm) Square Electrode ("Squircle") T2 = 2" (5cm) Round Electrode T3 = 2″ x 4″ (5cm x 10cm) Rectanqu lar Electrode T4 = 2" x 4" (5cm x 10cm) Oval Electrode T5 = 1.25" (3cm) Round Electrode T6 = 2.75" (7cm) Round Electrode T7 = 1.5" (4cm) Square Electrode T8 = 1.5" x 2.5" (4cm x 6cm) Rectanqular Electrode T9 = 1.5" x 2.5" (4cm x 6cm) Oval Electrode T 10= 1.5" x 3.5" (4cm x 9cm) Rectangular Electrode T 1 = 3" x 5" (7.5cm x 13cm) Rectangular Electrode T12= 2″ (5 0mm) Square Electrode Common/Usual Name: Cutaneous Electrode Classification Names: Electrode, Cutaneous Classification Panel: Neurology Panel/Product Code: 882.1320 / GXY {1}------------------------------------------------ K112312 #### 3. Legally Marketed Device to Which Equivalence is Claimed Marketed Device #1 Proprietary Name: Lifecare Electrodes, K083302 Common/Usual Name: Lifecare Neurostimulation Electrodes Classification Names: Electrode, Cutaneous Classification Panel: Neurology Panel/Product Code: Neurology / GXY #### 4. Intended Use / Indications For Use To conduct electrical stimulation from commercially available nerve stimulation devices to the patient's skin. Single patient use - re-usable. Self adhering and re-positionable. Over the counter use. #### 5. Device Description Theratrode electrodes are constructed as a layered assembly comprised of four components: - A patient contacting layer of hydrogel material which has been tested and . found to be bio-compatible with humans and provides both the electrically conductive medium necessary to aid in the transmission of electrical current to the patient plus the adhesive properties necessary to maintain sufficient contact with the patient's skin. - A carbon dispersion pad middle layer that evenly distributes the electrical . current across the surface of the electrode, - A non-conductive top layer of various materials such as spun lace (fabric), . polyethylene or polypropylene foam or other similar materials that form a protective and flexible top layer to the electrode. - A wire or conductive carbon fiber lead wire which is glued to the assembly . of the middle and top layer and terminates in a .080" (2mm) female connector common to the electrotherapy industry and which mates with the plurality of commercially available nerve stimulation devices on the market today. The female connector complies with IEC60601-1 Sub clause 56 3(c). Theratrode electrodes are non-sterile and are intended for multiple use by a single patient to apply electrical stimulation. Theratrode's construction is equivalent to the predicate device's construction. #### 6. Basis for Substantial Equivalence {2}------------------------------------------------ K112312 We have examined the Theratrode electrodes in comparison to the Predicate Device and determined the Theratrode electrodes to be Substantially Equivalent to the Predicate Device because they both: - . Are constructed in the same manner, - are constructed of same or similar materials, . - have identical indications for use, and . - have very similar performance characteristics (see bench test data . included in this submission). Any minor visual, dimensional or labeling differences between the Predicate device and the Theratrode electrodes do not pose any risk to their safe and effective use. - 7. Differences between the Marketed Device and the Theratrode family of cutaneous electrodes. There are no significant differences between the Theratrode Electrodes and the Marketed device. #### 8. Bench Testing (non-clinical) Comparison testing was performed between the Predicate Device and the Theratrode Electrode. There are no established performance standards for cutaneous electrodes so we chose to measure the electrical conductivity (inverse of impedance) as an appropriate measure of the electrode's performance. The comparison tests conclusively prove that Theratrode is at a minimum as efficient at conducting electricity as the Predicate device. Theratrode's hydrogel has passed three biocompatibility tests; skin irritation, cytotoxicity and delaved contact sensitization #### 9. Conclusion Through careful examination of the construction, materials, indications for use and performance we conclude the Theratrode electrodes are Substantially Equivalent to the Predicate Device. {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three lines forming the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -- WO66-G609 Silver Spring, MD 20993-0002 Phoenix Medical Devices, LLC c/o Mr. Jim Klett President 2458 Alton Parkway Irvine, CA 92606 Re: K112312 Trade/Device Name: Theratrode Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: II Product Code: GXY Dated: November 23, 2011 Received: November 29, 2011 Dear Mr. Klett: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical JAN 2 5 2012 {4}------------------------------------------------ ### Page 2 - Mr. Jim Klett device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. . If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Deboral Fills Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): 《//23/2 Device Name: Theratrode Model Numbers: T1 = 2" (50.5mm) Square Electrode (Squircle Shape) T2 = 2" (5cm) Round Electrode T3 = 2" x 4" (5cm x 10cm) Rectangular Electrode T4 = 2" x 4" (5cm x 10cm) Oval Electrode T5 = 1.25" (3cm) Round Electrode T6 = 2.75" (7cm) Round Electrode T7 = 1.5" (4cm) Square Electrode T8 = 1.5" x 2.5" (4cm x 6cm) Rectangular Electrode T9 = 1.5" x 2.5" (4cm x 6cm) Oval Electrode T10= 1.5" x 3.5" (4cm x 9cm) Rectangular Electrode T11= 3" x 5" (7.5cm x 13cm) Rectangular Electrode T12 = 2" (50mm) Square Electrode Indications for Use: To conduct electrical stimulation from common varieties of electrical stimulators to the patients skin. Single patient use - re-usable. Self adhering and re-positionable. Over the counter use. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Eyaluation (ODE) (Division Sign-Off) Division of Ophthalmic, Neurological and Ear. Nose and Throat Devices 510(k) Number K112312 Page of Phoenix Medical Devices 510(k) Submission for Theratrode 1-800-689-9892 Page 18 of 100
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...